Skip to content

STRATAFIX SXPP1A413Recalls & reported problems

Ethicon, LLC · Violet Antibacterial Polydioxanone, Knotless Tissue Control Device · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10705031245044), received between Mar 12, 2019 and Sep 30, 2026 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Material Separation” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation4
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Foreign Body In Patient1
  • No Consequences Or Impact To Patient1
  • Pneumothorax1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Material Separation

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948615 ↗
  • MalfunctionAug 18, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50300750 ↗
  • InjuryJul 27, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC LIVER RESECTION ON AN UNKNOWN DATE AND A BARBED SUTURE WAS USED FOR LIVER RETRACTION. POST-OP, THE PATIENT HAD A PNEUMOTHORAX. THE SURGEON INITIALLY ACCOUNTED THE PNEUMOTHORAX TO THE USE OF THE BARBED SUTURE, AS THE LIVER RETRACTION METHOD REQUIRES THE SUTURE TO BE FIXATED TO THE PERITONEUM / DIAPHRAGM, WHICH COULD POTENTIALLY CAUSE A DEFECT IN THE DIAPHRAGM. CLARIFICATION QUESTIONS WERE ASKED, SUCH AS HOW CAN A PNEUMOTHORAX OCCUR AND WAS THERE REALLY A DEFECT IN THE DIAPHRAGM. THE SURGEON CLARIFIED THAT THERE WAS NO DEFECT FOUND AND CONCLUDED IT MIGHT NOT HAVE BEEN RELATED TO THE BARBED SUTURE. HOWEVER, THERE ARE STILL SOME CONCERNS REGARDING THIS TECHNIQUE AFTER THIS INCIDENT. THE PATIENT HAS HEALED AND WAS DISCHARGED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Pneumothorax

    MAUDE report 17405696 ↗
  • MalfunctionMar 17, 2022Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC HERNIA REPAIR PROCEDURE ON (B)(6) 2022 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE DISCONNECTED FROM SUTURE DURING ENTERING IT THROUGH THE TROCAR. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13792026 ↗
  • MalfunctionMar 12, 2019Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A ROBOTIC INGUINAL HERNIA PROCEDURE ON AN UNKNOWN DATE IN 2019 AND BARBED SUTURE WAS USED. THE NEEDLE POPPED OFF THE SUTURE DURING CLOSURE OF THE PERITONEUM. THE NEEDLE WAS RETRIEVED FROM THE PATIENT. THE CASE WAS COMPLETED WITH ANOTHER LIKE DEVICE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8412635 ↗

Other SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices

All SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page