Deltec 21-3256-24Recalls & reported problems

Icu Medical, Inc. · Class 2

All models of Deltec (97)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10610586029790), received between Feb 9, 2017 and Feb 26, 2026 — classified as 2 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Difficult or Delayed Separation” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult or Delayed Separation3
  • Dull, Blunt3
  • Fail-Safe Did Not Operate2
  • Break1
  • Fluid/Blood Leak1
  • Protective Measures Problem1

Patient problems

  • Anxiety3
  • Needle Stick/Puncture3
  • Pain3
  • Exposure to Body Fluids2
  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 26, 2026Fluid/Blood Leak

    IT WAS REPORTED THAT WHILE ADMINISTERING AN INFUSION TO THE PATIENT, THE DOCTOR FOUND THAT THE DISPOSABLE IMPLANTABLE DRUG DELIVERY DEVICE'S INFUSION NEEDLE WAS DAMAGED, WITH BLOOD LEAKING, AND IT COULD NOT BE USED ON THE PATIENT, AND IT WAS PROMPTLY REPLACED WITH A NEW ONE. THERE WAS PATIENT INVOLVEMENT AND NO HARM/ ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24455308 ↗
  • MalfunctionJan 10, 2025Break; Break

    IT WAS REPORTED THAT THE NEEDLE SNAPPED DURING USE. THERE WAS PATIENT INVOLVEMENT. THE NEW PORT NEEDLE WAS USED TO ACCESS THE PATIENT'S IMPLANTED PORT. UPON DISPOSING OF THE INDUCER IN THE SHARPS BIN, THE SAFETY COVER, WHICH WAS SUPPOSED TO COVER THE GUIDE NEEDLE, POPPED BACK, EXPOSING THE NEEDLE AND RESULTING IN THE NEEDLE STABBING THE HEALTHCARE WORKER'S FINGER. THE DETAILS OF THE INJURY INDICATE A SHARPS INJURY SUSTAINED BY THE HEALTHCARE WORKER. NO PATIENT INJURY WAS REPORTED.

    Patient problems: Needle Stick/Puncture

    MAUDE report 21124083 ↗
  • InjuryMay 4, 2023Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    IT WAS REPORTED THAT THE NURSE WAS DE-ACCESSING A PORT AND THE SAFETY MECHANISM FAILED, RESULTING IN NEEDLESTICK AND BODILY FLUID EXPOSURE. THERE WAS A CLINICAL INJURY. PATIENT HAD BLOODBORNE DISEASE, RESULTING IN EXPOSURE OF THE HEALTHCARE TEAM MEMBER. TREATMENT INCLUDED MEDICAL PROPHYLAXIS FOR BLOODBORNE PATHOGEN. THE PATIENT DID NOT RECEIVE ANY MEDICAL TREATMENT. THE EVENT IS LISTED AS ONGOING DUE TO CONTINUED MEDICAL PROPHYLAXIS OF HEALTHCARE TEAM MEMBER. THE PATIENT INFORMATION WAS UNKNOWN.

    Patient problems: Exposure to Body Fluids; Needle Stick/Puncture

    MAUDE report 16869834 ↗
  • InjuryMay 2, 2023Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    IT WAS REPORTED THAT THE NURSE WAS DE-ACCESSING A PORT AND THE SAFETY MECHANISM FAILED, RESULTING IN NEEDLESTICK AND BODILY FLUID EXPOSURE. THERE WAS A CLINICAL INJURY. PATIENT HAD BLOODBORNE DISEASE, RESULTING IN EXPOSURE OF THE HEALTHCARE TEAM MEMBER. TREATMENT INCLUDED MEDICAL PROPHYLAXIS FOR BLOODBORNE PATHOGEN. THE PATIENT DID NOT RECEIVE ANY MEDICAL TREATMENT. THE EVENT IS LISTED AS ONGOING DUE TO CONTINUED MEDICAL PROPHYLAXIS OF HEALTHCARE TEAM MEMBER. THE PATIENT INFORMATION WAS UNKNOWN. TIME OF EVENT WAS 8:30.

    Patient problems: Exposure to Body Fluids; Needle Stick/Puncture

    MAUDE report 16856000 ↗
  • MalfunctionJul 20, 2022Dull, Blunt; Difficult or Delayed Separation; Dull, Blunt; Difficult or Delayed Separation

    IT WAS REPORTED THAT THE CANNULAS ARE PARTLY BLUNT, SO REPEATED PUNCTURING WAS NECESSARY AND THE SAFETY CLOSURE, WHICH MUST BE REMOVED AFTER PUNCTURE, WAS SO TIGHTLY ATTACHED THAT A STRONG PULL WAS NECESSARY. THERE WAS NO PERMANENT DAMAGE AND NO EXTRAVASATION, BUT THE REPEATED PUNCTURING OF THE CANNULA CAUSED PAIN FOR THE PATIENTS AND FEAR OF FURTHER PUNCTURING DUE TO THE TIGHT PULLING OF THE CLOSURE. NO INVASIVE INJURIES OCCURRED IN RELATION WITH PATIENTS.

    Patient problems: Anxiety; Pain

    MAUDE report 15066898 ↗
  • MalfunctionJul 20, 2022Dull, Blunt; Difficult or Delayed Separation; Dull, Blunt; Difficult or Delayed Separation

    IT WAS REPORTED THAT THE CANNULAS ARE PARTLY BLUNT, SO REPEATED PUNCTURING WAS NECESSARY AND THE SAFETY CLOSURE, WHICH MUST BE REMOVED AFTER PUNCTURE, WAS SO TIGHTLY ATTACHED THAT A STRONG PULL WAS NECESSARY. THERE WAS NO PERMANENT DAMAGE AND NO EXTRAVASATION, BUT THE REPEATED PUNCTURING OF THE CANNULA CAUSED PAIN FOR THE PATIENTS AND FEAR OF FURTHER PUNCTURING DUE TO THE TIGHT PULLING OF THE CLOSURE. NO INVASIVE INJURIES OCCURRED IN RELATION WITH PATIENTS.

    Patient problems: Anxiety; Pain

    MAUDE report 15066804 ↗
  • MalfunctionJul 20, 2022Dull, Blunt; Difficult or Delayed Separation; Dull, Blunt; Difficult or Delayed Separation

    IT WAS REPORTED THAT THE CANNULAS ARE PARTLY BLUNT, SO REPEATED PUNCTURING WAS NECESSARY AND THE SAFETY CLOSURE, WHICH MUST BE REMOVED AFTER PUNCTURE, WAS SO TIGHTLY ATTACHED THAT A STRONG PULL WAS NECESSARY. THERE WAS NO PERMANENT DAMAGE AND NO EXTRAVASATION, BUT THE REPEATED PUNCTURING OF THE CANNULA CAUSED PAIN FOR THE PATIENTS AND FEAR OF FURTHER PUNCTURING DUE TO THE TIGHT PULLING OF THE CLOSURE. NO INVASIVE INJURIES OCCURRED IN RELATION WITH PATIENTS.

    Patient problems: Anxiety; Pain

    MAUDE report 15066765 ↗
  • MalfunctionFeb 9, 2017Protective Measures Problem; Protective Measures Problem

    IT WAS REPORTED THAT A DELTEC® GRIPPER MICRO® BLUNT CANNULA SAFETY DEVICE FAILED WHILE A PATIENT PORT WAS BEING ACCESSED. THIS INCIDENT OCCURRED WITH A NURSE AND LEFT THE NEEDLE SLIGHTLY EXPOSED. NO PATIENT OR CLINICIAN INJURY WAS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6315686 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →