Skip to content

AVANTI 402604XRecalls & reported problems

Cordis Corporation · SI AVANTI+ .035 4F W/MINI GW · Class 2

All models of AVANTI (19)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705032000529), received between Jan 4, 2024 and Jan 4, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K970392); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 4, 2024Fracture; Fracture

    AS REPORTED, THE ACCOUNT STATED THAT THE INTRODUCER WIRE OF TWO 4F AVANTI SHEATH INTRODUCER (SI) WITH MINI GUIDEWIRE (GW) COULD NOT GO THROUGH THE HEMOSTATIC VALVE FROM EITHER SIDE. THE INTRODUCER WIRE COULD GO THROUGH THE DISTAL END OF THE SHEATH. THE PROBLEM WAS NOTICED AT THE BEGINNING AND LATER ON WHILE USING THE SECOND DEVICE. THERE WAS NO REPORTED PATIENT INJURY. MULTIPLE ATTEMPTS TO OBTAIN SUPPLEMENTAL INFORMATION WERE MADE; HOWEVER, ADDITIONAL EVENT DETAILS WERE NOT PROVIDED. THE DEVICES WERE RECEIVED FOR ANALYSIS. ANALYSIS OF THE SECOND DEVICE WAS COMPLETED AND THE DISTAL END OF THE GUIDEWIRE WAS INSPECTED USING A VISION SYSTEM CONFIRMING AN UNRAVEL CONDITION. ALSO, A SEPARATED CONDITION ON THE CORE WIRE WAS OBSERVED ON THE SECOND DEVICE. THE CORE WIRE IS SEPARATED IN TWO PIECES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18445935 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K970392. One submission often covers many models, so check the product and lot details in each record.

  • Z-2079-2021 · Jun 9, 2021 · Terminated
    Avanti+ Catheter Sheath Introducer - Product Usage: Is indicated for use in arterial and venous procedures requiring percutaneous introduction of intravascular devices.
  • Z-1174-2016 · Dec 31, 2015 · Terminated
    CORDIS AVANTI + CATHETER SHEATH INTRODUCER SYSTEM The product includes a Catheter Sheath Introducer (CSI) and a vessel dilator. A Catheter Sheath Introducer facilitates percutaneo

Other DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION devices

All DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page