SUPERTORQUE 532508Recalls & reported problems
Cordis Corporation · CATH ST 5F MP A2 (I) 125CM 2SH · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705032011709), received between Jan 2, 2019 and Jan 2, 2019 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Crack” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Crack2
- Difficult to Remove1
Patient problems
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 2, 2019Crack; Crack
DURING USE OF THE ANGIOGRAPHIC MULTIPURPOSE SUPER TORQUE CATHETER FOR TRANSARTERIAL CHEMOEMBOLIZATION, THE PHYSICIAN REPORTED THAT THE CATHETER WAS STUCK IN THE ARTERY AND THEN THE HUB WAS BROKEN (CRACKED) DURING INFUSION. ADDITIONAL INFORMATION WAS RECEIVED AND THERE WAS RESISTANCE WITH THE GUIDEWIRE. THERE WAS NO PATIENT INJURY. THE DEVICES WERE REMOVED SUCCESSFULLY BUT WITH SOME DIFFICULTY. IT HAPPENED WITH TWO CATHETERS WITH THE SAME LOT NUMBERS. THE TARGET WAS ARTERIA HEPATICA WHICH WAS SLIGHTLY TORTUOUS AND THE LESION WAS TREATED FOR TOTAL CHRONIC OCCLUSION IN THE PAST THREE MONTH. THE DEVICES WERE PREPPED PER THE INSTRUCTION FOR USE (IFU) WITH NO DEFECTS NOTED. A RADIAL APPROACH WAS USED. A CONTRALATERAL APPROACH WASN'T USED. THERE WERE NO DIFFICULTY TRACKING THE DEVICE TOWARDS THE LESION OR THROUGH THE VESSEL. THE CATHETER HAS NEVER BEEN IN AN ACUTE BEND. THE DEVICE WAS STORED IN THE CATH LAB PER THE IFU. THE PRODUCT WAS NOT RETURNED BUT IS AVAILABLE AS THE LAB HAS THE SAMPLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8207918 ↗MalfunctionJan 2, 2019Crack; Difficult to Remove; Crack; Difficult to Remove
DURING USE OF THE ANGIOGRAPHIC MULTIPURPOSE SUPER TORQUE CATHETER FOR TRANSARTERIAL CHEMOEMBOLIZATION, THE PHYSICIAN REPORTED THAT THE CATHETER WAS STUCK IN THE ARTERY AND THEN THE HUB WAS BROKEN (CRACKED) DURING INFUSION. ADDITIONAL INFORMATION WAS RECEIVED AND THERE WAS RESISTANCE WITH THE GUIDEWIRE. THERE WAS NO PATIENT INJURY. THE DEVICES WERE REMOVED SUCCESSFULLY BUT WITH SOME DIFFICULTY. IT HAPPENED WITH TWO CATHETERS WITH THE SAME LOT NUMBERS. THE TARGET WAS ARTERIA HEPATICA WHICH WAS SLIGHTLY TORTUOUS AND THE LESION WAS TREATED FOR TOTAL CHRONIC OCCLUSION IN THE PAST THREE MONTH. THE DEVICES WERE PREPPED PER THE INSTRUCTION FOR USE (IFU) WITH NO DEFECTS NOTED. A RADIAL APPROACH WAS USED. A CONTRALATERAL APPROACH WASN'T USED. THERE WERE NO DIFFICULTY TRACKING THE DEVICE TOWARDS THE LESION OR THROUGH THE VESSEL. THE CATHETER HAS NEVER BEEN IN AN ACUTE BEND. THE DEVICE WAS STORED IN THE CATH LAB PER THE IFU. THE PRODUCT WAS NOT RETURNED BUT IS AVAILABLE AS THE LAB HAS THE SAMPLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8207916 ↗
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H74908526031Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page