TEMPO 451457V0Recalls & reported problems

Cordis Corporation · CATH TEMPO 4F RIM 65CM · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10705032008174), received between Aug 4, 2025 and Sep 15, 2025 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Crack” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Crack2
  • Degraded2
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjurySep 15, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    AS REPORTED, A 4F 65CM RIM TEMPO DIAGNOSTIC CATHETER TORE WHEN WITHDRAWN OVER THE BIFURCATION AND HAD TO BE RETRIEVED WITH A SNARE. THERE WAS NO REPORTED PATIENT INJURY. DURING USE, THE DEVICE CRACKED ; THERE WAS NO DEVICE SEPARATION, AND THE CATHETER DID NOT BECOME ENTRAPPED. SNARE RETRIEVAL WAS IDENTIFIED AS THE ANTICIPATED METHOD; HOWEVER, NO SEPARATION OCCURRED. THE PRODUCT WAS STORED/HANDLED AND PREPPED PER THE IFU, WITH NO DAMAGE NOTED PRIOR TO OPENING AND NO DIFFICULTY REMOVING FROM PACKAGING OR DURING PREPARATION. THE DEVICE WAS REMOVED FROM STERILE PACKAGING BY THE HUB AND WAS NOT TORQUED/STEERED BY THE HUB. TARGET VESSEL CHARACTERISTICS INCLUDED MILD CALCIFICATION. THE ACCESS SITE VESSEL CHARACTERISTICS WERE NOT REPORTED. PROCEDURAL IMAGING (CD) IS NOT AVAILABLE. THE DEVICE WAS RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23051580 ↗
  • MalfunctionAug 4, 2025Crack; Degraded; Crack; Degraded

    AS REPORTED, TWO 4F RIM 65CM TEMPO DIAGNOSTIC CATHETERS WERE FOUND TO BE STARTING TO CRACK, IDENTIFIED ACROSS TWO SEPARATE LOT NUMBERS. THERE WAS NO REPORTED PATIENT INJURY. THERE WERE DAMAGES FOUND ON THE DEVICE OR DEVICE PACKAGING PRIOR TO OPENING. THE CRACKS ON THE DEVICES WERE FOUND BEFORE USE, WHILE ON THE SHELF. THE DEVICE WAS STORED AND PREPPED IN ACCORDANCE WITH THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO REPORTED PATIENT INJURY. THE DEVICE WAS NOT TORQUED OR "STEERED" BY THE HUB AS IT WAS NEVER OPENED. THE USER WAS TRAINED IN THE USE OF THE DEVICE. THOUGH ONLY 2 DEVICES WERE FOUND TO HAVE CRACKS, 7 DEVICES TOTAL ARE BEING RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22709393 ↗
  • MalfunctionAug 4, 2025Crack; Degraded; Crack; Degraded

    AS REPORTED, TWO 4F RIM 65CM TEMPO DIAGNOSTIC CATHETERS WERE FOUND TO BE STARTING TO CRACK, IDENTIFIED ACROSS TWO SEPARATE LOT NUMBERS. THERE WAS NO REPORTED PATIENT INJURY. THERE WERE DAMAGES FOUND ON THE DEVICE OR DEVICE PACKAGING PRIOR TO OPENING. THE CRACKS ON THE DEVICES WERE FOUND BEFORE USE, WHILE ON THE SHELF. THE DEVICE WAS STORED AND PREPPED IN ACCORDANCE WITH THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO REPORTED PATIENT INJURY. THE DEVICE WAS NOT TORQUED OR "STEERED" BY THE HUB AS IT WAS NEVER OPENED. THE USER WAS TRAINED IN THE USE OF THE DEVICE. THOUGH ONLY 2 DEVICES WERE FOUND TO HAVE CRACKS, 7 DEVICES TOTAL ARE BEING RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22708557 ↗

Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices

All CATHETER, INTRAVASCULAR, DIAGNOSTIC devices →