BLAKE 2214Recalls & reported problems
Ethicon Inc. · Blake Silicone Drain, 10 mm flat, full fluted · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10705031003507), received between Oct 28, 2019 and Aug 23, 2024 — classified as 7 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Break” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break7
Patient problems
- Foreign Body In Patient7
- No Clinical Signs, Symptoms or Conditions2
- Insufficient Information1
- No Code Available1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
InjuryAug 23, 2024Break; Break
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN SPINAL PROCEDURE ON AN UNKNOWN DATE IN (B)(6)2024 AND A DRAIN WAS USED. POST OP, THE DRAIN CAME APART AT THE Y JUNCTION PREMATURELY. PATIENT NEEDED TO BE RETURNED TO THE OR TO REMOVE THE DRAIN. NO FURTHER INFORMATION WAS PROVIDED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Foreign Body In Patient
MAUDE report 20047913 ↗InjuryAug 23, 2024Break; Break
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN SPINAL PROCEDURE ON AN UNKNOWN DATE IN 2024 AND A DRAIN WAS USED. POST OP, THE DRAIN CAME APART AT THE Y JUNCTION PREMATURELY. PATIENT NEEDED TO BE RETURNED TO THE OR TO REMOVE THE DRAIN. NO FURTHER INFORMATION WAS PROVIDED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Foreign Body In Patient
MAUDE report 20046833 ↗MalfunctionAug 22, 2024Break; Break
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN SPINAL PROCEDURE ON AN UNKNOWN DATE IN 2024 AND A DRAIN WAS USED. POST OP, THE DRAIN CAME APART AT THE Y JUNCTION PREMATURELY. PATIENT NEEDED TO BE RETURNED TO THE OR TO REMOVE THE DRAIN. NO FURTHER INFORMATION WAS PROVIDED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 20041998 ↗InjuryAug 22, 2024Break; Break
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN SPINAL PROCEDURE ON (B)(6) 2024 AND A DRAIN WAS USED. POST OP, THE DRAIN CAME APART AT THE Y JUNCTION PREMATURELY. PATIENT NEEDED TO BE RETURNED TO THE OR TO REMOVE THE DRAIN. NO FURTHER INFORMATION WAS PROVIDED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Foreign Body In Patient
MAUDE report 20040479 ↗InjuryJul 30, 2024
THE NEUROSURGEON WAS AT BEDSIDE TO REMOVE A BULB SUCTION DRAIN (BLAKE DRAIN) FROM A ONE-DAY POST-SURGICAL PATIENT. THE PATIENT HAD A L5-S1 TRANSFORAMINAL LUMBAR INTERBODY FUSION AND RIGHT L4-5 LAMINOTOMY/FORAMINOTOMY WITH INTRAPROCEDURAL DRAIN PLACEMENT ON (B)(6) 2024 AND WAS BEING PREPARED FOR DISCHARGE. UPON REMOVING THE SUTURES OF THE DRAIN AND GENTLY PULLING ON THE DEVICE, THE DRAIN "CAME APART, LEAVING A PORTION OF THE DRAIN INSIDE THE PATIENT" NEAR THE POINT OF ENTRY. THE NEUROSURGEON ATTEMPTED TO REMOVE THE DRAIN FRAGMENT FROM THE PATIENT BUT WAS UNSUCCESSFUL. THE PATIENT REQUIRES RETURN TO THE OPERATING ROOM FOR SURGICAL REMOVAL OF MEDICAL DEVICE. THE PATIENT HAD NOT BEEN NPO AND THEREFORE WAS REQUIRED TO STAY IN THE HOSPITAL FOR AN ADDITIONAL DAY TO SAFELY COMPLETE INTERVENTION.
Patient problems: Foreign Body In Patient
MAUDE report 19877196 ↗InjuryJul 30, 2024
THE PATIENT WAS ONE DAY POSTOPERATIVE L3-4, L4-5 LAMINECTOMY AND WAS BEING PREPARED FOR DISCHARGE. THE NEUROSURGICAL PHYSICIAN ASSISTANT WAS AT BEDSIDE TO REMOVE BULB SUCTION DRAIN (BLAKE DRAIN) THAT HAD BEEN PLACED INTRAOPERATIVELY. WHILE REMOVING BLAKE DRAIN, THE DRAIN TUBING BROKE NEAR THE POINT OF ENTRY WITH A PORTION OF THE DRAIN REMAINING WITHIN THE INCISION. THE NEUROSURGEON WAS IMMEDIATELY NOTIFIED AND ARRIVED AT BEDSIDE BUT WAS UNABLE TO RETRIEVE THE DRAIN FRAGMENT. THE PATIENT WAS RETURNED TO THE OPERATING ROOM FOR SAFE RETRIEVAL OF RETAINED MEDICAL DEVICE.
Patient problems: Insufficient Information
MAUDE report 19876898 ↗MalfunctionDec 1, 2023Break; Break
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE IN 2023 AND A DRAIN WAS USED. CATHETER SEPARATED WITHIN PATIENT AND HAD TO BE REMOVED DURING PROCEDURE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18245320 ↗InjuryNov 30, 2023Break; Break
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE IN 2023 AND A DRAIN WAS USED. THE DRAIN SEPARATED AND DISLODGED IN THE PATIENT. PATIENT HAD TO GO TO THE CATH LAB TO HAVE THE DISLODGED END OF THE CATHETER REMOVED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Foreign Body In Patient
MAUDE report 18236058 ↗InjuryOct 28, 2019Break; Break
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2019 AND A DRAIN WAS USED. ON (B)(6) 2019 THEY WENT TO REMOVE THE DRAIN BEDSIDE AND IT SNAPPED OFF. THE DRAIN WAS SURGICALLY REMOVED ON (B)(6) 2019. A NEW DRAIN WAS NOT NEEDED. THE PATIENT WAS DISCHARGED ON (B)(6) 2019. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Foreign Body In Patient; No Code Available; No Consequences Or Impact To Patient
MAUDE report 9246804 ↗
Other CATHETER, IRRIGATION devices
- Navarre Opti-Drain NOD8LPTBard Access Systems, Inc. · CATHETER, IRRIGATION901 reports
- Navarre NNU8LPTBard Access Systems, Inc. · CATHETER, IRRIGATION736 reports
- BLAKE 2227Ethicon Inc. · CATHETER, IRRIGATION331 reports
- Navarre NNU6LPTBard Access Systems, Inc. · CATHETER, IRRIGATION211 reports
- Navarre Opti-Drain NOD10LPTBard Access Systems, Inc. · CATHETER, IRRIGATION178 reports
- Navarre NNU10LPTBard Access Systems, Inc. · CATHETER, IRRIGATION127 reports
- BLAKE 2233Ethicon Inc. · CATHETER, IRRIGATION114 reports
- Vasoshield VH-5001Maquet Cardiovascular LLC · Catheter, Irrigation1 recall27 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page