Skip to content

STRATAFIX SXPP1B411Recalls & reported problems

Ethicon, LLC · Knotless Tissue Control Device, Violet Unidirectional (Antibacterial) Synthetic Absorbable Device · Class 2 · Implant

Quick summary

This Ethicon StrataFix barbed suture is an absorbable surgical thread used to close tissue without knots. The 34 reports mostly describe the suture breaking, fraying or coming apart from the needle during surgery, along with wound reopening, redness along the suture tracks, small abscesses, and infection. There were23 injuries and no deaths, and no recalls name this device. Reports dropped sharply, from 12 in the earlier 12 months to 2 in the most recent.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
37
Deaths reported
0
Injuries reported
25
Recalls naming it
0

Summary

The FDA's MAUDE database contains 37 adverse event reports that list this exact device (DI 10705031236141), received between Mar 5, 2019 and Sep 30, 2026 — classified as 25 injuries, 12 malfunctions.

The most frequently reported specific device problem is “Break” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 5 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 5 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 3 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem20
  • Break9
  • Material Integrity Problem5
  • Material Separation3
  • Delivered as Unsterile Product1
  • Positioning Problem1

Patient problems

  • Wound Dehiscence12
  • Unspecified Infection10
  • No Clinical Signs, Symptoms or Conditions9
  • Abscess6
  • Purulent Discharge6
  • Hypersensitivity/Allergic reaction5
  • Insufficient Information5
  • Skin Inflammation/ Irritation5

Recent adverse event reports

The 25 most recent of 37 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Unspecified Infection

    MAUDE report 50947350 ↗
  • InjurySep 30, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Unspecified Infection

    MAUDE report 50947333 ↗
  • MalfunctionSep 28, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923737 ↗
  • InjuryFeb 20, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PANNICULECTOMY PROCEDURE ON (B)(6) 2026 DUE TO EXCESSIVE PANNICULUS AND INTERTRIGO CANDIDIASIS AND BARBED SUTURE WAS USED ON THE FAT AND SKIN. POST-OP, THE PATIENT EXPERIENCED A SUBCLINICAL HEMATOMA. THE PATIENT WENT HOME POST OP DAY 1, REMOVED DRAIN, SKIN DIED FOLLOWING WEEK. THE SKIN DIED, ON POST OP DAY 10 THE PATIENT CAME TO ER AND UNDERWENT WOUND DEBRIDEMENT. THE SUTURE BEGAN BREAKING DOWN DURING THE SECOND PROCEDURE. THE PATIENT HAS RECOVERED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Necrosis

    MAUDE report 24400378 ↗
  • InjuryFeb 20, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PANNICULECTOMY PROCEDURE ON (B)(6) 2026 DUE TO INTERTRIGO CANDIDIASIS AND BARBED SUTURE WAS USED ON THE FAT. POST-OP, THE PATIENT EXPERIENCED A HEMATOMA/HEMORRHAGE. THE PATIENT WAS GIVEN A BLOOD TRANSFUSION ON POST OP DAY 1 AND DEVELOPED RAPID RESPONSE HYPERTENSION AND WAS TRANSFERRED TO STEP DOWN. THE PATIENT HAS RECOVERED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; High Blood Pressure/ Hypertension

    MAUDE report 24400368 ↗
  • InjuryJun 26, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY ON AN UNKNOWN DATE AND A BARBED SUTURE WAS USED. AT 2-3 MONTHS POST-OP, THE PATIENT EXPERIENCED IRRITATION, SMALL ABSCESS, REDNESS AND INFECTION FOLLOWING SUTURE TRACK MARKS. THE REDNESS FOLLOWS THE SUTURE BITE TRACK AND THE SURGEON QUESTIONED IF IT IS A PATIENT ALLERGY OR SOMETHING ELSE. THE SURGEON HAS GONE BACK TO USING A DIFFERENT SUTURE TO CLOSE THE SUBCUTANEOUS LAYER. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Abscess; Erythema; Hypersensitivity/Allergic reaction; Skin Inflammation/ Irritation; Unspecified Infection

    MAUDE report 22341256 ↗
  • InjuryJun 26, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY ON AN UNKNOWN DATE AND A BARBED SUTURE WAS USED. AT 2-3 MONTHS POST-OP, THE PATIENT EXPERIENCED IRRITATION, SMALL ABSCESS, REDNESS AND INFECTION FOLLOWING SUTURE TRACK MARKS. THE REDNESS FOLLOWS THE SUTURE BITE TRACK AND THE SURGEON QUESTIONED IF IT IS A PATIENT ALLERGY OR SOMETHING ELSE. THE SURGEON HAS GONE BACK TO USING A DIFFERENT SUTURE TO CLOSE THE SUBCUTANEOUS LAYER. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Abscess; Erythema; Hypersensitivity/Allergic reaction; Skin Inflammation/ Irritation; Unspecified Infection

    MAUDE report 22341254 ↗
  • InjuryJun 26, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY ON AN UNKNOWN DATE AND A BARBED SUTURE WAS USED. AT 2-3 MONTHS POST-OP, THE PATIENT EXPERIENCED IRRITATION, SMALL ABSCESS, REDNESS AND INFECTION FOLLOWING SUTURE TRACK MARKS. THE REDNESS FOLLOWS THE SUTURE BITE TRACK AND THE SURGEON QUESTIONED IF IT IS A PATIENT ALLERGY OR SOMETHING ELSE. THE SURGEON HAS GONE BACK TO USING A DIFFERENT SUTURE TO CLOSE THE SUBCUTANEOUS LAYER. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Abscess; Erythema; Hypersensitivity/Allergic reaction; Skin Inflammation/ Irritation; Unspecified Infection

    MAUDE report 22341253 ↗
  • InjuryJun 26, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A BARBED SUTURE WAS USED. AT 2-3 MONTHS POST-OP, THE PATIENT EXPERIENCED IRRITATION, SMALL ABSCESS, REDNESS AND INFECTION FOLLOWING SUTURE TRACK MARKS. THE REDNESS FOLLOWS THE SUTURE BITE TRACK AND THE SURGEON QUESTIONED IF IT IS A PATIENT ALLERGY OR SOMETHING ELSE. THE SURGEON HAS GONE BACK TO USING A DIFFERENT SUTURE TO CLOSE THE SUBCUTANEOUS LAYER. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Abscess; Erythema; Hypersensitivity/Allergic reaction; Skin Inflammation/ Irritation; Unspecified Infection

    MAUDE report 22341252 ↗
  • MalfunctionJun 2, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2025 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, WHEN SURGEON WENT TO USE THE SUTURE THE THREAD DETACHED FROM THE NEEDLE. HAD TO OPEN ANOTHER SUTURE TO COMPLETE CLOSURE. THERE WERE NO ADVERSE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22127875 ↗
  • MalfunctionApr 1, 2025Break; Break

    IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. "THE SUTURE CONTINUED TO BREAK WHEN THE SURGEON WAS TRYING A TIE A KNOT. HAD TO OPEN ANOTHER PACKAGE." THERE WERE NO ADVERSE PATIENT CONSEQUENCES. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21741280 ↗
  • InjuryJan 27, 2025Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT HIP SURGERY ON AN UNKNOWN DATE AND A BARBED SUTURE WAS USED. POST-OP, THE PATIENT EXPERIENCED AN ISSUE WITH NON-ABSORPTION OF THE SUTURE IN THE SUBCUTICULAR LAYER OF TISSUE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Insufficient Information; Wound Dehiscence

    MAUDE report 21241729 ↗
  • InjuryJan 27, 2025Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT HIP SURGERY ON AN UNKNOWN DATE AND A BARBED SUTURE WAS USED. POST-OP, THE PATIENT EXPERIENCED AN ISSUE WITH NON-ABSORPTION OF THE SUTURE IN THE SUBCUTICULAR LAYER OF TISSUE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Insufficient Information; Wound Dehiscence

    MAUDE report 21241727 ↗
  • InjuryJan 27, 2025Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT HIP SURGERY ON AN UNKNOWN DATE AND A BARBED SUTURE WAS USED. POST-OP, THE PATIENT EXPERIENCED AN ISSUE WITH NON-ABSORPTION OF THE SUTURE IN THE SUBCUTICULAR LAYER OF TISSUE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Insufficient Information; Wound Dehiscence

    MAUDE report 21241726 ↗
  • InjuryJan 23, 2025Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ORTHOPEDIC PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. THE PATIENT EXPERIENCED ISSUES WITH NON-ABSORPTION OF THE SUTURE IN THE SUBCUTICULAR LAYER OF TISSUE. THE PATIENT WAS BROUGHT BACK TO THE OPERATING ROOM FOR IRRIGATION AND DEBRIDEMENT OF THE INCISION SITE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Insufficient Information; Wound Dehiscence

    MAUDE report 21222739 ↗
  • InjuryJan 23, 2025Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ORTHOPEDIC PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. THE PATIENT EXPERIENCED ISSUES WITH NON-ABSORPTION OF THE SUTURE IN THE SUBCUTICULAR LAYER OF TISSUE. THE PATIENT WAS BROUGHT BACK TO THE OR FOR IRRIGATION AND DEBRIDEMENT OF THE INCISION SITE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 21222648 ↗
  • MalfunctionNov 18, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL HIP ARTHROPLASTY PROCEDURE ON (B)(6) 2024 AND BARBED SUTURE WAS USED. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20711982 ↗
  • InjuryNov 20, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL HIP PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. THE SURGEON RECENTLY SWITCHED TO CLOSING ALL HIS TOTAL HIPS WITH THE BARBED SUTURE. THE SURGEON REPORTED THAT HE HAD TO DISCONTINUE USE BECAUSE A PATIENT HAD ¿PUS OUT¿. THE PATIENT WAS TREATED AND THEY ARE FINE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Purulent Discharge

    MAUDE report 18179275 ↗
  • InjuryNov 20, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL HIP PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. THE SURGEON RECENTLY SWITCHED TO CLOSING ALL HIS TOTAL HIPS WITH THE BARBED SUTURE. THE SURGEON REPORTED THAT HE HAD TO DISCONTINUE USE BECAUSE A PATIENT HAD ¿PUS OUT¿. THE PATIENT WAS TREATED AND THEY ARE FINE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Purulent Discharge

    MAUDE report 18179274 ↗
  • InjuryNov 20, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL HIP PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. THE SURGEON RECENTLY SWITCHED TO CLOSING ALL HIS TOTAL HIPS WITH THE BARBED SUTURE. THE SURGEON REPORTED THAT HE HAD TO DISCONTINUE USE BECAUSE A PATIENT HAD ¿PUS OUT¿. THE PATIENT WAS TREATED AND THEY ARE FINE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Purulent Discharge

    MAUDE report 18179273 ↗
  • InjuryAug 23, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN PROCEDURE ON A KNEE PROSTHESES AND A BARBED SUTURE WAS USED. POST-OP, 5 DAYS AFTER SURGERY, IT BEGINS TO DRAIN FLUID FROM THE WOUND (WHICH IS NORMAL), BUT IN THIS CASE A LOT OF FLUID AND TISSUE HAS COME OUT, IT SEEMED INFECTED. A FEW WEEKS AGO A BLUE LIQUID CAME OUT. THE PATIENT DID NOT REQUIRE A REVISION SURGERY OR REMOVAL OF HARDWARE, HOWEVER, IT REQUIRED DRAINING. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Fluid Discharge; Hypersensitivity/Allergic reaction; Inflammation; Purulent Discharge; Unspecified Infection

    MAUDE report 17606913 ↗
  • MalfunctionDec 28, 2022Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2022 AND BARBED SUTURE WAS USED. THE THREAD HAS COME LOOSE FROM THE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16063275 ↗
  • MalfunctionDec 20, 2022Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A HYSTERECTOMY PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. DURING A UTERUS CLOSURE, THE SUTURE BROKE OFF THE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16019654 ↗
  • MalfunctionDec 9, 2022Break; Break

    IT WAS REPORTED A PATIENT UNDERWENT A SPINAL FUSION ON (B)(6) 2022 AND BARBED SUTURE WAS USED. THE CUSTOMER WAS USING A SUTURE AND IT SNAPPED. ON THE THIRD PASS OF THE SUTURE BROKE. THEY CUT OUT THE SUTURE. CASE COMPLETED WITH ANOTHER TYPE OF SUTURE. DEVICE IS NOT AVAILABLE FOR RETURN. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15949706 ↗
  • InjuryFeb 2, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A RIGHT TOTAL KNEE ARTHROSCOPY ON (B)(6) 2021 AND A BARBED SUTURE WAS USED. ON (B)(6) 2021, THE PATIENT CALLED THE OFFICE OF THE PROVIDER WITH CONCERNS ABOUT THEIR INCISION. ON (B)(6) 2021, THE PATIENT WAS SEEN IN THE CLINIC AND HAD WOUND DEHISCENCE AT THE PROXIMAL AND DISTAL ENDS OF THE INCISION. THE PATIENT POSSIBLY EXPERIENCED LOCAL REDNESS AND PUS. THE PATIENT WAS INSTRUCTED TO APPLY BETADINE TWICE A DAY. ON (B)(6) 2022, THE PATIENT WAS SEEN IN THE OFFICE AND IT WAS DETERMINED THAT IRRIGATION AND DEBRIDEMENT WAS NOT NECESSARY. THE SUTURE TAILS WERE REMOVED IN THE AREAS OF DEHISCENCE. THE INCISION IS CURRENTLY HEALING WITHOUT ISSUE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Purulent Discharge; Skin Inflammation/ Irritation; Wound Dehiscence

    MAUDE report 13423839 ↗

Other SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices

All SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.