MONOCRYL Y935HRecalls & reported problems
Ethicon, LLC · Undyed Monofilament (Poliglecaprone 25), Absorbable Surgical Suture · Class 2 · Implant
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10705031059894), received between Mar 5, 2018 and Jun 11, 2025 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Material Separation” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K964072); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- May 10, 2024Open, Classified
Issue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .
Z-2148-2024 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Material Separation4
- Break2
- Detachment Of Device Component2
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 11, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE SALES REP BY THE SURGEON THAT THE SKIN SUTURE BROKE MID CLOSURE. HAS HAPPENED MULTIPLE TIMES WITH THIS SUTURE LOT NUMBER. THE PROCEDURE WAS DELAYED BY TWO MINUTES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO ADVERSE CONSEQUENCES FOR THE PATIENT. DEVICE WAS RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22187066 ↗MalfunctionAug 10, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE DISTRIBUTOR BY THE CUSTOMER THAT THE NEEDLES ARE PULLING OFF THE SUTURE. THERE WERE 3 NEEDLES TAKE CONSECUTIVELY FROM THE BOX THAT PULLED OFF THE SUTURE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15198569 ↗MalfunctionAug 10, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE DISTRIBUTOR BY THE CUSTOMER THAT THE NEEDLES ARE PULLING OFF THE SUTURE. THERE WERE 3 NEEDLES TAKE CONSECUTIVELY FROM THE BOX THAT PULLED OFF THE SUTURE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15198567 ↗MalfunctionJul 14, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN ORTHOPEDIC PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE T WAS REPORTED BY THE NEEDLES ARE PULLING FREE FROM THE SUTURE. THIS HAPPENED MULTIPLE TIMES. THEY USED A LIKE DEVICE TO COMPLETE THE PROCEDURE WITH NO PATIENT HARM. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15023543 ↗MalfunctionJul 12, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE DISTRIBUTOR BY THE CUSTOMER THAT THE NEEDLES ARE PULLING OFF THE SUTURE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14995608 ↗MalfunctionDec 17, 2018Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE. IT IS UNKNOWN IF ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8170123 ↗MalfunctionMar 5, 2018Detachment Of Device Component; Detachment Of Device Component
IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS GRAFT PROCEDURE ON (B)(6) 2018 AND SUTURE WAS USED. DURING THE PROCEDURE, AFTER PASSING THRU THE SUBCUTICULAR TISSUE, THE SUTURE POPPED OFF AT THE SWAGE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7312790 ↗MalfunctionMar 5, 2018Detachment Of Device Component; Detachment Of Device Component
IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS GRAFT PROCEDURE ON (B)(6) 2018 AND SUTURE WAS USED. DURING THE PROCEDURE, AFTER PASSING THRU THE SUBCUTICULAR TISSUE, THE SUTURE POPPED OFF AT THE SWAGE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7312788 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K964072. One submission often covers many models, so check the product and lot details in each record.
- Z-0487-2025 · Oct 8, 2024 · Open, ClassifiedMONOCRYL" (poliglecaprone 25) Suture REF Y936H MONOCRYL" Sutures are indicated for use in general soft tissue approximation and/or ligation, but not for use in cardiovascular and
Other SUTURE, ABSORBABLE, SYNTHETIC devices
- MONOCRYL Y426HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC141 reports
- MONOCRYL Y936HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC1 recall66 reports
- MONOCRYL Y923HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC111 reports
- MONOCRYL Y427HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC1 recall53 reports
- MONOCRYL Y496GEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC86 reports
- MONOCRYL Y943GEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC77 reports
- MONOCRYL Y416HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC1 recall13 reports
- MONOCRYL Y359HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC1 recall5 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.