Level 1 L-370Recalls & reported problems

Icu Medical, Inc. · Class 2

Quick summary

This ICU Medical Level 1 fluid warmer warms fluids for infusion. Reports mostly describe leaks at connections and warming fluid crossing into the patient side of the tubing, plus air bubbles found during testing. There are 24 reports that mention this identifier, with no deaths and 2 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
24
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 10695085002819), received between Sep 7, 2017 and Feb 3, 2026 — classified as 2 injuries, 22 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (7 reports).

11 other recalls cover devices cleared under the same FDA submission (K911383); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak7
  • Air/Gas in Device4
  • Mechanical Problem4
  • Connection Problem3
  • Deformation Due to Compressive Stress3
  • Leak/Splash2
  • Difficult to Insert1
  • Improper Flow or Infusion1

Patient problems

  • No Clinical Signs, Symptoms or Conditions20
  • Insufficient Information2
  • No Consequences Or Impact To Patient2
  • No Patient Involvement2
  • No Information1

Recent adverse event reports

The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 3, 2026Fluid/Blood Leak

    IT WAS REPORTED THAT THE HOTLINE 3 RECIRCULATING FLUID WENT INTO THE PATIENT SIDE OF THE TUBING. THE DEVICE WAS BROKEN OFF AND WENT TO THE PATIENT. THERE WAS NO EXTERNAL LEAKING. THERE WAS PATIENT INVOLVEMENT, AND THERE WAS NO REPORTED HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24250309 ↗
  • MalfunctionDec 31, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS STATED THAT DEVICE HOTLINE LEAKED. THERE WAS NO REPORTED PATIENT INVOLVEMENT AND NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23936647 ↗
  • InjuryAug 25, 2025Improper Flow or Infusion; Improper Flow or Infusion

    IT WAS REPORTED THAT DURING SURGERY, A CAROTID ENDARTERECTOMY WAS CANCELLED SECONDARY TO A DISPOSABLE HOTLINE TUBING DEFECT THAT LED TO POSSIBLE DELIVERY OF WARMING WATER THROUGH THE IV TUBING INSTEAD OF BLOOD INTO THE PATIENT. PER REPORTER, THE WARMING WATER USED IN THE HOTLINE BLOOD WARMING DEVICE WAS NOT STERILE AND THIS BEING INJECTED INTO THE PATIENT'S BLOOD STREAM DUE TO THE TUBING DEFECT, WHICH IS LIFE THREATENING. REPORTER STATED THE TUBING HAD A DEFECT WHERE-IN RATHER THAN DELIVERING THE WARMED BLOOD, IT WAS DELIVERING THE FLUID/WATER BEING USED TO WARM THE BLOOD. THE SURGICAL CASE WAS ABORTED IMMEDIATELY.

    Patient problems: Insufficient Information

    MAUDE report 22889122 ↗
  • MalfunctionDec 27, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT WHEN THE PRODUCT WAS BEING USED, FLUID WAS LEAKING. STATUS OF THE PRODUCT AT THE TIME OF THE EVENT IS UNKNOWN. THERE WAS UNKNOWN PATIENT INVOLVEMENT, BUT NO PATIENT HARM/ADVERSE EVENT WAS BEING REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21028757 ↗
  • MalfunctionOct 18, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE HOTLINE FLUID TUBING WAS LEAKING AT THE CONNECTION POINT BETWEEN THE TWO LOCKING TABS AND THE SOCKET. TO ENSURE IT WAS NOT A PROBLEM WITH THE BLOOD WARMING MACHINE, THE NURSES TRIED TWO ADDITIONAL MACHINES, BUT THE LEAK PERSISTED. HOWEVER, WHEN THEY SWITCHED TO A NEW TUBING SET, THE ISSUE WAS RESOLVED, INDICATING THAT THE ORIGINAL SET WAS DEFECTIVE. THERE WAS NO PATIENT INVOLVEMENT, AND NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20481706 ↗
  • MalfunctionOct 31, 2023Air/Gas in Device; Air/Gas in Device

    IT WAS REPORTED THAT DURING A BENCH TEST, THERE WERE AIR BUBBLES NOTED IN THE TUBING. THERE WAS NO PATIENT INJURY. DEVICE PULLED FROM USE. EVENT REPORTED AS "STILL ON GOING".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18039381 ↗
  • MalfunctionOct 11, 2023Air/Gas in Device; Air/Gas in Device

    IT WAS REPORTED THAT DURING A BENCH TEST, AIR BUBBLES WERE NOTED IN HOTLINE TUBING. NO INJURY OR ADVERSE EFFECTS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17912039 ↗
  • MalfunctionOct 11, 2023Air/Gas in Device; Air/Gas in Device

    IT WAS REPORTED THAT DURING A BENCH TEST, AIR BUBBLES WERE NOTED IN HOTLINE TUBING. NO INJURY OR ADVERSE EFFECTS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17911850 ↗
  • MalfunctionOct 10, 2023Air/Gas in Device; Air/Gas in Device

    IT WAS REPORTED THAT DURING A BENCH TEST, AIR BUBBLES WERE NOTED IN HOTLINE TUBING. NO INJURY OR ADVERSE EFFECTS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17905715 ↗
  • MalfunctionNov 18, 2022Fluid/Blood Leak; Connection Problem; Fluid/Blood Leak; Connection Problem

    ADDITIONAL INFORMATION RECEIVED ON 09-DEC-2022 VIA EMAIL AND ATTACHED TO COMPLAINT OBJECT: THE CUSTOMER REPORTED THAT WE NEED TO DISREGARD THIS ISSUE. THEY FOUND FROM THE TECHNICAL DEPARTMENT THAT THEY NEED TO MAKE SURE TO PUSH IN THE LINE ALL THE WAY AND THIS WAS VERY DIFFICULT TO DO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15829304 ↗
  • MalfunctionNov 18, 2022Difficult to Insert; Leak/Splash; Connection Problem; Difficult to Insert; Leak/Splash; Connection Problem

    IT WAS REPORTED THAT THE CUSTOMER WAS HAVING ISSUES WITH THE CONNECTION NOT BEING COMPLETELY CONNECTED. THE UNIT WAS LEAKING FLUID DUE TO THE NEW TUBING. THIS HAS BEEN HAPPENING OFTEN WITH THE NEW SYSTEM. THERE HAS BEEN NO REPORT OF PATIENT INVOLVEMENT OR NO OBSERVABLE CLINICAL SYMPTOMS OR A CHANGE IN SYMPTOMS IDENTIFIED IN THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15828826 ↗
  • MalfunctionNov 18, 2022Fluid/Blood Leak; Connection Problem; Fluid/Blood Leak; Connection Problem

    IT WAS REPORTED THAT THE CUSTOMER WAS HAVING ISSUES WITH THE CONNECTION NOT BEING COMPLETELY CONNECTED. THE UNIT WAS LEAKING FLUID DUE TO THE NEW TUBING. THIS HAS BEEN HAPPENING OFTEN WITH THE NEW SYSTEM. THERE HAS BEEN NO REPORT OF PATIENT INVOLVEMENT OR NO OBSERVABLE CLINICAL SYMPTOMS OR A CHANGE IN SYMPTOMS IDENTIFIED IN THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15828656 ↗
  • MalfunctionOct 31, 2022Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THE DEVICE ROUTINELY EXPERIENCED THE CONNECTIVE TUBING (@ TOP OF THE L-370 SETS) KINKING OFF, WHEN ATTACHED TO EITHER AN IV ADMIN. SET OR BLOOD SET, IN MULTIPLE AREAS. MULTIPLE OCCASIONS, SINCE BEGAN USING HL-390 AND L-370 OVER LAST FEW MONTHS. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15699677 ↗
  • MalfunctionOct 31, 2022Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THE DEVICE ROUTINELY EXPERIENCED THE CONNECTIVE TUBING (@ TOP OF THE L-370 SETS) KINKING OFF, WHEN ATTACHED TO EITHER AN IV ADMIN. SET OR BLOOD SET, IN MULTIPLE AREAS. MULTIPLE OCCASIONS, SINCE BEGAN USING HL-390 AND L-370 OVER LAST FEW MONTHS. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15698599 ↗
  • MalfunctionOct 28, 2022Deformation Due to Compressive Stress; Deformation Due to Compressive Stress

    IT WAS REPORTED THAT THE DEVICE ROUTINELY EXPERIENCED THE CONNECTIVE TUBING (AT THE TOP OF THE SETS) KINKING OFF, WHEN ATTACHED TO EITHER AN INTRAVENOUS THERAPY (IV) ADMINISTRATION SET OR BLOOD SET, IN MULTIPLE AREAS. THIS ISSUE HAS OCCURRED ON MULTIPLE OCCASIONS, SINCE BEGINNING USE OF THE SETS OVER LAST FEW MONTHS. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15692896 ↗
  • MalfunctionOct 28, 2022Deformation Due to Compressive Stress; Deformation Due to Compressive Stress

    IT WAS REPORTED THAT THE DEVICE ROUTINELY EXPERIENCED THE CONNECTIVE TUBING (AT THE TOP OF THE SETS) KINKING OFF, WHEN ATTACHED TO EITHER AN INTRAVENOUS THERAPY (IV) ADMINISTRATION SET OR BLOOD SET, IN MULTIPLE AREAS. THIS ISSUE HAS OCCURRED ON MULTIPLE OCCASIONS, SINCE BEGINNING USE OF THE SETS OVER LAST FEW MONTHS. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15692124 ↗
  • MalfunctionOct 21, 2022Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THE DEVICE ROUTINELY EXPERIENCED THE CONNECTIVE TUBING (@ TOP OF THE L-370 SETS) KINKING OFF, WHEN ATTACHED TO EITHER AN IV ADMIN. SET OR BLOOD SET, IN MULTIPLE AREAS. MULTIPLE OCCASIONS, SINCE BEGAN USING HL-390 AND L-370 OVER LAST FEW MONTHS. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15649924 ↗
  • MalfunctionOct 21, 2022Inability to Irrigate; Inability to Irrigate

    IT WAS REPORTED THAT THE PHYSICIAN HAD TO SQUEEZE THE BULB ON THE BLOOD TUBING TO MAKE THE BLOOD INFUSE FASTER. THE CONNECTION LINE BECAME CRIMPED MAKING IT DIFFICULT TO INFUSE A BOLUS OF BLOOD TO THE PATIENT. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15649919 ↗
  • MalfunctionOct 20, 2022Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THE DEVICE ROUTINELY EXPERIENCED THE CONNECTIVE TUBING (@ TOP OF THE L-370 SETS) KINKING OFF, WHEN ATTACHED TO EITHER AN IV ADMIN. SET OR BLOOD SET, IN MULTIPLE AREAS. MULTIPLE OCCASIONS, SINCE BEGAN USING HL-390 AND L-370 OVER LAST FEW MONTHS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15638368 ↗
  • MalfunctionFeb 11, 2022Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT DURING A DURING A CRANIECTOMY,VENTRICULOSTOMY CASE THE CRNA NOTICED THAT THE STERILE WATER IN THE FLUID WARMER CONTAINER LOOKED BLOOD TINGED, THE RESULT OF SOME BACKED UP BLOOD IN THE IV TUBING. THE FLUID WARMER DEVICE AND TUBING WERE DISCONNECTED FROM THE PATIENT AND EXAMINED. IT APPEARED THAT THE FLUID WARMER TUBING WAS DEFECTIVE AND ALLOWED MIXTURE OF THE STERILE WATER FROM THE WARMER WITH THE PATIENT'S IV FLUID. THIS WAS CONFIRMED BY CONNECTING A NEW SET TO THE FLUID WARMER AND NOT ENCOUNTERING THE SAME ISSUE. NO PATIENT HARM OR INJURY WAS REPORTED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 13518700 ↗
  • MalfunctionApr 26, 2021Fluid/Blood Leak; Fluid/Blood Leak

    INFORMATION WAS RECEIVED REGARDING A LEVEL 1 HOTLINE DISPOSABLE. IT WAS REPORTED THAT THE INNER TUBE OF THE DEVICE WAS LEAKING, THE WARMER ITSELF DID NOT MALFUNCTION. HOWEVER, DURING THE CUSTOMER'S EVALUATION, THEY FOUND THAT THE WATER TANK WAS CRACKED. THE PATIENT WAS NO HAR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11730086 ↗
  • MalfunctionNov 26, 2020Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT A SMITHS MEDICAL HOTLINE TUBING OBSERVED WITH KINKS. IT WAS NOTED THESE ARE KINKED WITHIN THE PACKAGING. AT LEAST ONE INSTANCE WAS WHILE IN USE WITH A PATIENT. THE FLUID WOULD NOT RUN BECAUSE OF IT.

    Patient problems: No Consequences Or Impact To Patient; No Patient Involvement

    MAUDE report 10912736 ↗
  • MalfunctionNov 26, 2020Deformation Due to Compressive Stress; Deformation Due to Compressive Stress

    ADDITIONAL EVENT INFORMATION: REPORTER STATED THE TUBING WAS BLOCKED AND FLUID WOULD NOT RUN. DEVICE WAS NOT USED WITH PATIENT.

    Patient problems: No Consequences Or Impact To Patient; No Patient Involvement

    MAUDE report 10912735 ↗
  • InjurySep 7, 2017Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT DURING USE, THE INNER TUBING OF A LEVEL 1® HOTLINE® 3 BLOOD AND FLUID WARMER WAS DEFECTIVE AND LITTLE TO NO FLUID FLOWED THROUGH THE INNER TUBING. THE EVENT OUTCOME WAS REPORTED AS "OTHER SERIOUS (IMPORTANT MEDICAL EVENT)", BUT NO INFORMATION ON THE MEDICAL EVENT WAS PROVIDED. NO PERMANENT INJURY WAS REPORTED.

    Patient problems: No Information

    MAUDE report 6848250 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K911383. One submission often covers many models, so check the product and lot details in each record.

  • Z-2759-2026 · Jun 24, 2026 · Open, Classified
    Hotline Fluid Warming Set, single use device, is designed to warm blood and intravenous fluids and deliver them to the patient's intravenous access site at normothermic temperature
  • Z-1443-2014 · Mar 17, 2014 · Terminated
    Smiths Medical HOTLINE Fluid Warming Set L-70. Product Usage: intended for the delivery of warmed blood and intravenous fluids at normothermic temperatures under gravity conditio
  • Z-1444-2014 · Mar 17, 2014 · Terminated
    Smiths Medical HOTLINE Fluid Warming Set L-80. Product Usage: intended for the delivery of warmed blood and intravenous fluids at normothermic temperatures under gravity condit
  • Z-1979-2012 · Jun 27, 2012 · Terminated
    Smiths Medical HOTLINE 3 Blood and Fluid Warmer, Reorder Numbers: HL-390 Product Usage: The Micro-switch is used on the HOTLINE 3 Blood and Fluid Warmer. This product, for use in
  • Z-0428-2012 · Nov 21, 2011 · Terminated
    Level 1 Normothermic I.V. Fluid Administration Sets, Product Code DI-60HL , Product Usage: Single patient use sets designed for use by trained medical professionals, in conjuncti
  • Z-0426-2012 · Nov 21, 2011 · Terminated
    HOTLlNE Disposable Administration Sets, Product Code L-80 Product Usage: Single patient use Sets designed for use by trained medical professionals, in conjunction with HOTLlNE R
  • Z-0425-2012 · Nov 21, 2011 · Terminated
    HOTLlNE Disposable Administration Sets, Product Code L-370 Product Usage: Single patient use Sets designed for use by trained medical professionals, in conjunction with HOTLlNE
  • Z-0424-2012 · Nov 21, 2011 · Terminated
    HOTLlNE Disposable Administration Sets, Product Code L-270 Product Usage: Single patient use Sets designed for use by trained medical professionals, in conjunction with HOTLlNE
  • Z-0422-2012 · Nov 21, 2011 · Terminated
    HOTLlNE Disposable Administration Sets, Product Codes L-70 Product Usage: Single patient use Sets designed for use by trained medical professionals, in conjunction with HOTLlNE R
  • Z-0423-2012 · Nov 21, 2011 · Terminated
    HOTLlNE Disposable Administration Sets, Product Codes L-70NI Product Usage: Single patient use Sets designed for use by trained medical professionals, in conjunction with HOTLlNE
  • Z-0427-2012 · Nov 21, 2011 · Terminated
    Level 1 Normothermic I.V. Fluid Administration Sets, Product Code D-60HL , Product Usage: Single patient use sets designed for use by trained medical professionals, in conjunctio

Other Warmer, thermal, infusion fluid devices

All Warmer, thermal, infusion fluid devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.