Skip to content

TEMPO 451401V5Recalls & reported problems

Cordis Corporation · CATH TEMPO 4F STR 65CM 5SH · Class 2

All models of TEMPO (56)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705032007641), received between Nov 10, 2025 and Nov 10, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Degraded” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Degraded2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 10, 2025Degraded; Degraded

    AS REPORTED, THE BODY OF A 5F TEMPO ANGIOGRAPHIC CATHETERS WAS PEELING OFF, APPEARING AS THOUGH IT WAS DECOMPOSING. THERE WAS NO REPORTED PATIENT INJURY. IT WAS FURTHER REPORTED THAT ADDITIONAL DEVICES FROM THE SAME LOTS AND SKUS EXHIBITED SIMILAR PEELING, WITH APPROXIMATELY NINETEEN DEVICES AFFECTED. IT WAS CONFIRMED THAT THE REMAINING DEVICES WERE STERILE AND EXHIBITED THE SAME FAILURE. DEVICES ARE STORED IN THE WAREHOUSE ALONG WITH OTHER MEDICAL DEVICES. OZONE IS NOT USED TO DISINFECT, AND IT IS NOT USED IN ANY AREAS SUCH AS PROCEDURE OR STORAGE ROOMS. FOUR IMAGES WERE PROVIDED, SHOWING BLUE FLAKES ON A BLOOD-SOAKED SURGICAL TOWEL, CRACKS ON THE CATHETER BODY AND STRAIN RELIEF, AND YELLOWISH DISCOLORATION TO THE HUB. THE DEVICES WILL BE RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23509151 ↗
  • MalfunctionNov 10, 2025Degraded; Degraded

    STERILE DEVICE WAS RECEIVED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23509089 ↗

Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices

All CATHETER, INTRAVASCULAR, DIAGNOSTIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page