MEDFUSION 4000-0106-00Recalls & reported problems
Icu Medical, Inc. · Class 2
All models of Medfusion (16)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 15019517070750), received between Nov 7, 2020 and Jun 17, 2026 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Device Alarm System” (3 reports).
2 FDA recall records name this device's identifier, 2 still open.
14 other recalls cover devices cleared under the same FDA submission (K111386); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Dec 19, 2023Class IIOpen, Classified
Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.
Z-1541-2024 · Root cause: Process control · Recall details · FDA record ↗
- Aug 16, 2023Class IIOpen, Classified
A force sensor in the occlusion detector may drift out of calibration leading to increased occlusion detection times, false occlusion alarms, or a System Failure Alarm. If the force sensor calibration shift is large enough, the pump will display a System Failure Alarm (including Force Sensor BGND Test, Force Sensor Bridge Test, or Force Sensor Test). However, if the calibration shift is not large enough to trigger a System Failure Alarm, the threshold to detect an occlusion may increase, increasing the time to occlusion detection, or the threshold may decrease, leading to false occlusion alarms. Although shifts in the force sensor calibration may occur over time with any device, an increased potential for such shifts has been reported in devices produced before April 2022 due to mechanical interference between parts of the plunger head assembly. Out of an abundance of caution, we are notifying all customers of this potential issue.
Z-0079-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Alarm System3
- Failure to Infuse3
- Circuit Failure1
- Defective Component1
- Device Sensing Problem1
- Failure to Sense1
- Output Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 17, 2026Output Problem
IT WAS REPORTED THAT PUMP HAD MOTOR RATE ERROR, THERE WERE NO PATIENT INVOLVEMENT AND NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25534479 ↗MalfunctionMar 30, 2026Failure to Infuse
IT WAS REPORTED THAT THE PUMP HAD FLASH MEMORY ERROR, THERE WAS NO PATIENT INVOLVEMENT AND NO PATIENT HARM. INVESTIGATION FINDINGS INCLUDED A TRAVEL REVERSE ERROR, WHICH IS A REPORTABLE MALFUNCTION THAT COULD RESULT IN INTERRUPTION OF THERAPY IF RECURRING DURING DEVICE USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24734561 ↗MalfunctionMar 26, 2026Failure to Infuse
IT WAS REPORTED THAT THE PUMP EXHIBITED A PRIMARILY AUDIBLE ALARM. DURING PHYSICAL INSPECTION, IT WAS FOUND THAT THERE WAS A PRIMARILY AUDIBLE ALARM BGND TEST IN THE EVENT HISTORY LOG. THERE WAS NO PATIENT INVOLVEMENT, NO INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24705202 ↗MalfunctionMar 12, 2026Failure to Sense
IT WAS REPORTED THAT THE FORCE SENSOR REQUIRED REPLACEMENT. THERE WAS NO PATIENT INVOLVEMENT AND NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24579854 ↗MalfunctionNov 3, 2025Device Sensing Problem; Device Sensing Problem
IT WAS REPORTED THAT A SYRINGE NOT IN PLACE ERROR MESSAGE WAS DISPLAYED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23452253 ↗MalfunctionJun 26, 2025Defective Component; Failure to Infuse; Defective Component; Failure to Infuse
IT WAS REPORTED THAT THE PUMP EXHIBITED PRIMARY AUDIBLE ALARM. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22338651 ↗MalfunctionAug 25, 2022Device Alarm System; Device Alarm System
IT WAS REPORTED THAT THE PUMP DISPLAYED 'MOTOR DRIVE ALARM" MULTIPLE TIMES. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15296183 ↗MalfunctionFeb 10, 2022Circuit Failure; Circuit Failure
IT WAS REPORTED THAT THE DEVICE HAD INTERCONNECT BOARD DAMAGES. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13510046 ↗MalfunctionMar 24, 2021Device Alarm System; Device Alarm System
INFORMATION WAS RECEIVED INDICATING THAT A SMITHS MEDICAL MEDFUSION PUMP HAD A PRIMARY AUDIBLE ALARM ERROR CODE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11563936 ↗MalfunctionNov 7, 2020Device Alarm System; Device Alarm System
IT WAS REPORTED THAT THE DEVICE DISPLAYED A TRAVEL COURSE ERROR. NO PATIENT INJURY OR COMPLICATIONS WERE REPORTED IN RELATION TO THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10806252 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K111386. One submission often covers many models, so check the product and lot details in each record.
- Z-0864-2024 · Dec 19, 2023 · Open, Classifiedsmiths medical Medfusion Model 4000 syringe pump
- Z-1543-2024 · Dec 19, 2023 · Open, ClassifiedSmiths Medical Medfusion Guide Barrel Clamp, Item number G6000716
- Z-0272-2023 · Oct 10, 2022 · Open, ClassifiedMedfusion Model 4000 Syringe Infusion Pump, with software versions V1.6.0 or V1.6.1 that have implemented the Smart Pump Programming (SPP) feature in a clinical treatment environme
- Z-1271-2022 · Apr 18, 2022 · Open, ClassifiedMedfusion Syringe Pump Model: Model 4000-0100-50, Model 4000-0101-249, Model 4000-0101-50, Model 4000-0101-51, Model 4000-0101-78, Model 4000-0105-249, Model 4000-0105-51, Model 4
- Z-2734-2020 · Jun 26, 2020 · Open, ClassifiedMedfusion Syringe Pump Model 4000; Software Versions V1.5.0, V1.5.1, V1.6.0, and V1.6.1
- Z-2735-2020 · Jun 26, 2020 · Open, ClassifiedMedfusion Syringe Pump Model 3500, Software Version V6.0. Infusion pump.
- Z-0610-2020 · Oct 28, 2019 · TerminatedMedfusion Syringe Pump, Model 4000. The Model 4000 pump is a small, lightweight and portable syringe infusion pump. The pump offers a variety of delivery modes programmable to me
- Z-2554-2019 · Oct 15, 2018 · TerminatedMedfusion Model 4000 Syringe Infusion Pump, with PharmGuard Server Software (PGS). Sold under the following names: MEDFUSION 4000 Pump V1.0; MEDFUSION 4000 Pump V1.1, v1.1.1,
- Z-2662-2018 · Apr 13, 2018 · TerminatedMedfusion Syringe Infusion Pump, containing the Barrel Clamp Guide mechanism
- Z-1134-2018 · Nov 13, 2017 · TerminatedMedfusion Syringe Pump 4000 Series The Medfusion Syringe Infusion Pumps are indicated for the following uses: 1) Administration of fluids requiring precisely controlled infusion r
- Z-0313-2018 · Aug 22, 2017 · TerminatedMedfusion, Syringe Infusion Pump, Software Version V1.1, Rx Only, Manufactured by Smiths Medical ASD, Inc., 1500 Eureka Park, Lower Pemberton, Ashford, Kent, TN25 4BF, UK. In the
- Z-0653-2022 · Dec 3, 2015 · Open, ClassifiedMedfusion Model 3500 and 4000 Syringe Infusion Pumps and Medfusion 3010 and 3010A Syringe Infusion Pumps serviced during the same time period in which a main board was replaced usi
- Z-0540-2014 · Nov 22, 2013 · TerminatedMedfusion 4000 Syringe Infusion Pumps Medfusion 4000 Syringe Infusion Pumps are designed for use in care areas for the administration of fluids requiring precisely controlled infu
- Z-2107-2012 · Feb 14, 2012 · TerminatedSmiths Medical, Medfusion Syringe Infusion Pump, Model 4000, Software Version V1.0 & V1.1, Reorder Numbers 4000-0100-50, 4000-0100-95, 4000-0101-50, 4000-0101-51, and 4000-0101-78
Other ACCESSORIES, PUMP, INFUSION devices
- Medfusion 4000-0106-01Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION2 recalls4,390 reports
- Medfusion 4000-0105-51Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION2 recalls2,882 reports
- CADD 21-2111-0401-51Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION4 recalls461 reports
- CADD 21-2112-0402-78Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION3 recalls0 reports
- Medfusion 3500-0600-00Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION2 recalls9 reports
- Medfusion 3500-0600-51Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION2 recalls2 reports
- Medfusion 4000-0105-78Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION2 recalls0 reports
- Sapphire 1636202ICU Medical, Inc. · ACCESSORIES, PUMP, INFUSION1 recall3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page