Skip to content

TEMPO 451532H0Recalls & reported problems

Cordis Corporation · CATH TEMPO 5F SIM 3 100CM · Class 2

All models of TEMPO (56)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10705032008600), received between Jan 28, 2026 and Jan 28, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Crack” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 3 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Crack3
  • Degraded2
  • Detachment of Device or Device Component2

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 28, 2026Crack; Degraded; Detachment of Device or Device Component

    AS REPORTED, THE DISTAL PORTION AT THE CURVE OF A 5F 100 CM TEMPO CORONARY GUIDING CATHETER WAS OBSERVED TO HAVE A CRACK UPON OPENING THE PACKAGE AND SUBSEQUENTLY BROKE WHEN EXAMINED. THE DEVICE WAS NOT USED, AND ANOTHER TEMPO PRODUCT WAS SELECTED. THERE WAS NO REPORTED PATIENT INJURY. THE NURSE AND PHYSICIAN NOTED THAT THE CATHETER FELT VERY DRY, AND SIMILAR FINDINGS WERE OBSERVED WITH THREE CATHETERS FROM TWO DIFFERENT BATCHES, WHILE A CATHETER FROM ANOTHER BATCH WAS FOUND TO BE INTACT. THE PRODUCTS WERE PROPERLY STORED AND OPENED IN STERILE FIELD. NO DEVICE DAMAGE WAS OBSERVED PRIOR TO OPENING THE PACKAGE, AND NO DIFFICULTY WAS ENCOUNTERED IN REMOVING THE DEVICE FROM ITS STERILE PACKAGING. THE DEVICE WAS PREPPED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU), AND NO ISSUES WERE EXPERIENCED DURING PREPARATION. TWO IMAGES WERE PROVIDED SHOWING A CRACK AT THE DISTAL PORTION OF THE CATHETER. THE DEVICES WILL BE RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24201092 ↗
  • MalfunctionJan 28, 2026Crack; Degraded; Detachment of Device or Device Component

    AS REPORTED, THE DISTAL PORTION AT THE CURVE OF A 5F 100 CM TEMPO CORONARY GUIDING CATHETER WAS OBSERVED TO HAVE A CRACK UPON OPENING THE PACKAGE AND SUBSEQUENTLY BROKE WHEN EXAMINED. THE DEVICE WAS NOT USED, AND ANOTHER TEMPO PRODUCT WAS SELECTED. THERE WAS NO REPORTED PATIENT INJURY. THE NURSE AND PHYSICIAN NOTED THAT THE CATHETER FELT VERY DRY, AND SIMILAR FINDINGS WERE OBSERVED WITH THREE CATHETERS FROM TWO DIFFERENT BATCHES, WHILE A CATHETER FROM ANOTHER BATCH WAS FOUND TO BE INTACT. THE PRODUCTS WERE PROPERLY STORED AND OPENED IN STERILE FIELD. NO DEVICE DAMAGE WAS OBSERVED PRIOR TO OPENING THE PACKAGE, AND NO DIFFICULTY WAS ENCOUNTERED IN REMOVING THE DEVICE FROM ITS STERILE PACKAGING. THE DEVICE WAS PREPPED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU), AND NO ISSUES WERE EXPERIENCED DURING PREPARATION. TWO IMAGES WERE PROVIDED SHOWING A CRACK AT THE DISTAL PORTION OF THE CATHETER. THE DEVICES WILL BE RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24201070 ↗
  • MalfunctionJan 28, 2026Crack

    AS REPORTED, THE DISTAL PORTION AT THE CURVE OF A 5F 100 CM TEMPO CORONARY GUIDING CATHETER WAS OBSERVED TO HAVE A CRACK UPON OPENING THE PACKAGE AND SUBSEQUENTLY BROKE WHEN EXAMINED. THE DEVICE WAS NOT USED, AND ANOTHER TEMPO PRODUCT WAS SELECTED. THERE WAS NO REPORTED PATIENT INJURY. THE NURSE AND PHYSICIAN NOTED THAT THE CATHETER FELT VERY DRY, AND SIMILAR FINDINGS WERE OBSERVED WITH THREE CATHETERS FROM TWO DIFFERENT BATCHES, WHILE A CATHETER FROM ANOTHER BATCH WAS FOUND TO BE INTACT. THE PRODUCTS WERE PROPERLY STORED AND OPENED IN STERILE FIELD. NO DEVICE DAMAGE WAS OBSERVED PRIOR TO OPENING THE PACKAGE, AND NO DIFFICULTY WAS ENCOUNTERED IN REMOVING THE DEVICE FROM ITS STERILE PACKAGING. THE DEVICE WAS PREPPED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU), AND NO ISSUES WERE EXPERIENCED DURING PREPARATION. TWO IMAGES WERE PROVIDED SHOWING A CRACK AT THE DISTAL PORTION OF THE CATHETER. THE DEVICES WILL BE RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24200982 ↗

Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices

All CATHETER, INTRAVASCULAR, DIAGNOSTIC devices →