Stratafix Spiral SXPD2B426Recalls & reported problems
Surgical Specialties México, S. de R.L. de C.V. · 2V-4 3-0 PDO 24X24 · Class 2 · Implant
All models of Stratafix Spiral (24)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10705031207936), received between Sep 30, 2021 and Jan 7, 2026 — classified as 8 injuries.
The most frequently reported specific device problem is “Material Protrusion/Extrusion” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Protrusion/Extrusion7
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Ulcer5
- Unspecified Infection5
- Fluid Discharge1
- Impaired Healing1
- Post Operative Wound Infection1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryJan 7, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED ON (B)(6) 2025 THAT POST-OPERATIVELY, THE PATIENT EXPERIENCED EXUDATION OF THE WOUND. THE SURGEON RE-SUTURED TO CLOSE THE INCISION. NO OTHER INTERVENTION REQUIRED. IT DOES NOT APPEAR THAT THE ADVERSE REACTION WAS RELATED TO ANY SUTURE DEFECT OR MALFUNCTION.
Patient problems: Fluid Discharge
MAUDE report 23992163 ↗InjuryOct 21, 2021Material Protrusion/Extrusion
THE PATIENT HAD C-SECTION SURGERY IN (B)(6) 2021 DUE TO SCARRED UTERUS. AND WAS DISCHARGED 3 DAYS AFTER OPERATION. (B)(6), THE PATIENT CAME TO THE HOSPITAL FOR REEXAMINATION AND COMPLAINED OF POOR WOUND HEALING, UNABSORBED SUTURE, AND THE APPEARANCE OF THE SUTURE CAME OUT FROM SKIN, AND THE HOSPITAL REMOVED THE UNABSORBED SUTURE AND ASKED FOR CLARIFICATION FROM COMPANY. CLINICAL FEEDBACK THAT IT MAYBE RELATED TO SUTURE ABSORPTION DEGRADATION TOO SLOW AND FOREIGN BODY REACTION.
Patient problems: Impaired Healing
MAUDE report 12672752 ↗InjuryOct 1, 2021Material Protrusion/Extrusion
POST-OP INFECTION. THIS CASE IS FROM HEALTH AUTHORITY. IT WAS REPORTED THAT THE PATIENT WAS TREATED WITH A JOHNSON & JOHNSON SURGICAL KNOT- FREE SUTURE FOR THE CAESAREAN SECTION INCISION SUTURE. A MONTH PLUS LATER, IT WAS FOUND THAT THE WOUND SEWED HAD BEEN INFECTED. THE THREAD TIP OF THE SUTURE HAD BEEN EXTRUDED OUT OF THE ABDOMINAL INCISION, WHERE ALSO HAD AN ULCER. THE INCISION WAS CURRENTLY HEALED AFTER CUTTING OFF THE EXTRUDED THREAD TIP. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: Ulcer; Unspecified Infection
MAUDE report 12561865 ↗InjuryOct 1, 2021Material Protrusion/Extrusion
POST-OP INFECTION. THIS CASE IS FROM HEALTH AUTHORITY. IT WAS REPORTED THAT THE PATIENT WAS TREATED WITH A JOHNSON & JOHNSON SURGICAL KNOT- FREE SUTURE FOR THE CAESAREAN SECTION INCISION SUTURE. THREE MONTHS PLUS LATER, IT WAS FOUND THAT THE WOUND SEWED HAD BEEN INFECTED. THE THREAD TIP OF THE SUTURE HAD BEEN EXTRUDED OUT OF THE ABDOMINAL INCISION, WHERE ALSO HAD AN ULCER. THE INCISION WAS CURRENTLY HEALED AFTER CUTTING OFF THE EXTRUDED THREAD TIP. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: Ulcer; Unspecified Infection
MAUDE report 12561809 ↗InjuryOct 1, 2021Material Protrusion/Extrusion
POST-OP INFECTION. THIS CASE IS FROM HEALTH AUTHORITY. IT WAS REPORTED THAT THE PATIENT WAS TREATED WITH A JOHNSON & JOHNSON SURGICAL KNOT- FREE SUTURE FOR THE CAESAREAN SECTION INCISION SUTURE. ABOUT TWO MONTHS PLUS LATER, IT WAS FOUND THAT THE WOUND SEWED HAD BEEN INFECTED. THE THREAD TIP OF THE SUTURE HAD BEEN EXTRUDED OUT OF THE ABDOMINAL INCISION, WHERE ALSO HAD AN ULCER. THE INCISION WAS CURRENTLY HEALED AFTER CUTTING OFF THE EXTRUDED THREAD TIP. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: Ulcer; Unspecified Infection
MAUDE report 12561774 ↗InjuryOct 1, 2021Material Protrusion/Extrusion
POST-OP INFECTION. THIS CASE IS FROM HEALTH AUTHORITY. IT WAS REPORTED THAT THE PATIENT WAS TREATED WITH A JOHNSON & JOHNSON SURGICAL KNOT- FREE SUTURE FOR THE CAESAREAN SECTION INCISION SUTURE. ABOUT TWO MONTHS PLUS LATER, IT WAS FOUND THAT THE WOUND SEWED HAD BEEN INFECTED. THE THREAD TIP OF THE SUTURE HAD BEEN EXTRUDED OUT OF THE ABDOMINAL INCISION, WHERE ALSO HAD AN ULCER. THE INCISION WAS CURRENTLY HEALED AFTER CUTTING
Patient problems: Ulcer; Unspecified Infection
MAUDE report 12561748 ↗InjuryOct 1, 2021Material Protrusion/Extrusion
POST-OP INFECTION. THIS CASE IS FROM HEALTH AUTHORITY. IT WAS REPORTED THAT THE PATIENT WAS TREATED WITH A JOHNSON & JOHNSON SURGICAL KNOT- FREE SUTURE FOR THE CAESAREAN SECTION INCISION SUTURE. ABOUT FOUR MONTHS PLUS LATER, IT WAS FOUND THAT THE WOUND SEWED HAD BEEN INFECTED. THE THREAD TIP OF THE SUTURE HAD BEEN EXTRUDED OUT OF THE ABDOMINAL INCISION, WHERE ALSO HAD AN ULCER. THE INCISION WAS CURRENTLY HEALED AFTER CUTTING OFF THE EXTRUDED THREAD TIP. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: Ulcer; Unspecified Infection
MAUDE report 12561682 ↗InjurySep 30, 2021Material Protrusion/Extrusion
POST-OP INFECTION. THIS CASE IS FROM HEALTH AUTHORITY. IT WAS REPORTED THAT THE PATIENT WAS TREATED WITH A JOHNSON & JOHNSON SURGICAL KNOT- FREE SUTURE FOR THE CAESAREAN SECTION INCISION SUTURE. ABOUT A MONTH PLUS LATER, IT WAS FOUND THAT THE WOUND SEWED HAD BEEN INFECTED. THE THREAD TIP OF THE SUTURE HAD BEEN EXTRUDED OUT OF THE ABDOMINAL INCISION, WHERE ALSO HAD AN ULCER. THE INCISION WAS CURRENTLY HEALED AFTER CUTTING OFF THE EXTRUDED THREAD TIP. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: Post Operative Wound Infection
MAUDE report 12558783 ↗
Other SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices
- STRATAFIX SXPP1A405Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE617 reports
- STRATAFIX SXPP1A400Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE485 reports
- STRATAFIX SXPP1A401Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE446 reports
- STRATAFIX SXPP1A301Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE443 reports
- STRATAFIX SXPP2B405Ethicon, LLC · Suture, surgical, absorbable, polydioxanone406 reports
- STRATAFIX SXPP1A406Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE339 reports
- STRATAFIX SXPP1A404Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 death319 reports
- STRATAFIX SXPP1B420Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 death229 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page