Deltec 21-2961-24Recalls & reported problems
Icu Medical, Inc. · Class 2
All models of Deltec (97)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 10610586027697), received between Mar 23, 2017 and Feb 10, 2026 — classified as 1 injury, 12 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (4 reports).
1 other recall cover devices cleared under the same FDA submission (K021999); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak4
- Obstruction of Flow2
- Air/Gas in Device1
- Complete Blockage1
- Crack1
- Defective Device1
- Difficult to Remove1
- Failure to Prime1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- No Consequences Or Impact To Patient1
- Post Operative Wound Infection1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 10, 2026Fluid/Blood Leak
IT WAS STATED THAT THERE WAS AN EQUIPMENT MALFUNCTION ENCOUNTERED IN THE ONCOLOGY DAY HOSPITAL (HDJ) DURING A CHEMOTHERAPY SESSION. THE NURSE NOTICED A DROP-BY-DROP LEAKAGE AT THE LEVEL OF THE GRIPPER VALVE. FURTHER ON, IT WAS NOTICED THAT THE CHEMOTHERAPY SOLUTION LEAKED ONTO THE PATIENT¿S CLOTHING. THE INFUSION WAS STOPPED AND THE PATIENT¿S SKIN WAS DECONTAMINATED. AT THAT TIME, NO SKIN REACTION HAS BEEN OBSERVED. THE PATIENT WAS SEEN AGAIN, THE NEXT DAY FOR SKIN DECONTAMINATION. THERE WAS PATIENT INVOLVEMENT AND NO CLINICAL CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24303085 ↗MalfunctionApr 21, 2025Complete Blockage; Complete Blockage
IT WAS REPORTED THAT THE SALINE WON'T FLOW. THE CUSTOMER TRIED TO PUSH IT OUT, BUT THE SYRINGE PLUNGER IS TOO HARD. THE EVENT OCCURRED BEFORE PATIENT USE AT THE FACILITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21875998 ↗MalfunctionDec 6, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS STATED THAT PATIENT ON CHEMO WAS SENT HOME WITH BAXTER INFUSER PUMP, AND IT WAS FOUND OUT THAT CHEMO WAS LEAKING UNDER THE PAC GRIPPER. TESTING ON THE DEVICE WAS PERFORMED USING FLUSHING WITH NORMAL SALINE TO FIND THE LEAK AND SMALL LEAK UNDER GRIPPER WAS FOUND. THERE WAS A PATIENT INVOLVEMENT, AND NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20864912 ↗MalfunctionSep 25, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT DURING THE CLEANING OF THE DEVICE WITH SALINE, THE NURSE NOTICED A LEAK AT THE Y-SITE. THE PLASTIC TUBE APPEARS TO BE BROKEN ON HALF OF THE DIAMETER OF THE RIGID PLASTIC BASE. THERE WAS NO PATIENT INVOLVEMENT, AND NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20301361 ↗MalfunctionSep 18, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT WHEN FLUSHING GRIPPER, NOTED SOLUTION TO BE POOLING ON PROXIMAL PORT MEMBRANE ON OUTSIDE. THERE WAS PATIENT INVOLVEMENT, BUT NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20251080 ↗MalfunctionSep 5, 2023Crack; Air/Gas in Device; Crack; Air/Gas in Device
IT WAS REPORTED THAT DURING PATIENT USE IT WAS NOTICED THAT THE Y SITE PART WAS CRACKED. WHEN THE NEEDLE WAS REMOVED, HEPARIN WAS LOCKED, BUT THE AIR BUBBLED IN AND COULD NOT BE RELEASED FROM THE SYRINGE. THE MEDICATION BEING INFUSED AT THE TIME WAS CHEMO. NO ADVERSE PATIENT EFFECTS WERE REPORTED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17684883 ↗MalfunctionMay 11, 2023Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED, THAT THE CUSTOMER INFUSED MEDICAL FLUID (PHYSIOLOGICAL SALINE SOLUTION) INTO THE PATIENT THROUGH THE PRODUCT USING A SYRINGE. BUT BACKFLOW OF MEDICAL FLUID WAS NOT OBSERVED. SO THE CUSTOMER REMOVED THE PRODUCT FROM THE PATIENT, AND PASSED MEDICAL FLUID THROUGH THE PRODUCT WITH THE SYRINGE CONNECTED. HOWEVER, AT THAT TIME, MEDICAL FLUID WAS UNABLE TO BE DRAWN (ONLY AIR BUBBLE WERE DRAWN). NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16916265 ↗MalfunctionSep 11, 2022Failure to Prime; Failure to Prime
IT WAS REPORTED THAT THE DEVICE FAILED TO PRIME DURING PRE-USE CHECK. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15397870 ↗MalfunctionSep 10, 2022Flushing Problem; Flushing Problem
IT WAS REPORTED THAT THE CUSTOMER PERFORMED A PRIMING WITHOUT PROBLEM AND WAS ABLE TO CONFIRM A BLOOD RETURN FROM THE PATIENT. HOWEVER, UPON FLUSHING, HE NOTICED THAT THE MEDICAL FLUID WAS UNABLE TO BE INFUSED. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15397560 ↗MalfunctionSep 6, 2022Defective Device; Defective Device
IT WAS REPORTED THAT THE SYRINGE CONNECTED TO THE PRODUCT AT A PRESSURE CHECK WAS CAUSING MORE AMOUNT OF AIR BEING DRAWN. THE CUSTOMER SUSPECTED LEAKAGE OF AIR OCCURRED WITH THE DEVICE. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15363123 ↗MalfunctionAug 23, 2022Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED THAT DURING A PRE-USE CHECK, THE CUSTOMER NOTICED THE INFUSION ROUTE WAS OCCLUDED. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15277970 ↗InjuryAug 4, 2022Use of Device Problem; Infusion or Flow Problem; Use of Device Problem; Infusion or Flow Problem
IT WAS REPORTED THAT DURING THE USE OF THE PRODUCT, THE SKIN AT THE PUNCTURE SITE BECAME SUPPURATED. ADDITIONALLY THE REPORTER STATED "I RECOGNIZE THAT IT IS A PROBLEM WITH THE TECHNIQUE, BUT I WOULD LIKE YOU TO ANALYZE THE PRODUCT AS WELL FOR GOOD LIQUID FLOW."
Patient problems: Post Operative Wound Infection
MAUDE report 15163218 ↗MalfunctionMar 23, 2017Difficult to Remove; Protective Measures Problem; Difficult to Remove; Protective Measures Problem
IT WAS REPORTED THAT A DELTEC® GRIPPER PLUS® SAFETY NEEDLE WAS DIFFICULT TO TAKE OUT OF THE PORT DUE TO TIGHTNESS AND WHEN IT WAS REMOVED FORCEFULLY, THE SAFETY DID NOT FUNCTION. THIS WAS OBSERVED IMMEDIATELY UPON USE WITH A PATIENT. IT WAS NOTED THAT THE DEVICE WAS CHECK IN ACCORDANCE WITH ITS INSTRUCTIONS FOR USE. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6427449 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K021999. One submission often covers many models, so check the product and lot details in each record.
- Z-0253-2020 · Jul 15, 2019 · TerminatedPort full Access Tray Product Code: DT19330
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page