Skip to content

MONOCRYL Y935HRecalls & reported problems

Ethicon, LLC · Undyed Monofilament (Poliglecaprone 25), Absorbable Surgical Suture · Class 2 · Implant

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10705031059894), received between Mar 5, 2018 and Jun 11, 2025 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Material Separation” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K964072); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 10, 2024Open, Classified

    Issue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .

    Z-2148-2024 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Material Separation4
  • Break2
  • Detachment Of Device Component2

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient3

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 11, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE SALES REP BY THE SURGEON THAT THE SKIN SUTURE BROKE MID CLOSURE. HAS HAPPENED MULTIPLE TIMES WITH THIS SUTURE LOT NUMBER. THE PROCEDURE WAS DELAYED BY TWO MINUTES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO ADVERSE CONSEQUENCES FOR THE PATIENT. DEVICE WAS RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22187066 ↗
  • MalfunctionAug 10, 2022Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE DISTRIBUTOR BY THE CUSTOMER THAT THE NEEDLES ARE PULLING OFF THE SUTURE. THERE WERE 3 NEEDLES TAKE CONSECUTIVELY FROM THE BOX THAT PULLED OFF THE SUTURE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15198569 ↗
  • MalfunctionAug 10, 2022Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE DISTRIBUTOR BY THE CUSTOMER THAT THE NEEDLES ARE PULLING OFF THE SUTURE. THERE WERE 3 NEEDLES TAKE CONSECUTIVELY FROM THE BOX THAT PULLED OFF THE SUTURE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15198567 ↗
  • MalfunctionJul 14, 2022Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ORTHOPEDIC PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE T WAS REPORTED BY THE NEEDLES ARE PULLING FREE FROM THE SUTURE. THIS HAPPENED MULTIPLE TIMES. THEY USED A LIKE DEVICE TO COMPLETE THE PROCEDURE WITH NO PATIENT HARM. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15023543 ↗
  • MalfunctionJul 12, 2022Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE DISTRIBUTOR BY THE CUSTOMER THAT THE NEEDLES ARE PULLING OFF THE SUTURE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14995608 ↗
  • MalfunctionDec 17, 2018Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE. IT IS UNKNOWN IF ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8170123 ↗
  • MalfunctionMar 5, 2018Detachment Of Device Component; Detachment Of Device Component

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS GRAFT PROCEDURE ON (B)(6) 2018 AND SUTURE WAS USED. DURING THE PROCEDURE, AFTER PASSING THRU THE SUBCUTICULAR TISSUE, THE SUTURE POPPED OFF AT THE SWAGE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7312790 ↗
  • MalfunctionMar 5, 2018Detachment Of Device Component; Detachment Of Device Component

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS GRAFT PROCEDURE ON (B)(6) 2018 AND SUTURE WAS USED. DURING THE PROCEDURE, AFTER PASSING THRU THE SUBCUTICULAR TISSUE, THE SUTURE POPPED OFF AT THE SWAGE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7312788 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K964072. One submission often covers many models, so check the product and lot details in each record.

  • Z-0487-2025 · Oct 8, 2024 · Open, Classified
    MONOCRYL" (poliglecaprone 25) Suture REF Y936H MONOCRYL" Sutures are indicated for use in general soft tissue approximation and/or ligation, but not for use in cardiovascular and

Other SUTURE, ABSORBABLE, SYNTHETIC devices

All SUTURE, ABSORBABLE, SYNTHETIC devices →