TEMPO 451515H0Recalls & reported problems

Cordis Corporation · CATH TEMPO 5F BER II 100CM · Class 2

All models of TEMPO (56)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10705032008471), received between Mar 25, 2019 and Dec 10, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Crack” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Crack3
  • Degraded1
  • Leak/Splash1

Patient problems

  • No Known Impact Or Consequence To Patient3
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 10, 2025Degraded; Degraded

    AS REPORTED, THE BLUE COVER OF TWO 5F PITGTAIL 90CM 5 SIDE HOLES AND ONE 5F BERENSTEIN II 100CM TEMPO DIAGNOSTIC CATHETERS WERE PEELING. THERE WAS NO REPORTED PATIENT INJURY AS THEY WERE NOT USED ON A PATIENT. THE DEVICES WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23767804 ↗
  • MalfunctionApr 5, 2019Crack; Crack

    AS REPORTED, A QUANTITY OF THREE 5F TEMPO BER II 100CM DIAGNOSTIC CATHETERS WERE BEING USED FOR A THORACIC ENDOVASCULAR ANEURYSM REPAIR (TEVAR), THE FIRST CATHETER WAS FOUND TO BE LEAKING WHEN BEING FLUSHED AND THE OTHER TWO BROKE WHEN BEING PUT ON THE WIRE. THERE WAS NO REPORTED PATIENT INJURY. THE SECOND AND THIRD CATHETERS WERE NOT USED INSIDE THE PATIENT. A NEW CATHETER WAS USED TO COMPLETE THE PROCEDURE. AT THE TIME ALL OF THE CATHETERS WERE IN DATE, A FEW WERE ABOUT TO EXPIRE. ANOTHER CORDIS 5F TEMPO BER 100CM CATHETER WAS USED TO COMPLETE THE PROCEDURE. NO ANOMALIES WERE NOTED WHEN THE FIRST DEVICE WAS REMOVED FROM THE PACKAGING. THE FIRST DEVICE WAS NOT RESTERILIZED. THE FIRST DEVICE WAS FLUSHED WITH HEPARINIZED SALINE FOR THE INSERTION OF THE GUIDEWIRE. UPON FLUSHING THE FIRST DEVICE, IT WAS NOTICED THAT THERE WAS A CLEAN BREAK MID-SHAFT OF THE CATHETER. THE SECOND DEVICE WAS REMOVED FROM THE PACKAGING AND WAS FLEXED TO MAKE SURE IT WAS STRONG, IT SNAPPED IN FOUR PLACES. NO FURTHER ANOMALIES WERE NOTED DURING PREP OF THE SECOND DEVICE. THE SECOND DEVICE WAS DISCARDED. THE THIRD DEVICE WAS REMOVED FROM THE PACKAGING AND WAS FLEXED TO MAKE SURE IT WAS STRONG, IT SNAPPED IN F

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8488704 ↗
  • MalfunctionApr 5, 2019Crack; Crack

    AS REPORTED, A QUANTITY OF THREE 5F TEMPO BER II 100CM DIAGNOSTIC CATHETERS WERE BEING USED FOR A THORACIC ENDOVASCULAR ANEURYSM REPAIR (TEVAR), THE FIRST CATHETER WAS FOUND TO BE LEAKING WHEN BEING FLUSHED AND THE OTHER TWO BROKE WHEN BEING PUT ON THE WIRE. THERE WAS NO REPORTED PATIENT INJURY. THE SECOND AND THIRD CATHETERS WERE NOT USED INSIDE THE PATIENT. A NEW CATHETER WAS USED TO COMPLETE THE PROCEDURE. AT THE TIME ALL OF THE CATHETERS WERE IN DATE, A FEW WERE ABOUT TO EXPIRE. ANOTHER CORDIS 5F TEMPO BER 100CM CATHETER WAS USED TO COMPLETE THE PROCEDURE. NO ANOMALIES WERE NOTED WHEN THE FIRST DEVICE WAS REMOVED FROM THE PACKAGING. THE FIRST DEVICE WAS NOT RESTERILIZED. THE FIRST DEVICE WAS FLUSHED WITH HEPARINIZED SALINE FOR THE INSERTION OF THE GUIDEWIRE. UPON FLUSHING THE FIRST DEVICE, IT WAS NOTICED THAT THERE WAS A CLEAN BREAK MID-SHAFT OF THE CATHETER. THE SECOND DEVICE WAS REMOVED FROM THE PACKAGING AND WAS FLEXED TO MAKE SURE IT WAS STRONG, IT SNAPPED IN FOUR PLACES. NO FURTHER ANOMALIES WERE NOTED DURING PREP OF THE SECOND DEVICE. THE SECOND DEVICE WAS DISCARDED. THE THIRD DEVICE WAS REMOVED FROM THE PACKAGING AND WAS FLEXED TO MAKE SURE IT WAS STRONG, IT SNAPPED IN F

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8488629 ↗
  • MalfunctionMar 25, 2019Crack; Leak/Splash; Crack; Leak/Splash

    AS REPORTED, A QUANTITY OF THREE 5F TEMPO BER II 100CM DIAGNOSTIC CATHETERS WERE BEING USED FOR A THORACIC ENDOVASCULAR ANEURYSM REPAIR (TEVAR), THE FIRST CATHETER WAS FOUND TO BE LEAKING WHEN BEING FLUSHED AND THE OTHER TWO BROKE WHEN BEING PUT ON THE WIRE. THERE WAS NO REPORTED PATIENT INJURY. THE SECOND AND THIRD CATHETERS WERE NOT USED INSIDE THE PATIENT. A NEW CATHETER WAS USED TO COMPLETE THE PROCEDURE. AT THE TIME ALL OF THE CATHETERS WERE IN DATE, A FEW WERE ABOUT TO EXPIRE. ANOTHER CORDIS 5F TEMPO BER 100CM CATHETER WAS USED TO COMPLETE THE PROCEDURE. NO ANOMALIES WERE NOTED WHEN THE FIRST DEVICE WAS REMOVED FROM THE PACKAGING. THE FIRST DEVICE WAS NOT RESTERILIZED. THE FIRST DEVICE WAS FLUSHED WITH HEPARINIZED SALINE FOR THE INSERTION OF THE GUIDEWIRE. UPON FLUSHING THE FIRST DEVICE, IT WAS NOTICED THAT THERE WAS A CLEAN BREAK MID-SHAFT OF THE CATHETER. THE SECOND DEVICE WAS REMOVED FROM THE PACKAGING AND WAS FLEXED TO MAKE SURE IT WAS STRONG, IT SNAPPED IN FOUR PLACES. NO FURTHER ANOMALIES WERE NOTED DURING PREP OF THE SECOND DEVICE. THE SECOND DEVICE WAS DISCARDED. THE THIRD DEVICE WAS REMOVED FROM THE PACKAGING AND WAS FLEXED TO MAKE SURE IT WAS STRONG, IT SNAPPED IN F

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8448035 ↗

FDA clearance and classification

Cleared through Special 510(k) K181836 on Aug 8, 2018 (decision: Substantially Equivalent). Submitted by Cordis, A Cardinal Health Company as “4F Infiniti Angiographic Catheter, 4F & 5F Nylex Angiography Catheters, 4F & 5F Tempo Angiography Catheters”. The FDA review took 29 days. FDA record ↗

The FDA regulates product code DQO (catheter, intravascular, diagnostic) under 21 CFR 870.1200 as a Class II device. 21 CFR 870.1200 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the TEMPO 451515H0

Has the TEMPO 451515H0 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10705032008471), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the TEMPO 451515H0?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Mar 25, 2019 and Dec 10, 2025. In those reports, the most common specific device problems are “Crack” (3), “Degraded” (1) and “Leak/Splash” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the TEMPO 451515H0?

None of the reports are classified as deaths, 0 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the TEMPO 451515H0 FDA approved?

It was cleared through Special 510(k) K181836 on Aug 8, 2018. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the TEMPO 451515H0, and is it still sold?

The FDA's device database lists Cordis Corporation as the company responsible for it (brand TEMPO, model 451515H0). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.