STRATAFIX SXPP1A304Recalls & reported problems
Ethicon, LLC · Violet Antibacterial Polydioxanone, Knotless Tissue Control Device · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10705031245297), received between Feb 3, 2023 and Aug 17, 2026 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Material Separation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation2
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Abscess1
- Inflammation1
- Insufficient Information1
- Unspecified Infection1
- Wound Dehiscence1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 17, 2026Material Separation
It was reported that a patient underwent an unknown orthopedic procedure on an unknown date and barbed suture was used. During an unknown orthopedic surgery, the suture was detached from the needle easily when trying to grasp it with the needle holder. The product could not be used. It was not a control release needle. Another product was used to complete the case. There were no adverse consequences to the patient. Additional information was requested.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50277842 ↗MalfunctionAug 4, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. DURING SURGERY, THE NEEDLE HAD BEEN DETACHED FROM THE SUTURE WHEN TRYING TO USE IT IN OPERATION. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22707985 ↗InjuryFeb 3, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A SUBTOTAL STOMACH-PRESERVING PANCREATODUODENECTOMY PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. THE SUTURE METHOD WAS CONTINUOUS SUTURE. THE PATIENT DEVELOPED A SEVERE SURGICAL SITE INFECTION. ABOUT 3 WEEKS AFTER THE SURGERY, THE WOUND WAS DEHISCENT AND IT WAS FOUND THAT THE PRODUCT WAS BROKEN. THE SURGEON STATED THEY ARE AWARE OF THE RISK OF DEHISCENCE BECAUSE OF THE WOUND SSI, BUT THE PRODUCT BROKE 3 WEEKS AFTER SURGERY AND THE PATIENT WAS A 81 YEAR-OLD WOMAN WHO HAS NO STRONG ABDOMINAL MUSCLES. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Abscess; Inflammation; Insufficient Information; Unspecified Infection; Wound Dehiscence
MAUDE report 16292994 ↗
Other SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices
- STRATAFIX SXPP1A405Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE617 reports
- STRATAFIX SXPP1A400Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE485 reports
- STRATAFIX SXPP1A401Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE446 reports
- STRATAFIX SXPP1A301Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE443 reports
- STRATAFIX SXPP2B405Ethicon, LLC · Suture, surgical, absorbable, polydioxanone406 reports
- STRATAFIX SXPP1A406Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE339 reports
- STRATAFIX SXPP1A404Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 death319 reports
- STRATAFIX SXPP1B420Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 death229 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.