ETHICON M649GRecalls & reported problems
Ethicon, LLC · Monofilament 316L Stainless Steel, Nonabsorbable Surgical Suture · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Ethicon steel surgical suture, a nonabsorbable stainless steel monofilament, is used to close tissue during surgery. Most of the 37 reports describe the suture separating from its needle, the needle tip breaking off during surgery, or contamination or hair found inside the sterile packaging, usually noticed before or during the procedure. There are 3 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 36
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 36 adverse event reports that list this exact device (DI 10705031045415), received between Feb 5, 2018 and Sep 4, 2026 — classified as 2 injuries, 34 malfunctions.
The most frequently reported specific device problem is “Material Separation” (21 reports).
8 other recalls cover devices cleared under the same FDA submission (K170767, K946173); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation21
- Break4
- Device Contaminated During Manufacture or Shipping4
- Appropriate Device Problem Term/Code Not Available2
- Defective Component2
- Material Integrity Problem2
- Adverse Event Without Identified Device or Use Problem1
- Delivered as Unsterile Product1
Patient problems
- No Clinical Signs, Symptoms or Conditions28
- No Consequences Or Impact To Patient5
- Erythema1
- Foreign Body In Patient1
- Hemorrhage/Blood Loss/Bleeding1
- Itching Sensation1
- Not Applicable1
- Rash1
Recent adverse event reports
The 25 most recent of 36 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 4, 2026Device Contaminated During Manufacture or Shipping
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50613510 ↗MalfunctionJan 22, 2026Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED BY A SALES REP THAT, IT WAS FOUND THAT HERE HAD BEEN CONTAMINATION INSIDE OF THE PACKAGE. ANOTHER COMPETITIVE PRODUCT WAS USED TO COMPLETE THE CASE. NO SAMPLE WILL BE RETURNED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24152112 ↗MalfunctionJan 21, 2026Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED BY A SALES REP THAT, IT WAS FOUND THAT HERE HAD BEEN CONTAMINATION INSIDE OF THE PACKAGE. ANOTHER COMPETITIVE PRODUCT WAS USED TO COMPLETE THE CASE. NO SAMPLE WILL BE RETURNED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24137121 ↗MalfunctionJan 15, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN CARDIOVASCULAR SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS FOUND THAT THE WIRE WAS DETACHED. NO TROUBLES WERE CAUSED TO THE PATIENT AND THE DOCTOR BECAUSE OF THIS ISSUE. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24086650 ↗MalfunctionOct 21, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A LEFT ATRIAL APPENDAGE LIGATION+MECHANICAL VALVE REPLACEMENT OF AORTIC VALVE+MECHANICAL VALVE REPLACEMENT OF MITRAL VALVE (PRESERVING SUBVALVULAR STRUCTURE) +TRICUSPID ANNULOPLASTY+EPICARDIAL ELECTRODE IMPLANTATION+PERICARDIAL FENESTRATION PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, STEEL WIRE WAS USED TO FIX STERNUM DURING LEFT ATRIAL APPENDAGE LIGATION+MECHANICAL VALVE REPLACEMENT OF AORTIC VALVE+MECHANICAL VALVE REPLACEMENT OF MITRAL VALVE (PRESERVING SUBVALVULAR STRUCTURE) +TRICUSPID ANNULOPLASTY+EPICARDIAL ELECTRODE IMPLANTATION+PERICARDIAL FENESTRATION SURGERY. IN THE PROCESS OF FIXING STERNUM, WITHOUT FORCED PULLING BY EXTERNAL FORCE AND CUTTING OFF OF STEEL WIRE, THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED. THE SURGEON IMMEDIATELY STOPPED THE STERNAL FIXATION AND VERIFIED THE POSITION OF STEEL WIRE AND SUTURE NEEDLE TO CONFIRM THAT THERE WAS NO IN VIVO RESIDUE AND NO INJURY TO THE SURGEON; THEN QUICKLY REPLACED ANOTHER ONE AND CHECKED THE CONNECTION FIRMNESS TWICE AND COMPLETED THE FIXATION ACCORDING TO THE SPECIFICATIONS AFTER ADJUSTING THE OPERATION METHOD TO CONTROL THE FORCE. NO ADVERSE PATIENT CONS
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23349324 ↗MalfunctionJul 18, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT, DURING AN UNKNOWN SURGERY, THE NEEDLE WAS BROKEN DURING USE. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. WHEN WE SEND THE SAMPLE TO YOU, WE WILL LET YOU KNOW ITS RETURN DATE AND TRACKING NUMBER. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. AFFILIATE REFERENCE NUMBER. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22541723 ↗MalfunctionMay 2, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT THE SURGEON THAT THERE WAS HAIR WITHIN THE STERILE PACKAGING. THERE WAS NO PATIENT CONSEQUENCE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21951457 ↗MalfunctionOct 21, 2024Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN CARDIOVASCULAR SURGERY ON (B)(6) 2024 AND SUTURE WAS USED. DURING AN UNKNOWN CARDIOVASCULAR SURGERY, WHEN OPEN THE PACKAGE, THE NEEDLE WAS ALREADY DETACHED. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20494870 ↗MalfunctionOct 9, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A CARDIOVASCULAR PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, MULTIPLE NEEDLES CAME OF THE STEELWIRES. NOT FROM ALL OF THEM. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20412736 ↗MalfunctionOct 9, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A CARDIOVASCULAR PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, MULTIPLE NEEDLES CAME OF THE STEELWIRES. NOT FROM ALL OF THEM. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20412293 ↗MalfunctionMay 21, 2024Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON (B)(6) 2024 AND SUTURE WAS USED. WHEN THE BOX WAS OPENED, ONE PACKAGE WAS ALREADY BROKEN. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19358534 ↗MalfunctionMar 13, 2024Material Separation; Material Separation
IT WAS REPORTED THAT PATIENT UNDERWENT AN ORTHOPEDIC PROCEDURE ON (B)(6) 2024, AND SUTURE WAS USED. WHILE SUTURING THE SURGICAL WOUND FOR A FRACTURE PATIENT, THE DOCTOR FOUND THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAD HAPPENED IN THE NEWLY DISPENSED SUTURE. THEY CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18897716 ↗MalfunctionSep 5, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED BY A SALES REP THAT, THE SUTURE WAS EASILY DETACHED FROM THE NEEDLE. IT WAS NOT A CONTROL RELEASE NEEDLE. IT IS NOT KNOWN WHEN THE EVENT OCCURRED. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. WHEN WE SEND THE SAMPLE TO YOU. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17686624 ↗MalfunctionAug 14, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT, THE SUTURE WAS EASILY DETACHED FROM THE NEEDLE. IT WAS NOT A CONTROL RELEASE NEEDLE. IT IS NOT KNOWN WHEN THE EVENT OCCURRED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17546815 ↗MalfunctionFeb 1, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE NEEDLE WAS EASILY DETACHED FROM THE SUTURE. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16277639 ↗MalfunctionJan 6, 2023Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN CARDIOVASCULAR SURGERY, ON (B)(6) 2022 AND SUTURE WAS USED. WHEN PULLING THE NEEDLE AFTER PENETRATING THE STERNUM, THE NEEDLE CAME OFF IMMEDIATELY FROM THE STEEL JUNCTION. FURTHER DETAILS ARE NOT PROVIDED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16114325 ↗MalfunctionJan 6, 2023Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN CARDIOVASCULAR SURGERY, ON (B)(6) 2022 AND SUTURE WAS USED. WHEN PULLING THE NEEDLE AFTER PENETRATING THE STERNUM, THE NEEDLE CAME OFF IMMEDIATELY FROM THE STEEL JUNCTION. FURTHER DETAILS ARE NOT PROVIDED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16114318 ↗MalfunctionDec 22, 2022Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A CARDIOVASCULAR PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. BEFORE USE ON THE PATIENT, WHEN A SCRUB NURSE OPENED THE PACKAGE, IT WAS FOUND THAT THE NEEDLE HAD BEEN DETACHED FROM THE WIRE. IT WAS NOT A CONTROL RELEASE NEEDLE. FURTHER DETAILS ARE NOT PROVIDE. UPON EVALUATION OF THE RETURNED DEVICE, AFTER INVESTIGATION OF THE SAMPLE, SHORT INSERTION WAS OBSERVED IN THE STRAND. THE VISUAL INSPECTION CONCLUDED THAT THE COMBINATION OF THE NEEDLE/SUTURE WAS DETACHED FROM THE NEEDLE SINCE THE INSERTION END OF THE SUTURE DID NOT MEET THE REQUIREMENT. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. NO ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16039163 ↗MalfunctionSep 16, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN CARDIOVASCULAR PROCEDURE AND SUTURE WAS USED. WHEN THE SURGEON GRASPED THE NEEDLE TO USE THE WIRE, THE NEEDLE CAME OFF EASILY DURING INTERRUPTED SUTURING. ANOTHER LOT OF PRODUCT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15442141 ↗MalfunctionSep 16, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN CARDIOVASCULAR PROCEDURE AND SUTURE WAS USED. WHEN THE SURGEON GRASPED THE NEEDLE TO USE THE WIRE, THE NEEDLE CAME OFF EASILY DURING INTERRUPTED SUTURING. ANOTHER LOT OF PRODUCT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15442139 ↗MalfunctionAug 26, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN CARDIOVASCULAR PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED FOR STERNAL CLOSURE. THE NEEDLE WAS DETACHED FROM THE WIRE EASILY DURING USE. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION WILL BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15302509 ↗MalfunctionAug 26, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN CARDIOVASCULAR PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED FOR STERNAL CLOSURE. THE NEEDLE WAS DETACHED FROM THE WIRE EASILY DURING USE. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION WILL BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15302464 ↗MalfunctionApr 5, 2022Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. WHEN THE PACKAGE WAS OPENED, IT WAS FOUND THAT THE NEEDLE TIP HAD BEEN BROKEN ALREADY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14018255 ↗MalfunctionMar 9, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN CARDIOVASCULAR SURGERY ON (B)(6) 2022. AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE WAS DETACHED FROM THE SUTURE DURING USE, AND IT WAS NOT A CONTROL RELEASE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13716870 ↗MalfunctionJan 11, 2022Defective Component; Appropriate Device Problem Term/Code Not Available; Defective Component; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. BEFORE USE ON PATIENT, IT WAS REPORTED THAT THE SUTURE DIAMETER IS DIFFERENT FROM TWO DIFFERENT LOT. ON THE PRODUCT OUTER PACKAGING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13233235 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K170767, K946173. One submission often covers many models, so check the product and lot details in each record.
- Z-1516-2025 · Mar 7, 2025 · Open, ClassifiedETHIBOND EXCEL GRN 36IN(90CM) USP2-0(M3) D/A MH
- Z-1097-2025 · Dec 20, 2024 · Open, ClassifiedPERMA-HAND Silk Suture, REF: K872H, W723H, 623H, K833H; Perma-hand Silk Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovas
- Z-2486-2020 · Jun 10, 2020 · Open, ClassifiedPERMA-HAND/MERSILK SILK BLU and VIRGIN SILK BLU, Product Codes EH7457G EH7983G FV768 FV769 FV7771 U7058 U7059 W1782 W1784 W1819 W1820 W818 W819 W870
- Z-1841-2019 · Apr 19, 2019 · TerminatedETHIBOND EXCEL Polyester Suture-ETHIBOND Green 75cm USP1 Single Armed CT-1 Product Code: X425H Product Usage: ETHIBOND EXCEL Suture is intended for use in general soft tissue app
- Z-1842-2019 · Apr 19, 2019 · TerminatedETHIBOND EXCEL ETHIBOND Polvester Suture-Green 75cm USP1 Single Armed CTX Product Code: X865W Product Usage: ETHIBOND EXCEL Suture is intended for use in general soft tissue appr
- Z-0747-2019 · Dec 18, 2018 · TerminatedPERMA-HAND Silk Suture Size 5, Black 30" (75cm) Double Armed LR Needles, Product Code 485T Product Usage: PERMA-HAND* suture is indicated for use in general soft tissue approxima
- Z-1839-2015 · May 8, 2015 · TerminatedEthicon Fast Absorbing Plain Gut - MultiPass Needles, Sterile, Absorbable Surgical Suture, 5-0 (1.5 metric), 18" (45 cm), Conventional Cutting, Lot #JD6163, Exp: July 2019. Fast ab
- Z-0492-2015 · Oct 17, 2014 · TerminatedEthicon PERMA-HAND Silk siliconized black braided 10 strands per packet Ethicon LLC- Made in USA.
Other SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE devices
- ETHICON M660GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE102 reports
- ETHICON M650GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE68 reports
- ETHIBOND EXCEL X425HEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE1 recall12 reports
- Steel 88862224-49Covidien LP · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE1 recall3 reports
- ETHICON M655GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE46 reports
- ETHIBOND EXCEL MB66GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE42 reports
- Steel 88862424-89Covidien LP · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE39 reports
- ETHIBOND EXCEL X523HEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE26 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page