MEGADYNE MESE1Recalls & reported problems
Megadyne Medical Products, Inc. · Megadyne Smoke Evacuator · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Megadyne MESE1 surgical smoke evacuator removes smoke from operating rooms during electrosurgery. Its 65 reports mostly describe the unit switching on but not activating when the surgeon touches the pencil, plus units running continuously and quietly on their own, along with loud clicking and premature filter wear. Two injuries are reported, including burns through gloves. No deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 64
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 64 adverse event reports that list this exact device (DI 10614559104545), received between Oct 1, 2020 and Jun 23, 2025 — classified as 2 injuries, 62 malfunctions.
The most frequently reported specific device problem is “Difficult or Delayed Activation” (59 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult or Delayed Activation59
- Ventilation Problem in Device Environment9
- Noise, Audible7
- Adverse Event Without Identified Device or Use Problem6
- Appropriate Device Problem Term/Code Not Available4
- Arcing at Paddles1
- Self-Activation or Keying1
Patient problems
- No Clinical Signs, Symptoms or Conditions58
- Partial thickness (Second Degree) Burn4
- Burn(s)1
- No Consequences Or Impact To Patient1
- Superficial (First Degree) Burn1
- Unspecified Tissue Injury1
Recent adverse event reports
The 25 most recent of 64 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 23, 2025Difficult or Delayed Activation; Difficult or Delayed Activation
IT WAS REPORTED THAT SINCE (B)(6) 2025 THE DEVICE CONTINUOUSLY RUNS WHILE IN OPEN MODE, WHEN CONNECTED VIA DIRECT CONNECT TO FT10 GENERATOR. TO LESSON THE SOUND THEY PUT THE UNIT IN LAP MODE. TO KEEP THE UNIT FROM PULLING AIR ALL THE TIME THEY LEAVE THE RED CAP ON THE FLUID TRAP TO SAVE THE LIFE OF THEIR FILTER. THERE WERE NO PATIENT ADVERSE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22306514 ↗MalfunctionMar 5, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A CADAVER LAB THE BOVIE WERE AROUND 30 CUT 30 COAG. THE SURGEON GOT BURNED THROUGH THEIR GLOVES.
Patient problems: Partial thickness (Second Degree) Burn
MAUDE report 21528882 ↗MalfunctionMar 5, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A CADAVER LAB THE BOVIE WERE AROUND 30 CUT 30 COAG. THE SURGEON GOT BURNED THROUGH THEIR GLOVES.
Patient problems: Partial thickness (Second Degree) Burn
MAUDE report 21528701 ↗MalfunctionMar 5, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A CADAVER LAB THE BOVIE WERE AROUND 30 CUT 30 COAG. THE SURGEON GOT BURNED THROUGH THEIR GLOVES.
Patient problems: Partial thickness (Second Degree) Burn
MAUDE report 21528635 ↗MalfunctionJan 22, 2025Difficult or Delayed Activation; Appropriate Device Problem Term/Code Not Available; Noise, Audible; Difficult or Delayed Activation; Appropriate Device Problem Term/Code Not Available; Noise, Audible
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE A NEW MESE1 EVACUATOR UNIT WAS MAKING LOUD NOISES AND RANDOM CLICKING SOUNDS. ALSO PULLING AIR WHEN PENCIL WAS NOT BEING ACTIVATED IN OPEN MODE USING THE CONNECT CABLE ATTACHED TO THE FT10 MEDTRONIC ESU. UNIT USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21207323 ↗MalfunctionJan 21, 2025Difficult or Delayed Activation; Difficult or Delayed Activation
IT WAS REPORTED THAT THE SMOKE EVACUATOR IS ON AND NO ONE IS IN THE ROOM. NOTHING IS ATTACHED TO THE SMOKE EVACUATOR AND AS IT IS SO QUIET THE UNIT CONTINUES TO RUN AND NO ONE NOTICES. THIS IS CAUSING FILTER LIFE TO END PREMATURELY. EVERY TIME THEY UNPLUG FROM THE WALL OUR UNIT IS A FLIP SWITCH AND THEY ARE USE TO A PUSH BUTTON. IF THE SMOKE EVACUATOR IS TURNED ON FIRST AND THEN TURN ON THE BOVIE, THEN IT JUST RUNS. SOME OF THE NURSES DON'T HEAR IT RUNNING, AND THEY COULD RUN A UNIT FOR FOUR HOURS NOT KNOWING IT'S ON. ON JUST THAT BOVIE IT HAS TO GET A BASE LINE WHEN THE CABLE IS ON. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21196677 ↗MalfunctionJan 21, 2025Difficult or Delayed Activation; Difficult or Delayed Activation
IT WAS REPORTED THAT THE SMOKE EVACUATOR IS ON AND NO ONE IS IN THE ROOM. NOTHING IS ATTACHED TO THE SMOKE EVACUATOR AND AS IT IS SO QUIET THE UNIT CONTINUES TO RUN AND NO ONE NOTICES. THIS IS CAUSING FILTER LIFE TO END PREMATURELY. EVERY TIME THEY UNPLUG FROM THE WALL OUR UNIT IS A FLIP SWITCH AND THEY ARE USE TO A PUSH BUTTON. IF THE SMOKE EVACUATOR IS TURNED ON FIRST AND THEN TURN ON THE BOVIE, THEN IT JUST RUNS. SOME OF THE NURSES DON'T HEAR IT RUNNING, AND THEY COULD RUN A UNIT FOR FOUR HOURS NOT KNOWING IT'S ON. ON JUST THAT BOVIE IT HAS TO GET A BASE LINE WHEN THE CABLE IS ON. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21196676 ↗MalfunctionJan 21, 2025Difficult or Delayed Activation; Difficult or Delayed Activation
IT WAS REPORTED THAT THE SMOKE EVACUATOR IS ON AND NO ONE IS IN THE ROOM. NOTHING IS ATTACHED TO THE SMOKE EVACUATOR AND AS IT IS SO QUIET THE UNIT CONTINUES TO RUN AND NO ONE NOTICES. THIS IS CAUSING FILTER LIFE TO END PREMATURELY. EVERY TIME THEY UNPLUG FROM THE WALL OUR UNIT IS A FLIP SWITCH AND THEY ARE USE TO A PUSH BUTTON. IF THE SMOKE EVACUATOR IS TURNED ON FIRST AND THEN TURN ON THE BOVIE, THEN IT JUST RUNS. SOME OF THE NURSES DON'T HEAR IT RUNNING, AND THEY COULD RUN A UNIT FOR FOUR HOURS NOT KNOWING IT'S ON. ON JUST THAT BOVIE IT HAS TO GET A BASE LINE WHEN THE CABLE IS ON. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21196671 ↗MalfunctionJan 21, 2025Difficult or Delayed Activation; Difficult or Delayed Activation
IT WAS REPORTED THAT THE SMOKE EVACUATOR IS ON AND NO ONE IS IN THE ROOM. NOTHING IS ATTACHED TO THE SMOKE EVACUATOR AND AS IT IS SO QUIET THE UNIT CONTINUES TO RUN AND NO ONE NOTICES. THIS IS CAUSING FILTER LIFE TO END PREMATURELY. EVERY TIME THEY UNPLUG FROM THE WALL OUR UNIT IS A FLIP SWITCH AND THEY ARE USE TO A PUSH BUTTON. IF THE SMOKE EVACUATOR IS TURNED ON FIRST AND THEN TURN ON THE BOVIE, THEN IT JUST RUNS. SOME OF THE NURSES DON'T HEAR IT RUNNING, AND THEY COULD RUN A UNIT FOR FOUR HOURS NOT KNOWING IT'S ON. ON JUST THAT BOVIE IT HAS TO GET A BASE LINE WHEN THE CABLE IS ON. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21196667 ↗MalfunctionJan 21, 2025Difficult or Delayed Activation; Difficult or Delayed Activation
IT WAS REPORTED THAT THE SMOKE EVACUATOR IS ON AND NO ONE IS IN THE ROOM. NOTHING IS ATTACHED TO THE SMOKE EVACUATOR AND AS IT IS SO QUIET THE UNIT CONTINUES TO RUN AND NO ONE NOTICES. THIS IS CAUSING FILTER LIFE TO END PREMATURELY. EVERY TIME THEY UNPLUG FROM THE WALL OUR UNIT IS A FLIP SWITCH AND THEY ARE USE TO A PUSH BUTTON. IF THE SMOKE EVACUATOR IS TURNED ON FIRST AND THEN TURN ON THE BOVIE, THEN IT JUST RUNS. SOME OF THE NURSES DON'T HEAR IT RUNNING, AND THEY COULD RUN A UNIT FOR FOUR HOURS NOT KNOWING IT'S ON. ON JUST THAT BOVIE IT HAS TO GET A BASE LINE WHEN THE CABLE IS ON. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21196562 ↗MalfunctionDec 9, 2024Difficult or Delayed Activation; Difficult or Delayed Activation
IT WAS REPORTED THAT DURING A CANCER SURGERY, WHEN YOU START USING THE EVACUATOR, SOMETIMES IT STAYS ON CONTINUOUSLY, AND IN CUT-OFF MODE THE EVACUATION DOES NOT WORK. THE SURGERY ENDED WITH THE SAME PRODUCT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20877532 ↗MalfunctionOct 22, 2024Difficult or Delayed Activation; Noise, Audible; Difficult or Delayed Activation; Noise, Audible
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE A RATTLING SOUND LIKE A METALLIC SOUND MAY SOUND WHEN FLUSHING. WHEN USED IN CONJUNCTION WITH OTHER COMPANIES' ELECTROSURGERY UNITS, SMOKE MAY REMAIN FLUE ALL THE TIME. TURNING THE POWER ON AND OFF DOES NOT FIX IT. IT IS THE SAME FOR CONNECTING CABLES, 2403J AND RF SENSORS AS OTHER ELECTROSURGICAL EQUIPMENT. IT REMAINS IN SMOKE EXHAUST ALL THE TIME, ESPECIALLY WHEN IT IS OPEN. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20504197 ↗MalfunctionSep 10, 2024Difficult or Delayed Activation; Difficult or Delayed Activation
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THAT AT THE 90 DAY EVALUATION THE RF SENSOR SELF ACTIVATES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20179603 ↗MalfunctionMay 22, 2024Difficult or Delayed Activation; Noise, Audible; Difficult or Delayed Activation; Noise, Audible
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE DEVICE IS VERY NOISY WHEN IN USE AND SUCKING AIR IN EVEN WHEN NOT ATTACHED TO DIATHERMY MACHINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19370535 ↗MalfunctionApr 4, 2024Difficult or Delayed Activation; Difficult or Delayed Activation
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE SMOKE FLUE DOES NOT STOP OR THE SMOKE FLUE SWITCH DOES NOT TURN ON. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19042820 ↗MalfunctionJan 11, 2024Difficult or Delayed Activation; Noise, Audible; Difficult or Delayed Activation; Noise, Audible
IT WAS REPORTED THAT DURING THE COMPREHENSIVE INSPECTION WE CONNECTED IT TO MEGA POWER1000 AND CONFIRMED THAT IT DID NOT STOP AFTER IT WAS CONNECTED. WE HAVE CONFIRMED THAT THE PUMP IS MAKING AN ABNORMAL NOISE. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT. THE CUSTOMER DOES NOT WANT TO PROVIDE ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18492131 ↗MalfunctionOct 5, 2023Difficult or Delayed Activation; Difficult or Delayed Activation
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE MESE1 WAS MALFUNCTION. WHEN I POWER ON THE UNIT WITHOUT RF SENSOR AND TRY TO TEST THE FUNCTIONAL BY PRESSED THE MANUAL BUTTON MESE1 WAS FUNCTION NORMALLY. BUT IF I POWER ON THE UNIT WITH RF SENSOR TOGETHER WITH MEGEN1 AND TELESCOPING SMOKE EVACUATION SOFT TISSUE DISSECTOR; I TRY TO PRESSED THE DISSECTOR BY HANDSWITCH MESE1 DOESN?T FUNCTION. IF I POWER ON THE UNIT AND CONNECT RF SENSOR AT THE BACK OF THE MESE1 THE MESE1 WILL FUNCTION AUTOMATICALLY WITHOUT PRESSING THE DISSECTOR. SERIAL NUMBER : (B)(6). THERE WERE NO PATIENT STATUS/ OUTCOME / CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17875860 ↗MalfunctionAug 14, 2023Difficult or Delayed Activation; Noise, Audible; Difficult or Delayed Activation; Noise, Audible
IT WAS REPORTED THAT THE UNIT IS EXCESSIVELY LOUD AND WILL NOT POWER OFF DURING ACTIVATION OF BOVIE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17542685 ↗MalfunctionAug 14, 2023Difficult or Delayed Activation; Ventilation Problem in Device Environment; Appropriate Device Problem Term/Code Not Available; Difficult or Delayed Activation; Ventilation Problem in Device Environment; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING A TESTING THE SHORT PLUG WAS NOT RECOGNIZED. THE MESE1 IS ACTING AS DESIGNED AND THE DEVICE DOES NOT NEED TO BE RETURNED FOR SERVICE. WE SET THE GENERATORS CURRENT SENSOR SENSITIVITY UP AND THAT WORKED. THE ACTIVATION SEEMED NORMAL. BUT THEN IT SEEMED A BIT SLOW WITH THE CUT BUTTON. THEY TURNED IT UP A NOTCH, IT KEPT ON GOING. THEY TURNED IT DOWN, IT KEPT ON GOING. THEY PUT IT BACK TO SETTING 3, AS WE STARTED OUT, AND IT WORKED LIKE A CHARM. THEN, THE MACHINE STARTED TO TURN ON RANDOMLY. SO THE PENCIL WAY PLUGGED IN, BUT LAYING DOWN ON THE TABLE, NO ACTIVATIONS WERE DONE, BUT DESPITE, THE MESE1 WAS ACTIVATED MULTIPLE TIMES.¿ ¿THEY ALSO HAD A PLUG/WORK AROUND INSTEAD OF A MEGASOFT/RETURN ELEKTRO.¿
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17542111 ↗MalfunctionFeb 14, 2023Difficult or Delayed Activation; Noise, Audible; Difficult or Delayed Activation; Noise, Audible
IT WAS REPORTED THE DURING AN UNKNOWN PROCEDURE THERE ARE STRANGE "TIK TAK" SOUNDS HEARD ESPECIALLY IN LAP MODE (LV1-5). THE SMOKE EVACUATOR WOULD ACTIVATE AUTOMATICALLY EVEN THOUGH THE ENERGY CAPITAL/ DIATHERMY HAS NOT ACTIVATED AND THE SMOKE EVACUATOR WOULD NOT STOP AFTER RUN TIME HAS REACHED/ MANUAL STOP BUTTON IS PRESSED. IT IS NOISY WHEN OPERATING IN OPEN MODE (LV3-5).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16371862 ↗MalfunctionFeb 2, 2023Difficult or Delayed Activation; Adverse Event Without Identified Device or Use Problem; Ventilation Problem in Device Environment; Difficult or Delayed Activation; Adverse Event Without Identified Device or Use Problem; Ventilation Problem in Device Environment
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE FLOW RATE IS OUT OF SPEC. THE UNIT WAS TOO SENSITIVE SO WHEN THE AUTOACTIVATION FUNCTION CAME ON WHEN IT IS NOT SUPPOSED TO, AND DEFLATED THE PNEUMOPERITONEUM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16283528 ↗InjuryFeb 2, 2023Arcing at Paddles; Adverse Event Without Identified Device or Use Problem; Arcing at Paddles; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE RIGHT AFTER USE WHEN THE DIATHERMY TIP WAS NOT ACTIVATED, HCP WAS PUTTING THE PENCIL DOWN WHEN IT BRUSHED AGAINST HER ASSISTANT¿S GOWN AND BURNT THROUGH HIS GOWN AND HE SUFFERED MINOR BURNS ON HIS FOREARM. THE ASSISTANT IS NOW FINE, MEDICAL TREATMENT WAS SOUGHT IMMEDIATELY. HCP ALSO MENTIONED WITNESSING SPARKS AT THE JUNCTURE WHERE THE DIATHERMY TOUCHES THE FORCEPS DURING USE. THE SETTINGS OF THE ESU ARE AS FOLLOWS: CUT 20 AND COAG 35.
Patient problems: Partial thickness (Second Degree) Burn; Superficial (First Degree) Burn
MAUDE report 16283070 ↗MalfunctionJan 6, 2023Difficult or Delayed Activation; Difficult or Delayed Activation
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE DEVICE WAS TRIGGERING CONSTANTLY. THE CHANGED RF AND CONNECT CABLES BUT THE SAME ISSUES CONTINUED THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16112302 ↗MalfunctionNov 29, 2022Difficult or Delayed Activation; Noise, Audible; Difficult or Delayed Activation; Noise, Audible
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE POWER SWITCH OF THE MEGADYNE SMOKE EVACUATOR WAS OUT OF ORDER. THE SMOKE EVACUATOR REMAINS ON/ ACTIVATED AFTER USING THE POWER SWITCH TO TURN OFF THE MACHINE. NURSE ALSO COMPLAINED THAT THE OPERATIONAL NOISE LEVEL FOR THIS MACHINE IS HIGHER THAN WHEN THEY FIRST USED IT AND THE NOISE LEVEL DISTRACTED DOCTOR¿S FOCUS ON THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15877420 ↗MalfunctionNov 1, 2022Difficult or Delayed Activation; Difficult or Delayed Activation
IT WAS REPORTED THAT DURING A GASTRIC BYPASS THE DEMONSTRATION OF NEW EQUIPMENT (MESE1), IT WAS FOUNDED THAT THE SENSING OF ENERGY PRODUCT OPERATION WAS NOT PERFORMED IN OPEN AND LAP MODES, SO THE DEMONSTRATION WAS STOPPED. WE CHANGED AND CONNECTED THE RF SENSOR, BUT IT DIDN'T WORK. WHEN WE MANUALLY OPERATED MESE1 WITHOUT CONNECTING THE RF SENSOR, THE EQUIPMENT WORKED, BUT WHEN WE CONNECTED THE RF SENSOR, IT STARTED WORKING AUTOMATICALLY. IT IS OUT OF CONTROL, WE CAN NOT STOP AND START. THE EVENT OCCURRED DURING STERILE PROCESSING. THE EVENT OCCURRED DURING FIELD INSPECTION. THE EVENT OCCURRED DURING INTERNAL SERVICE ACTIVITIES SUCH AS CALIBRATION. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15707590 ↗
FDA clearance and classification
Cleared through 510(k) K200250 on Jun 26, 2020 (decision: Substantially Equivalent). Submitted by Megadyne Medical Products, Inc. as “Megadyne Smoke Evacuator, Megadyne Filter, Megadyne Fluid Trap, Megadyne Connect Cable, 1 m and 2.1 m, Megadyne RF Sensor”. The FDA review took 144 days. FDA record ↗
The FDA regulates product code FYD (apparatus, exhaust, surgical) under 21 CFR 878.5070 as a Class II device. 21 CFR 878.5070 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the MEGADYNE MESE1
Has the MEGADYNE MESE1 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10614559104545), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the MEGADYNE MESE1?
The FDA's MAUDE database lists 64 adverse event reports for this device, received between Oct 1, 2020 and Jun 23, 2025. In those reports, the most common specific device problems are “Difficult or Delayed Activation” (59), “Ventilation Problem in Device Environment” (9) and “Noise, Audible” (7); the most common patient problems are “Partial thickness (Second Degree) Burn” (4), “Burn(s)” (1) and “Superficial (First Degree) Burn” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the MEGADYNE MESE1?
None of the reports are classified as deaths, 2 as injuries and 62 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the MEGADYNE MESE1 FDA approved?
It was cleared through 510(k) K200250 on Jun 26, 2020. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the MEGADYNE MESE1, and is it still sold?
The FDA's device database lists Megadyne Medical Products, Inc. as the company responsible for it (brand MEGADYNE, model MESE1). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other APPARATUS, EXHAUST, SURGICAL devices
- Neptune 0703001000Stryker Corporation · APPARATUS, EXHAUST, SURGICAL2 deaths185 reports
- CONMED BUFFALO FILTER 52560Conmed Corporation · APPARATUS, EXHAUST, SURGICAL1 recall0 reports
- RapidVac SE3690Covidien LP · APPARATUS, EXHAUST, SURGICAL15 reports
- RapidVac SE3695Covidien LP · APPARATUS, EXHAUST, SURGICAL10 reports
- MEGADYNE 2255JMegadyne Medical Products, Inc. · APPARATUS, EXHAUST, SURGICAL4 reports
- ZIP Pen 252510Megadyne Medical Products, Inc. · APPARATUS, EXHAUST, SURGICAL4 reports
- NEBULAE SRS Laparoscopic Surgical Pump 7-700-00Northgate Technologies Inc. · Apparatus, Exhaust, Surgical1 death1 report
- Bovie SE02-220Aspen Surgical Products, Inc. · Apparatus, Exhaust, Surgical1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page