ETHICON M655GRecalls & reported problems
Ethicon, LLC · Monofilament 316L Stainless Steel, Nonabsorbable Surgical Suture · Class 2 · Implant
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Ethicon steel surgical suture (M655G) is a nonabsorbable stainless steel thread used to close tissue during surgery. The 46 reports mostly describe the suture or its needle breaking during operations, sometimes leaving a fragment inside the patient, along with material separating or fraying, black debris on the needle, damaged packaging, and product arriving unsterile. Two reports describe injury; no deaths were reported. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 46
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 46 adverse event reports that list this exact device (DI 10705031045491), received between Jul 9, 2018 and Mar 13, 2026 — classified as 2 injuries, 44 malfunctions.
The most frequently reported specific device problem is “Break” (26 reports).
10 other recalls cover devices cleared under the same FDA submission (K170767, K946173); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break26
- Material Separation12
- Device Contaminated During Manufacture or Shipping3
- Delivered as Unsterile Product2
- Adverse Event Without Identified Device or Use Problem1
- Material Frayed1
- Packaging Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions32
- No Consequences Or Impact To Patient12
- Foreign Body In Patient1
- Insufficient Information1
- Not Applicable1
- Pain1
- Unspecified Infection1
Recent adverse event reports
The 25 most recent of 46 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 13, 2026Packaging Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT THE PRIMARY PACKAGE OF THE SUTURE WAS FOUND DAMAGED PRIOR TO USE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24590221 ↗MalfunctionDec 22, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, SUTURE BREAKING AT NEEDLE SITE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23869473 ↗MalfunctionJul 17, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THERE WAS BLACK DEBRIS NOTICED OUTSIDE THE RUBBER PROTECTOR OF THE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22531634 ↗MalfunctionJan 16, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN CORONARY ARTERY BYPASS PROCEDURE ON (B)(6) 2025 AND A SUTURE WAS USED. DURING THE SURGERY, THE NEEDLE SUDDENLY BROKE WHEN SUTURING THE STERNUM. THE BROKEN NEEDLE AND FRAGMENTS WERE REMOVED IN TIME AND SPLICED TOGETHER COMPLETELY TO CONFIRM THAT THERE WERE NO FALLEN FRAGMENTS. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Insufficient Information
MAUDE report 21172311 ↗MalfunctionMay 25, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6) 2024 AND STERNAL WIRE WAS PLACED UPON CLOSING THE STERNOTOMY. INCIDENCE OF THE WIRES BREAKING INSIDE THE PATIENT WAS REPORTED INTRA OPERATIVELY. THE BREAKAGE DID NOT OCCUR WHERE THE SURGEON WAS TYING/TIGHTENING THE WIRES, BUT INSTEAD SNAPPING/BREAKING WHERE THEY WERE PLACED IN TISSUE/BONE. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19398695 ↗InjuryMay 25, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6) 2024 AND STERNAL WIRE WAS PLACED UPON CLOSING THE STERNOTOMY. THE PATIENT HAD TO BE BROUGHT BACK POST OP BECAUSE ONE WIRE WAS STICKING THE PATIENT AND CAUSING PAIN. THE BREAKAGE DID NOT OCCUR WHERE THE SURGEON WAS TYING/TIGHTENING THE WIRE, BUT INSTEAD SNAPPING/BREAKING WHERE IT WAS PLACED IN TISSUE/BONE. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Pain
MAUDE report 19398694 ↗MalfunctionMay 25, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON 28-APR-2024 AND STERNAL WIRE WAS PLACED UPON CLOSING THE STERNOTOMY. INCIDENCE OF THE WIRES BREAKING INSIDE THE PATIENT WAS REPORTED INTRA OPERATIVELY. THE BREAKAGE DID NOT OCCUR WHERE THE SURGEON WAS TYING/TIGHTENING THE WIRES, BUT INSTEAD SNAPPING/BREAKING WHERE THEY WERE PLACED IN TISSUE/BONE. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19398693 ↗MalfunctionMar 27, 2024Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PASSAGE OF THE STEEL WIRE THROUGH THE PATIENT'S STERNUM FOR OSTEOSYNTHESIS, THE TIP OF THE NEEDLE HAD BROKEN OFF AND FALLEN INTO THE PATIENT. THE TIP OF THE NEEDLE WAS FOUND. NO PATIENT CONSEQUENCE WAS REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18986219 ↗MalfunctionJan 4, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE IN 2023 AND SUTURE WAS USED. THE DOCTOR EXPERIENCED THIS SUTURE CONTINUOUSLY TO BREAK UNINTENTIONALLY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18443083 ↗MalfunctionDec 5, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE IN 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SALES REP REPORTED THAT THE STERNAL WIRES (ITEM # M655) HAVE BEEN BREAKING WHILE TWISTING AND APPLYING SLIGHT PRESSURE. THE SURGEON AND PA SIMPLY USED OTHER WIRES TO COMPLETE THE STERNAL CLOSURE WITHOUT ANY HARM TO THE PATIENT. NO DELAYS AND NO FRAGMENTS MADE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18269851 ↗MalfunctionDec 5, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE IN 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SALES REP REPORTED THAT THE STERNAL WIRES (ITEM # M655) HAVE BEEN BREAKING WHILE TWISTING AND APPLYING SLIGHT PRESSURE. THE SURGEON AND PA SIMPLY USED OTHER WIRES TO COMPLETE THE STERNAL CLOSURE WITHOUT ANY HARM TO THE PATIENT. NO DELAYS AND NO FRAGMENTS MADE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18269848 ↗MalfunctionDec 5, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE IN 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SALES REP REPORTED THAT THE STERNAL WIRES (ITEM # M655) HAVE BEEN BREAKING WHILE TWISTING AND APPLYING SLIGHT PRESSURE. THE SURGEON AND PA SIMPLY USED OTHER WIRES TO COMPLETE THE STERNAL CLOSURE WITHOUT ANY HARM TO THE PATIENT. NO DELAYS AND NO FRAGMENTS MADE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18269821 ↗MalfunctionAug 7, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE SALES REP THAT A NEEDLE POPPED OFF OF A STEEL STERNAL WIRE. COORDINATOR STATED THAT NEEDLE AND PACKAGING ARE AVAILABLE FOR RETURN. UNKNOWN IF THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17477319 ↗MalfunctionJul 24, 2023Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A SUTURE WAS USED. A HAIR WAS FOUND ON A PACKAGE OF 7 STEEL WIRES. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17379694 ↗MalfunctionJul 24, 2023Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A SUTURE WAS USED. A HAIR WAS FOUND ON A PACKAGE OF 7 STEEL WIRES. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17378848 ↗MalfunctionJun 12, 2023Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2023 AND SUTURE WAS USED. BEFORE USE ON THE PATIENT, IT WAS REPORTED BY THE DISTRIBUTOR THAT WHEN THEY RECEIVED THE SHIPMENT THE PRODUCTS INSIDE WERE SIGNIFICANTLY DAMAGED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17107839 ↗MalfunctionJun 12, 2023Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2023 AND SUTURE WAS USED. BEFORE USE ON THE PATIENT, IT WAS REPORTED BY THE DISTRIBUTOR THAT WHEN THEY RECEIVED THE SHIPMENT THE PRODUCTS INSIDE WERE SIGNIFICANTLY DAMAGED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17107838 ↗MalfunctionMar 20, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS GRAFT PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. THE SURGEON WAS CLOSING THE STERNUM AND ON HIS FIRST PASS WITH THE NEEDLE THRU THE MENUBRIUM, HE PULLED UP ON THE NEEDLE AND THE NEEDLE BROKE. THEY GOT A SECOND STERNAL WIRE, HE TOOK TWO BITES, ONE ON EACH SIDE OF THE MENUBRIUM, PULLED UP ON THE NEEDLE AND THE SECOND NEEDLE BROKE. THEY THEN GOT A THIRD STERNAL WIRE, TOOK TWO PASSES, PULLED UP AND THE NEEDLE POPPED OFF THE WIRE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16576367 ↗MalfunctionMar 20, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS GRAFT PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. THE SURGEON WAS CLOSING THE STERNUM AND ON HIS FIRST PASS WITH THE NEEDLE THRU THE MENUBRIUM, HE PULLED UP ON THE NEEDLE AND THE NEEDLE BROKE. THEY GOT A SECOND STERNAL WIRE, HE TOOK TWO BITES, ONE ON EACH SIDE OF THE MENUBRIUM, PULLED UP ON THE NEEDLE AND THE SECOND NEEDLE BROKE. THEY THEN GOT A THIRD STERNAL WIRE, TOOK TWO PASSES, PULLED UP AND THE NEEDLE POPPED OFF THE WIRE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16576329 ↗MalfunctionMar 20, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS GRAFT PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. THE SURGEON WAS CLOSING THE STERNUM AND ON HIS FIRST PASS WITH THE NEEDLE THRU THE MENUBRIUM, HE PULLED UP ON THE NEEDLE AND THE NEEDLE BROKE. THEY GOT A SECOND STERNAL WIRE, HE TOOK TWO BITES, ONE ON EACH SIDE OF THE MENUBRIUM, PULLED UP ON THE NEEDLE AND THE SECOND NEEDLE BROKE. THEY THEN GOT A THIRD STERNAL WIRE, TOOK TWO PASSES, PULLED UP AND THE NEEDLE POPPED OFF THE WIRE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16576263 ↗MalfunctionJan 6, 2023Material Frayed; Material Frayed
IT WAS REPORTED THAT A PATIENT UNDERWENT AN AORTIC DISSECTION PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. THE DEBRIS WAS FOUND, WHICH FELL OFF FROM THE SUTURE DURING THE PROCEDURE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16114958 ↗MalfunctionDec 29, 2022Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. IN THE PROCESS OF USE, IT IS FOUND THAT THE DREGS OF SUTURE PRODUCTS FALL. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16067865 ↗MalfunctionOct 13, 2022Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, THE STERNAL WIRE/NEEDLE TIP BROKE OFF. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. NO ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15597343 ↗MalfunctionOct 13, 2022Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, THE STERNAL WIRE/NEEDLE TIP BROKE OFF. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. NO ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15597342 ↗MalfunctionOct 13, 2022Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, THE STERNAL WIRE/NEEDLE TIP BROKE OFF. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. NO ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15597340 ↗
FDA clearance and classification
Cleared through Special 510(k) K170767 on Nov 1, 2017 (decision: Substantially Equivalent). Submitted by ETHICON, Inc. as “Surgical Stainless Steel Suture, Stainless Steel Suture”. The FDA review took 232 days. FDA record ↗
Cleared through 510(k) K946173 on Jan 9, 1995 (decision: Substantially Equivalent). Submitted by ETHICON, Inc. as “Stainless Steel, Gut, Poly (Glycolide/L-Lactide) Polyamide, Polypropylene, Poly (Ethylene Terephthalate) And Silk”. The FDA review took 21 days. FDA record ↗
The FDA regulates product code GAQ (suture, nonabsorbable, steel, monofilament and multifilament, sterile) under 21 CFR 878.4495 as a Class II device. 21 CFR 878.4495 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K170767, K946173. One submission often covers many models, so check the product and lot details in each record.
- Z-0284-2027 · Sep 23, 2026 · Open, ClassifiedETHILON Nylon Suture, Product Code 697G
- Z-0153-2027 · Sep 8, 2026 · Open, ClassifiedPERMA-HAND Silk Black Braided Sterile Nonabsorbable Surgical Suture, 1X18" TF-5, Product Code: N271H
- Z-1516-2025 · Mar 7, 2025 · Open, ClassifiedETHIBOND EXCEL GRN 36IN(90CM) USP2-0(M3) D/A MH
- Z-1097-2025 · Dec 20, 2024 · Open, ClassifiedPERMA-HAND Silk Suture, REF: K872H, W723H, 623H, K833H; Perma-hand Silk Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovas
- Z-2486-2020 · Jun 10, 2020 · Open, ClassifiedPERMA-HAND/MERSILK SILK BLU and VIRGIN SILK BLU, Product Codes EH7457G EH7983G FV768 FV769 FV7771 U7058 U7059 W1782 W1784 W1819 W1820 W818 W819 W870
- Z-1842-2019 · Apr 19, 2019 · TerminatedETHIBOND EXCEL ETHIBOND Polvester Suture-Green 75cm USP1 Single Armed CTX Product Code: X865W Product Usage: ETHIBOND EXCEL Suture is intended for use in general soft tissue appr
- Z-1841-2019 · Apr 19, 2019 · TerminatedETHIBOND EXCEL Polyester Suture-ETHIBOND Green 75cm USP1 Single Armed CT-1 Product Code: X425H Product Usage: ETHIBOND EXCEL Suture is intended for use in general soft tissue app
- Z-0747-2019 · Dec 18, 2018 · TerminatedPERMA-HAND Silk Suture Size 5, Black 30" (75cm) Double Armed LR Needles, Product Code 485T Product Usage: PERMA-HAND* suture is indicated for use in general soft tissue approxima
- Z-1839-2015 · May 8, 2015 · TerminatedEthicon Fast Absorbing Plain Gut - MultiPass Needles, Sterile, Absorbable Surgical Suture, 5-0 (1.5 metric), 18" (45 cm), Conventional Cutting, Lot #JD6163, Exp: July 2019. Fast ab
- Z-0492-2015 · Oct 17, 2014 · TerminatedEthicon PERMA-HAND Silk siliconized black braided 10 strands per packet Ethicon LLC- Made in USA.
Common questions about the ETHICON M655G
Has the ETHICON M655G been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705031045491), based on the FDA recall file dated Nov 27, 2026. However, 10 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the ETHICON M655G?
The FDA's MAUDE database lists 46 adverse event reports for this device, received between Jul 9, 2018 and Mar 13, 2026. In those reports, the most common specific device problems are “Break” (26), “Material Separation” (12) and “Device Contaminated During Manufacture or Shipping” (3); the most common patient problems are “Foreign Body In Patient” (1), “Not Applicable” (1) and “Pain” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ETHICON M655G?
None of the reports are classified as deaths, 2 as injuries and 44 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the ETHICON M655G FDA approved?
It was cleared through 510(k) K946173 on Jan 9, 1995; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the ETHICON M655G MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the ETHICON M655G, and is it still sold?
The FDA's device database lists Ethicon, LLC as the company responsible for it (brand ETHICON, model M655G). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE devices
- ETHICON M660GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE100 reports
- ETHICON M650GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE67 reports
- ETHIBOND EXCEL X425HEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE1 recall12 reports
- Steel 88862224-49Covidien LP · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE1 recall3 reports
- ETHIBOND EXCEL MB66GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE40 reports
- Steel 88862424-89Covidien LP · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE39 reports
- ETHICON M649GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE35 reports
- ETHIBOND EXCEL X523HEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE26 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page