Skip to content

AVANTI 504505XRecalls & reported problems

Cordis Corporation · ACCS VESSEL DILATOR 5F .038GW · Class 2

All models of AVANTI (19)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705032009836), received between Jun 14, 2024 and Jun 14, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K970392); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 14, 2024Leak/Splash; Detachment of Device or Device Component; Leak/Splash; Detachment of Device or Device Component

    AS REPORTED, AFTER INSERTION INTO THE PATIENT A 5F .038¿ VESSEL DILATOR HAD TO BE REMOVED AS THERE WAS AN INCOMPLETE CONNECTION AT THE VALVE. THERE WAS NO REPORTED PATIENT INJURY. RESISTANCE WAS FELT PROXIMAL AT THE SIDE ARM/VALVE WHEN THREADING THE WIRE. AFTER INSERTION INTO THE PATIENT, A ¿PARALUMINATE¿ (WATER LEAKED OUT NEXT TO THE LOCK CONNECTION PART). THERE WAS SIMPLY A VALVE LEAK. THE INTENDED PROCEDURE WAS REPORTED TO BE A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) FOR WHICH A CONTRALATERAL APPROACH WAS NOT USED. THE DEVICE WAS NOT PULLED FROM THE PACKAGING BY THE HUB. THE PRODUCT WAS STORED, HANDLED, AND PREPPED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO DIFFICULTY EXPERIENCED IN PREPPING THE DEVICE. THE VESSEL DILATOR SNAPPED INTO PLACE AT THE HUB. THERE WERE NO DIFFICULTIES ENCOUNTERED DURING THE INSERTION PROCESS OR DURING THE WITHDRAWAL PROCESS. THERE WAS NO EXCESS FORCE USED DURING THE PROCEDURE. ONLY WHEN THE WIRE WAS THREADED THROUGH DID A RESISTANCE BECOME APPARENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19537882 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K970392. One submission often covers many models, so check the product and lot details in each record.

  • Z-2079-2021 · Jun 9, 2021 · Terminated
    Avanti+ Catheter Sheath Introducer - Product Usage: Is indicated for use in arterial and venous procedures requiring percutaneous introduction of intravascular devices.
  • Z-1174-2016 · Dec 31, 2015 · Terminated
    CORDIS AVANTI + CATHETER SHEATH INTRODUCER SYSTEM The product includes a Catheter Sheath Introducer (CSI) and a vessel dilator. A Catheter Sheath Introducer facilitates percutaneo

Other DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION devices

All DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION devices →