Rapicide PA Test Strips ML02-0118Recalls & reported problems

Medivators Inc. · Rapicide PA Test Strips · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10677964064071), received between May 9, 2018 and Jun 26, 2018 — classified as 2 injuries.

The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Improper or Incorrect Procedure or Method2
  • Shelf Life Exceeded2

Patient problems

  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 26, 2018Shelf Life Exceeded; Improper or Incorrect Procedure or Method; Shelf Life Exceeded; Improper or Incorrect Procedure or Method

    A FACILITY REPORTED USING EXPIRED RAPICIDE PA HIGH LEVEL DISINFECTANT TEST STRIPS FOR CHECKING THE MINIMUM REQUIRED CONCENTRATION (MRC) OF RAPICIDE PA HLD USED IN THEIR AUTOMATED ENDOSCOPE REPROCESSOR. HLD USED WITH THEIR AER WAS NOT PROPERLY VERIFIED FOR MRC, THUS THERE IS POTENTIAL THAT ENDOSCOPES WERE NOT ADEQUATELY HIGH-LEVEL DISINFECTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7639885 ↗
  • InjuryMay 9, 2018Shelf Life Exceeded; Improper or Incorrect Procedure or Method; Shelf Life Exceeded; Improper or Incorrect Procedure or Method

    MEDIVATORS FIELD SERVICE ENGINEER REPORTED, WHILE ON SITE SERVICING THE FACILITY'S ADVANTAGE PLUS AUTOMATED ENDOSCOPE REPROCESSOR, THE FACILITY HAD BEEN USING EXPIRED TEST STRIPS TO CHECK THE MINIMUM REQUIRED CONCENTRATION (MRC) OF RAPICIDE PA HIGH LEVEL DISINFECTANT USED IN THE AER. THERE IS POTENTIAL THAT HLD USED WITH THEIR AER WAS NOT PROPERLY VERIFIED FOR MRC, THUS THERE IS POTENTIAL THAT ENDOSCOPES WERE NOT ADEQUATELY HIGH-LEVEL DISINFECTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7498577 ↗

Other Indicator, Physical/Chemical Sterilization Process devices