BCI 9004051Recalls & reported problems
Icu Medical, Inc. · Class 2
Quick summary
This ICU Medical BCI carbon-dioxide analyzer measures carbon dioxide in breathing gases. The 24 reports mostly describe power and battery problems, repeated occlusion alarms, and erratic or missing readings, with no deaths and no injuries mentioned. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 24
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 10610586036873), received between Feb 9, 2017 and May 25, 2023 — classified as 24 malfunctions.
The most frequently reported specific device problem is “Device Alarm System” (5 reports).
1 other recall cover devices cleared under the same FDA submission (K970209); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Device Alarm System5
- Calibration Problem4
- Power Problem3
- Failure to Cycle2
- Incorrect, Inadequate or Imprecise Result or Readings2
- Overheating of Device2
- Unable to Obtain Readings2
- Battery Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions16
- Insufficient Information5
- No Consequences Or Impact To Patient2
- No Known Impact Or Consequence To Patient2
- No Patient Involvement1
Recent adverse event reports
The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 25, 2023Power Problem; Unexpected Shutdown; Power Problem; Unexpected Shutdown
IT WAS REPORTED THAT THE DEVICE WONT STAY ON. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17004216 ↗MalfunctionSep 29, 2022Device Alarm System; Defective Component; Device Alarm System; Defective Component
ADDITIONAL INFORMATION WAS RECEIVED VIA EMAIL ON 02-OCT-2022. EVENT DATE WAS UPDATED. NO PATIENT WAS INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15505129 ↗MalfunctionSep 22, 2022Power Problem; Power Problem
IT WAS REPORTED THAT THE POWER LIGHTS WERE FLASHING. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15467077 ↗MalfunctionSep 14, 2022Erratic or Intermittent Display; Erratic or Intermittent Display
IT WAS REPORTED THAT THE DEVICE HAD ERRATIC READINGS. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15422572 ↗MalfunctionSep 2, 2022Device Alarm System; Battery Problem; Device Alarm System; Battery Problem
IT WAS REPORTED THAT THE DEVICE WAS SHOWING CONSTANT OCCLUSION ERROR AND BATTERY DIES FAST. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15349735 ↗MalfunctionSep 1, 2022Device Alarm System; Mechanical Problem; Device Alarm System; Mechanical Problem
IT WAS REPORTED THAT THERE WAS AN OCCLUSION ISSUE, PARTS HAD BEEN REPLACED BUT STILL HAVING THIS ISSUE MONTHS LATER. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15334282 ↗MalfunctionSep 1, 2022Device Alarm System; Incorrect, Inadequate or Imprecise Result or Readings; Failure to Read Input Signal; Device Alarm System; Incorrect, Inadequate or Imprecise Result or Readings; Failure to Read Input Signal
IT WAS REPORTED THAT THE DEVICE COULD NOT COLLECT DATA AND THE OCCLUSION MESSAGE COMES UP. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15334164 ↗MalfunctionAug 31, 2022Break; Unable to Obtain Readings; Break; Unable to Obtain Readings
IT WAS REPORTED THAT THE DEVICE WAS NOT READING CORRECTLY AND CABLE WAS CUT OFF. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15328812 ↗MalfunctionAug 31, 2022Data Problem; Data Problem
IT WAS REPORTED THAT THERE WAS AN INACCURATE DATA ISSUE WITH THE REPORTED DEVICE. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15326710 ↗MalfunctionAug 31, 2022Inaccurate Information; Inaccurate Information
IT WAS REPORTED THAT THE DEVICE WAS READING INCORRECTLY. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15326498 ↗MalfunctionApr 5, 2022Defective Alarm; Unintended System Motion; Defective Alarm; Unintended System Motion
IT WAS REPORTED THAT THE FAN CONSTANTLY RUNS AND ALARMS. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14015617 ↗MalfunctionMar 30, 2022Calibration Problem; Calibration Problem
IT WAS REPORTED THE DEVICE WOLD CALIBRATE CORRECTLY WITH 10% CO2, READING IN THE TWENTIES WHEN HOOKED UP TO A PATIENT. CONFIRMED THE EVENT WAS NOT DUE TO THE PATIENT, BY USING A SECONDARY (WORKING) DEVICE. NOT ABLE TO DO A FACTORY CALIBRATION, WOULD LIKE TO SEND IT IN. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13950161 ↗MalfunctionFeb 24, 2022Unable to Obtain Readings; Calibration Problem; Unable to Obtain Readings; Calibration Problem
IT WAS REPORTED THAT THERE IS AN UNSTABLE CO2 READING AND CALIBRATION. NO ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13595016 ↗MalfunctionNov 13, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THE COMPORT AND SIGNALS ARE INCORRECT ON A SMITHS MEDICAL CAPNOGRAPH. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 12808294 ↗MalfunctionNov 7, 2021Failure to Cycle; Failure to Cycle
IT WAS REPORTED THAT NO WAVE AND NOT RESPONDING TO BREATHS, GIVING INTERMITTENT ISSUES. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 12763904 ↗MalfunctionNov 6, 2021Failure to Cycle; Failure to Cycle
IT WAS REPORTED THAT A SMITHS MEDICAL CAPNOGRAPH HAS NO WAVE AND NOT RESPONDING TO BREATHS (LOUD WHEN TUNED ON). NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12755557 ↗MalfunctionOct 30, 2021Device Alarm System; Device Alarm System
WHILE CALIBRATING IT GAVE AN OCCLUSION ERROR.
Patient problems: Insufficient Information
MAUDE report 12725431 ↗MalfunctionOct 15, 2021Material Deformation; Material Deformation
IT WAS REPORTED THE CASING WAS BROKEN AND THE SCREWS WEREN'T FITTING. NO ADVERSE AFFECTS.
Patient problems: Insufficient Information
MAUDE report 12643475 ↗MalfunctionOct 15, 2021No Display/Image; No Display/Image
IT WAS REPORTED THAT WHILE IN USE WITH A PATIENT THE DEVICE STARTING HAVING A BLANK SCREEN. NO ADVERSE AFFECTS TO PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12643382 ↗MalfunctionAug 18, 2021Power Problem; Power Problem
INFORMATION RECEIVED A SMITHS MEDICAL PATIENT MONITORING/BCI CAPNOGRAPHY MONITOR CAPNOCHECK SLEEP - 9004 IS MAKING NOISE AND SHUTTING OFF ON ITS OWN. NO FURTHER INFORMATION.
Patient problems: Insufficient Information
MAUDE report 12337788 ↗MalfunctionApr 26, 2019Calibration Problem; Calibration Problem
INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL BCI CAPNOCHECK SLEEP OXIMETER "WILL NOT HOLD 1VOLT CALIBRATION". NO ADVERSE EFFECTS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8557474 ↗MalfunctionJan 10, 2019Low Readings; Calibration Problem; Low Readings; Calibration Problem
INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL BCI CAPNOCHECK SLEEP OXIMETER DISPLAYED LOW CO2 READINGS AFTER CALIBRATION. THE REPORTER NOTED TWO DIFFERENT GAS SOURCES WERE USED. NO ADVERSE EFFECTS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8237128 ↗MalfunctionFeb 24, 2018Overheating of Device; Electronic Property Issue; Overheating of Device; Electronic Property Issue
IT WAS REPORTED THAT WHEN PLUGGING INTO THE BACK OF A BCI® CAPNOCHECK® SLEEP CAPNOGRAPH/OXIMETER, ELECTRICAL CURRENTS WERE SEEN COMING OUT OF IT. REPORTER STATED THE MACHINE WOULD NOT TURN ON AND IT WAS HOT WHILE RUNNING. NO INJURY WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7294507 ↗MalfunctionFeb 9, 2017Overheating of Device; Smoking; Overheating of Device; Smoking
IT WAS REPORTED THAT A BCI® CAPNOCHECK® SLEEP CAPNOGRAPH/OXIMETER WAS HOT AND SMOKING. THE DEVICE'S GREEN POWER LIGHT DID NOT FUNCTION. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement
MAUDE report 6315691 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K970209. One submission often covers many models, so check the product and lot details in each record.
- Z-1014-2013 · Feb 6, 2013 · TerminatedSmiths Medical, BCI Remote Alarm Cable (Normally Open & Normally Closed), Reorder Numbers: WW3398NO & WW3398NC. For use with respiration carbon dioxide measurement equipment. Pr
Other ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE devices
- Alaris 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall2 deaths8,444 reports
- Alaris 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall607 reports
- BD Alaris™ ETCO2 Module, Model 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE485 reports
- BD Alaris™ ETCO2 Module, Model 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE471 reports
- NA GF-210RNihon Kohden Corporation · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE386 reports
- Trilogy IT2110X21BRespironics, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE5 recalls30 reports
- Trilogy Trilogy Evo Ventilator, GermanyRespironics, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE5 recalls26 reports
- Trilogy Evo Trilogy Evo Ventilator with oxygen blending model, GermanyRespironics, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE5 recalls9 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.