Skip to content

NYLEX 526530Recalls & reported problems

Cordis Corporation · CATH NYLEX 5F PIG TR 65CM 8SH · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705032010740), received between May 8, 2019 and May 8, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 8, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    AS REPORTED, DURING THE INJECTION OF THE CONTRAST PRODUCT WITH AN AUTOMATIC INJECTOR (40CC TO 20CC/SECOND, MAXIMUM PRESSURE 723 PSI) TO PERFORM A PULMONARY ANGIOGRAPHY, THE 5F PIGTAIL CATHETER (CATH NYLEX 5F PIG TR 65CM 8SH) BROKE AT 4CM FROM THE BASE WHICH CAUSED A DELAY IN MANAGING THE PATIENT. THE DEVICE WAS RECEIVED IN TWO PIECES. (B)(6) CONFIRMED THAT THE HUB HAD SEPARATED FROM THE BODY SHAFT OF THE CATHETER.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8590821 ↗

Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices

All CATHETER, INTRAVASCULAR, DIAGNOSTIC devices →