This Ethicon Stratafix barbed suture is a violet, absorbable surgical thread used to close soft tissue and skin without knots. The 38 reports mostly describe the thread breaking during surgery or the needle pulling away from the thread, so surgeons switched to another suture to finish the operation; a few mention the wound reopening slightly days later. There are 3 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
39
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 39 adverse event reports that list this exact device (DI 10705031227743), received between Mar 5, 2018 and Aug 20, 2026 — classified as 3 injuries, 36 malfunctions.

The most frequently reported specific device problem is “Break” (31 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 3 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 4 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break31
  • Material Separation7
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions24
  • No Consequences Or Impact To Patient12
  • Wound Dehiscence3
  • Foreign Body Reaction2

Recent adverse event reports

The 25 most recent of 39 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 20, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50348573 ↗
  • MalfunctionJan 13, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT WOUND CLOSURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. THE SUTURE BROKE WHEN SEWING IN THE SURGERY. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24060080 ↗
  • MalfunctionJan 13, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT WOUND CLOSURE IN (B)(6) 2025 AND BARBED SUTURE WAS USED. THE SUTURE BROKE WHEN SEWING IN THE SURGERY. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24059734 ↗
  • MalfunctionJan 13, 2026Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT WOUND CLOSURE IN (B)(6) 2025 AND BARBED SUTURE WAS USED. THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24057104 ↗
  • MalfunctionJan 13, 2026Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT WOUND CLOSURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY DURING USE ON THE PATIENT. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24056957 ↗
  • InjuryMay 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY ON (B)(6) 2025 AND BARBED SUTURE WAS USED ON THE DERMIS. THE WOUND WHERE THE SUTURES HAD BEEN PERFORMED WAS OPENED A LITTLE A FEW DAYS AFTER THE SURGERY. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THE PATIENT IS BEING FOLLOWED UP. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Wound Dehiscence

    MAUDE report 21971743 ↗
  • MalfunctionDec 6, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY ON (B)(6) 2024 AND BARBED SUTURE WAS USED. DURING A TOTAL KNEE ARTHROPLASTY: TKA, WHEN THE SUTURE WHICH WAS USED FOR THE SUBCUTANEOUS OF THE KNEE WAS USED AS A FIXED SUTURING, THE SUTURE (THE BASE OF THE TAB) WAS BROKEN. THE STRENGTH OF THE TRACTION AND OTHER DETAILS ARE UNKNOWN. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20864163 ↗
  • MalfunctionNov 21, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN ORTHOPEDIC PROCEDURE ON (B)(6) 2024 AND BARBED SUTURE WAS USED. DURING AN UNKNOWN ORTHOPEDIC SURGERY, A SUTURE BREAKAGE OCCURRED ON (B)(6). THE DOCTOR SAID THAT THE SUTURE PART NEAR THE FIXATION TAB AND NEAR THE NEEDLE BROKE AND NOT USED IMPROPERLY, SUCH AS STRONG TRACTION. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20749967 ↗
  • MalfunctionNov 21, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY PROCEDURE ON (B)(6) 2024 AND BARBED SUTURE WAS USED. DURING A TOTAL KNEE ARTHROPLASTY: TKA, ON BOTH (B)(6), WHEN IT WAS FIXED AND SUTURED BY THE SUTURE USED ON THE SUBCUTANEOUS OF THE KNEE, THE SUTURE (OR THE BASE OF THE FIXATION TAB) WAS BROKEN. THE STRENGTH OF THE TRACTION ETC. IS UNKNOWN. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20749962 ↗
  • MalfunctionNov 21, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY PROCEDURE ON (B)(6) 2024 AND BARBED SUTURE WAS USED. DURING A TOTAL KNEE ARTHROPLASTY: TKA, ON BOTH (B)(6), WHEN IT WAS FIXED AND SUTURED BY THE SUTURE USED ON THE SUBCUTANEOUS OF THE KNEE, THE SUTURE (OR THE BASE OF THE FIXATION TAB) WAS BROKEN. THE STRENGTH OF THE TRACTION ETC. IS UNKNOWN. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20749958 ↗
  • MalfunctionJul 24, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN ORTHOPEDIC SURGERY ON (B)(6) 2024 AND A BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE CAME OFF FROM THE NEEDLE WITH USE. IT WAS A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19822013 ↗
  • MalfunctionSep 28, 2023Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TKA: TOTAL KNEE ARTHROPLASTY PROCEDURE ON (B)(6) 2023 AND BARBED SUTURE WAS USED. THE SUTURE WAS BROKEN AROUND THE SWAGE PART DURING CONTINUOUS SUTURING ON SUBCUTANEOUS TISSUE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17835205 ↗
  • MalfunctionJun 19, 2023Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ORTHOPEDIC PROCEDURE ON (B)(6) 2023 AND A BARBED SUTURE WAS USED. A CROSS STITCH WAS BROKEN DURING USE. THE PRODUCT WAS USED ON SUBCUTANEOUS DERMIS. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17159926 ↗
  • MalfunctionMar 31, 2023Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ORTHOPEDIC PROCEDURE ON (B)(6) 2022, AND BARBED SUTURE WAS USED. THE SUTURE BROKE 1~2MM BEFORE THE FIXATION TAB DURING CONTINUOUS SUTURE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16654791 ↗
  • MalfunctionMar 8, 2023Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE REPLACEMENT PROCEDURE ON (B)(6)2023 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, WHEN PUTTING THE FIRST CROSS STITCH ON FASCIA, THE FIXATION TAB WAS BROKEN. ANOTHER SUTURE WAS USED TO CLOSE THE WOUND. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16501650 ↗
  • InjuryNov 4, 2022Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A BILATERAL TOTAL KNEE ARTHROPLASTY ON (B)(6) 2022 AND A BARBED SUTURE WAS USED FOR SUBCUTANEOUS DERMAL SUTURE. POST-OP, TWO MONTHS LATER, THE PATIENT EXPERIENCED A SYMPTOM THAT SEEMED TO BE A FOREIGN SUBSTANCE REACTION. THE WOUND BECAME WHITE AND DEHISCENT. THE SUTURE BROKE AND DEHISCENCE OCCURRED. THEREFORE, REMOVAL OF THE SUTURE AND EPIDERMAL RE-SUTURING WAS PERFORMED WITH NYLON. THE AREA WHERE THE SUTURE WAS REMOVED WAS RECOVERING. THE PATIENT IS BEING FOLLOWED UP ON AN OUTPATIENT BASIS AND IS DOING WELL. IT WAS OPINED THAT SINCE THE SYMPTOMS IMPROVED AFTER REMOVAL, THERE IS A POSSIBILITY THAT THE DEVICE WAS THE CAUSE OF THE EVENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Foreign Body Reaction; Wound Dehiscence

    MAUDE report 15734093 ↗
  • InjuryNov 4, 2022Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A BILATERAL TOTAL KNEE ARTHROPLASTY ON (B)(6) 2022 AND A BARBED SUTURE WAS USED FOR SUBCUTANEOUS DERMAL SUTURE. POST-OP, TWO MONTHS LATER, THE PATIENT EXPERIENCED A SYMPTOM THAT SEEMED TO BE A FOREIGN SUBSTANCE REACTION. THE WOUND BECAME WHITE AND DEHISCENT. THE SUTURE BROKE AND DEHISCENCE OCCURRED. THEREFORE, REMOVAL OF THE SUTURE AND EPIDERMAL RE-SUTURING WAS PERFORMED WITH NYLON. THE AREA WHERE THE SUTURE WAS REMOVED WAS RECOVERING. THE PATIENT IS BEING FOLLOWED UP ON AN OUTPATIENT BASIS AND IS DOING WELL. IT WAS OPINED THAT SINCE THE SYMPTOMS IMPROVED AFTER REMOVAL, THERE IS A POSSIBILITY THAT THE DEVICE WAS THE CAUSE OF THE EVENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Foreign Body Reaction; Wound Dehiscence

    MAUDE report 15732716 ↗
  • MalfunctionJul 25, 2022Material Separation; Material Separation

    IT WAS REPORTED BY A SALES REP THAT, DURING AN UNKNOWN CARDIOVASCULAR SURGERY, THE SUTURE WAS DETACHED FROM THE NEEDLE DURING SUTURING. IT WAS NOT A CONTROL RELEASE NEEDLE. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. WHEN WE SEND THE SAMPLE TO YOU, WE WILL LET YOU KNOW ITS RETURN DATE AND TRACKING NUMBER. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. AFFILIATE REFERENCE NUMBER IS (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15100425 ↗
  • MalfunctionMay 5, 2022Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. THE SUTURE BREAKAGE OCCURRED DURING CONTINUOUS SUTURING OF SUBCUTANEOUS AND DERMAL TISSUES TOGETHER. THE SUTURE WAS BROKEN AT THE END OF SUTURING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14287054 ↗
  • MalfunctionNov 7, 2021Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE CONNECTION PART BETWEEN THE SUTURE AND THE NEEDLE BROKE DURING USE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12760628 ↗
  • MalfunctionJun 16, 2021Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY ON AN UNKNOWN DATE AND A BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE WHILE USE ON THE SUBCUTIS. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12007662 ↗
  • MalfunctionJun 16, 2021Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY ON AN UNKNOWN DATE AND A BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE WHILE USE ON THE SUBCUTIS. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12007658 ↗
  • MalfunctionApr 9, 2021Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE REPLACEMENT PROCEDURE ON (B)(6) 2021 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT SUTURE BREAKAGE OCCURRED 2 TIMES IN A ROW DURING CONTINUOUS SUTURING ON THE SUBCUTIS. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11641806 ↗
  • MalfunctionApr 9, 2021Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE REPLACEMENT PROCEDURE ON (B)(6) 2021 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT SUTURE BREAKAGE OCCURRED 2 TIMES IN A ROW DURING CONTINUOUS SUTURING ON THE SUBCUTIS. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11641804 ↗
  • MalfunctionMar 17, 2021Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ORTHOPEDIC PROCEDURE ON (B)(6) 2021 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, TWO (2) SUTURES BROKE IN A ROW DURING WOUND CLOSURE THOUGH THEY WERE USED WITH THE USUAL WAY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11509660 ↗