STRATAFIX SXPP1B453Recalls & reported problems

Ethicon, LLC · Knotless Tissue Control Device, Violet Unidirectional (Antibacterial) Synthetic Absorbable Device · Class 2 · Implant

All models of STRATAFIX (180)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Ethicon Stratafix knotless suture is an absorbable thread used to close tissue. Its 79 reports, with no deaths and 5 injuries, mostly describe malfunctions (74 reports), usually the suture breaking during sewing or the needle detaching from it. Several reports say the surgeon switched to another suture to finish the operation. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
81
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 81 adverse event reports that list this exact device (DI 10705031240827), received between Jan 2, 2018 and Sep 29, 2026 — classified as 5 injuries, 76 malfunctions.

The most frequently reported specific device problem is “Break” (48 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 5 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 4 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 5 reportsMay 2025: 0 reportsJun 2025: 3 reportsJul 2025: 1 reportsAug 2025: 2 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 1 reportsMay 2026: 7 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 3 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break48
  • Material Separation18
  • Detachment Of Device Component4
  • Positioning Failure4
  • Device Markings/Labelling Problem2
  • Adverse Event Without Identified Device or Use Problem1
  • Defective Component1
  • Device Contaminated During Manufacture or Shipping1

Patient problems

  • No Clinical Signs, Symptoms or Conditions68
  • No Consequences Or Impact To Patient8
  • Wound Dehiscence4
  • Failure to Anastomose2
  • Not Applicable2
  • Pocket Erosion2

Recent adverse event reports

The 25 most recent of 81 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50936021 ↗
  • MalfunctionSep 29, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935927 ↗
  • MalfunctionSep 29, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935810 ↗
  • MalfunctionMay 13, 2026Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A BARBED SUTURE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE HAPPENED IN THE SURGERY. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25164006 ↗
  • MalfunctionMay 13, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND A BARBED SUTURE WAS USED. THE SUTURE BROKE WHEN SEWING IN SURGERY. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25163723 ↗
  • MalfunctionMay 13, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A SUTURE WAS USED. THE LOOP HAD BEEN MISSING WHEN THE PACKAGE OF THE NEWLY DISPENSED PRODUCT WAS JUST OPENED. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25163401 ↗
  • MalfunctionMay 13, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND A BARBED SUTURE WAS USED. THE SUTURE BROKE WHEN SEWING IN SURGERY. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25163235 ↗
  • MalfunctionMay 13, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A BARBED SUTURE WAS USED. THE SUTURE BROKE WHEN SEWING IN SURGERY. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25163190 ↗
  • MalfunctionMay 13, 2026Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A ROBOTIC VENTRAL HERNIA PROCEDURE ON (B)(6) 2026 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE SALES REP THAT WHILE THE SURGEON WAS SUTURING MESH, IT WAS REPORTED THAT THE NEEDLE DETACHED FROM THE SUTURE. THE DETACHMENT OCCURRED DURING USE, AND THE SUTURE COULD NO LONGER BE UTILIZED AS INTENDED. NO PATIENT CONSEQUENCES. DEVICE IS AVAILABLE FOR RETURNED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25161978 ↗
  • MalfunctionMay 13, 2026Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A ROBOTIC VENTRAL HERNIA PROCEDURE ON (B)(6) 2026 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE SALES REP THAT WHILE THE SURGEON WAS SUTURING MESH, IT WAS REPORTED THAT THE NEEDLE DETACHED FROM THE SUTURE. THE DETACHMENT OCCURRED DURING USE, AND THE SUTURE COULD NO LONGER BE UTILIZED AS INTENDED. NO PATIENT CONSEQUENCES. DEVICE IS AVAILABLE FOR RETURNED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25161972 ↗
  • MalfunctionApr 30, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A HERNIA PROCEDURE ON (B)(6) 2026 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW PERITONEUM. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25044276 ↗
  • InjuryMar 5, 2026Break

    IT WAS REPORTED BY A SALES REP THAT, DURING AN ORTHOPEDIC SURGERY AT THE WARDS/ICU, SXPP1B453 WAS USED FOR THE DERMIS IN A TRAUMATIC PELVIC CASE. (PRINEO WAS USED FOR THE EPIDERMIS.) AT THE OUTPATIENT VISIT AFTER SURGERY, DERMAL DEHISCENCE WAS CONFIRMED. THE PRODUCT WAS USED ON THE PELVIS. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. NO SAMPLE WILL BE RETURNED. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. AFFILIATE REFERENCE NUMBER: (B)(4).

    Patient problems: Pocket Erosion; Wound Dehiscence

    MAUDE report 24525078 ↗
  • InjuryMar 5, 2026Break

    IT WAS REPORTED A PATIENT UNDERWENT AN ORTHOPEDIC PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. THE PRODUCT WAS USED ON THE SCAPULA. THE INCISION WAS A ¿V-SHAPED¿ (ANGLED) INCISION ALONG THE SCAPULA. TWO SUTURES WERE USED, ONE ON EACH SIDE. WOUND DEHISCENCE OCCURRED ON THE RIGHT SCAPULAR SIDE, WHEREAS NO DEHISCENCE OCCURRED ON THE LEFT SCAPULAR SIDE. THE DOCTOR POINTED OUT THAT THE POSSIBILITY OF A BURN ON THE RIGHT SHOULDER AND SUBCUTANEOUS DEGLOVING. SUTURE CONDITION: POSSIBILITY DUE TO PROGRESSING HYDROLYSIS, THE DISSOLVED SUTURE WAS NOT PURPLE BUT NEARLY TRANSPARENT. THE BROKEN SUTURE IS IN A CONDITION OF PROTRUDING LOOSELY FROM THE WOUND. AT AN OUTPATIENT VISIT, ADDITIONAL SUTURING WAS PERFORMED IN THE TREATMENT ROOM. THE PATIENT¿S CURRENT CONDITION IS STABLE, WITH NO INFECTION AFTER THE WOUND DEHISCENCE AND NO OTHER PARTICULAR ISSUES. DOCTOR¿S COMMENTS: THERE WAS A BURN ON THE RIGHT SHOULDER AND POSSIBLE SUBCUTANEOUS DEGLOVING, WHICH MAY HAVE MADE THE TISSUE FRAGILE AND CONTRIBUTED TO THE EVENT. DIFFERENT DOCTORS PERFORMED THE SUTURING ON THE RIGHT AND LEFT SIDES, AND DIFFERENCES IN TECHNIQUE MAY HAVE INFLUENCED THE OUTCOME. EXCESSIVE TENSION DURING SUTURING MIGHT

    Patient problems: Pocket Erosion; Wound Dehiscence

    MAUDE report 24525077 ↗
  • MalfunctionDec 31, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN PROSTATECTOMY PROCEDURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE HAPPENED IN SURGERY. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23935325 ↗
  • MalfunctionDec 16, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. THE SURGEON REPORTED TO THE SALES REP THAT DURING A ROBOTIC VENTRAL HERNIA PROCEDURE, A SUTURE SNAPPED AND BROKE. A SECOND SUTURE WAS OPENED AND USED TO SUCCESSFULLY COMPLETE THE CASE. NO PATIENT CONSEQUENCES WERE REPORTED. THE DEVICE IS AVAILABLE FOR RETURN. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23819108 ↗
  • MalfunctionNov 26, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A HERNIA REPAIR PROCEDURE ON (B)(6) 2025 AND A BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED. ¿

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23658360 ↗
  • MalfunctionNov 18, 2025Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN CARDIOVASCULAR SURGERY ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT THE PRODUCT WAS LABELED AS 22CM SO FAR, BUT THIS WAS LABELED AS 23CM. IS THERE ANY DIFFERENCE IN THE CONTENTS? IT IS UNKNOWN WHETHER THE CONTENT WAS DIFFERENT OR NOT. PLEASE SEE THE ATTACHED PHOTO FOR THE PACKAGE. IT WAS NOT A CONTROL RELEASE NEEDLE. NO SAMPLE WILL BE RETURNED. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23585722 ↗
  • MalfunctionSep 18, 2025Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. THE SUTURE WAS LABELED AS 23CM IN THE CATALOG AND THE ALUMINUM PACKAGE IS ORIGINALLY LABELED AS 22CM BUT SOME ALUMINUM PACKAGES WERE LABELED AS 23CM. THERE WAS NO PROBLEM WITH THE PRODUCT DUE TO THE LABEL OF THE ALUMINUM PACKAGE. THIS EVENT WAS FOUND BEFORE USE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23083484 ↗
  • MalfunctionAug 25, 2025Positioning Failure; Defective Component; Positioning Failure; Defective Component

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. AT 06:25, THE DOCTOR USED THE SUTURE TO REPAIR THE DUODENAL PERFORATION FOR THE PATIENT WITH DUODENAL PERFORATION. AFTER TYING THE KNOT, THE KNOT WAS LOOSE. AFTER OPENING THE SUTURE PACKAGE, THE DOCTOR TOUCHED IT AND FOUND NO BARB. THE EFFECT OF SUTURE ON PERFORATION WAS OBSERVED DURING SUTURE. NO ABNORMALITY WAS FOUND AND CONTINUED TO USE IT. CLOSELY OBSERVE PATIENT'S WOUND CONDITION. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22885376 ↗
  • MalfunctionAug 22, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22873532 ↗
  • MalfunctionJul 15, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC PROCEDURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. THE SUTURE WAS DETACHED FROM THE NEEDLE WHEN INSERTING INTO THE ABDOMINAL CAVITY AND IT COULD NOT BE USED. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22512610 ↗
  • MalfunctionJun 30, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A RADICAL PROSTATECTOMY PROCEDURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22357831 ↗
  • MalfunctionJun 13, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN ROBOTIC UNKNOWN COLORECTAL PROCEDURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE WHILE BEING PASSED THROUGH THE ROBOT PORT. SUTURE WAS HANDED UP WITH A NEEDLE DRIVER THAT HAS NO SHARP PARTS AND BROKE ABOUT THREE CENTIMETERS FROM THE NEEDLE. UNKNOWN AS TO HOW THE PROCEDURE WAS COMPLETED. NO PRODUCT RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22205565 ↗
  • MalfunctionJun 11, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC RETROPERITONEAL REPAIR OF UMBILICAL HERNIA PROCEDURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE GAP AROUND THE UMBILICAL FORAMEN WAS FULLY EXPANDED, AND THE UMBILICAL RING WAS CLOSED WITH CONTINUOUS SUTURE CLOSURE. THE PERITONEUM WAS CLOSED WITH CONTINUOUS SUTURE CLOSURE. WHEN USING SUTURE CLOSURE, THE SUTURE BROKE AND WAS PROMPTLY REMOVED AND CHECKED FOR INTEGRITY. A NEW SUTURE WAS REPLACED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22188515 ↗
  • MalfunctionApr 30, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A SPINAL ORTHOPEDICS SURGERY ON (B)(6) 2025 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21936494 ↗

FDA clearance and classification

The FDA regulates product code NEW (suture, surgical, absorbable, polydioxanone) under 21 CFR 878.4840 as a Class II device. 21 CFR 878.4840 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the STRATAFIX SXPP1B453

Has the STRATAFIX SXPP1B453 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10705031240827), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the STRATAFIX SXPP1B453?

The FDA's MAUDE database lists 81 adverse event reports for this device, received between Jan 2, 2018 and Sep 29, 2026. In those reports, the most common specific device problems are “Break” (48), “Material Separation” (18) and “Detachment Of Device Component” (4); the most common patient problems are “Wound Dehiscence” (4), “Failure to Anastomose” (2) and “Not Applicable” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the STRATAFIX SXPP1B453?

None of the reports are classified as deaths, 5 as injuries and 76 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the STRATAFIX SXPP1B453 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Safe: it poses no known hazards in any MRI environment.

Who makes the STRATAFIX SXPP1B453, and is it still sold?

The FDA's device database lists Ethicon, LLC as the company responsible for it (brand STRATAFIX, model SXPP1B453). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices

All SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page