STRATAFIX SXPP1A403Recalls & reported problems
Ethicon, LLC · Violet Antibacterial Polydioxanone, Knotless Tissue Control Device · Class 2 · Implant
Quick summary
This Ethicon Stratafix knotless suture is an absorbable barbed suture used to close tissue without knots. Of its 171 reports, 13 are filed as injuries, and many describe the suture breaking or the needle pulling off during surgery. The reports mention no deaths, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 173
- Deaths reported
- 0
- Injuries reported
- 13
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 173 adverse event reports that list this exact device (DI 10705031219328), received between Dec 8, 2017 and Sep 30, 2026 — classified as 13 injuries, 160 malfunctions.
The most frequently reported specific device problem is “Break” (100 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break100
- Material Separation47
- Adverse Event Without Identified Device or Use Problem8
- Material Integrity Problem8
- Delivered as Unsterile Product5
- Appropriate Device Problem Term/Code Not Available4
- Packaging Problem4
- Device Contaminated During Manufacture or Shipping1
Patient problems
- No Clinical Signs, Symptoms or Conditions140
- No Consequences Or Impact To Patient22
- Unspecified Infection6
- Wound Dehiscence6
- Inflammation4
- Not Applicable4
- Erythema2
- Swelling2
Recent adverse event reports
The 25 most recent of 173 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50950668 ↗MalfunctionSep 24, 2026Break
It was reported that a patient underwent a Total Knee Arthroplasty procedure on (b)(6) 2026, and barbed suture was used. It was reported that the suture was broken when the surgeon attempted to sew the tissue. Changed another one to complete the surgery. There is no report on patient's injury. No additional information could be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50882768 ↗MalfunctionSep 15, 2026Material Integrity Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50779317 ↗MalfunctionAug 17, 2026Material Separation
It was reported that a patient underwent a heart valve procedure on an unknown date and suture was used. The problem of pulling off suture needle happened in surgery (PX83H/10D5R6 ). Changed to another one to continue the surgery (SXPP1A403/10933A ) , when suturing the fascial layer with suture during chest closure, the problem of pulling off suture needle happened. Changed to another one to complete the surgery. No adverse patient consequences were reported. Additional information was requested.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50287086 ↗MalfunctionAug 5, 2026Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50146102 ↗MalfunctionJul 27, 2026Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50011421 ↗MalfunctionJul 13, 2026Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25801902 ↗MalfunctionJun 1, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE BROKE TWICE DURING SURGERY AND CANNOT BE USED ANYMORE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THE LOT OF IP IS UNAVAILABLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25355523 ↗MalfunctionMay 21, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25240946 ↗MalfunctionMay 11, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A BYPASS PROCEDURE ON (B)(6) 2026 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED WHEN SEWING MUSCLE FASCIA IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25143357 ↗MalfunctionMay 4, 2026Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE WAS FOUND OXIDATION WHEN JUST OPEN THE PACKAGE. THE LOT OF IP IS UNAVAILABLE. ENCLOSED PLEASE FIND DEFECT PHOTO. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25072613 ↗MalfunctionMay 1, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING HANDLING PRIOR TO USE ON PATIENT, THE SUTURE PULLED OFF THE NEEDLE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. UPON EVALUATION OF THE RETURNED DEVICE, A SHORT INSERTION LENGTH WAS OBSERVED AT THE SUTURE END, AND NO REMNANT WAS FOUND WITHIN THE NEEDLE BARREL HOLE. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25058932 ↗MalfunctionApr 27, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON (B)(6) 2026 AND BARBED SUTURE WAS USED. THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24998018 ↗MalfunctionApr 20, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A TENSION REDUCING SUTURE SURGERY ON (B)(6) 2026 AND BARBED SUTURE WAS USED. DURING THE PATIENT'S SURGERY, THE DOCTOR PERFORMED A TENSION REDUCING SUTURE OPERATION, AND THE LEAD SURGEON USED SUTURE TO PERFORM SKIN SUTURING. WHEN THE SUTURE PENETRATES THE SKIN, THE TAIL END OF THE SUTURE SUDDENLY BREAKS. AFTER DISCOVERING ABNORMAL CONSUMABLES, MEDICAL STAFF IMMEDIATELY SUSPENDED THE OPERATION AND EVALUATED THE FRACTURE SITUATION TOGETHER WITH THE SURGEON. ACCORDING TO THE ON-SITE DOCTOR'S JUDGMENT, THE REMAINING USABLE LENGTH OF THE BROKEN SUTURE IS SUFFICIENT TO COMPLETE THE REMAINING SUTURING OPERATION, AND BY ADJUSTING THE SUTURING METHOD REASONABLY, THE SURGERY CAN BE COMPLETED WITHOUT INCREASING THE PATIENT'S ADDITIONAL TRAUMA AND COST. TO ENSURE SURGICAL SAFETY AND PATIENT INTERESTS, THE MEDICAL TEAM CONFIRMED THE FEASIBILITY OF THE SURGICAL PLAN ON SITE AND CONTINUED TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24926915 ↗MalfunctionApr 20, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND BARBED SUTURE WAS USED. IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY . CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24926626 ↗MalfunctionApr 15, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A HYSTERECTOMY SURGERY ON (B)(6) 2026 AND BARBED SUTURE WAS USED. THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY . CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24885427 ↗MalfunctionApr 8, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN CARDIOVASCULAR PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. DURING AN UNKNOWN CARDIOVASCULAR SURGERY, THE SUTURE WAS BROKEN DURING SUTURING THE TISSUE. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24818888 ↗MalfunctionApr 1, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND A BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE PULL OFF ISSUE HAPPENED WITH USE. THE DEVICE WAS REPLACED WITH THE SAME BATCH OF SUTURE FOR NORMAL USE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS EXCEEDS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24757505 ↗MalfunctionMar 25, 2026Break
IT WAS REPORTED A PATIENT UNDERWENT A MYOMECTOMY PROCEDURE ON (B)(6) 2026 AND BARBED SUTURE WAS USED. THE SUTURE GOT SPLIT WHEN IT WAS USED TO SUTURE THE WOUND, AND THEN THE SUTURE WAS BROKEN WHEN THE DOCTOR PULLED IT. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24685854 ↗MalfunctionMar 18, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT, THE SUTURE WAS BROKEN DURING USE. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24626803 ↗MalfunctionMar 13, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND BARBED SUTURE WAS USED. THE SUTURE GOT SPLIT IN THE SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24598262 ↗MalfunctionMar 13, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND BARBED SUTURE WAS USED. THE SUTURE GOT SPLIT IN THE SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24591634 ↗MalfunctionMar 5, 2026Material Separation
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND BARBED SUTURE WAS USED. THE PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THE NEEDLE DID NOT FELL INTO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24518121 ↗MalfunctionMar 5, 2026Material Separation
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND BARBED SUTURE WAS USED. THE PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THE NEEDLE DID NOT FELL INTO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24518105 ↗MalfunctionMar 5, 2026Material Separation
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(60 2026 AND BARBED SUTURE WAS USED. THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THE NEEDLE DID NOT FELL INTO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24518071 ↗
Other SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices
- STRATAFIX SXPP1A405Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE617 reports
- STRATAFIX SXPP1A400Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE485 reports
- STRATAFIX SXPP1A401Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE446 reports
- STRATAFIX SXPP1A301Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE443 reports
- STRATAFIX SXPP2B405Ethicon, LLC · Suture, surgical, absorbable, polydioxanone406 reports
- STRATAFIX SXPP1A406Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE339 reports
- STRATAFIX SXPP1A404Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 death319 reports
- STRATAFIX SXPP1B420Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 death229 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.