STRATAFIX SXPP1B409Recalls & reported problems

Ethicon, LLC · Knotless Tissue Control Device, Violet Unidirectional (Antibacterial) Synthetic Absorbable Device · Class 2 · Implant

All models of STRATAFIX (181)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Ethicon STRATAFIX barbed suture is a knotless, absorbable violet surgical suture used to close tissue without tying knots. The 48 reports mostly describe the suture breaking during surgery, the barbs not holding so the suture slipped back or unraveled while a wound was being closed, and the needle detaching from the suture. Six reports involve injuries; no deaths were reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
49
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 49 adverse event reports that list this exact device (DI 10705031236127), received between Mar 7, 2018 and Sep 16, 2026 — classified as 6 injuries, 43 malfunctions.

The most frequently reported specific device problem is “Break” (27 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 3 reportsFeb 2024: 0 reportsMar 2024: 3 reportsApr 2024: 5 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 6 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 3 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break27
  • Positioning Failure13
  • Material Separation5
  • Adverse Event Without Identified Device or Use Problem2
  • Defective Component2
  • Positioning Problem2
  • Component Misassembled1
  • Failure To Adhere Or Bond1

Patient problems

  • No Clinical Signs, Symptoms or Conditions39
  • No Consequences Or Impact To Patient5
  • Wound Dehiscence3
  • Insufficient Information2
  • Not Applicable2
  • Unspecified Infection2
  • Sepsis1

Recent adverse event reports

The 25 most recent of 49 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 16, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50794380 ↗
  • MalfunctionApr 8, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND BARBED SUTURE WAS USED. BARBED SUTURE BREAKAGE WHILE DOING THE SURGERY. THERE WAS A 25 MINUTE SURGICAL DELAY REPORTED. THERE WERE NO PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24820193 ↗
  • MalfunctionApr 8, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND BARBED SUTURE WAS USED. BARBED SUTURE BREAKAGE WHILE DOING THE SURGERY. THERE WAS A 25 MINUTES SURGICAL DELAY REPORTED. THERE WERE NO PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24820192 ↗
  • MalfunctionApr 8, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND BARBED SUTURE WAS USED. BARBED SUTURE BREAKAGE WHILE DOING THE SURGERY. THERE WAS A 25 MINUTES SURGICAL DELAY REPORTED. THERE WERE NO PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24820191 ↗
  • MalfunctionNov 25, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE STRATAFIX SUTURE BROKE WHILE THE WOUND WAS BEING CLOSED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23647412 ↗
  • MalfunctionNov 27, 2024Positioning Failure; Positioning Failure

    IT WAS REPORTED A PATIENT UNDERWENT TOTAL LAPAROSCOPIC HYSTERECTOMY PROCEDURE ON AN UNKNOWN DATE IN 2024 AND BARBED SUTURE WAS USED. DURING A TLH AND WHEN CLOSING THE VAGINAL VAULT SURGEON FOUND THAT THE BARBS WEREN¿T HOLDING AND ACTUALLY THE SUTURE SLIPPED BACK. NO REPORTED PATIENT CONSEQUENCES. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20796156 ↗
  • MalfunctionNov 26, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDUR EON (B)(6) 2024 AND SUTURE WAS USED. WHILE SUTURING THE NEEDLE GETS DETACHED FROM THE SWAGING END. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20787033 ↗
  • MalfunctionOct 22, 2024Positioning Failure; Positioning Failure

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC VAGINAL HYSTERECTOMY PROCEDURE IN (B)(6) 2024 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE SURGEON CLOSING VAULT WITH STRATAFIX & AT TERMINATION OF SUTURING THE VAULT JUST 'UNRAVELED'. IT WAS LIKE THERE WAS NO ANCHORS ON THE SUTURE. SURGEON COMPLETED CLOSURE OF THE VAULT WITH AN UNINTERRUPTED VICRYL SUTURE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Wound Dehiscence

    MAUDE report 20505696 ↗
  • MalfunctionAug 8, 2024Positioning Failure; Positioning Failure

    IT WAS REPORTED A PATIENT UNDERWENT A LAP MYOMECTOMY ON (B)(6) 2024 AND BARBED SUTURE WAS USED. AS PER HCP SUTURE IS NOT SUPPORTING HOLDING STRENGTH DURING MYOMECTOMY PROCEDURE. BARB NOT HOLDING TISSUE PROPERLY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19937809 ↗
  • MalfunctionAug 8, 2024Positioning Failure; Positioning Failure

    IT WAS REPORTED A PATIENT UNDERWENT A LAP MYOMECTOMY ON (B)(6) 2024 AND BARBED SUTURE WAS USED. AS PER HCP SUTURE IS NOT SUPPORTING HOLDING STRENGTH DURING MYOMECTOMY PROCEDURE. BARB NOT HOLDING TISSUE PROPERLY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19937806 ↗
  • MalfunctionAug 8, 2024Positioning Failure; Positioning Failure

    IT WAS REPORTED A PATIENT UNDERWENT A LAP MYOMECTOMY ON (B)(6) 2024 AND BARBED SUTURE WAS USED. AS PER HCP SUTURE IS NOT SUPPORTING HOLDING STRENGTH DURING MYOMECTOMY PROCEDURE. BARB NOT HOLDING TISSUE PROPERLY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19937799 ↗
  • MalfunctionAug 8, 2024Positioning Failure; Positioning Failure

    IT WAS REPORTED A PATIENT UNDERWENT A LAP MYOMECTOMY ON (B)(6) 2024 AND BARBED SUTURE WAS USED. AS PER HCP SUTURE IS NOT SUPPORTING HOLDING STRENGTH DURING MYOMECTOMY PROCEDURE. BARB NOT HOLDING TISSUE PROPERLY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19937798 ↗
  • MalfunctionAug 8, 2024Positioning Failure; Positioning Failure

    IT WAS REPORTED A PATIENT UNDERWENT A LAP MYOMECTOMY ON (B)(6) 2024 AND BARBED SUTURE WAS USED. AS PER HCP SUTURE IS NOT SUPPORTING HOLDING STRENGTH DURING MYOMECTOMY PROCEDURE. BARB NOT HOLDING TISSUE PROPERLY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19937797 ↗
  • MalfunctionAug 8, 2024Positioning Failure; Positioning Failure

    IT WAS REPORTED A PATIENT UNDERWENT A LAP MYOMECTOMY ON (B)(6) 2024 AND BARBED SUTURE WAS USED. AS PER HCP SUTURE IS NOT SUPPORTING HOLDING STRENGTH DURING MYOMECTOMY PROCEDURE. BARB NOT HOLDING TISSUE PROPERLY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19937789 ↗
  • MalfunctionApr 12, 2024Break; Break

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A LAPAROSCOPIC MYOMECTOMY ON (B)(6) 2024 AND BARBED SUTURE WAS USED. THE SUTURES HAD BROKEN DURING SURGERY. THERE WAS NO DELAY DUE TO THE EVENT, AND THE FRAGMENTS GENERATED WERE EASILY REMOVED WITHOUT REQUIRING ADDITIONAL INTERVENTION. NO PATIENT CONSEQUENCE WAS REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19100956 ↗
  • MalfunctionApr 12, 2024Break; Break

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A LAPAROSCOPIC MYOMECTOMY ON (B)(6) 2024 AND BARBED SUTURE WAS USED. THE SUTURES HAD BROKEN DURING SURGERY. THERE WAS NO DELAY DUE TO THE EVENT, AND THE FRAGMENTS GENERATED WERE EASILY REMOVED WITHOUT REQUIRING ADDITIONAL INTERVENTION. NO PATIENT CONSEQUENCE WAS REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19100215 ↗
  • MalfunctionApr 12, 2024Break; Break

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A LAPAROSCOPIC MYOMECTOMY ON (B)(6) 2024 AND BARBED SUTURE WAS USED. THE SUTURES HAD BROKEN DURING SURGERY. THERE WAS NO DELAY DUE TO THE EVENT, AND THE FRAGMENTS GENERATED WERE EASILY REMOVED WITHOUT REQUIRING ADDITIONAL INTERVENTION. NO PATIENT CONSEQUENCE WAS REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19100176 ↗
  • MalfunctionApr 12, 2024Break; Break

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A LAPAROSCOPIC MYOMECTOMY ON (B)(6) 2024 AND BARBED SUTURE WAS USED. THE SUTURES HAD BROKEN DURING SURGERY. THERE WAS NO DELAY DUE TO THE EVENT, AND THE FRAGMENTS GENERATED WERE EASILY REMOVED WITHOUT REQUIRING ADDITIONAL INTERVENTION. NO PATIENT CONSEQUENCE WAS REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19100156 ↗
  • MalfunctionApr 12, 2024Break; Break

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A LAPAROSCOPIC MYOMECTOMY ON (B)(6) 2024 AND BARBED SUTURE WAS USED. THE SUTURES HAD BROKEN DURING SURGERY. THERE WAS NO DELAY DUE TO THE EVENT, AND THE FRAGMENTS GENERATED WERE EASILY REMOVED WITHOUT REQUIRING ADDITIONAL INTERVENTION. NO PATIENT CONSEQUENCE WAS REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19100152 ↗
  • InjuryMar 20, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A MYOMECTOMY PROCEDURE ON (B)(6) 2024 AND BARBED SUTURE WAS USED. THE SUTURE NEEDLE BROKE DURING THE PROCEDURE. THE CASE WAS MOVED TO AN OPEN PROCEDURE. THE PROCEDURE WAS EXTENDED BY 50 MINUTES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 18944698 ↗
  • InjuryMar 20, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A MYOMECTOMY PROCEDURE ON (B)(6) 2024 AND BARBED SUTURE WAS USED. THE SUTURE NEEDLE BROKE DURING THE PROCEDURE. PIECES MANAGED TO RETRIEVE WITHIN LAP MODE. THE NEEDLE FRAGMENTS WERE REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION. THE SURGERY WAS DELAYED 20 MINUTES DUE TO THIS EVENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18944697 ↗
  • InjuryMar 20, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A MYOMECTOMY PROCEDURE ON (B)(6) 2024 AND BARBED SUTURE WAS USED. THE SUTURE NEEDLE BROKE DURING THE PROCEDURE. XRAY HAD TO BE USED AND PROCEDURE MOVED TO OPEN. THE PROCEDURE WAS EXTENDED BY 2 HOURS. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 18943077 ↗
  • MalfunctionJan 31, 2024Positioning Failure; Positioning Failure

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN VAGINAL PROCEDURE ON (B)(6) 2024 AND BARBED SUTURE WAS USED. DURING THE CLOSURE OF THE VAGINAL DOME, THE SURGEON STATES THAT THE SUTURE DEVICE DID NOT ANCHOR IN THE TISSUE, SHOWING THAT THE THREAD DOES NOT HAVE ANCHORS FORMED. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18615772 ↗
  • MalfunctionJan 29, 2024Positioning Failure; Defective Component; Positioning Failure; Defective Component

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2023 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE DOCTOR REPORTS THAT THE SPIRAL STRATAFIX DEVICE HAS NO ANCHORS THAT ALLOW IT TO HOLD ON TO THE TISSUE "IT'S LIKE IT'S USING A NORMAL PDS". NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18594394 ↗
  • MalfunctionJan 29, 2024Positioning Failure; Defective Component; Positioning Failure; Defective Component

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON 05/01/2023 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE DOCTOR REPORTS THAT THE SPIRAL STRATAFIX DEVICE HAS NO ANCHORS THAT ALLOW IT TO HOLD ON TO THE TISSUE "IT'S LIKE IT'S USING A NORMAL PDS". NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18594392 ↗

Other SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices

All SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices →