Why it was recalled
Incorrect cannula of the sheath introducer (smaller than intended).
FDA root cause category: Mixed-up of materials/components
What the company is doing
The firm, Cordis, sent an 'Urgent MEDICAL DEVICE CORRECTION (Field Safety Notice) dated 12/31/2015 to its customers. The notice described the product, problem and actions to be taken. The customers were instructed to do the following: 1) Read the Description of the problem section carefully to fully understand the issue involved. 2) Check your inventory to determine if you have any remaining affected product in your possession. Check all storage and usage locations. The purpose is to identify the product, not to remove the product. 3) Keep a copy of this notice with any affected product. 4) Review, complete, sign and return the enclosed Acknowledgement Form directly to Cordis by fax to (908) 429-8287 Alternate fax (904) 928-5077. number or scan and email to [email protected]. 5) Share this letter with anyone who needs to be informed in your facility and in any facility where potentially affected devices may have been transferred. 6) Maintain awareness of this notice until all affected product has been consumed. If vessel dilator to sheath incompatibility is experienced, you can report the incident through the standard complaint process (1-800-327-7714 Option 3). For any health care professional with medical concerns, please contact Cordis to speak to a clinician at 1-800-327-7714 Option 3, Monday through Friday from 9:00 AM to 5:00 PM EST. For questions related to the Correction that are not adequately addressed in this letter, please contact the Cordis Field Action Coordinator at (786) 313-2365. For questions related to the Acknowledgement Form and its return, please contact Cordis QA at (786) 313-8730."
Recall details
- Recall number
- Z-1174-2016
- Status
- Terminated
- Initiated by the company
- Dec 31, 2015
- Posted by the FDA
- Mar 15, 2016
- Recalling firm
- Cordis Corporation
- Quantity
- 1,985 units
- Product code
- DRE
- FDA submissions
- K970392
- Distribution
- US Nationwide Distribution including states of:AL, AZ, CA, CO, CT, FL, GA, IN, KY, LA, MA, MD, MI, MN, MO, MT, NC, NH, NJ,NY, OH, OR, PA, RI, TN, TX, VA, WA, WV, and Hawaii.
Product description
CORDIS AVANTI + CATHETER SHEATH INTRODUCER SYSTEM The product includes a Catheter Sheath Introducer (CSI) and a vessel dilator. A Catheter Sheath Introducer facilitates percutaneous entry of an intravascular device. A vessel dilator facilitates the percutaneous entry of the CSI by forming an atraumatic transition from the skin through the subcutaneous tissue to the vessel.
Code information (models, lots, serial numbers)
Catalog No. 504656X (GTIN: 20705032010389) Lot No. 17305580, Expiration date: 2018-07-31.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- AVANTI 504656XCordis Corporation · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION1 recall2 reports
Product lines
Other devices under the same FDA submission
27 devices cleared under K970392 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- AVANTI 504607XCordis US Corp. · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION20 reports
- AVANTI 504606XCordis Corporation · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION19 reports
- AVANTI 504608XCordis Corporation · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION19 reports
- AVANTI 504605XCordis US Corp. · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION16 reports
- AVANTI 504611XCordis US Corp. · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION14 reports
- AVANTI 504604XCordis US Corp. · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION7 reports
- AVANTI 402606XCordis Corporation · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION4 reports
- AVANTI 402605XCordis Corporation · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION3 reports
- AVANTI 504605SCordis US Corp. · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION3 reports
- AVANTI 504609XCordis US Corp. · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION3 reports
- AVANTI 402605PCordis Corporation · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION2 reports
- AVANTI 402607XCordis Corporation · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page