STRATAFIX SXPP1A409Recalls & reported problems

Ethicon, LLC · Violet Antibacterial Polydioxanone, Knotless Tissue Control Device · Class 2 · Implant

Adverse event reports
12
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10705031227729), received between Nov 16, 2018 and Sep 24, 2026 — classified as 5 injuries, 7 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem7
  • Break5
  • Positioning Failure2

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Impaired Healing3
  • Necrosis2
  • Purulent Discharge2
  • Swelling/ Edema2
  • Wound Dehiscence2
  • Abscess1
  • Cellulitis1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that a patient underwent a right total hip replacement surgery on (b)(6) 2026 and a barbed suture was used. The patient was admitted due to developmental dysplasia of the right hip joint and underwent total hip replacement surgery on the right side. The doctor used suture during the operation, and the procedure went smoothly. After the surgery, the patient did not exercise and clean the affected hip joint in a timely manner, and did not effectively protect the affected hip joint. The incision was not properly cleaned and protected. On (b)(6) 2026, the patient and their family came to the outpatient department of the hospital for further treatment. The outpatient department admitted the patient with a diagnosis of "poor wound healing after surgery". After admission, relevant examinations were completed, and symptomatic supportive treatment such as anti infection was given to the patient. On (b)(6) 2026, under general anesthesia, a right hip skin and subcutaneous necrotic tissue resection and debridement surgery, as well as a right hip partial prosthesis revision surgery, were performed. The surgical process was smooth and the patient's condition was steadily controll

    Patient problems: Impaired Healing; Necrosis

    MAUDE report 50889589 ↗
  • InjurySep 18, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Cellulitis; Discomfort; Purulent Discharge

    MAUDE report 50820432 ↗
  • MalfunctionJun 2, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ORTHOPEDIC PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. DURING AN UNKNOWN ORTHOPEDIC SURGERY, THE SUTURES WERE BROKEN REPEATEDLY. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO SAMPLE WILL BE RETURNED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25365901 ↗
  • InjuryApr 17, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LUMBAR SPINE POSTERIOR ROBOT ASSISTED DISCECTOMY, DECOMPRESSION, BONE GRAFT FUSION, AND INTERNAL FIXATION SURGERY UNDER GENERAL ANESTHESIA ON (B)(6) 2026 AND A BARBED SUTURE WAS USED. THE PATIENT WAS ADMITTED WITH LUMBAR DISC HERNIATION AND UNDERWENT ROUTINE PREOPERATIVE EXAMINATIONS TO RULE OUT SURGICAL CONTRAINDICATIONS. THE PATIENT UNDERWENT SURGERY AND THE DOCTOR USED A FUSION CAGE FOR INTERBODY BONE GRAFTING DURING THE OPERATION. INTRAOPERATIVE USE OF FLUID GELATIN TO STOP BLEEDING. THE PATIENT WAS GIVEN NUTRITIONAL SUPPORT AND ANTI INFECTIVE TREATMENT AFTER SURGERY, AND REMOVED THE WOUND DRAINAGE TUBE WHEN THE DRAINAGE VOLUME DECREASES. POST-OP, THE PATIENT EXPERIENCED POOR WOUND HEALING AND WOUND SECRETION. ON THE (B)(6) DAY AFTER SURGERY, THE PATIENT FOUND A SMALL AMOUNT OF YELLOW WHITE PURULENT FLUID OOZING OUT FROM THE WOUND DURING DRESSING CHANGE, AND THE WOUND PARTIALLY CRACKED, WITH POOR WOUND HEALING. ON THE (B)(6) DAY AFTER SURGERY, THE PATIENT WAS READMITTED AND THE DOCTOR PRESCRIBED DAILY DRESSING CHANGES. BASED ON THE RESULTS OF METAGENOMIC TESTING, THE PATIENT CONTINUED WITH ANTI INFECTIVE TREATMENT. ADJUST THE TREATMEN

    Patient problems: Impaired Healing; Purulent Discharge; Wound Dehiscence

    MAUDE report 24912098 ↗
  • MalfunctionAug 6, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE PROCEDURE ON (B)(6) 2025 AND A BARBED SUTURE WAS USED. DURING THE PROCEDURE, WHILE SUTURING THE SUBCUTANEOUS LAYER, THE NEEDLE BENT AFTER PROBABLY 10 CM. AND THE TIP BROKE WHILE MANIPULATION WITH NEEDLE HOLDER. THE SURGERY WAS NOT DELAYED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22726455 ↗
  • MalfunctionAug 6, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE PROCEDURE ON (B)(6) 2025 AND A BARBED SUTURE WAS USED. DURING THE PROCEDURE, WHILE SUTURING THE SUBCUTANEOUS LAYER, THE NEEDLE BENT AFTER PROBABLY 10 CM. AND THE TIP BROKE WHILE MANIPULATION WITH NEEDLE HOLDER. THE SURGERY WAS NOT DELAYED. ANOTHER LIKE DEVICE WAS USED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22726451 ↗
  • MalfunctionJul 16, 2025Positioning Failure; Adverse Event Without Identified Device or Use Problem; Positioning Failure; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PLASTIC SURGERY ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. THE PATIENT DID THREAD LIFTING SURGERY, AND FEEDBACK POSTOPERATIVE FACIAL SWELLING, UNEVEN AND SUSPECT THE BARB WAS NOT HANGED INTO TISSUE. AND SUSPECT THE SUTURE IS SXPP1A409 OR SXPP1A101. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Swelling/ Edema

    MAUDE report 22523244 ↗
  • MalfunctionJul 16, 2025Positioning Failure; Adverse Event Without Identified Device or Use Problem; Positioning Failure; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PLASTIC SURGERY ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. THE PATIENT DID THREAD LIFTING SURGERY, AND FEEDBACK POSTOPERATIVE FACIAL SWELLING, UNEVEN AND SUSPECT THE BARB WAS NOT HANGED INTO TISSUE. AND SUSPECT THE SUTURE IS SXPP1A409 OR SXPP1A101. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Swelling/ Edema

    MAUDE report 22523199 ↗
  • InjuryJan 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT TREATMENT FOR A FRACTURE OF THE RIGHT TIBIAL PLATEAU ON (B)(6)2024 AND BARBED SUTURE WAS USED. THE PATIENT WAS DISCHARGED SMOOTHLY. ON THE 9TH DAY AFTER SURGERY, THE PATIENT WAS READMITTED DUE TO POOR WOUND HEALING. ON THE 17TH DAY AFTER SURGERY, A NECROTIC TISSUE RESECTION AND DEBRIDEMENT WERE PERFORMED, AND THE BLOODY EXUDATE WAS REDUCED. THE WOUND WAS NOT SUTURED AND THE PATIENT WAS DISCHARGED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Fluid Discharge; Hemorrhage/Blood Loss/Bleeding; Impaired Healing; Necrosis

    MAUDE report 21171678 ↗
  • MalfunctionJan 30, 2023Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A HERNIA REPAIR PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. DURING USE ON THE PATIENT IT WAS NOTED THAT THE NEEDLE BROKE. THE NEEDLE BENT AND GOT BROKEN WHILE USING IN ABDOMINAL WALL AT PERITONEUM. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 16264631 ↗
  • MalfunctionApr 13, 2022Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN CARDIOVASCULAR PROCEDURE ON (B)(6) 2022 AND BARBED SUTURE WAS USED. THE NEEDLE WAS BROKEN AT THE MIDDLE OF THE BODY PART DURING CONTINUOUS SUTURING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14106277 ↗
  • InjuryNov 16, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT CRANIOTOMY BRAIN TUMOR RESECTION PROCEDURE ON (B)(6) 2018 AND BARBED SUTURE WAS USED DURING CLOSURE. THE PATIENT WAS DISCHARGED (B)(6) FROM THE FACILITY. THE PATIENT EXPERIENCED UNKNOWN SYMPTOMS AND WAS DIAGNOSED WITH AN ABSCESS ON UNKNOWN DATE AND WOUND DEHISCENCE. A SECOND PROCEDURE WAS PERFORMED ON UNKNOWN DATE BETWEEN (B)(6) AND (B)(6). THE SURGEON REPORTED THAT THE ABSCESS DEVELOPED ON THE TAB PORTION OF THE SUTURE. THE SUTURE WAS REMOVED AND IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. THE PATIENT RECEIVED ANTIBIOTICS AND IS REPORTED AS NOT GETTING WORSE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Abscess; Not Applicable; Wound Dehiscence

    MAUDE report 8079183 ↗

Other SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices

All SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.