Skip to content

STRATAFIX SXPP2B203Recalls & reported problems

Ethicon, LLC · Stratafix (Spiral PDS Plus), Bidirectional · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Ethicon Stratafix Spiral PDS Plus barbed surgical suture is used to close skin and soft tissue, mainly in knee and joint replacement surgery. The 28 reports mostly describe the suture breaking during wound closure, or the material separating, sometimes leaving a broken needle tip or metallic fragment in the patient; a few mention the barbs not holding as well as expected and wounds reopening. Three reports describe injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
27
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 27 adverse event reports that list this exact device (DI 10705031465169), received between Nov 4, 2022 and May 6, 2026 — classified as 2 injuries, 25 malfunctions.

The most frequently reported specific device problem is “Break” (13 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 3 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break13
  • Material Separation9
  • Adverse Event Without Identified Device or Use Problem2
  • Positioning Failure2

Patient problems

  • No Clinical Signs, Symptoms or Conditions24
  • Wound Dehiscence2
  • Foreign Body In Patient1

Recent adverse event reports

The 25 most recent of 27 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 6, 2026

    IT WAS A CASE OF RIGHT KNEE ARTHROPLASTY REVISION, DR. [REDACTED] WAS THE ONE WHO CLOSED THE FIRST AND SECOND LAYER OF THE SKIN AND LET HIS PA [PHYSICIAN ASSISTANT] CLOSED THE REST. WHILE CLOSING, THE PA NOTICED THAT THE TIP OF THE NEEDLE IS BROKEN. DR. [REDACTED] WAS NOTIFIED. X-RAY CALLED FOR BROKEN ITEM, CHARGE NURSE WAS NOTIFIED ALSO. X-RAY WAS DONE AND READ BY DR. [REDACTED] AND THE REPORT WAS POSITIVE FOR METALLIC FOREIGN BODY OVER THE MEDIAL ASPECT OF THE MEDIAL FEMORAL CONDYLE. DR. [REDACTED] OPEN THE SURGICAL SITE TO LOCATE THE BROKEN NEEDLE AND WHILE DOING IT HE IS USING THE SUCTION. SECOND X-RAY WAS DONE AND PER RADIOLOGY REPORT THE BROKEN NEEDLE HAS BEEN REMOVED. DR. [REDACTED] AND DR. [REDACTED] ALSO HAS A TELEPHONE CONVERSATION REGARDING THE RESULT.

    Patient problems: Foreign Body In Patient

    MAUDE report 25092677 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY (TKA) ON (B)(6) 2026 AND BARBED SUTURE WAS USED. IT WAS REPORTED THAT ORTHOPEDIC SURGEON, WAS USING BARBED SUTURE IN A TOTAL KNEE PROCEDURE. HE HAD TWO STRANDS BREAK ON HIM WHILE ATTEMPTING TO CLOSE THE WOUND. PROCEDURE WAS DELAYED BY 10 MINUTES. NO PATIENT HARM WAS REPORTED. DEVICES ARE AVAILABLE FOR RETURN. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24790038 ↗
  • MalfunctionApr 6, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY (TKA) ON (B)(6) 2026 AND BARBED SUTURE WAS USED. IT WAS REPORTED THAT ORTHOPEDIC SURGEON, WAS USING BARBED SUTURE IN A TOTAL KNEE PROCEDURE. HE HAD TWO STRANDS BREAK ON HIM WHILE ATTEMPTING TO CLOSE THE WOUND. PROCEDURE WAS DELAYED BY 10 MINUTES. NO PATIENT HARM WAS REPORTED. DEVICES ARE AVAILABLE FOR RETURN. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24790036 ↗
  • MalfunctionJun 13, 2024Positioning Failure; Positioning Failure

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE REPLACEMENT ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. THE SURGICAL FIRST ASSIST STATED THAT THE BARBS ON THE BARBED SUTURE DID NOT SEEM AS EFFECTIVE AS USUAL. NO PATIENT HARM. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19534565 ↗
  • MalfunctionJun 13, 2024Positioning Failure; Positioning Failure

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE REPLACEMENT ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. THE SURGICAL FIRST ASSIST STATED THAT THE BARBS ON THE BARBED SUTURE DID NOT SEEM AS EFFECTIVE AS USUAL. NO PATIENT HARM. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19534564 ↗
  • MalfunctionFeb 9, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL JOINT PROCEDURE IN 2024 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE SALES REP REPORTED THAT 3 EACHES OF SXPP2B203 BROKE DURING A TOTAL JOINT CASE. ALL 3 SUTURES WERE SXPP2B203 LOT #TDBBZZ. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. NO ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18676443 ↗
  • MalfunctionFeb 9, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL JOINT PROCEDURE IN 2024 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE SALES REP REPORTED THAT 3 EACHES OF SXPP2B203 BROKE DURING A TOTAL JOINT CASE. ALL 3 SUTURES WERE SXPP2B203 LOT #TDBBZZ. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. NO ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18676440 ↗
  • MalfunctionFeb 9, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL JOINT PROCEDURE IN 2024 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE SALES REP REPORTED THAT 3 EACHES OF SXPP2B203 BROKE DURING A TOTAL JOINT CASE. ALL 3 SUTURES WERE SXPP2B203 LOT #TDBBZZ. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. NO ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18676438 ↗
  • MalfunctionOct 12, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL JOINT PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. DURING THE PROCEDURE THE NEEDLE WAS RANDOMLY POPPING OFF OF THE SWAGE. THE DOCTOR WAS NOT EXCESSIVELY PULLING OR MISHANDLING THE SUTURE. HE'S NOT BEING EXTREMELY MINDFUL OF HOW HE PULLS THE STITCH THROUGH TO PREVENT THIS FROM HAPPENING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17927139 ↗
  • MalfunctionOct 12, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL JOINT PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. DURING THE PROCEDURE THE NEEDLE WAS RANDOMLY POPPING OFF OF THE SWAGE. THE DOCTOR WAS NOT EXCESSIVELY PULLING OR MISHANDLING THE SUTURE. HE'S NOT BEING EXTREMELY MINDFUL OF HOW HE PULLS THE STITCH THROUGH TO PREVENT THIS FROM HAPPENING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17926869 ↗
  • MalfunctionOct 12, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL JOINT PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. DURING THE PROCEDURE THE NEEDLE WAS RANDOMLY POPPING OFF OF THE SWAGE. THE DOCTOR WAS NOT EXCESSIVELY PULLING OR MISHANDLING THE SUTURE. HE'S NOT BEING EXTREMELY MINDFUL OF HOW HE PULLS THE STITCH THROUGH TO PREVENT THIS FROM HAPPENING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17926538 ↗
  • MalfunctionAug 9, 2023Break; Break

    IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE IN 2023 AND BARBED SUTURE WAS USED. SUTURE BROKE IN THE MIDDLE AFTER THE VERY FIRST THROW. THERE WERE NO PATIENT CONSEQUENCES. DEVICE NOT AVAILABLE FOR RETURN. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17494439 ↗
  • MalfunctionAug 9, 2023Break; Break

    IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE IN 2023 AND BARBED SUTURE WAS USED. SUTURE BROKE IN THE MIDDLE AFTER THE VERY FIRST THROW. THERE WERE NO PATIENT CONSEQUENCES. DEVICE NOT AVAILABLE FOR RETURN. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17494432 ↗
  • MalfunctionAug 9, 2023Break; Break

    IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE IN 2023 AND BARBED SUTURE WAS USED. SUTURE BROKE IN THE MIDDLE AFTER THE VERY FIRST THROW. THERE WERE NO PATIENT CONSEQUENCES. DEVICE NOT AVAILABLE FOR RETURN. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17494426 ↗
  • MalfunctionMar 20, 2023Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2023 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, MULTIPLE #1 STRATAFIX SUTURES HAVE BEEN BREAKING IN THE CENTER OF THE SUTURE. IT IS DURING THE FIRST TWO TO THREE THROWS WHEN PULLING TO SET THE SUTURE. UNKNOWN IF THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16579902 ↗
  • MalfunctionMar 20, 2023Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2023 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, MULTIPLE #1 STRATAFIX SUTURES HAVE BEEN BREAKING IN THE CENTER OF THE SUTURE. IT IS DURING THE FIRST TWO TO THREE THROWS WHEN PULLING TO SET THE SUTURE. UNKNOWN IF THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16579901 ↗
  • MalfunctionMar 20, 2023Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2023 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, MULTIPLE #1 STRATAFIX SUTURES HAVE BEEN BREAKING IN THE CENTER OF THE SUTURE. IT IS DURING THE FIRST TWO TO THREE THROWS WHEN PULLING TO SET THE SUTURE. UNKNOWN IF THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16579900 ↗
  • MalfunctionMar 20, 2023Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2023 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, MULTIPLE #1 STRATAFIX SUTURES HAVE BEEN BREAKING IN THE CENTER OF THE SUTURE. IT IS DURING THE FIRST TWO TO THREE THROWS WHEN PULLING TO SET THE SUTURE. UNKNOWN IF THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16579899 ↗
  • MalfunctionMar 20, 2023Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2023 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, MULTIPLE #1 STRATAFIX SUTURES HAVE BEEN BREAKING IN THE CENTER OF THE SUTURE. IT IS DURING THE FIRST TWO TO THREE THROWS WHEN PULLING TO SET THE SUTURE. UNKNOWN IF THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16579898 ↗
  • MalfunctionFeb 7, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE POPPED OFF THE SUTURE WHILE SUTURING CAPSULE CLOSED. IT WAS UNKNOWN IF HE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16310878 ↗
  • MalfunctionFeb 7, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE POPPED OFF THE SUTURE WHILE SUTURING CAPSULE CLOSED. IT WAS UNKNOWN IF HE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16310875 ↗
  • MalfunctionFeb 7, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE POPPED OFF THE SUTURE WHILE SUTURING CAPSULE CLOSED. IT WAS UNKNOWN IF HE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16310872 ↗
  • MalfunctionFeb 7, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE POPPED OFF THE SUTURE WHILE SUTURING CAPSULE CLOSED. IT WAS UNKNOWN IF HE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16310868 ↗
  • MalfunctionFeb 7, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE POPPED OFF THE SUTURE WHILE SUTURING CAPSULE CLOSED. IT WAS UNKNOWN IF HE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16310867 ↗
  • MalfunctionFeb 7, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE POPPED OFF THE SUTURE WHILE SUTURING CAPSULE CLOSED. IT WAS UNKNOWN IF HE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16310866 ↗

Other Suture, surgical, absorbable, polydioxanone devices

All Suture, surgical, absorbable, polydioxanone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page