STRATAFIX SXPP1B450Recalls & reported problems
Ethicon, LLC · Knotless Tissue Control Device, Violet Unidirectional (Antibacterial) Synthetic Absorbable Device · Class 2 · Implant
All models of STRATAFIX (180)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Ethicon Stratafix knotless suture has 207 adverse event reports: no deaths and 36 injuries. Most reports describe the suture breaking or separating, and some describe needles popping off during robotic hysterectomies, with surgeons finishing those operations using other sutures. Some report wounds reopening and bleeding. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 207
- Deaths reported
- 0
- Injuries reported
- 35
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 207 adverse event reports that list this exact device (DI 10705031236219), received between Feb 13, 2018 and Sep 17, 2026 — classified as 35 injuries, 172 malfunctions.
The most frequently reported specific device problem is “Material Separation” (82 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation82
- Break72
- Adverse Event Without Identified Device or Use Problem31
- Positioning Failure13
- Device Markings/Labelling Problem7
- Material Integrity Problem3
- Failure To Adhere Or Bond2
- Material Frayed2
Patient problems
- No Clinical Signs, Symptoms or Conditions140
- No Consequences Or Impact To Patient24
- Wound Dehiscence21
- Hemorrhage/Blood Loss/Bleeding14
- Adhesion(s)4
- Granuloma4
- Insufficient Information4
- Pain4
Recent adverse event reports
The 25 most recent of 207 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 17, 2026Break
It was reported that a patient underwent a robotic hysterectomy procedure on an unknown date and barbed suture was used. during a robotic hysterectomy procedure, several sutures broke during use. The issue reportedly occurred three times on the same day (Tuesday) 2 x in the same patient and other for different patient. It also occurred on at least one occasion during the previous week. The reporter was unsure whether the issue may have occurred one to two additional times that day. No adverse patient consequences were reported. Additional information was requested
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50803055 ↗MalfunctionSep 17, 2026Break
It was reported that a patient underwent a robotic hysterectomy procedure on an unknown date and barbed suture was used. during a robotic hysterectomy procedure, several sutures broke during use. The issue reportedly occurred three times on the same day (Tuesday) 2 x in the same patient and other for different patient. It also occurred on at least one occasion during the previous week. The reporter was unsure whether the issue may have occurred one to two additional times that day. No adverse patient consequences were reported. Additional information was requested
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50802877 ↗MalfunctionSep 17, 2026Break
It was reported that a patient underwent a robotic hysterectomy procedure on an unknown date and barbed suture was used. during a robotic hysterectomy procedure, several sutures broke during use. The issue reportedly occurred three times on the same day (Tuesday) 2 x in the same patient and other for different patient. It also occurred on at least one occasion during the previous week. The reporter was unsure whether the issue may have occurred one to two additional times that day. No adverse patient consequences were reported. Additional information was requested
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50802779 ↗MalfunctionAug 3, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50115197 ↗MalfunctionJun 15, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A HYSTERECTOMY PROCEDURE ON (B)(6) 2026 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT ON (B)(6) 2026 MULTIPLE STRANDS OF STRATAFIX SPIRAL NEEDLES SEPARATED AT THE SWAGE DURING TWO DIFFERENT ROBOTIC HYSTERECTOMIES WITH SURGEON. A TOTAL OF 7 SUTURE STRANDS, 3 DURING ONE PROCEDURE, 4 DURING ANOTHER, HAD THE NEEDLES EASILY POP-OFF. THE PROCEDURES WERE COMPLETED SUCCESSFULLY USING ADDITIONAL STRATAFIX SUTURES AND THERE WAS NO ADVERSE PATIENT IMPACT. STAFF DID NOT RETAIN ANY OF THE BROKEN SUTURES OR NEEDLES, NOR DID THEY DOCUMENT THE LOT. NO DELAY OR PATIENT HARM WAS REPORTED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25503640 ↗MalfunctionJun 15, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A HYSTERECTOMY PROCEDURE ON (B)(6) 2026 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT ON (B)(6) 2026 MULTIPLE STRANDS OF STRATAFIX SPIRAL NEEDLES SEPARATED AT THE SWAGE DURING TWO DIFFERENT ROBOTIC HYSTERECTOMIES WITH SURGEON. A TOTAL OF 7 SUTURE STRANDS, 3 DURING ONE PROCEDURE, 4 DURING ANOTHER, HAD THE NEEDLES EASILY POP-OFF. THE PROCEDURES WERE COMPLETED SUCCESSFULLY USING ADDITIONAL STRATAFIX SUTURES AND THERE WAS NO ADVERSE PATIENT IMPACT. STAFF DID NOT RETAIN ANY OF THE BROKEN SUTURES OR NEEDLES, NOR DID THEY DOCUMENT THE LOT. NO DELAY OR PATIENT HARM WAS REPORTED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25503633 ↗MalfunctionJun 15, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A HYSTERECTOMY PROCEDURE ON (B)(6) 2026 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT ON (B)(6) 2026 MULTIPLE STRANDS OF STRATAFIX SPIRAL NEEDLES SEPARATED AT THE SWAGE DURING TWO DIFFERENT ROBOTIC HYSTERECTOMIES WITH SURGEON. A TOTAL OF 7 SUTURE STRANDS, 3 DURING ONE PROCEDURE, 4 DURING ANOTHER, HAD THE NEEDLES EASILY POP-OFF. THE PROCEDURES WERE COMPLETED SUCCESSFULLY USING ADDITIONAL STRATAFIX SUTURES AND THERE WAS NO ADVERSE PATIENT IMPACT. STAFF DID NOT RETAIN ANY OF THE BROKEN SUTURES OR NEEDLES, NOR DID THEY DOCUMENT THE LOT. NO DELAY OR PATIENT HARM WAS REPORTED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25503631 ↗MalfunctionJun 15, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A ROBOTIC HYSTERECTOMY ON (B)(6) 2026 AND BARBED SUTURE WAS USED. IT WAS REPORTED THAT ON (B)(6) 2026 MULTIPLE STRANDS OF BARBED SPIRAL NEEDLES SEPARATED AT THE SWAGE DURING TWO DIFFERENT ROBOTIC HYSTERECTOMIES WITH SURGEON. A TOTAL OF 7 SUTURE STRANDS, 3 DURING ONE PROCEDURE, 4 DURING ANOTHER, HAD THE NEEDLES EASILY POP-OFF. THE PROCEDURES WERE COMPLETED SUCCESSFULLY USING ADDITIONAL BARBED SUTURES AND THERE WAS NO ADVERSE PATIENT IMPACT. STAFF DID NOT RETAIN ANY OF THE BROKEN SUTURES OR NEEDLES, NOR DID THEY DOCUMENT THE LOT. NO DELAY OR PATIENT HARM WAS REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25503045 ↗MalfunctionJun 15, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A ROBOTIC HYSTERECTOMY ON (B)(6) 2026 AND BARBED SUTURE WAS USED. IT WAS REPORTED THAT ON (B)(6) 2026 MULTIPLE STRANDS OF BARBED SPIRAL NEEDLES SEPARATED AT THE SWAGE DURING TWO DIFFERENT ROBOTIC HYSTERECTOMIES WITH SURGEON. A TOTAL OF 7 SUTURE STRANDS, 3 DURING ONE PROCEDURE, 4 DURING ANOTHER, HAD THE NEEDLES EASILY POP-OFF. THE PROCEDURES WERE COMPLETED SUCCESSFULLY USING ADDITIONAL BARBED SUTURES AND THERE WAS NO ADVERSE PATIENT IMPACT. STAFF DID NOT RETAIN ANY OF THE BROKEN SUTURES OR NEEDLES, NOR DID THEY DOCUMENT THE LOT. NO DELAY OR PATIENT HARM WAS REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25502613 ↗MalfunctionMay 20, 2026Break; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A MYOMECTOMY PROCEDURE ON (B)(6) 2026 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, SUTURE BROKEN FROM THE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25228890 ↗MalfunctionMay 4, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT, DURING AN UNKNOWN SURGERY, THE SUTURE WAS DETACHED FROM THE NEEDLE DURING USE. 3 CASES IN TOTAL. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO SAMPLE WILL BE RETURNED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25070898 ↗MalfunctionMay 4, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. DURING AN UNKNOWN SURGERY, THE SUTURE WAS DETACHED FROM THE NEEDLE DURING USE. 3 CASES IN TOTAL. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO SAMPLE WILL BE RETURNED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25067976 ↗MalfunctionMay 4, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. DURING AN UNKNOWN SURGERY, THE SUTURE WAS DETACHED FROM THE NEEDLE DURING USE. 3 CASES IN TOTAL. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO SAMPLE WILL BE RETURNED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25067972 ↗MalfunctionApr 28, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT, DURING AN UNKNOWN SURGERY, THE SUTURE WAS DETACHED FROM THE NEEDLE EASILY DURING USE. IT WAS NOT A CONTROL RELEASE NEEDLE. LOT NUMBER IS UNKNOWN. THERE WAS NO EFFECT ON THE PATIENT. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO SAMPLE WILL BE RETURNED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25018349 ↗MalfunctionApr 21, 2026Material Integrity Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND BARBED SUTURE WAS USED. WHEN THE PACKAGE WAS OPENED, IT WAS FOUND THAT THE NEEDLE HAD BEEN DETACHED. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24945329 ↗MalfunctionApr 21, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A ROBOTIC BILATERAL INGUINAL AND UMBILICAL HERNIA REPAIR ON (B)(6) 2026 AND BARBED SUTURE WAS USED. DURING A ROBOTIC BILATERAL INGUINAL AND UMBILICAL HERNIA REPAIR, THE SURGEON TOOK TWO PASSES INTO MEDTRONIC DEXTILE ANATOMICAL MESH WITH THE STRATAFIX SXPP1B432 SUTURE. AS HE WAS TIGHTENING THE SUTURE, IT BROKE AND HE NEEDED TO REMOVE THE SUTURE STRANDS. HE ATTEMPTED TO USE A NEW STRATAFIX TWO MORE TIMES AND THE SAME THING OCCURRED. HE USED A VLOC SUTURE TO COMPLETE THE INGUINAL HERNIA REPAIRS. HE THEN USED AN SXPP1B450 STRATAFIX ON THE UMBILICAL HERNIA REPAIR AND AGAIN THE SUTURE BROKE AFTER A COUPLE OF PASSES. THE SURGEON SWITCHED TO VLOC TO COMPLETE THE SURGERY. HE IS A VERY EXPERIENCED ROBOTIC SURGEON. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THE DEVICES WERE RETURNING. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24940252 ↗MalfunctionApr 21, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A ROBOTIC BILATERAL INGUINAL AND UMBILICAL HERNIA REPAIR ON (B)(6) 2026 AND BARBED SUTURE WAS USED. DURING A ROBOTIC BILATERAL INGUINAL AND UMBILICAL HERNIA REPAIR, THE SURGEON TOOK TWO PASSES INTO MEDTRONIC DEXTILE ANATOMICAL MESH WITH THE STRATAFIX SXPP1B432 SUTURE. AS HE WAS TIGHTENING THE SUTURE, IT BROKE AND HE NEEDED TO REMOVE THE SUTURE STRANDS. HE ATTEMPTED TO USE A NEW STRATAFIX TWO MORE TIMES AND THE SAME THING OCCURRED. HE USED A VLOC SUTURE TO COMPLETE THE INGUINAL HERNIA REPAIRS. HE THEN USED AN SXPP1B450 STRATAFIX ON THE UMBILICAL HERNIA REPAIR AND AGAIN THE SUTURE BROKE AFTER A COUPLE OF PASSES. THE SURGEON SWITCHED TO VLOC TO COMPLETE THE SURGERY. HE IS A VERY EXPERIENCED ROBOTIC SURGEON. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THE DEVICES WERE RETURNING. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24940242 ↗MalfunctionApr 21, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A ROBOTIC BILATERAL INGUINAL AND UMBILICAL HERNIA REPAIR ON (B)(6) 2026 AND BARBED SUTURE WAS USED. DURING A ROBOTIC BILATERAL INGUINAL AND UMBILICAL HERNIA REPAIR, THE SURGEON TOOK TWO PASSES INTO MEDTRONIC DEXTILE ANATOMICAL MESH WITH THE STRATAFIX SXPP1B432 SUTURE. AS HE WAS TIGHTENING THE SUTURE, IT BROKE AND HE NEEDED TO REMOVE THE SUTURE STRANDS. HE ATTEMPTED TO USE A NEW STRATAFIX TWO MORE TIMES AND THE SAME THING OCCURRED. HE USED A VLOC SUTURE TO COMPLETE THE INGUINAL HERNIA REPAIRS. HE THEN USED AN SXPP1B450 STRATAFIX ON THE UMBILICAL HERNIA REPAIR AND AGAIN THE SUTURE BROKE AFTER A COUPLE OF PASSES. THE SURGEON SWITCHED TO VLOC TO COMPLETE THE SURGERY. HE IS A VERY EXPERIENCED ROBOTIC SURGEON. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THE DEVICES WERE RETURNING. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24940241 ↗MalfunctionApr 21, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A ROBOTIC BILATERAL INGUINAL AND UMBILICAL HERNIA REPAIR ON (B)(6) 2026 AND BARBED SUTURE WAS USED. DURING A ROBOTIC BILATERAL INGUINAL AND UMBILICAL HERNIA REPAIR, THE SURGEON TOOK TWO PASSES INTO MEDTRONIC DEXTILE ANATOMICAL MESH WITH THE STRATAFIX SXPP1B432 SUTURE. AS HE WAS TIGHTENING THE SUTURE, IT BROKE AND HE NEEDED TO REMOVE THE SUTURE STRANDS. HE ATTEMPTED TO USE A NEW STRATAFIX TWO MORE TIMES AND THE SAME THING OCCURRED. HE USED A VLOC SUTURE TO COMPLETE THE INGUINAL HERNIA REPAIRS. HE THEN USED AN SXPP1B450 STRATAFIX ON THE UMBILICAL HERNIA REPAIR AND AGAIN THE SUTURE BROKE AFTER A COUPLE OF PASSES. THE SURGEON SWITCHED TO VLOC TO COMPLETE THE SURGERY. HE IS A VERY EXPERIENCED ROBOTIC SURGEON. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THE DEVICES WERE RETURNING. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24940173 ↗MalfunctionMar 31, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN HYSTERECTOMY PROCEDURE ON (B)(6) 2026 AND BARBED SUTURE WAS USED. IT WAS REPORTED BY A SALES REP THAT, DURING TOTAL HYSTERECTOMY, THE SUTURE WAS DETACHED FROM THE NEEDLE EASILY WHEN THE PRODUCT WAS USED. IT WAS A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24745326 ↗InjuryFeb 6, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC HYSTERECTOMY ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. THE PATIENT REPORTED VAGINAL BLEEDING WEEKS OUT REQUIRING EXTRA CLINIC VISITS OR ER VISITS WITH POSSIBLE FULGURATION OF TISSUE. THE MALE PARTNER POTENTIALLY MAY HAVE HAD PENIS PAIN/ CUT DUE TO BARBS DURING INTERCOURSE. THE PATIENT MAY HAVE HAD RESIDUAL SUTURE REQUIRING MULTIPLE VISITS FOR CUTTING THE SUTURE WEEKS AND MONTHS AFTER SURGERY. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 24279219 ↗InjuryFeb 6, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC HYSTERECTOMY ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. THE PATIENT REPORTED VAGINAL BLEEDING WEEKS OUT REQUIRING EXTRA CLINIC VISITS OR ER VISITS WITH POSSIBLE FULGURATION OF TISSUE. THE MALE PARTNER POTENTIALLY MAY HAVE HAD PENIS PAIN/ CUT DUE TO BARBS DURING INTERCOURSE. THE PATIENT MAY HAVE HAD RESIDUAL SUTURE REQUIRING MULTIPLE VISITS FOR CUTTING THE SUTURE WEEKS AND MONTHS AFTER SURGERY. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 24279198 ↗InjuryFeb 6, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC HYSTERECTOMY ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. THE PATIENT REPORTED VAGINAL BLEEDING WEEKS OUT REQUIRING EXTRA CLINIC VISITS OR EMERGENCY ROOM VISITS WITH POSSIBLE FULGURATION OF TISSUE. THE MALE PARTNER POTENTIALLY MAY HAVE HAD PENIS PAIN/ CUT DUE TO BARBS DURING INTERCOURSE. THE PATIENT MAY HAVE HAD RESIDUAL SUTURE REQUIRING MULTIPLE VISITS FOR CUTTING THE SUTURE WEEKS AND MONTHS AFTER SURGERY. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 24279197 ↗InjuryJan 23, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC HYSTERECTOMY ON (B)(6) 2025 AND BARBED SUTURE WAS USED ON THE VAGINAL CUFF. THE PATIENT EXPERIENCED MODERATE BLEEDING 3 WEEKS POSTOP. MULTIPLE OFFICE VISITS. NO DEHISCENCE BUT TREATED GRANULATION TISSUE. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Granuloma; Hemorrhage/Blood Loss/Bleeding
MAUDE report 24169019 ↗InjuryJan 23, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC HYSTERECTOMY ON (B)(6) 2025 AND BARBED SUTURE WAS USED ON THE VAGINAL CUFF. THE PATIENT EXPERIENCED MODERATE BLEEDING AT 4 WEEKS POSTOP, NO DEHISCENCE BUT TINY AMT GRANULATION TISSUE TREATED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Granuloma; Hemorrhage/Blood Loss/Bleeding
MAUDE report 24169018 ↗
FDA clearance and classification
The FDA regulates product code NEW (suture, surgical, absorbable, polydioxanone) under 21 CFR 878.4840 as a Class II device. 21 CFR 878.4840 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the STRATAFIX SXPP1B450
Has the STRATAFIX SXPP1B450 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705031236219), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the STRATAFIX SXPP1B450?
The FDA's MAUDE database lists 207 adverse event reports for this device, received between Feb 13, 2018 and Sep 17, 2026. In those reports, the most common specific device problems are “Material Separation” (82), “Break” (72) and “Positioning Failure” (13); the most common patient problems are “Wound Dehiscence” (21), “Hemorrhage/Blood Loss/Bleeding” (14) and “Adhesion(s)” (4). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the STRATAFIX SXPP1B450?
None of the reports are classified as deaths, 35 as injuries and 172 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the STRATAFIX SXPP1B450 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Safe: it poses no known hazards in any MRI environment.
Who makes the STRATAFIX SXPP1B450, and is it still sold?
The FDA's device database lists Ethicon, LLC as the company responsible for it (brand STRATAFIX, model SXPP1B450). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices
- STRATAFIX SXPP1A405Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE611 reports
- STRATAFIX SXPP1A400Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE478 reports
- STRATAFIX SXPP1A401Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE442 reports
- STRATAFIX SXPP1A301Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE442 reports
- STRATAFIX SXPP2B405Ethicon, LLC · Suture, surgical, absorbable, polydioxanone400 reports
- STRATAFIX SXPP1A406Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE334 reports
- STRATAFIX SXPP1A404Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 death319 reports
- STRATAFIX SXPP1B420Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 death230 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page