TEMPO 451503H5Recalls & reported problems
Cordis Corporation · CATH TEMPO 5F PIG 100CM 5SH · Class 2
All models of TEMPO (56)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 10705032008273), received between Oct 27, 2020 and Sep 30, 2026 — classified as 1 injury, 15 malfunctions.
The most frequently reported specific device problem is “Device Contamination with Chemical or Other Material” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Device Contamination with Chemical or Other Material6
- Detachment of Device or Device Component4
- Crack2
- Delivered as Unsterile Product1
- Device-Device Incompatibility1
- Difficult to Remove1
- Material Puncture/Hole1
Patient problems
- No Clinical Signs, Symptoms or Conditions15
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Crack
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949105 ↗MalfunctionJul 27, 2026Detachment of Device or Device Component
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50025079 ↗MalfunctionJan 16, 2026Device-Device Incompatibility
AS REPORTED, THE PORT OF A 5F PIG 100CM TEMPO DIAGNOSTIC CATHETER WAS FOUND TO BE UNEVEN AND COULD NOT CONNECT TO THE Y-CONNECTOR WHILE PREPARING TO PERFORM A PROCEDURE. THE CATHETER WAS IMMEDIATELY REPLACED, AND THE REPLACEMENT FUNCTIONED NORMALLY. THERE WAS NO REPORTED PATIENT INJURY. THE ISSUE DELAYED THE PATIENT¿S TREATMENT. THE DEVICE WILL BE RETURNED FOR EVALUATION. THE HOSPITAL REPORTED THIS CASE AS AN ADVERSE EVENT TO THE CHINA NMPA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24097399 ↗MalfunctionNov 14, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, THE PROXIMAL SEGMENT OF THE TUBING OF A 5F 100CM 5 SIDE HOLE (SH) PIGTAIL (PIG) TEMPO ANGIOGRAPHIC CATHETER WAS FOUND TO BE INCOMPLETE BEFORE THE PATIENT'S INTERVENTIONAL PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY. THE EVENT WAS REPORTED TO THE CHINA NATIONAL MEDICAL PRODUCTS ADMINISTRATION (NMPA) AS AN ADVERSE EVENT. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23553547 ↗MalfunctionJul 28, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
AS REPORTED, THE PRODUCT PACKAGING OF THE 5F 100CM 5 SIDE-HOLE (SH) PIGTAILED TEMPO DIAGNOSTIC CATHETER WAS DAMAGED, AFFECTING STERILITY. THERE WERE NO REPORTS OF PATIENT INJURY. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22630693 ↗InjuryJun 24, 2025Difficult to Remove; Difficult to Remove
AS REPORTED, THE 5F 100CM PIGTAILED TEMPO DIAGNOSTIC CATHETER DID NOT PASS SMOOTHLY DURING THE PROCEDURE. THE CATHETER WAS REPEATEDLY INTRODUCED SEVERAL TIMES IN THE RIGHT COMMON FEMORAL VEIN AND SAPHENOUS VEIN, AND A KINK OCCURRED IN THE TEMPO CATHETER AT THE PROXIMAL END OF THE LEFT FEMORAL VEIN PUNCTURE SHEATH. THE NEXT OPERATION COULD NOT BE PERFORMED. IN ORDER NOT TO DELAY THE PATIENT'S TREATMENT, THE EXTRACORPOREAL SECTION OF THE TEMPO CATHETER WAS CUT OFF, THE 8F VASCULAR SHEATH WAS REPLACED, THE KINKED TEMPO CATHETER WAS WITHDRAWN AND WAS REPLACED WITH A NEW TEMPO CATHETER. THE INTERVENTION WAS SUCCESSFULLY COMPLETED. THE PATIENT RETURNED TO THE WARD SAFELY. THERE WERE NO REPORTS OF PATIENT INJURY. THE PATIENT'S RIGHT SAPHENOUS VEIN AND FEMORAL VEIN THROMBOSIS HAD BEEN INEFFECTIVE WITH CONSERVATIVE MEDICAL TREATMENT AND THE PATIENT HAD UNDERGONE INFERIOR VENA CAVA FILTER PLACEMENT AND THROMBUS ASPIRATION. THE TEMPO CATHETER WAS INTRODUCED FOR IMAGING TO UNDERSTAND THE THROMBUS SITUATION, BUT THAT IS WHEN IT WAS DISCOVERED THAT THE CATHETER COULD NOT PASS SMOOTHLY. THE HOSPITAL REPORTED THIS CASE AS AN ADVERSE EVENT TO CHINA NMPA. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22318395 ↗MalfunctionNov 14, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, WHEN THE 5F PIGTAIL 100CM 5-SIDEHOLE (SH) TEMPO DIAGNOSTIC CATHETER WAS TAKEN OUT OF THE PACKAGE, THE CATHETER SEPARATED INTO TWO PARTS. THERE WAS NO REPORTED PATIENT INJURY, AS THE DEVICE WAS NOT USED IN THE PATIENT. IT WAS STORED CORRECTLY LIKE ALL THE PRODUCTS THAT ARE ALWAYS USED IN THIS HOSPITAL, WITH NO DAMAGES FOUND ON THE DEVICE PACKAGING PRIOR TO OPENING. THE DEVICE WAS STORED WHERE ALL MEDICAL PRODUCTS ARE STORED, IN A WAREHOUSE WITH ACCESS CONTROL VIA PASSWORD. NO PHOTOS OF THE DEVICE WERE AVAILABLE AS THEY WERE DISCARDED. THE DEVICE WILL NOT BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20684022 ↗MalfunctionOct 24, 2024Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
AS REPORTED, A WHITE WAXY SUBSTANCE WAS FOUND IN THE TUBES OF TWO 5F 100 CM 5-SIDE HOLE (SH) TEMPO DIAGNOSTIC CATHETERS DURING PREP AND THE HARDNESS OF THE TUBE BODY WAS SIGNIFICANTLY INCREASED COMPARED WITH BEFORE. DUE TO THE ISSUE, A THIRD PIGTAIL ANGIOGRAPHIC TUBE WAS USED TO COMPLETE THE OPERATION. THERE WAS NO REPORTED PATIENT INJURY. THE DISCOVERY OF THIS ISSUE WAS MADE DURING A LOWER LIMB ARTERY INTERVENTIONAL SURGERY. THE DEVICE WAS STORED AND PREPPED IN ACCORDANCE WITH THE INSTRUCTIONS FOR USE (IFU). THERE WERE NO DAMAGES FOUND ON THE DEVICE PACKAGING. A NON-CORDIS GUIDEWIRE WAS USED IN THE PATIENT, BUT THE CONTRAST CATHETER WITH WHITE WAX WAS NOT IMPLANTED IN THE PATIENT. THE PRODUCTS WERE DISCARDED AT SITE AND WILL NOT BE RETURNED. A PICTURE WAS RECEIVED FOR REVIEW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20520634 ↗MalfunctionOct 24, 2024Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
AS REPORTED, A WHITE WAXY SUBSTANCE WAS FOUND IN THE TUBES OF TWO 5F 100 CM 5-SIDE HOLE (SH) TEMPO DIAGNOSTIC CATHETERS DURING PREP AND THE HARDNESS OF THE TUBE BODY WAS SIGNIFICANTLY INCREASED COMPARED WITH BEFORE. DUE TO THE ISSUE, A THIRD PIGTAIL ANGIOGRAPHIC TUBE WAS USED TO COMPLETE THE OPERATION. THERE WAS NO REPORTED PATIENT INJURY. THE DISCOVERY OF THIS ISSUE WAS MADE DURING A LOWER LIMB ARTERY INTERVENTIONAL SURGERY. THE DEVICE WAS STORED AND PREPPED IN ACCORDANCE WITH THE INSTRUCTIONS FOR USE (IFU). THERE WERE NO DAMAGES FOUND ON THE DEVICE PACKAGING. A NON-CORDIS GUIDEWIRE WAS USED IN THE PATIENT, BUT THE CONTRAST CATHETER WITH WHITE WAX WAS NOT IMPLANTED IN THE PATIENT. THE PRODUCTS WERE DISCARDED AT SITE AND WILL NOT BE RETURNED. A PICTURE WAS RECEIVED FOR REVIEW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20520584 ↗MalfunctionOct 7, 2024Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
AS REPORTED, DURING THE ASSEMBLY OF A 5F 100CM 5 SIDEHOLE (SH) TEMPO PIGTAIL DIAGNOSTIC CATHETER, WHEN THE GUIDEWIRE ENTERED THE CATHETER, THE OPERATOR OBSERVED AN OBVIOUS WHITE WAXY SUBSTANCE, WHICH DISSOLVED AFTER FLUSHING WITH WATER AND WHICH WAS REMOVED WHEN INSERTING THE GUIDEWIRE. THE OPERATOR THEN SWITCHED TO ANOTHER GUIDEWIRE OF THE SAME MODEL, AND THE SAME CONDITION WAS OBSERVED. THE GUIDEWIRE USED WAS NON-CORDIS, AND THERE WAS A LITTLE RESISTANCE NOTED WHEN INSERTING THE GUIDEWIRE, BUT IT WAS NOT OBVIOUS. A THIRD GUIDEWIRE WAS THEN USED, AND IT WAS USED NORMALLY. THERE WAS NO REPORTED PATIENT INJURY. THE DEVICE WAS STORED AND PREPPED ACCORDING TO THE IFU, AND THERE WAS NO DAMAGE TO THE DEVICE OR PACKAGING PRIOR TO USE. THE STERILE POUCH WAS RECEIVED INTACT, AND THERE WAS NO OBVIOUS DAMAGE TO THE OUTER COATING OF THE CATHETER. THE TWO CONTRAST CATHETERS HAD BEEN DISCARDED BY THE OPERATOR AND COULD NOT BE RETURNED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20384857 ↗MalfunctionOct 7, 2024Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
AS REPORTED, DURING THE ASSEMBLY OF A 5F 100CM 5 SIDEHOLE (SH) TEMPO PIGTAIL DIAGNOSTIC CATHETER, WHEN THE GUIDEWIRE ENTERED THE CATHETER, THE OPERATOR OBSERVED AN OBVIOUS WHITE WAXY SUBSTANCE, WHICH DISSOLVED AFTER FLUSHING WITH WATER AND WHICH WAS REMOVED WHEN INSERTING THE GUIDEWIRE. THE OPERATOR THEN SWITCHED TO ANOTHER GUIDEWIRE OF THE SAME MODEL, AND THE SAME CONDITION WAS OBSERVED. THE GUIDEWIRE USED WAS NON-CORDIS, AND THERE WAS A LITTLE RESISTANCE NOTED WHEN INSERTING THE GUIDEWIRE, BUT IT WAS NOT OBVIOUS. A THIRD GUIDEWIRE WAS THEN USED, AND IT WAS USED NORMALLY. THERE WAS NO REPORTED PATIENT INJURY. THE DEVICE WAS STORED AND PREPPED ACCORDING TO THE IFU, AND THERE WAS NO DAMAGE TO THE DEVICE OR PACKAGING PRIOR TO USE. THE STERILE POUCH WAS RECEIVED INTACT, AND THERE WAS NO OBVIOUS DAMAGE TO THE OUTER COATING OF THE CATHETER. THE TWO CONTRAST CATHETERS HAD BEEN DISCARDED BY THE OPERATOR AND COULD NOT BE RETURNED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20384767 ↗MalfunctionSep 27, 2024Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
AS REPORTED, WHEN A 5F 100 CM TEMPO PIGTAIL CATHETER WAS ASSEMBLED OUTSIDE THE PATIENT, A WHITE WAX-LIKE SUBSTANCE, MOVEABLE, SOLUBLE IN WATER, WAS FOUND AT THE TIP OF THE CATHETER THAT WAS BEING USED FOR PREOPERATIVE IMAGING OF VENA CAVA FILTER IMPLANTATION. THE CATHETER WAS DISCARDED AND A NEW TEMPO PIGTAIL CATHETER WAS USED FOR IMAGING. THERE WAS NO REPORTED PATIENT INJURY. THE DEVICE WAS STORED AS PER THE INSTRUCTIONS FOR USE (IFU) AT ROOM TEMPERATURE AWAY FROM LIGHT. NO ACTUAL PRODUCT DAMAGE WAS OBSERVED. THERE WAS NO DAMAGE TO THE SHIPPING BOX, OUTER PRODUCT BOX, OR INNER PRODUCT POUCH. THE STERILE POUCH RECEIVED WAS NOT OPEN OR COMPROMISED. THE INNER PACKAGE SEAL WAS NOT OPENED IN ANTICIPATION OF USE AND THEN RESHELVED. A PICTURE WAS PROVIDED FOR REVIEW. IN THE PHOTO, A CATHETER IS OBSERVED, AND ON THE DISTAL END OF THE CATHETER, A WHITE LIKE MATERIAL IS SEEN COMING OUT OF THE END. NO OTHER OUTSTANDING DETAILS OR ANOMALIES COULD BE OBSERVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20326916 ↗MalfunctionJul 2, 2024Material Puncture/Hole; Material Puncture/Hole
AS REPORTED, DURING PREOPERATIVE PREPARATION, A BROKEN HOLE WAS FOUND AT THE PROXIMAL END OF A 5F TEMPO PIGTAIL ANGIOGRAPHIC CATHETER AND SALINE SPRAYED OUT ABNORMALLY. THE CATHETER WAS REPLACED. THERE WAS NO REPORTED PATIENT INJURY. THE HOSPITAL REPORTED IT AS AN ADVERSE EVENT TO THE CHINA NMPA DIRECTLY. PLEASE UPGRADE THIS CASE TO AE. MULTIPLE ATTEMPTS TO OBTAIN SUPPLEMENTAL INFORMATION WERE MADE; HOWEVER, ADDITIONAL EVENT DETAILS WERE NOT PROVIDED. THE DEVICE WAS NOT RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19661783 ↗MalfunctionMay 21, 2024Crack; Crack
AS REPORTED, THE BODY SHAFT OF THE 5F 100 CM TEMPO PIGTAIL ANGIOGRAPHY CATHETER WAS FOUND "FRACTURED" BEFORE CLINICAL USE, WHEN THE PACKAGE WAS OPENED, AFFECTING ITS USE. THE DEVICE WAS NOT USED IN THE PATIENT. ANOTHER TEMPO CATHETER WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY. THE PRODUCT WAS STORED AND HANDLED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO DEVICE DAMAGE NOTICED PRIOR TO OPENING THE PACKAGE OR DIFFICULTY REMOVING THE DEVICE FROM THE STERILE PACKAGING. THE DEVICE WAS NOT PREPPED AFTER THE DAMAGE WAS FOUND. NEITHER THE PRODUCT NOR ANY OF THE OTHER DEVICES USED WITH IT HAD BEEN RE-STERILIZED. THE PRODUCT WILL NOT BE RETURNED. THE HOSPITAL REPORTED IT AS AN ADVERSE EVENT TO THE CHINA NMPA DIRECTLY. PLEASE UPGRADE THIS CASE TO AE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19363264 ↗MalfunctionFeb 9, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, THE PATIENT WAS ADMITTED TO THE HOSPITAL FOR ACUTE CEREBRAL INFARCTION. THE PATIENT WAS SENT TO THE INTERVENTION ROOM FOR AORTIC ARCH + TOTAL CEREBRAL ANGIOGRAPHY AFTER ADMISSION. DURING THE PATIENT'S IMAGING PROCEDURE, THE DISTAL TIP OF TEMPO 5F 100CM PIG ANGIOGRAPHY CATHETER FRACTURED AND DISLODGED, RESULTING IN THE STUMP OF THE TUBE BEING LEFT IN THE PATIENT'S VESSEL. THE PIECE WAS IN STENOTIC SEGMENT OF THE COMMON CAROTID ARTERY. IT CAUSED INJURY TO THE PATIENT AND SERIOUSLY JEOPARDIZED THE PATIENT'S LIFE. THE SURGEON IMMEDIATELY PERFORMED ENDOVASCULAR CATHETER STUMP REMOVAL AFTER DISCOVERY. THE DISTAL TIP PIECE WAS REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THE PATIENT STATUS IS KNOWN TO BE GOOD. THE DEVICE WAS STORED AND HANDLED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THE DEVICE WAS INSPECTED AND PREPPED ACCORDING TO THE IFU. THERE WAS NOTHING UNUSUAL NOTED ABOUT THE DEVICE PRIOR TO USE. THERE WAS NO DAMAGE NOTED TO THE DEVICE NOTICED AFTER OPENING THE PACKAGE. THE DEVICE DID NOT PASS THROUGH ANY ACUTE BENDS. THE DEVICE WILL NOT BE RETURNED FOR EVALUATION AS IT WAS DISCARDED. THE HOSPITAL REPORTED IT AS AN ADVERSE EVENT TO THE CHINA
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18676656 ↗MalfunctionOct 27, 2020Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
DURING PREP, A 5F 100CM PIGTAIL (PIG) 5 SIDE HOLES (SH) TEMPO CATHETER TIP WAS CONTAMINATED. THERE WERE DIFFICULTIES REMOVING THE PRODUCT FROM THE PACKAGING. THE PACKAGING DID NOT REMOVE. THERE WAS NO REPORTED PATIENT INJURY. ANOTHER TEMPO CATHETER WAS USED TO COMPLETE THE PROCEDURE. THE TARGET LESION IS THE CAROTID ARTERY WITH NO TORTUOSITY, CALCIFICATION, AND TWENTY PERCENT STENOSIS. THE DEVICE WAS HANDLED ACCORDING TO THE INSTRUCTION FOR USE (IFU). THERE WERE NO DAMAGES TO THE DEVICE PACKAGING. THE DEVICE WILL BE RETURNED FOR EVALUATION. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10744626 ↗
FDA clearance and classification
Cleared through Special 510(k) K181836 on Aug 8, 2018 (decision: Substantially Equivalent). Submitted by Cordis, A Cardinal Health Company as “4F Infiniti Angiographic Catheter, 4F & 5F Nylex Angiography Catheters, 4F & 5F Tempo Angiography Catheters”. The FDA review took 29 days. FDA record ↗
The FDA regulates product code DQO (catheter, intravascular, diagnostic) under 21 CFR 870.1200 as a Class II device. 21 CFR 870.1200 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the TEMPO 451503H5
Has the TEMPO 451503H5 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705032008273), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the TEMPO 451503H5?
The FDA's MAUDE database lists 16 adverse event reports for this device, received between Oct 27, 2020 and Sep 30, 2026. In those reports, the most common specific device problems are “Device Contamination with Chemical or Other Material” (6), “Detachment of Device or Device Component” (4) and “Crack” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the TEMPO 451503H5?
None of the reports are classified as deaths, 1 as an injury and 15 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the TEMPO 451503H5 FDA approved?
It was cleared through Special 510(k) K181836 on Aug 8, 2018. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the TEMPO 451503H5, and is it still sold?
The FDA's device database lists Cordis Corporation as the company responsible for it (brand TEMPO, model 451503H5). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H74908526031Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page