MONOCRYL Y496GRecalls & reported problems
Ethicon, LLC · Undyed Monofilament (Poliglecaprone 25), Absorbable Surgical Suture · Class 2 · Implant
Quick summary
This Ethicon Monocryl absorbable suture is used to close wounds and tissue. The 85 reports mostly describe the suture material separating or breaking and needles popping off during procedures, with no deaths and 1 injury reported. No recalls name this device. Reports have fallen recently, with none in the past 12 months compared with 9 the year before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 86
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 86 adverse event reports that list this exact device (DI 10705031059375), received between Feb 13, 2018 and Sep 30, 2026 — classified as 1 injury, 85 malfunctions.
The most frequently reported specific device problem is “Material Separation” (51 reports).
2 other recalls cover devices cleared under the same FDA submission (K964072); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation51
- Break27
- Detachment Of Device Component3
- Detachment of Device or Device Component3
- Component Misassembled1
- Dull, Blunt1
- Positioning Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions45
- No Consequences Or Impact To Patient39
- Insufficient Information1
- Unspecified Infection1
Recent adverse event reports
The 25 most recent of 86 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Break
It was reported that a patient underwent an unknown procedure on an unknown date and suture was used. It was reported that the suture broke off at the end. The device was returning. There were no patient consequences reported. Additional information was requested.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50946225 ↗MalfunctionAug 12, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE SALES REP THAT THE NEEDLE WAS POPPING OFF. DEVICE IS AVAILABLE FOR RETURN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22786433 ↗MalfunctionAug 12, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE SALES REP THAT THE NEEDLE WAS POPPING OFF. DEVICE IS AVAILABLE FOR RETURN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22786432 ↗MalfunctionAug 12, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE SALES REP THAT THE NEEDLE WAS POPPING OFF. DEVICE IS AVAILABLE FOR RETURN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22786418 ↗MalfunctionAug 12, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE SALES REP THAT THE NEEDLE WAS POPPING OFF. DEVICE IS AVAILABLE FOR RETURN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22786414 ↗MalfunctionAug 4, 2025Component Misassembled; Component Misassembled
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THERE IS A DOUBLE NEEDLE FOUND ON A UNIDIRECTIONAL SUTURE. SURGICAL DELAY UNKNOWN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22707229 ↗MalfunctionJun 24, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE SALES REP THAT THE SUTURE BROKE OFF THE NEEDLE WHILE IN USE. DEVICES ARE NOT AVAILABLE FOR RETURN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22319458 ↗MalfunctionJun 24, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE SALES REP THAT THE SUTURE BROKE OFF THE NEEDLE WHILE IN USE. . DEVICES ARE NOT AVAILABLE FOR RETURN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22319457 ↗MalfunctionMay 8, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE TIP OF THE NEEDLE POPPED OFF THE SUTURE. ANOTHER LIKE DEVICE WAS USED TO CONTINUE WITH THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21989804 ↗MalfunctionDec 5, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT THE THEIR WAS A RUPTURE OF THE TIP OF THE NEEDLE WHEN SUTURING ORGAN. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20852438 ↗MalfunctionSep 5, 2024Break; Break
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE IN 2024 AND SUTURE WAS USED. THE SUTURES THEY HAVE ARE BREAKING ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20153034 ↗MalfunctionSep 5, 2024Break; Break
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE IN 2024 AND SUTURE WAS USED. THE SUTURES THEY HAVE ARE BREAKING ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20152999 ↗MalfunctionJul 12, 2024Material Separation; Material Separation
IT WAS REPORTED THAT PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024, AND SUTURE WAS USED. THE NEEDLE HAD SNAPPED OFF OF THE SUTURE. THEY WERE ABLE TO RETRIEVE THE NEEDLE AND THERE WAS NO PATIENT HARM. THEY GOT A NEW SUTURE TO COMPLETE THE CASE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19732056 ↗InjuryApr 23, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE TIP OF THE NEEDLE CAME OFF AND IT WAS NOT FOUND. A FLAT PLATE OF ABDOMEN WAS PERFORMED AND CAME BACK NEGATIVE. THE PATIENT WAS RE-OPENED, NOTHING WAS FOUND. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Insufficient Information
MAUDE report 19166050 ↗MalfunctionFeb 9, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SALES REP REPORTED THAT FOUR OF THE NEEDLES WERE PULLED AWAY WHEN IT WAS BEING USED. NO PATIENT HARM REPORTED. NO ADDITIONAL INFORMATION GIVEN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18675736 ↗MalfunctionFeb 9, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SALES REP REPORTED THAT FOUR OF THE NEEDLES WERE PULLED AWAY WHEN IT WAS BEING USED. NO PATIENT HARM REPORTED. NO ADDITIONAL INFORMATION GIVEN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18675735 ↗MalfunctionFeb 9, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SALES REP REPORTED THAT FOUR OF THE NEEDLES WERE PULLED AWAY WHEN IT WAS BEING USED. NO PATIENT HARM REPORTED. NO ADDITIONAL INFORMATION GIVEN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18675733 ↗MalfunctionFeb 9, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SALES REP REPORTED THAT FOUR OF THE NEEDLES WERE PULLED AWAY WHEN IT WAS BEING USED. NO PATIENT HARM REPORTED. NO ADDITIONAL INFORMATION GIVEN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18675729 ↗MalfunctionFeb 8, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SALES REP REPORTED THAT FOUR OF THE NEEDLES WERE PULLED AWAY WHEN IT WAS BEING USED IN A PROCEDURE. NO ADDITIONAL INFORMATION GIVEN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18666958 ↗MalfunctionFeb 8, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SALES REP REPORTED THAT FOUR OF THE NEEDLES WERE PULLED AWAY WHEN IT WAS BEING USED IN A PROCEDURE. NO ADDITIONAL INFORMATION GIVEN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18666956 ↗MalfunctionFeb 8, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SALES REP REPORTED THAT FOUR OF THE NEEDLES WERE PULLED AWAY WHEN IT WAS BEING USED IN A PROCEDURE. NO ADDITIONAL INFORMATION GIVEN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18666952 ↗MalfunctionFeb 8, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SALES REP REPORTED THAT FOUR OF THE NEEDLES WERE PULLED AWAY WHEN IT WAS BEING USED IN A PROCEDURE. NO ADDITIONAL INFORMATION GIVEN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18666951 ↗MalfunctionFeb 5, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. THE SUTURE FELL OUT OF THE NEEDLE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18640849 ↗MalfunctionJan 31, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION ON (B)(6)2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE POPPED OFF THE SUTURE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18615377 ↗MalfunctionJan 31, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE POPPED OFF THE SUTURE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18615374 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K964072. One submission often covers many models, so check the product and lot details in each record.
- Z-0487-2025 · Oct 8, 2024 · Open, ClassifiedMONOCRYL" (poliglecaprone 25) Suture REF Y936H MONOCRYL" Sutures are indicated for use in general soft tissue approximation and/or ligation, but not for use in cardiovascular and
- Z-2148-2024 · May 10, 2024 · Open, ClassifiedMONOCRYL (poliglecaprone 25) Suture - Intended for use in general soft tissue approximation and/or ligation where an absorbable material is indicated PRODUCT CODE: Y315H Y345
Other SUTURE, ABSORBABLE, SYNTHETIC devices
- MONOCRYL Y426HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC141 reports
- MONOCRYL Y936HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC1 recall66 reports
- MONOCRYL Y923HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC111 reports
- MONOCRYL Y427HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC1 recall53 reports
- MONOCRYL Y943GEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC77 reports
- MONOCRYL Y416HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC1 recall13 reports
- MONOCRYL Y935HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC1 recall8 reports
- MONOCRYL Y359HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC1 recall5 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.