Deltec 21-3369-24Recalls & reported problems
Icu Medical, Inc. · Class 2
All models of Deltec (97)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10610586033025), received between Jan 13, 2023 and Jan 13, 2023 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Loose or Intermittent Connection” (5 reports).
2 other recalls cover devices cleared under the same FDA submission (K072657); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Loose or Intermittent Connection5
- Obstruction of Flow5
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 13, 2023Loose or Intermittent Connection; Obstruction of Flow; Loose or Intermittent Connection; Obstruction of Flow
IT WAS REPORTED THAT THE MICROCLAVE UNSCREWED FROM THE GRIPPER THROUGHOUT THE TREATMENT WITH AN INFUSOR. IT OCCURRED WITH 4 PATIENTS AND AT THE END OF THE TREATMENT THERE WAS A LOT OF BLEEDBACK OBSTRUCTING THE LINE. THERE WAS SPILLAGE OF DRUG AND BLOOD THROUGH THE CONNECTION OF THE MICROCLAVE AND GRIPPER. THERE WAS BLOOD LOSS, BLEEDBACK, CHEMO, AND BIOHAZARD. THE DEVICE WAS NOT REPROCESSED OR RE-STERILIZED. THERE IS A USED CONTAMINATED SAMPLE AND SISTER SAMPLES AVAILABLE FOR EVALUATION. FIVE PICTURES OF THE PRODUCT ISSUE WERE PROVIDED. THE EVENT OCCURRED DURING PATIENT USE; HOWEVER, THERE WAS NO NEED FOR MEDICAL INTERVENTION, NO DELAY IN THERAPY, NO ADVERSE EVENT AND NO ONE WAS HARMED AS A RESULT OF THIS EVENT. NO PATIENT INJURY WAS REPORTED. ADDITIONAL INFORMATION RECEIVED ON 20-DEC-2022: RECEIVED AN EMAIL FROM THE SALES REPRESENTATIVE ON 20-DEC-2022, STATING THAT THE MICROCLAVE WAS UNSCREWING FROM THE LUER OF THE GRIPPER WHILE THE INFUSER TREATMENT WAS PASSING. WHEN THE MICROCLAVE WAS LOOSENED FROM THE GRIPPER CONNECTION, BLOOD FLEW BACK AS THE GRIPPER WAS OPENED. THE BLOOD REFLUXED FROM THE PORT-A-CATH INTO THE ELASTOMER WHEN THE DRUG FINISHED PASSING. THE GRIPPER WAS PLACED FOR
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16157506 ↗MalfunctionJan 13, 2023Loose or Intermittent Connection; Obstruction of Flow; Loose or Intermittent Connection; Obstruction of Flow
IT WAS REPORTED THAT THE MICROCLAVE UNSCREWED FROM THE GRIPPER THROUGHOUT THE TREATMENT WITH AN INFUSOR. IT OCCURRED WITH 4 PATIENTS AND AT THE END OF THE TREATMENT THERE WAS A LOT OF BLEEDBACK OBSTRUCTING THE LINE. THERE WAS SPILLAGE OF DRUG AND BLOOD THROUGH THE CONNECTION OF THE MICROCLAVE AND GRIPPER. THERE WAS BLOOD LOSS, BLEEDBACK, CHEMO, AND BIOHAZARD. THE DEVICE WAS NOT REPROCESSED OR RE-STERILIZED. THERE IS A USED CONTAMINATED SAMPLE AND SISTER SAMPLES AVAILABLE FOR EVALUATION. FIVE PICTURES OF THE PRODUCT ISSUE WERE PROVIDED. THE EVENT OCCURRED DURING PATIENT USE; HOWEVER, THERE WAS NO NEED FOR MEDICAL INTERVENTION, NO DELAY IN THERAPY, NO ADVERSE EVENT AND NO ONE WAS HARMED AS A RESULT OF THIS EVENT. NO PATIENT INJURY WAS REPORTED. ADDITIONAL INFORMATION RECEIVED ON 20-DEC-2022: RECEIVED AN EMAIL FROM THE SALES REPRESENTATIVE ON 20-DEC-2022, STATING THAT THE MICROCLAVE WAS UNSCREWING FROM THE LUER OF THE GRIPPER WHILE THE INFUSER TREATMENT WAS PASSING. WHEN THE MICROCLAVE WAS LOOSENED FROM THE GRIPPER CONNECTION, BLOOD FLEW BACK AS THE GRIPPER WAS OPENED. THE BLOOD REFLUXED FROM THE PORT-A-CATH INTO THE ELASTOMER WHEN THE DRUG FINISHED PASSING. THE GRIPPER WAS PLACED FOR
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16157469 ↗MalfunctionJan 13, 2023Loose or Intermittent Connection; Obstruction of Flow; Loose or Intermittent Connection; Obstruction of Flow
IT WAS REPORTED THAT THE MICROCLAVE UNSCREWED FROM THE GRIPPER THROUGHOUT THE TREATMENT WITH AN INFUSOR. IT OCCURRED WITH 4 PATIENTS AND AT THE END OF THE TREATMENT THERE WAS A LOT OF BLEEDBACK OBSTRUCTING THE LINE. THERE WAS SPILLAGE OF DRUG AND BLOOD THROUGH THE CONNECTION OF THE MICROCLAVE AND GRIPPER. THERE WAS BLOOD LOSS, BLEEDBACK, CHEMO AND BIOHAZARD. THE DEVICE WAS NOT REPROCESSED OR RE-STERILIZED. THERE IS A USED CONTAMINATED SAMPLE AND SISTER SAMPLES AVAILABLE FOR EVALUATION. FIVE PICTURES OF THE PRODUCT ISSUE WERE PROVIDED. THE EVENT OCCURRED DURING PATIENT USE; HOWEVER, THERE WAS NO NEED FOR MEDICAL INTERVENTION, NO DELAY IN THERAPY, NO ADVERSE EVENT AND NO ONE WAS HARMED AS A RESULT OF THIS EVENT. NO PATIENT INJURY WAS REPORTED. ADDITIONAL INFORMATION RECEIVED ON 20-DEC-2022: RECEIVED AN EMAIL FROM THE SALES REPRESENTATIVE ON 20-DEC-2022, STATING THAT THE MICROCLAVE WAS UNSCREWING FROM THE LUER OF THE GRIPPER WHILE THE INFUSER TREATMENT WAS PASSING. WHEN THE MICROCLAVE WAS LOOSENED FROM THE GRIPPER CONNECTION, BLOOD FLEW BACK AS THE GRIPPER WAS OPENED. THE BLOOD REFLUXED FROM THE PORT-A-CATH INTO THE ELASTOMER WHEN THE DRUG FINISHED PASSING. THE GRIPPER WAS PLACED FOR
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16157445 ↗MalfunctionJan 13, 2023Loose or Intermittent Connection; Obstruction of Flow; Loose or Intermittent Connection; Obstruction of Flow
IT WAS REPORTED THAT THE MICROCLAVE UNSCREWED FROM THE GRIPPER THROUGHOUT THE TREATMENT WITH AN INFUSOR. IT OCCURRED WITH 4 PATIENTS AND AT THE END OF THE TREATMENT THERE WAS A LOT OF BLEEDBACK OBSTRUCTING THE LINE. THERE WAS SPILLAGE OF DRUG AND BLOOD THROUGH THE CONNECTION OF THE MICROCLAVE AND GRIPPER. THERE WAS BLOOD LOSS, BLEEDBACK, CHEMO, AND BIOHAZARD. THE DEVICE WAS NOT REPROCESSED OR RE-STERILIZED. THERE IS A USED CONTAMINATED SAMPLE AND SISTER SAMPLES AVAILABLE FOR EVALUATION. FIVE PICTURES OF THE PRODUCT ISSUE WERE PROVIDED. THE EVENT OCCURRED DURING PATIENT USE; HOWEVER, THERE WAS NO NEED FOR MEDICAL INTERVENTION, NO DELAY IN THERAPY, NO ADVERSE EVENT, AND NO ONE WAS HARMED AS A RESULT OF THIS EVENT. NO PATIENT INJURY WAS REPORTED. ADDITIONAL INFORMATION RECEIVED ON 20-DEC-2022: RECEIVED AN EMAIL FROM THE SALES REPRESENTATIVE ON 20-DEC-2022, STATING THAT THE MICROCLAVE WAS UNSCREWING FROM THE LUER OF THE GRIPPER WHILE THE INFUSER TREATMENT WAS PASSING. WHEN THE MICROCLAVE WAS LOOSENED FROM THE GRIPPER CONNECTION, BLOOD FLEW BACK AS THE GRIPPER WAS OPENED. THE BLOOD REFLUXED FROM THE PORT-A-CATH INTO THE ELASTOMER WHEN THE DRUG FINISHED PASSING. THE GRIPPER WAS PLACED FO
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16157380 ↗MalfunctionJan 13, 2023Loose or Intermittent Connection; Obstruction of Flow; Loose or Intermittent Connection; Obstruction of Flow
IT WAS REPORTED THAT THE MICROCLAVE UNSCREWED FROM THE GRIPPER THROUGHOUT THE TREATMENT WITH AN INFUSOR. IT OCCURRED WITH 4 PATIENTS AND AT THE END OF THE TREATMENT THERE WAS A LOT OF BLEEDBACK OBSTRUCTING THE LINE. THERE WAS SPILLAGE OF DRUG AND BLOOD THROUGH THE CONNECTION OF THE MICROCLAVE AND GRIPPER. THERE WAS BLOOD LOSS, BLEEDBACK, CHEMO AND BIOHAZARD. THE DEVICE WAS NOT REPROCESSED OR RE-STERILIZED. THERE IS A USED CONTAMINATED SAMPLE AND SISTER SAMPLES AVAILABLE FOR EVALUATION. FIVE PICTURES OF THE PRODUCT ISSUE WERE PROVIDED. THE EVENT OCCURRED DURING PATIENT USE; HOWEVER, THERE WAS NO NEED FOR MEDICAL INTERVENTION, NO DELAY IN THERAPY, NO ADVERSE EVENT AND NO ONE WAS HARMED AS A RESULT OF THIS EVENT. NO PATIENT INJURY WAS REPORTED. ADDITIONAL INFORMATION RECEIVED ON 20-DEC-2022: RECEIVED AN EMAIL FROM THE SALES REPRESENTATIVE ON 20-DEC-2022, STATING THAT THE MICROCLAVE WAS UNSCREWING FROM THE LUER OF THE GRIPPER WHILE THE INFUSER TREATMENT WAS PASSING. WHEN THE MICROCLAVE WAS LOOSENED FROM THE GRIPPER CONNECTION, BLOOD FLEW BACK AS THE GRIPPER WAS OPENED. THE BLOOD REFLUXED FROM THE PORT-A-CATH INTO THE ELASTOMER WHEN THE DRUG FINISHED PASSING. THE GRIPPER WAS PLACED FOR
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16157325 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K072657. One submission often covers many models, so check the product and lot details in each record.
- Z-2499-2020 · Jun 11, 2020 · TerminatedPORT-A-CATH II POWER P.A.C. Low Profile" system.
- Z-0190-2020 · Jul 8, 2019 · TerminatedGRIPPER Needles sold as part of PORT-A-CATH II SINGLE LUMEN SYSTEM. Labeled as the following item descriptions: 1. TRAY, PAC, VEN, TI, PU, 1.9MM, UL, 6FR/INTRO 1/EA; 2.
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗