TEMPO 451515H0Recalls & reported problems
Cordis Corporation · CATH TEMPO 5F BER II 100CM · Class 2
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10705032008471), received between Mar 25, 2019 and Dec 10, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Crack” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Crack3
- Degraded1
- Leak/Splash1
Patient problems
- No Known Impact Or Consequence To Patient3
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 10, 2025Degraded; Degraded
AS REPORTED, THE BLUE COVER OF TWO 5F PITGTAIL 90CM 5 SIDE HOLES AND ONE 5F BERENSTEIN II 100CM TEMPO DIAGNOSTIC CATHETERS WERE PEELING. THERE WAS NO REPORTED PATIENT INJURY AS THEY WERE NOT USED ON A PATIENT. THE DEVICES WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23767804 ↗MalfunctionApr 5, 2019Crack; Crack
AS REPORTED, A QUANTITY OF THREE 5F TEMPO BER II 100CM DIAGNOSTIC CATHETERS WERE BEING USED FOR A THORACIC ENDOVASCULAR ANEURYSM REPAIR (TEVAR), THE FIRST CATHETER WAS FOUND TO BE LEAKING WHEN BEING FLUSHED AND THE OTHER TWO BROKE WHEN BEING PUT ON THE WIRE. THERE WAS NO REPORTED PATIENT INJURY. THE SECOND AND THIRD CATHETERS WERE NOT USED INSIDE THE PATIENT. A NEW CATHETER WAS USED TO COMPLETE THE PROCEDURE. AT THE TIME ALL OF THE CATHETERS WERE IN DATE, A FEW WERE ABOUT TO EXPIRE. ANOTHER CORDIS 5F TEMPO BER 100CM CATHETER WAS USED TO COMPLETE THE PROCEDURE. NO ANOMALIES WERE NOTED WHEN THE FIRST DEVICE WAS REMOVED FROM THE PACKAGING. THE FIRST DEVICE WAS NOT RESTERILIZED. THE FIRST DEVICE WAS FLUSHED WITH HEPARINIZED SALINE FOR THE INSERTION OF THE GUIDEWIRE. UPON FLUSHING THE FIRST DEVICE, IT WAS NOTICED THAT THERE WAS A CLEAN BREAK MID-SHAFT OF THE CATHETER. THE SECOND DEVICE WAS REMOVED FROM THE PACKAGING AND WAS FLEXED TO MAKE SURE IT WAS STRONG, IT SNAPPED IN FOUR PLACES. NO FURTHER ANOMALIES WERE NOTED DURING PREP OF THE SECOND DEVICE. THE SECOND DEVICE WAS DISCARDED. THE THIRD DEVICE WAS REMOVED FROM THE PACKAGING AND WAS FLEXED TO MAKE SURE IT WAS STRONG, IT SNAPPED IN F
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8488704 ↗MalfunctionApr 5, 2019Crack; Crack
AS REPORTED, A QUANTITY OF THREE 5F TEMPO BER II 100CM DIAGNOSTIC CATHETERS WERE BEING USED FOR A THORACIC ENDOVASCULAR ANEURYSM REPAIR (TEVAR), THE FIRST CATHETER WAS FOUND TO BE LEAKING WHEN BEING FLUSHED AND THE OTHER TWO BROKE WHEN BEING PUT ON THE WIRE. THERE WAS NO REPORTED PATIENT INJURY. THE SECOND AND THIRD CATHETERS WERE NOT USED INSIDE THE PATIENT. A NEW CATHETER WAS USED TO COMPLETE THE PROCEDURE. AT THE TIME ALL OF THE CATHETERS WERE IN DATE, A FEW WERE ABOUT TO EXPIRE. ANOTHER CORDIS 5F TEMPO BER 100CM CATHETER WAS USED TO COMPLETE THE PROCEDURE. NO ANOMALIES WERE NOTED WHEN THE FIRST DEVICE WAS REMOVED FROM THE PACKAGING. THE FIRST DEVICE WAS NOT RESTERILIZED. THE FIRST DEVICE WAS FLUSHED WITH HEPARINIZED SALINE FOR THE INSERTION OF THE GUIDEWIRE. UPON FLUSHING THE FIRST DEVICE, IT WAS NOTICED THAT THERE WAS A CLEAN BREAK MID-SHAFT OF THE CATHETER. THE SECOND DEVICE WAS REMOVED FROM THE PACKAGING AND WAS FLEXED TO MAKE SURE IT WAS STRONG, IT SNAPPED IN FOUR PLACES. NO FURTHER ANOMALIES WERE NOTED DURING PREP OF THE SECOND DEVICE. THE SECOND DEVICE WAS DISCARDED. THE THIRD DEVICE WAS REMOVED FROM THE PACKAGING AND WAS FLEXED TO MAKE SURE IT WAS STRONG, IT SNAPPED IN F
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8488629 ↗MalfunctionMar 25, 2019Crack; Leak/Splash; Crack; Leak/Splash
AS REPORTED, A QUANTITY OF THREE 5F TEMPO BER II 100CM DIAGNOSTIC CATHETERS WERE BEING USED FOR A THORACIC ENDOVASCULAR ANEURYSM REPAIR (TEVAR), THE FIRST CATHETER WAS FOUND TO BE LEAKING WHEN BEING FLUSHED AND THE OTHER TWO BROKE WHEN BEING PUT ON THE WIRE. THERE WAS NO REPORTED PATIENT INJURY. THE SECOND AND THIRD CATHETERS WERE NOT USED INSIDE THE PATIENT. A NEW CATHETER WAS USED TO COMPLETE THE PROCEDURE. AT THE TIME ALL OF THE CATHETERS WERE IN DATE, A FEW WERE ABOUT TO EXPIRE. ANOTHER CORDIS 5F TEMPO BER 100CM CATHETER WAS USED TO COMPLETE THE PROCEDURE. NO ANOMALIES WERE NOTED WHEN THE FIRST DEVICE WAS REMOVED FROM THE PACKAGING. THE FIRST DEVICE WAS NOT RESTERILIZED. THE FIRST DEVICE WAS FLUSHED WITH HEPARINIZED SALINE FOR THE INSERTION OF THE GUIDEWIRE. UPON FLUSHING THE FIRST DEVICE, IT WAS NOTICED THAT THERE WAS A CLEAN BREAK MID-SHAFT OF THE CATHETER. THE SECOND DEVICE WAS REMOVED FROM THE PACKAGING AND WAS FLEXED TO MAKE SURE IT WAS STRONG, IT SNAPPED IN FOUR PLACES. NO FURTHER ANOMALIES WERE NOTED DURING PREP OF THE SECOND DEVICE. THE SECOND DEVICE WAS DISCARDED. THE THIRD DEVICE WAS REMOVED FROM THE PACKAGING AND WAS FLEXED TO MAKE SURE IT WAS STRONG, IT SNAPPED IN F
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8448035 ↗
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H74908526031Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.