TEMPO 451415H0Recalls & reported problems
Cordis Corporation · CATH TEMPO 4F BER II 100CM · Class 2
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10705032007863), received between Jun 24, 2019 and Nov 5, 2025 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Degraded” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Degraded4
- Detachment of Device or Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 5, 2025Degraded; Degraded
AS REPORTED, A 5 F TEMPO AQUA 0.038" 65CM CATHETER BEGAN CRUMBLING DURING A PROCEDURE. THE CATHETER WAS NOT USED AT THE TIME OF NOTICING THE FLAKING. A DIFFERENT UNKNOWN CATHETER WAS USED. THE PATIENT LATER EXPIRED THAT DAY. IT IS STATED THAT THE TECH WIPED HIS HANDS AND NOTICED BLUE FLAKES ON HIS GLOVES AND CHECKED THE CATHETER AND NOTICED IT WAS "CRUMBLING", AND THAT NONE OF THE CATHETER FLAKED OFF IN THE PATIENT'S BODY, JUST ON THE STERILE FIELD. THE CATHETER DID NOT FLAKE OFF INSIDE THE PATIENT¿S BODY; THE FLAKING OCCURRED ON THE STERILE FIELD AND WAS NOTICED DURING THE PROCEDURE BEFORE BEING PASSED TO THE PHYSICIAN. THE DEVICE WAS PREPPED AND FLUSHED, WITH NO DIFFICULTIES ENCOUNTERED DURING PREPARATION. THERE WAS NO NEED TO REPLACE THE STERILE FIELD, GLOVES, OR GOWN AFTER NOTICING THE FLAKING. THE PROCEDURE WAS A PERIPHERAL INTERVENTION CONDUCTED ON AN OUTPATIENT BASIS FOR PERIPHERAL ARTERY DISEASE (PAD). THE PATIENT DIED DUE TO AN ACCESS SITE BLEED AND EXHIBITED HYPOTENSION PRIOR TO EXPIRING. THE DEVICE WAS NOT EXPOSED TO HEAT OR STORED IN A DARK LOCATION, AND THE FACILITY DOES NOT USE UV LIGHTING OR OZONE FOR STERILIZATION OR MICROBIAL REDUCTION. THE TARGET SITE WAS THE COMM
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23478275 ↗MalfunctionNov 5, 2025Degraded; Degraded
AS REPORTED, A 5 F TEMPO AQUA 0.038" 65CM CATHETER BEGAN CRUMBLING DURING A PROCEDURE. THE CATHETER WAS NOT USED AT THE TIME OF NOTICING THE FLAKING. A DIFFERENT UNKNOWN CATHETER WAS USED. THE PATIENT LATER EXPIRED THAT DAY. IT IS STATED THAT THE TECH WIPED HIS HANDS AND NOTICED BLUE FLAKES ON HIS GLOVES AND CHECKED THE CATHETER AND NOTICED IT WAS "CRUMBLING", AND THAT NONE OF THE CATHETER FLAKED OFF IN THE PATIENT'S BODY, JUST ON THE STERILE FIELD. THE CATHETER DID NOT FLAKE OFF INSIDE THE PATIENT¿S BODY; THE FLAKING OCCURRED ON THE STERILE FIELD AND WAS NOTICED DURING THE PROCEDURE BEFORE BEING PASSED TO THE PHYSICIAN. THE DEVICE WAS PREPPED AND FLUSHED, WITH NO DIFFICULTIES ENCOUNTERED DURING PREPARATION. THERE WAS NO NEED TO REPLACE THE STERILE FIELD, GLOVES, OR GOWN AFTER NOTICING THE FLAKING. THE PROCEDURE WAS A PERIPHERAL INTERVENTION CONDUCTED ON AN OUTPATIENT BASIS FOR PERIPHERAL ARTERY DISEASE (PAD). THE PATIENT DIED DUE TO AN ACCESS SITE BLEED AND EXHIBITED HYPOTENSION PRIOR TO EXPIRING. THE DEVICE WAS NOT EXPOSED TO HEAT OR STORED IN A DARK LOCATION, AND THE FACILITY DOES NOT USE UV LIGHTING OR OZONE FOR STERILIZATION OR MICROBIAL REDUCTION. THE TARGET SITE WAS THE COMM
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23478206 ↗MalfunctionNov 5, 2025Degraded; Degraded
AS REPORTED, A 5 F TEMPO AQUA 0.038" 65CM CATHETER BEGAN CRUMBLING DURING A PROCEDURE. THE CATHETER WAS NOT USED AT THE TIME OF NOTICING THE FLAKING. A DIFFERENT UNKNOWN CATHETER WAS USED. THE PATIENT LATER EXPIRED THAT DAY. IT IS STATED THAT THE TECH WIPED HIS HANDS AND NOTICED BLUE FLAKES ON HIS GLOVES AND CHECKED THE CATHETER AND NOTICED IT WAS "CRUMBLING", AND THAT NONE OF THE CATHETER FLAKED OFF IN THE PATIENT'S BODY, JUST ON THE STERILE FIELD. THE CATHETER DID NOT FLAKE OFF INSIDE THE PATIENT¿S BODY; THE FLAKING OCCURRED ON THE STERILE FIELD AND WAS NOTICED DURING THE PROCEDURE BEFORE BEING PASSED TO THE PHYSICIAN. THE DEVICE WAS PREPPED AND FLUSHED, WITH NO DIFFICULTIES ENCOUNTERED DURING PREPARATION. THERE WAS NO NEED TO REPLACE THE STERILE FIELD, GLOVES, OR GOWN AFTER NOTICING THE FLAKING. THE PROCEDURE WAS A PERIPHERAL INTERVENTION CONDUCTED ON AN OUTPATIENT BASIS FOR PERIPHERAL ARTERY DISEASE (PAD). THE PATIENT DIED DUE TO AN ACCESS SITE BLEED AND EXHIBITED HYPOTENSION PRIOR TO EXPIRING. THE DEVICE WAS NOT EXPOSED TO HEAT OR STORED IN A DARK LOCATION, AND THE FACILITY DOES NOT USE UV LIGHTING OR OZONE FOR STERILIZATION OR MICROBIAL REDUCTION. THE TARGET SITE WAS THE COMM
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23478080 ↗MalfunctionNov 5, 2025Degraded; Degraded
AS REPORTED, A 5 F TEMPO AQUA 0.038" 65CM CATHETER BEGAN CRUMBLING DURING A PROCEDURE. THE CATHETER WAS NOT USED AT THE TIME OF NOTICING THE FLAKING. A DIFFERENT UNKNOWN CATHETER WAS USED. THE PATIENT LATER EXPIRED THAT DAY. IT IS STATED THAT THE TECH WIPED HIS HANDS AND NOTICED BLUE FLAKES ON HIS GLOVES AND CHECKED THE CATHETER AND NOTICED IT WAS "CRUMBLING", AND THAT NONE OF THE CATHETER FLAKED OFF IN THE PATIENT'S BODY, JUST ON THE STERILE FIELD. THE CATHETER DID NOT FLAKE OFF INSIDE THE PATIENT¿S BODY; THE FLAKING OCCURRED ON THE STERILE FIELD AND WAS NOTICED DURING THE PROCEDURE BEFORE BEING PASSED TO THE PHYSICIAN. THE DEVICE WAS PREPPED AND FLUSHED, WITH NO DIFFICULTIES ENCOUNTERED DURING PREPARATION. THERE WAS NO NEED TO REPLACE THE STERILE FIELD, GLOVES, OR GOWN AFTER NOTICING THE FLAKING. THE PROCEDURE WAS A PERIPHERAL INTERVENTION CONDUCTED ON AN OUTPATIENT BASIS FOR PERIPHERAL ARTERY DISEASE (PAD). THE PATIENT DIED DUE TO AN ACCESS SITE BLEED AND EXHIBITED HYPOTENSION PRIOR TO EXPIRING. THE DEVICE WAS NOT EXPOSED TO HEAT OR STORED IN A DARK LOCATION, AND THE FACILITY DOES NOT USE UV LIGHTING OR OZONE FOR STERILIZATION OR MICROBIAL REDUCTION. THE TARGET SITE WAS THE COMM
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23475589 ↗InjuryJun 24, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, A 4F TEMPO BER II 100CM CATHETER DISTAL TIP BECAME DETACHED CAUSING EMBOLIZATION OF THE POSTERIOR TIBIAL ARTERY. ADDITIONALLY, THE INCIDENT HAS LED TO A PROCEDURE DELAY IN ORDER TO RETRIEVE THE CATHETER¿S TIP. THE TARGET LESION WAS THE SUPERFICIAL FEMORAL ARTERY (SFA). THERE WAS SEVERE CALCIFICATION. THERE WAS NOT ANY VESSEL TORTUOSITY. THERE WAS NOT ANY VESSEL ANGULATION. THE DEVICE WAS BEING USED TO TREAT CHRONIC TOTAL OCCLUSION (CTO). THE DEVICE WAS NOT RESTERILIZED. THERE WERE NOT ANY ANOMALIES NOTED WHEN REMOVED FROM THE PACKAGE. THERE WERE NOT ANY ANOMALIES NOTED DURING PREP. THE DEVICE WAS NO INSERTED THROUGH A STOPCOCK INSTEAD OF A HEMOSTATIC VALVE. RESISTANCE WAS MET WHILE ADVANCING THE DEVICE. EXCESSIVE TORQUING WAS REQUIRED. RESISTANCE WAS MET ONLY WHEN CROSSING THE OCCLUSION. THE DEVICE DID NOT KINK IN THE AREA OF SEPARATION. RESISTANCE WAS NOT MET WHILE WITHDRAWING THE DEVICE. THE DISTAL END SEPARATED AT THE TIP.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8728303 ↗
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H74908526031Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.