TEMPO 451406P0Recalls & reported problems
Cordis Corporation · CATH TEMPO 4F MP A1 (I) 125CM · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705032007757), received between Oct 12, 2020 and Nov 6, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Delivered as Unsterile Product” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Delivered as Unsterile Product1
- Difficult to Open or Remove Packaging Material1
- Tear, Rip or Hole in Device Packaging1
- Unsealed Device Packaging1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 6, 2024Delivered as Unsterile Product; Unsealed Device Packaging; Delivered as Unsterile Product; Unsealed Device Packaging
AS REPORTED, THE INNER POUCH PACKAGING OF A 4F MULTI-PURPOSE A1 (I) 125CM TEMPO DIAGNOSTIC CATHETER WAS DAMAGED. THE STERILE POUCH WAS RECEIVED OPEN/COMPROMISED. THE PRODUCT WAS UNUSED. IT IS NOT POSSIBLE THAT THE PRODUCT HAD BEEN PURPOSELY OPENED IN ANTICIPATION OF USE AND THEN RESHELVED. THE CONDITION WAS NOTED AFTER IT HAD BEEN RECEIVED AND STORED IN THE LAB PRIOR TO USE. IN THE LAB, IT WAS STORED IN THE CATH LAB AND MANAGED BY A SPECIALIST. THE ACTUAL PRODUCT WAS NOT DAMAGED. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20619758 ↗MalfunctionOct 12, 2020Tear, Rip or Hole in Device Packaging; Difficult to Open or Remove Packaging Material; Tear, Rip or Hole in Device Packaging; Difficult to Open or Remove Packaging Material
AS REPORTED, THE PACKAGING OF 125CM MULTIPURPOSE (MP) A1 TEMPO ANGIOGRAPHIC CATHETER THAT INCLUDES THE STERILE PACKAGING WAS DAMAGED. THERE WAS NO REPORTED PATIENT INJURY. THE DEVICE WAS STORED AND HANDLED PER THE INSTRUCTIONS FOR USE (IFU). THE OUTER PRODUCT BOX AND THE INNER STERILE POUCH WAS CRACKED. THE INTEGRITY OF THE STERILE POUCH WAS COMPROMISED. THE DAMAGE WAS NOTED AFTER IT HAD BEEN RECEIVED AND STORED IN THE LAB PRIOR TO USING. THE ACTUAL PRODUCT WAS NOT DAMAGED. THE PRODUCT WAS NOT USED IN PATIENT. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10663296 ↗
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H74908526031Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗