BLAKE 2211Recalls & reported problems
Ethicon Inc. · Blake Silicone Drain, 7 mm flat, full fluted · Class 1
All models of BLAKE (18)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10705031003477), received between Apr 8, 2019 and Jun 27, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Infusion or Flow Problem1
- Material Split, Cut or Torn1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 27, 2024Break; Break
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND DRAIN WAS USED. DURING THE PROCEDURE, THE 7 FR DRAIN IS PERFORMED THROUGH THE SKIN HOLE, WHICH IS WHERE IT BREAKS IN THE WHITE PART. NORMALLY, THIS DOES NOT HAPPEN IN THIS MANEUVER. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19627368 ↗MalfunctionOct 24, 2022Break; Break
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN GYNECOLOGICAL PROCEDURE ON (B)(6) 2022 AND A DRAIN WAS USED. A CASE OF TECHNO-SURVEILLANCE OF A FLAT SILICONE DRAINAGE SYSTEM FLAT 7MM STRIPED UNI IS REPORTED, WHICH AT THE TIME OF PASSING IT, SPLITS IN 2 AT THE UNION OF THE 2 SILICONES, SO IT DOES NOT WORK AND IT IS DECIDED TO USE ANOTHER OF THE SAME CHARACTERISTICS. NO REPORTED ADVERSE PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15660826 ↗MalfunctionDec 15, 2020Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED A PATIENT UNDERWENT AN INITIAL DIAGNOSIS FOR MALIGNANT BREAST TUMOR ON (B)(6) 2020, AND A DRAIN WAS USED. WHEN PLACED, IT UNRAVELED. ATTEMPTED TO ATTACH, BUT IT WOULD NOT CREATE VACUUM. NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11010659 ↗MalfunctionApr 8, 2019Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE AND A DRAIN WAS USED. A DRAIN WAS PULLED FROM A PATIENT IN THE CLINIC AND WHEN PULLED THE DRAIN WAS SPLIT. THE MANUFACTURER IS NOT CLEAR. THERE WAS NO ADVERSE CONSEQUENCE FOR THE PATIENT REPORTED. THE DEVICE IS NOT AVAILABLE FOR RETURN. ADDITIONAL INFORMATION REQUESTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8492149 ↗
Other CATHETER, IRRIGATION devices
- Navarre Opti-Drain NOD8LPTBard Access Systems, Inc. · CATHETER, IRRIGATION901 reports
- Navarre NNU8LPTBard Access Systems, Inc. · CATHETER, IRRIGATION736 reports
- BLAKE 2227Ethicon Inc. · CATHETER, IRRIGATION331 reports
- Navarre NNU6LPTBard Access Systems, Inc. · CATHETER, IRRIGATION211 reports
- Navarre Opti-Drain NOD10LPTBard Access Systems, Inc. · CATHETER, IRRIGATION178 reports
- Navarre NNU10LPTBard Access Systems, Inc. · CATHETER, IRRIGATION127 reports
- BLAKE 2233Ethicon Inc. · CATHETER, IRRIGATION114 reports
- Vasoshield VH-5001Maquet Cardiovascular LLC · Catheter, Irrigation1 recall27 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page