Medfusion 3500-0600-00Recalls & reported problems

Icu Medical, Inc. · Class 2

All models of Medfusion (16)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10610586041259), received between Oct 2, 2017 and Apr 13, 2023 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Device Alarm System” (3 reports).

2 FDA recall records name this device's identifier, 2 still open.

16 other recalls cover devices cleared under the same FDA submission (K111386); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Dec 19, 2023Class IIOpen, Classified

    Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.

    Z-1540-2024 · Root cause: Process control · Recall details · FDA record ↗

  2. Aug 16, 2023Class IIOpen, Classified

    A force sensor in the occlusion detector may drift out of calibration leading to increased occlusion detection times, false occlusion alarms, or a System Failure Alarm. If the force sensor calibration shift is large enough, the pump will display a System Failure Alarm (including Force Sensor BGND Test, Force Sensor Bridge Test, or Force Sensor Test). However, if the calibration shift is not large enough to trigger a System Failure Alarm, the threshold to detect an occlusion may increase, increasing the time to occlusion detection, or the threshold may decrease, leading to false occlusion alarms. Although shifts in the force sensor calibration may occur over time with any device, an increased potential for such shifts has been reported in devices produced before April 2022 due to mechanical interference between parts of the plunger head assembly. Out of an abundance of caution, we are notifying all customers of this potential issue.

    Z-0078-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Alarm System3
  • Insufficient Device Problem Information2
  • Mechanical Problem2
  • Device Displays Incorrect Message1
  • Failure to Sense1
  • Patient Device Interaction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Insufficient Information2
  • No Information2
  • No Patient Involvement2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 13, 2023Device Alarm System; Device Alarm System

    IT WAS REPORTED THE PUMP ALARMED PRIMARY AUDIBLE ALARM BACKGROUND TEST ERROR FOLLOWED BY AN ACU WATCHDOG FAILURE. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16733390 ↗
  • MalfunctionApr 12, 2023Device Alarm System; Device Alarm System

    IT WAS REPORTED THE PUMP EXHIBITED A PRIMARY AUDIBLE ALARM. NO PATIENT INJURY REPORTED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16728359 ↗
  • MalfunctionApr 22, 2022Device Alarm System; Device Alarm System

    INFORMATION WAS RECEIVED INDICATING THAT THE PUMP FAILED THE OCCLUSION TEST DUE TO "SYSTEM FAILURE: PRIMARY AUDIBLE ALARM BACKGROUND TEST." IT IS UNKNOWN IF THERE WAS ANY PATIENT INVOLVEMENT OR ANY ADVERSE EFFECTS.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 14190200 ↗
  • MalfunctionDec 4, 2021Mechanical Problem; Failure to Sense; Mechanical Problem; Failure to Sense

    INFORMATION WAS RECEIVED INDICATING THAT THIS SMITHS MEDICAL MEDFUSION 3500 PUMP EXHIBITED ISSUE WITH THE PLUNGER SENSOR, BARREL CLAMP SENSOR, AND SHAFT MAKING NOISES WHEN MOVING AROUND. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12930916 ↗
  • MalfunctionNov 2, 2020Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED THAT THIS SMITHS MEDICAL CADD MEDFUSION 3500 PUMP EXHIBITED MOTOR NOT WORKING. NO REPORTED ADVERSE EFFECTS.

    Patient problems: No Information

    MAUDE report 10776239 ↗
  • MalfunctionAug 17, 2020Patient Device Interaction Problem; Patient Device Interaction Problem

    INFORMATION WAS RECEIVED INDICATING THAT A SMITHS MEDICAL PUMP'S BUTTONS WERE NOT RESPONDING. THERE WAS NO REPORTED ADVERSE EVENTS.

    Patient problems: No Information

    MAUDE report 10417970 ↗
  • MalfunctionDec 6, 2017Device Displays Incorrect Message; Device Displays Incorrect Message

    INFORMATION WAS RECEIVED INDICATING THAT A MOTOR RATE ERROR OCCURRED ON A SMITHS MEDICAL MEDFUSION 3500 SYRINGE PUMP. THERE WERE NO REPORTED ADVERSE EFFECTS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7086773 ↗
  • MalfunctionOct 2, 2017Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THERE WAS A CLUTCH PLUNGER LEVER ISSUE OCCURRED ON A MEDFUSION® 3500 SYRINGE PUMP. THERE WAS NO REPORTED PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 6906961 ↗
  • MalfunctionOct 2, 2017Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THERE WAS A CLUTCH PLUNGER LEVER ISSUE OCCURRED ON A MEDFUSION® 3500 SYRINGE PUMP. THERE WAS NO REPORTED PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 6906945 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K111386. One submission often covers many models, so check the product and lot details in each record.

  • Z-0864-2024 · Dec 19, 2023 · Open, Classified
    smiths medical Medfusion Model 4000 syringe pump
  • Z-1543-2024 · Dec 19, 2023 · Open, Classified
    Smiths Medical Medfusion Guide Barrel Clamp, Item number G6000716
  • Z-1541-2024 · Dec 19, 2023 · Open, Classified
    Smiths Medical Medfusion Model 4000 Syringe pump, Item Numbers: a) 4000-0101-249, b) 4000-0101-50, c) 4000-0101-51, d) 4000-0105-249, e) 4000-0105-50, f) 4000-0105-51, g) 4
  • Z-0079-2024 · Aug 16, 2023 · Open, Classified
    Medfusion Syringe Pump, Model 4000-XXXX-XX: a) 0100-50, b) 0100-249, c) 0101-50, d) 0101-51, e) 0101-78, f) 0101-249, g) 0105-51, h) 0105-78, i) 0105-249, j) 0106-00, k)
  • Z-0272-2023 · Oct 10, 2022 · Open, Classified
    Medfusion Model 4000 Syringe Infusion Pump, with software versions V1.6.0 or V1.6.1 that have implemented the Smart Pump Programming (SPP) feature in a clinical treatment environme
  • Z-1271-2022 · Apr 18, 2022 · Open, Classified
    Medfusion Syringe Pump Model: Model 4000-0100-50, Model 4000-0101-249, Model 4000-0101-50, Model 4000-0101-51, Model 4000-0101-78, Model 4000-0105-249, Model 4000-0105-51, Model 4
  • Z-2735-2020 · Jun 26, 2020 · Open, Classified
    Medfusion Syringe Pump Model 3500, Software Version V6.0. Infusion pump.
  • Z-2734-2020 · Jun 26, 2020 · Open, Classified
    Medfusion Syringe Pump Model 4000; Software Versions V1.5.0, V1.5.1, V1.6.0, and V1.6.1
  • Z-0610-2020 · Oct 28, 2019 · Terminated
    Medfusion Syringe Pump, Model 4000. The Model 4000 pump is a small, lightweight and portable syringe infusion pump. The pump offers a variety of delivery modes programmable to me
  • Z-2554-2019 · Oct 15, 2018 · Terminated
    Medfusion Model 4000 Syringe Infusion Pump, with PharmGuard Server Software (PGS). Sold under the following names: MEDFUSION 4000 Pump V1.0; MEDFUSION 4000 Pump V1.1, v1.1.1,
  • Z-2662-2018 · Apr 13, 2018 · Terminated
    Medfusion Syringe Infusion Pump, containing the Barrel Clamp Guide mechanism
  • Z-1134-2018 · Nov 13, 2017 · Terminated
    Medfusion Syringe Pump 4000 Series The Medfusion Syringe Infusion Pumps are indicated for the following uses: 1) Administration of fluids requiring precisely controlled infusion r
  • Z-0313-2018 · Aug 22, 2017 · Terminated
    Medfusion, Syringe Infusion Pump, Software Version V1.1, Rx Only, Manufactured by Smiths Medical ASD, Inc., 1500 Eureka Park, Lower Pemberton, Ashford, Kent, TN25 4BF, UK. In the
  • Z-0653-2022 · Dec 3, 2015 · Open, Classified
    Medfusion Model 3500 and 4000 Syringe Infusion Pumps and Medfusion 3010 and 3010A Syringe Infusion Pumps serviced during the same time period in which a main board was replaced usi
  • Z-0540-2014 · Nov 22, 2013 · Terminated
    Medfusion 4000 Syringe Infusion Pumps Medfusion 4000 Syringe Infusion Pumps are designed for use in care areas for the administration of fluids requiring precisely controlled infu

…and 1 more.

Other ACCESSORIES, PUMP, INFUSION devices

All ACCESSORIES, PUMP, INFUSION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page