ZIP Pen 252510Recalls & reported problems

Megadyne Medical Products, Inc. · ZIP Pen Ergonomic Smoke Evacuation Pencil, E-Z Clean, Non-Stick, PTFE Coated, 10 Ft (3.04m) Tubing · Class 2

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10614559105337), received between Dec 14, 2019 and Dec 31, 2024 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Arcing of Electrodes” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Arcing of Electrodes2
  • Flaked1
  • Human-Device Interface Problem1
  • Material Separation1
  • Thermal Decomposition of Device1

Patient problems

  • Burn(s)1
  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 31, 2024Arcing of Electrodes; Human-Device Interface Problem; Arcing of Electrodes; Human-Device Interface Problem

    IT WAS REPORTED THAT THEY WERE REMOVING A LESION/CYST FROM THE PATIENTS NECK AT A AMBULATORY SURGERY SITE. WHEN THE UNIT WAS BEING ACTIVATED THE STERILE DRAPES CAUGHT FIRE AND THE FIRE SPREAD TO THE PATIENT. THEY PUT THE FIRE OUT AND PATIENT WAS TAKEN TO LOCAL ER BY AMBULANCE, PATIENT SUSTAINED BURNS. THAT IS ALL THE INFORMATION KNOWN TO THE BIOMED AT THIS TIME.

    Patient problems: Burn(s)

    MAUDE report 21047099 ↗
  • MalfunctionOct 5, 2023Arcing of Electrodes; Arcing of Electrodes

    IT WAS REPORTED THAT DURING A BIOPSY THAT A FIRE BALL CAME OUT OF PENCIL. THEY SWITCHED OUT TIP TO MEDTRONIC. NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17875907 ↗
  • InjuryMar 14, 2022Flaked; Material Separation; Flaked; Material Separation

    IT WAS REPORTED THAT DURING A INGUINAL HERNIA REPAIR, THE EZ-CLEAN COATING PARTIALLY DETACHED FROM THE ELECTRODE. IT WAS DURING THE TEST WITH SMOKE EVACUATOR PENCIL. PART OF THE PROCEDURE WAS TUNNELED, IT CAN NOT BE EXCLUDED THAT THE POSSIBILITY THAT THE PHYSICIAN TOUCHED THE ELECTRODE WITH A METAL INSTRUMENT. THE ELECTRODE WAS ONLY CLEANED WITH A DAMP CLOTH (NO ABRASIVE SPONGE OR KNIFE). THE CUSTOMER IS FAMILIAR WITH EZ-CLEAN ELECTRODES. HE HAS BEEN USING THEM FOR SEVERAL YEARS (VERSION 0012M). HE HAS NEVER EXPERIENCED THIS BEFORE. DISADVANTAGES TO THE PATIENT INCLUDED SURGICAL TIME INCREASED BY 10 MINUTES. THE COATING PARTICLES COULDN¿T ALL BE REMOVED AND ARE THEREFORE STILL IN THE INGUINAL CANAL OF THE PATIENT. THE ELECTRODE WAS IMMEDIATELY REMOVED FROM THE SURGICAL FIELD PICTURES WERE TAKEN OF THE PARTICLES JUST AFTER REMOVAL AND IN THE PATIENT. A NEW SMOKE EVACUATOR PEN WITH THE SAME REFERENCE AND LOT NUMBER WAS USED TO FINISH THE PROCEDURE. THE EZ-CLEAN ELECTRODE OF THIS PEN SHOWED NO PROBLEM.

    Patient problems: Foreign Body In Patient

    MAUDE report 13757690 ↗
  • MalfunctionDec 14, 2019Thermal Decomposition of Device

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE BOVIE PENCIL WAS LAID ON THE DRAPE AND IT BURNED A HOLE IN IT. THIS WAS NEAR THE PATIENT'S FACE. THE TIP WAS CHANGED, AND THE PROCEDURE WAS CONTINUED. THE BOVIE WAS LAID ON THE DRAPE BELOW THE PATIENT¿S FEET AND IT BURNED A HOLE IN IT AS WELL. THE PATIENT WAS NOT BURNED, AND THE CUT WAS SET ON 35 AND THE COAG WAS ALSO SET ON 35.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9470635 ↗

Other APPARATUS, EXHAUST, SURGICAL devices

All APPARATUS, EXHAUST, SURGICAL devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.