CADD 21-7109-24Recalls & reported problems
Icu Medical, Inc. · Class 2
All models of CADD (123)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10610586034497), received between Jul 31, 2018 and Aug 8, 2022 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak6
- Connection Problem1
- Insufficient Flow or Under Infusion1
Patient problems
- No Consequences Or Impact To Patient4
- No Clinical Signs, Symptoms or Conditions3
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 8, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT DURING ADMINISTRATION OF CHEMOTHERAPY THE MEDICATION CAME OUT OF THE FILTER. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15183524 ↗MalfunctionAug 5, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT DURING ADMINISTRATION OF CHEMOTHERAPY THE MEDICATION CAME OUT OF THE FILTER. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15176722 ↗MalfunctionMar 18, 2021Connection Problem; Connection Problem
INFORMATION WAS RECEIVED INDICATING THAT THE CADD EXTENSION SET HAD BROKEN OFF AND FOUND SEPARATED IN BED. THERE WERE NO ADVERSE EVENTS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11516562 ↗MalfunctionMar 26, 2019Fluid/Blood Leak; Fluid/Blood Leak
INFORMATION WAS RECEIVED THAT DURING USE OF THIS SMITHS MEDICAL CADD EXTENSION SETS, FLUID LEAKED AT THE FILTER WAS NOTED. IT WAS REPORTED THAT THE DEVICE WAS IN USE WITH PATIENT FOR INFUSION OF VELETRI. NO REPORTED ADVERSE EFFECTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8453795 ↗MalfunctionMar 26, 2019Fluid/Blood Leak; Fluid/Blood Leak
INFORMATION WAS RECEIVED THAT DURING USE OF THIS SMITHS MEDICAL CADD EXTENSION SETS, FLUID LEAKED AT THE FILTER WAS NOTED. IT WAS REPORTED THAT THE PATIENT HAD TO GO TO THE "HOSPITAL TO RECEIVE ADDITIONAL MEDICATION WITH A CRISIS AS PER INSUFFICIENT TREATMENT DUE TO FLUID LEAKAGE". ACCORDING TO THE REPORT, THE DEVICE WAS IN USE WITH PATIENT FOR INFUSION OF VELETRI. NO ADDITIONAL ADVERSE EFFECTS WERE REPORTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8453790 ↗MalfunctionMar 26, 2019Fluid/Blood Leak; Fluid/Blood Leak
INFORMATION WAS RECEIVED THAT DURING USE OF THIS SMITHS MEDICAL CADD EXTENSION SETS, FLUID LEAKED AT THE FILTER WAS NOTED. IT WAS REPORTED THAT THE DEVICE WAS IN USE WITH PATIENT FOR INFUSION OF VELETRI. NO REPORTED ADVERSE EFFECTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8453788 ↗MalfunctionMar 26, 2019Fluid/Blood Leak; Fluid/Blood Leak
INFORMATION WAS RECEIVED THAT DURING USE OF THIS SMITHS MEDICAL CADD EXTENSION SETS, FLUID LEAKED AT THE FILTER WAS NOTED. IT WAS REPORTED THAT THE DEVICE WAS IN USE WITH PATIENT FOR INFUSION OF VELETRI. NO REPORTED ADVERSE EFFECTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8453782 ↗MalfunctionJul 31, 2018Insufficient Flow or Under Infusion; Insufficient Flow or Under Infusion
INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL CADD EXTENSION SET HAD OVERINFUSED. THE PROGRAMMED SETTINGS WERE 2ML/ HR, WITH A TOTAL VOLUME OF 90 ML OVER 46 HOURS. 10 ML WERE REPORTEDLY REMAINING. TUBING WAS NOTED TO HAVE DISCONNECTED FROM 100 ML CASSETTE. NO PATIENT INJURY RESULTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7736744 ↗
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.