STRATAFIX SXPP1A443Recalls & reported problems

Ethicon, LLC · Violet Antibacterial Polydioxanone, Knotless Tissue Control Device · Class 2 · Implant

All models of STRATAFIX (181)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Ethicon Stratafix knotless barbed surgical suture dissolves over time to close tissue. Most of the 38 reports describe the suture breaking or coming apart, including during surgery and well before its expected strength duration, plus some reports of the material separating. Sixteen reports describe injuries such as wound reopening, skin tissue dying, joint dislocation, swelling and bleeding, and there are no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
38
Deaths reported
0
Injuries reported
16
Recalls naming it
0

Summary

The FDA's MAUDE database contains 38 adverse event reports that list this exact device (DI 10705031245280), received between Jul 20, 2019 and Apr 3, 2026 — classified as 16 injuries, 22 malfunctions.

The most frequently reported specific device problem is “Break” (15 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 7 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 3 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 3 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break15
  • Adverse Event Without Identified Device or Use Problem14
  • Material Separation6
  • Positioning Failure3

Patient problems

  • No Clinical Signs, Symptoms or Conditions21
  • Joint Dislocation7
  • Wound Dehiscence4
  • Swelling/ Edema3
  • Hemorrhage/Blood Loss/Bleeding2
  • Hypersensitivity/Allergic reaction2
  • Necrosis2
  • Unspecified Infection2

Recent adverse event reports

The 25 most recent of 38 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 3, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LAMINECTOMY/FUSION PROCEDURE ON (B)(6) 2026 AND A BARBED SUTURE WAS USED. POST-OP, THE PATIENT EXPERIENCED AN INFECTION AND DEHISCENCE. WHEN THE SURGEON WENT TO REOPEN THE INCISION, HE FOUND THAT THE SUTURE HAD ESSENTIALLY COMPLETELY DISSOLVED AND WAS SUPER FRAGILE WITHIN A MONTH, UNDER THE AVERAGE AMOUNT OF TIME THE SUTURE HOLDS STRENGTH FOR. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Fluid Discharge; Swelling/ Edema; Unspecified Infection; Wound Dehiscence

    MAUDE report 24776693 ↗
  • InjuryFeb 20, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PANNICULECTOMY PROCEDURE ON (B)(6) 2026 DUE TO EXCESSIVE PANNICULUS AND INTERTRIGO CANDIDIASIS AND BARBED SUTURE WAS USED ON THE FASCIA. POST-OP, THE PATIENT EXPERIENCED A SUBCLINICAL HEMATOMA. THE PATIENT WENT HOME POST OP DAY 1, REMOVED DRAIN, SKIN DIED FOLLOWING WEEK. THE SKIN DIED, ON POST OP DAY 10 THE PATIENT CAME TO ER AND UNDERWENT WOUND DEBRIDEMENT. THE SUTURE BEGAN BREAKING DOWN DURING THE SECOND PROCEDURE. THE PATIENT HAS RECOVERED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Necrosis

    MAUDE report 24400372 ↗
  • InjuryFeb 20, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PANNICULECTOMY PROCEDURE ON (B)(6) 2026 DUE TO INTERTRIGO CANDIDIASIS AND BARBED SUTURE WAS USED ON THE FASCIA. POST-OP, THE PATIENT EXPERIENCED A HEMATOMA/HEMORRHAGE. THE PATIENT WAS GIVEN A BLOOD TRANSFUSION ON POST OP DAY 1 AND DEVELOPED RAPID RESPONSE HYPERTENSION AND WAS TRANSFERRED TO STEP DOWN. THE PATIENT HAS RECOVERED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; High Blood Pressure/ Hypertension

    MAUDE report 24400369 ↗
  • MalfunctionDec 11, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. DURING AN UNKNOWN OPEN SURGERY AND WOUND CLOSURE, THE NEEDLE TIP WAS BROKEN, BUT THE NEEDLE TIP COULD ALSO BE RETRIEVED WITHOUT REMAINING NO PIECES IN THE PATIENT¿S BODY. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23778267 ↗
  • InjuryOct 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PANNICULECTOMY PROCEDURE ON AN UNKNOWN DATE AND A BARBED SUTURE WAS USED. POST-OP, THE PATIENT EXPERIENCED SKIN NECROSIS. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Necrosis; Wound Dehiscence

    MAUDE report 23425588 ↗
  • InjuryOct 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED BY THE PATIENT THAT THEY UNDERWENT A KNEE SURGERY ON AN UNKNOWN DATE AND A BARBED SUTURE WAS USED. THE PATIENT STATED THEY WERE ALLERGIC TO SOMETHING AND HAVING ALLERGY TESTING. ON (B)(6) 2025, THE PATIENT HAD SWELLING AND COULD NOT PUT WEIGHT ON IT. THE PATIENT STATED THEY MAY BE ALLERGIC TO NICKEL. THE SURGEON STATED THE PATIENT HAD A REACTION AND TOOK CARE OF IT WITH ANTIBIOTIC. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Hypersensitivity/Allergic reaction; Impaired Healing; Swelling/ Edema

    MAUDE report 23309355 ↗
  • InjuryOct 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED BY THE PATIENT THAT THEY UNDERWENT AN EYE SURGERY ON AN UNKNOWN DATE AND A BARBED SUTURE WAS USED. THE PATIENT STATED THEY WERE ALLERGIC TO SOMETHING, STITCH IN EYE HAVING ISSUES AND HAVING ALLERGY TESTING. ON JUNE 24TH, THE PATIENT HAD SWELLING AND COULD NOT PUT WEIGHT ON IT. THE PATIENT STATED THEY MAY BE ALLERGIC TO NICKEL. THE EYE SURGEON STATED THE PATIENT HAD A REACTION AND TOOK CARE OF IT WITH ANTIBIOTIC. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Hypersensitivity/Allergic reaction; Swelling/ Edema

    MAUDE report 23309349 ↗
  • InjuryFeb 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED IN THE DEEP FASCIA LAYER. THE SURGEON SAID THE PATIENT GOT AN INFECTION AFTERWARDS. WHEN HE REMOVED THE SUTURE PUS WAS COMING OUT FROM THE AREA WHERE THE BARBED SUTURE WAS LOCATED. THE SURGEON IS CONVINCED IT WAS BECAUSE IT WAS OF THE BARBED SUTURE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Purulent Discharge; Unspecified Infection

    MAUDE report 21332954 ↗
  • MalfunctionDec 26, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE REPLACEMENT PROCEDURE IN (B)(4) 2024 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, PER USER FACILITY MEDWATCH FORM, #(B)(4), DURING TOTAL KNEE ARTHROPLASTY; NEEDLE POPPED OFF SUTURE DURING USE FOR WOUND CLOSURE. NEEDLE WAS SECURED AND SET ASIDE. NEW SUTURE NEEDLE OBTAINED; SAME THING HAPPENED. ONCE MORE A NEW SUTURE NEEDLE WAS OBTAINED WITH THE SAME RESULT. THERE WAS NO PATIENT CONSEQUENCES REPORTED. DEVICE AVAILABILITY UNKNOWN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21020385 ↗
  • MalfunctionDec 26, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE REPLACEMENT PROCEDURE IN (B)(6) 2024 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, PER USER FACILITY MEDWATCH FORM, #(B)(4), DURING TOTAL KNEE ARTHROPLASTY; NEEDLE POPPED OFF SUTURE DURING USE FOR WOUND CLOSURE. NEEDLE WAS SECURED AND SET ASIDE. NEW SUTURE NEEDLE OBTAINED; SAME THING HAPPENED. ONCE MORE A NEW SUTURE NEEDLE WAS OBTAINED WITH THE SAME RESULT. THERE WAS NO PATIENT CONSEQUENCES REPORTED. DEVICE AVAILABILITY UNKNOWN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21020382 ↗
  • MalfunctionDec 26, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE REPLACEMENT PROCEDURE IN (B)(6) 2024 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, PER USER FACILITY MEDWATCH FORM, # (B)(4), DURING TOTAL KNEE ARTHROPLASTY; NEEDLE POPPED OFF SUTURE DURING USE FOR WOUND CLOSURE. NEEDLE WAS SECURED AND SET ASIDE. NEW SUTURE NEEDLE OBTAINED; SAME THING HAPPENED. ONCE MORE A NEW SUTURE NEEDLE WAS OBTAINED WITH THE SAME RESULT. THERE WAS NO PATIENT CONSEQUENCES REPORTED. DEVICE AVAILABILITY UNKNOWN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21020380 ↗
  • MalfunctionOct 28, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. THE SUTURE CAME OFF THE NEEDLE WHILE SUTURING UNKNOWN TISSUE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20550138 ↗
  • MalfunctionAug 14, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A KNEE PROCEDURE ON (B)(6) 2024 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, X2 SNAPPED SUTURE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19981376 ↗
  • MalfunctionAug 14, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A KNEE PROCEDURE ON (B)(6)2024 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, X2 SNAPPED SUTURE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19981373 ↗
  • InjuryMay 7, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY PROCEDURE ON (B)(6)2024 AND BARBED SUTURE WAS USED. THE PATIENT EXPERIENCED A DISLOCATED PATELLA POST OPERATIVELY. THE DISLOCATION WAS DISCOVERED BETWEEN 5 WEEKS POST OP AND 8 WEEK POST OP (35 DAY/ 56 DAY). THE PATIENT WAS TAKEN TO SURGERY FOR VMO REPAIR/ ADVANCEMENT AND MAY HAVE HAD A LATERAL RELEASE DEPENDING ON TRACKING OF THE PATELLA. REPAIR WAS DONE WITH INTERRUPTED NON ABSORBABLE SUTURE AS WELL AS A CONTINUOUS NON ABSORBABLE 1 V LOC AND THE PATIENT DID NOT HAVE A REOCCURRENCE OF DISLOCATING PATELLA FOLLOWING THE REPAIR. WHEN THE EXPOSURE WAS MADE THE ABSORBABLE SUTURE WAS COMPLETELY GONE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Joint Dislocation

    MAUDE report 19265891 ↗
  • InjuryMay 7, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY PROCEDURE ON (B)(6) 2024 AND BARBED SUTURE WAS USED. THE PATIENT EXPERIENCED A DISLOCATED PATELLA POST OPERATIVELY. THE DISLOCATION WAS DISCOVERED BETWEEN 5 WEEKS POST OP AND 8 WEEK POST OP (35 DAY/ 56 DAY). THE PATIENT WAS TAKEN TO SURGERY FOR VMO REPAIR/ ADVANCEMENT AND MAY HAVE HAD A LATERAL RELEASE DEPENDING ON TRACKING OF THE PATELLA. REPAIR WAS DONE WITH INTERRUPTED NON ABSORBABLE SUTURE AS WELL AS A CONTINUOUS NON ABSORBABLE 1 V LOC AND THE PATIENT DID NOT HAVE A REOCCURRENCE OF DISLOCATING PATELLA FOLLOWING THE REPAIR. WHEN THE EXPOSURE WAS MADE THE ABSORBABLE SUTURE WAS COMPLETELY GONE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Joint Dislocation

    MAUDE report 19265888 ↗
  • InjuryMay 7, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY PROCEDURE ON (B)(6) 2024 AND BARBED SUTURE WAS USED. THE PATIENT EXPERIENCED A DISLOCATED PATELLA POST OPERATIVELY. THE DISLOCATION WAS DISCOVERED BETWEEN 5 WEEKS POST OP AND 8 WEEK POST OP (35 DAY/ 56 DAY). THE PATIENT WAS TAKEN TO SURGERY FOR VMO REPAIR/ ADVANCEMENT AND MAY HAVE HAD A LATERAL RELEASE DEPENDING ON TRACKING OF THE PATELLA. REPAIR WAS DONE WITH INTERRUPTED NON ABSORBABLE SUTURE AS WELL AS A CONTINUOUS NON ABSORBABLE 1 V LOC AND THE PATIENT DID NOT HAVE A REOCCURRENCE OF DISLOCATING PATELLA FOLLOWING THE REPAIR. WHEN THE EXPOSURE WAS MADE THE ABSORBABLE SUTURE WAS COMPLETELY GONE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Joint Dislocation

    MAUDE report 19265887 ↗
  • InjuryMay 7, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY PROCEDURE ON 19(B)(6) 2023 AND BARBED SUTURE WAS USED. THE PATIENT EXPERIENCED A DISLOCATED PATELLA POST OPERATIVELY. THE DISLOCATION WAS DISCOVERED BETWEEN 5 WEEKS POST OP AND 8 WEEK POST OP (35 DAY/ 56 DAY). THE PATIENT WAS TAKEN TO SURGERY FOR VMO REPAIR/ ADVANCEMENT AND MAY HAVE HAD A LATERAL RELEASE DEPENDING ON TRACKING OF THE PATELLA. REPAIR WAS DONE WITH INTERRUPTED NON ABSORBABLE SUTURE AS WELL AS A CONTINUOUS NON ABSORBABLE 1 V LOC AND THE PATIENT DID NOT HAVE A REOCCURRENCE OF DISLOCATING PATELLA FOLLOWING THE REPAIR. WHEN THE EXPOSURE WAS MADE THE ABSORBABLE SUTURE WAS COMPLETELY GONE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Joint Dislocation

    MAUDE report 19265886 ↗
  • InjuryMay 7, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY PROCEDURE ON (B)(6) 2023 AND BARBED SUTURE WAS USED. THE PATIENT EXPERIENCED A DISLOCATED PATELLA POST OPERATIVELY. THE DISLOCATION WAS DISCOVERED BETWEEN 5 WEEKS POST OP AND 8 WEEK POST OP (35 DAY/ 56 DAY). THE PATIENT WAS TAKEN TO SURGERY FOR VMO REPAIR/ ADVANCEMENT AND MAY HAVE HAD A LATERAL RELEASE DEPENDING ON TRACKING OF THE PATELLA. REPAIR WAS DONE WITH INTERRUPTED NON ABSORBABLE SUTURE AS WELL AS A CONTINUOUS NON ABSORBABLE 1 V LOC AND THE PATIENT DID NOT HAVE A REOCCURRENCE OF DISLOCATING PATELLA FOLLOWING THE REPAIR. WHEN THE EXPOSURE WAS MADE THE ABSORBABLE SUTURE WAS COMPLETELY GONE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Joint Dislocation

    MAUDE report 19265885 ↗
  • InjuryMay 7, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY PROCEDURE ON (B)(6)2023 AND BARBED SUTURE WAS USED. THE PATIENT EXPERIENCED A DISLOCATED PATELLA POST OPERATIVELY. THE DISLOCATION WAS DISCOVERED BETWEEN 5 WEEKS POST OP AND 8 WEEK POST OP (35 DAY/ 56 DAY). THE PATIENT WAS TAKEN TO SURGERY FOR VMO REPAIR/ ADVANCEMENT AND MAY HAVE HAD A LATERAL RELEASE DEPENDING ON TRACKING OF THE PATELLA. REPAIR WAS DONE WITH INTERRUPTED NON ABSORBABLE SUTURE AS WELL AS A CONTINUOUS NON ABSORBABLE 1 V LOC AND THE PATIENT DID NOT HAVE A REOCCURRENCE OF DISLOCATING PATELLA FOLLOWING THE REPAIR. WHEN THE EXPOSURE WAS MADE THE ABSORBABLE SUTURE WAS COMPLETELY GONE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Joint Dislocation

    MAUDE report 19265884 ↗
  • InjuryMay 7, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY PROCEDURE ON (B)(6) 2024 AND BARBED SUTURE WAS USED. THE PATIENT EXPERIENCED A DISLOCATED PATELLA POST OPERATIVELY. THE DISLOCATION WAS DISCOVERED BETWEEN 5 WEEKS POST OP AND 8 WEEK POST OP (35 DAY/ 56 DAY). THE PATIENT WAS TAKEN TO SURGERY FOR VMO REPAIR/ ADVANCEMENT AND MAY HAVE HAD A LATERAL RELEASE DEPENDING ON TRACKING OF THE PATELLA. REPAIR WAS DONE WITH INTERRUPTED NON ABSORBABLE SUTURE AS WELL AS A CONTINUOUS NON ABSORBABLE 1 V LOC AND THE PATIENT DID NOT HAVE A REOCCURRENCE OF DISLOCATING PATELLA FOLLOWING THE REPAIR. WHEN THE EXPOSURE WAS MADE THE ABSORBABLE SUTURE WAS COMPLETELY GONE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Joint Dislocation

    MAUDE report 19265883 ↗
  • MalfunctionMar 12, 2024Break; Break

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN ABDOMINAL HYSTERECTOMY ON AN UNKNOWN DATE IN 2024, AND BARBED SUTURE WAS USED. DURING THE SURGERY, A PDS STRATAFIX SYMMETRIC BROKE CLOSE TO THE SWAGE. THERE WAS NO PATIENT HARM. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18884836 ↗
  • MalfunctionJul 21, 2023Material Separation; Material Separation

    IT WAS REPORTED A PATIENT UNDERWENT UNKNOWN ORTHOPEDIC SURGERY ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. DURING SURGERY, THE SURGEON APPLIED A STITCH ON MUSCULAR LAYER DURING WOUND CLOSURE, THE NEEDLE WAS DETACHED FROM THE SUTURE. IT WAS NOT A CONTROL RELEASE NEEDLE. SUTURE METHOD WAS CONTINUOUS SUTURE. FURTHER DETAILS ARE NOT PROVIDED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17372590 ↗
  • MalfunctionApr 6, 2023Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CLOSING KNEE JOINT CAPSULE PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. THE SUTURE RIPPED OUT ONE TO TWO MONTHS AFTER SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16691809 ↗
  • MalfunctionMar 23, 2023Positioning Failure; Positioning Failure

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN GYNECOLOGICAL PROCEDURE ON (B)(6) 2023 AND BARBED SUTURE WAS USED. DURING FACIA CLOSURE, THE FIXATION PAD WAS RIPPING THROUGH THE FACIA. SURGEON TRIED A SECOND TIME WITH THE SAME SUTURE STRAND WITH THE SAME OUTCOME. A COMPETITOR¿S SUTURE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16600249 ↗

Other SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices

All SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices →