Level 1 HL-390-38Recalls & reported problems
Icu Medical, Inc. · Class 2
All models of Level 1 (52)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10695085002826), received between Jul 24, 2024 and May 26, 2026 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (2 reports).
11 other recalls cover devices cleared under the same FDA submission (K911383); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Electrical /Electronic Property Problem2
- Overheating of Device2
- Temperature Problem2
- Circuit Failure1
- Device Sensing Problem1
- Display or Visual Feedback Problem1
- Fluid/Blood Leak1
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 26, 2026Device Sensing Problem
THE DEVICE WAS RETURNED AS IT WAS REPORTED THAT THE UNIT WOULD NOT RECOGNIZE THE CASSETTE AND IT WAS FOUND OUT DURING SELF-TEST. THE COMPLETED INVESTIGATION CONFIRMED THE REPORTED ISSUE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25279378 ↗MalfunctionJan 26, 2026Electrical /Electronic Property Problem
IT WAS REPORTED THAT THE DEVICE UNIT DID NOT ALARM DURING THE ALARM TEST AND OVER-TEMPERATURE TEST. THE UNIT DID NOT WARM UP PROPERLY. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24177672 ↗MalfunctionAug 13, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED THAT WHEN THE DEVICE WAS CONNECTED, IT BLEW A FUSE AND IT IS OLD. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22795039 ↗MalfunctionJul 28, 2025Overheating of Device; Overheating of Device
THE DEVICE HAD OVERTEMPT. THE EVENT OCCURRED DURING SELF TESTING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22637141 ↗MalfunctionJun 25, 2025Overheating of Device; Overheating of Device
IT WAS STATED THAT DEVICE DOES NOT PUMP-THEN OVERHEATS DURING PATIENT USE. THERE WAS PATIENT INVOLVEMENT, WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22332178 ↗MalfunctionApr 28, 2025Pumping Problem; Temperature Problem; Pumping Problem; Temperature Problem
IT WAS REPORTED THAT THE DEVICE EXHIBITED A HIGH TEMPERATURE. THERE WAS NO INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21921470 ↗MalfunctionMar 25, 2025Circuit Failure; Circuit Failure
IT WAS REPORTED THAT THE DEVICE FAILED THE PREVENTATIVE MAINTENANCE (PM) TEST. OVER TEMPERATURE TEST FAILED AND SCREEN FLICKERS. STATUS OF THE PRODUCT AT TIME OF EVENT WAS DURING ROUTINE PREVENTIVE MAINTENANCE INSPECTION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21691501 ↗MalfunctionSep 19, 2024Display or Visual Feedback Problem; Incorrect, Inadequate or Imprecise Result or Readings; Temperature Problem; Display or Visual Feedback Problem; Incorrect, Inadequate or Imprecise Result or Readings; Temperature Problem
IT WAS REPORTED THAT THE DISPLAY WAS ERRATIC. WHEN THE LID WAS OFF, IT WENT BLANK AND GLITCHED. IT HAD ERRATIC TEMPERATURE READINGS. THERE WAS NO PATIENT INVOLVEMENT AND NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20258683 ↗MalfunctionJul 24, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE DEVICE WAS LEAKING WATER NEAR THE CONNECTION WHERE YOU PUT THE TUBING SET. THE EVENT OCCURRED DURING PREVENTIVE MAINTENANCE INSPECTION IN THE BIOMED SHOP. THERE WAS NO PATIENT INVOLVEMENT AND NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19822392 ↗
FDA clearance and classification
Cleared through 510(k) K911383 on Oct 17, 1991 (decision: Substantially Equivalent). Submitted by Level 1 Technologies, Inc. as “Hotline Fluid Warmer”. The FDA review took 203 days. FDA record ↗
The FDA regulates product code LGZ (warmer, thermal, infusion fluid) under 21 CFR 880.5725 as a Class II device. 21 CFR 880.5725 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K911383. One submission often covers many models, so check the product and lot details in each record.
- Z-2759-2026 · Jun 24, 2026 · Open, ClassifiedHotline Fluid Warming Set, single use device, is designed to warm blood and intravenous fluids and deliver them to the patient's intravenous access site at normothermic temperature
- Z-1444-2014 · Mar 17, 2014 · TerminatedSmiths Medical HOTLINE Fluid Warming Set L-80. Product Usage: intended for the delivery of warmed blood and intravenous fluids at normothermic temperatures under gravity condit
- Z-1443-2014 · Mar 17, 2014 · TerminatedSmiths Medical HOTLINE Fluid Warming Set L-70. Product Usage: intended for the delivery of warmed blood and intravenous fluids at normothermic temperatures under gravity conditio
- Z-1979-2012 · Jun 27, 2012 · TerminatedSmiths Medical HOTLINE 3 Blood and Fluid Warmer, Reorder Numbers: HL-390 Product Usage: The Micro-switch is used on the HOTLINE 3 Blood and Fluid Warmer. This product, for use
- Z-0426-2012 · Nov 21, 2011 · TerminatedHOTLlNE Disposable Administration Sets, Product Code L-80 Product Usage: Single patient use Sets designed for use by trained medical professionals, in conjunction with HOTLlNE R
- Z-0422-2012 · Nov 21, 2011 · TerminatedHOTLlNE Disposable Administration Sets, Product Codes L-70 Product Usage: Single patient use Sets designed for use by trained medical professionals, in conjunction with HOTLlNE R
- Z-0428-2012 · Nov 21, 2011 · TerminatedLevel 1 Normothermic I.V. Fluid Administration Sets, Product Code DI-60HL , Product Usage: Single patient use sets designed for use by trained medical professionals, in conjuncti
- Z-0424-2012 · Nov 21, 2011 · TerminatedHOTLlNE Disposable Administration Sets, Product Code L-270 Product Usage: Single patient use Sets designed for use by trained medical professionals, in conjunction with HOTLlNE
- Z-0427-2012 · Nov 21, 2011 · TerminatedLevel 1 Normothermic I.V. Fluid Administration Sets, Product Code D-60HL , Product Usage: Single patient use sets designed for use by trained medical professionals, in conjunctio
- Z-0425-2012 · Nov 21, 2011 · TerminatedHOTLlNE Disposable Administration Sets, Product Code L-370 Product Usage: Single patient use Sets designed for use by trained medical professionals, in conjunction with HOTLlNE
- Z-0423-2012 · Nov 21, 2011 · TerminatedHOTLlNE Disposable Administration Sets, Product Codes L-70NI Product Usage: Single patient use Sets designed for use by trained medical professionals, in conjunction with HOTLlNE
Common questions about the Level 1 HL-390-38
Has the Level 1 HL-390-38 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10695085002826), based on the FDA recall file dated Nov 27, 2026. However, 11 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the Level 1 HL-390-38?
The FDA's MAUDE database lists 9 adverse event reports for this device, received between Jul 24, 2024 and May 26, 2026. In those reports, the most common specific device problems are “Electrical /Electronic Property Problem” (2), “Overheating of Device” (2) and “Temperature Problem” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Level 1 HL-390-38?
None of the reports are classified as deaths, 0 as injuries and 9 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Level 1 HL-390-38 FDA approved?
It was cleared through 510(k) K911383 on Oct 17, 1991. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Level 1 HL-390-38, and is it still sold?
The FDA's device database lists Icu Medical, Inc. as the company responsible for it (brand Level 1, model HL-390-38). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Warmer, thermal, infusion fluid devices
- Level 1 HL-90Icu Medical, Inc. · Warmer, thermal, infusion fluid991 reports
- Level 1 HL-390Icu Medical, Inc. · Warmer, thermal, infusion fluid634 reports
- Level 1 L-70Icu Medical, Inc. · Warmer, thermal, infusion fluid1 recall385 reports
- 3-Spike Disposable Set (1 Piece) 903-00006PBelmont Instrument Corporation · Warmer, Thermal, Infusion Fluid1 recall1 death97 reports
- Rapid Infuser RI-2, 120V, 750 ml/min 903-00039Belmont Instrument Corporation · Warmer, Thermal, Infusion Fluid132 reports
- Level 1 L-70NIIcu Medical, Inc. · Warmer, thermal, infusion fluid114 reports
- Rapid Infuser RI-2, 120V, 1000ml/min 903-00037Belmont Instrument Corporation · Warmer, Thermal, Infusion Fluid1 death75 reports
- Hyperthermia Pump Procedure Kit 902-00045Belmont Instrument Corporation · Warmer, Thermal, Infusion Fluid1 recall4 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page