SUPERTORQUE PLUS 533532Recalls & reported problems
Cordis Corporation · CATH F5.2+ MP A2 80CM 2SH · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705032012546), received between Jan 31, 2025 and Jan 31, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Material Fragmentation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Fragmentation2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 31, 2025Material Fragmentation; Material Fragmentation
AS REPORTED, TWO 5FR MP A2 SUPERTORQUE PLUS CATHETERS CRACKED AT THE HUB WHEN A SYRINGE WAS ATTACHED. THE DEVICES WERE NOT USED IN THE PATIENT AS THE CRACK WAS NOTICED WHEN THE SYRINGE WAS ATTACHED FOR FLUSHING. THE PROCEDURE WAS COMPLETED BY OPENING AND USING A NEW MPA2 CATHETER. THERE WAS NO REPORTED INJURY TO THE PATIENT. THE PRODUCTS WERE STORED PROPERLY ACCORDING TO THE INSTRUCTIONS FOR USE (IFU), AND THERE WAS NO DIFFICULTY REMOVING THEM FROM THE PACKAGING. NO OTHER DAMAGES WERE NOTED PRIOR TO INSERTING THE PRODUCTS INTO THE PATIENT. THE DEVICES PREPPED NORMALLY, WITH A SYRINGE ATTACHED TO THE CATHETERS (X2) FOR FLUSHING. THE INTENDED PROCEDURE WAS A RIGHT HEART CATHETERIZATION FOR BLOOD SAMPLING, TARGETING NORMAL PA¿S, WITH NO CALCIFICATION IN A PEDIATRIC PATIENT. THE VESSEL HAD NO TORTUOSITY. THE SYRINGE WAS NOT TORQUED AGAINST RESISTANCE WHEN ATTACHING TO THE CATHETER. AND THE SECOND DEVICE WILL NOT BE RETURNED FOR EVALUATION. THE FIRST DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21282405 ↗MalfunctionJan 31, 2025Material Fragmentation; Material Fragmentation
AS REPORTED, TWO 5FR MP A2 SUPERTORQUE PLUS CATHETERS CRACKED AT THE HUB WHEN A SYRINGE WAS ATTACHED. THE DEVICES WERE NOT USED IN THE PATIENT AS THE CRACK WAS NOTICED WHEN THE SYRINGE WAS ATTACHED FOR FLUSHING. THE PROCEDURE WAS COMPLETED BY OPENING AND USING A NEW MPA2 CATHETER. THERE WAS NO REPORTED INJURY TO THE PATIENT. THE PRODUCTS WERE STORED PROPERLY ACCORDING TO THE INSTRUCTIONS FOR USE (IFU), AND THERE WAS NO DIFFICULTY REMOVING THEM FROM THE PACKAGING. NO OTHER DAMAGES WERE NOTED PRIOR TO INSERTING THE PRODUCTS INTO THE PATIENT. THE DEVICES PREPPED NORMALLY, WITH A SYRINGE ATTACHED TO THE CATHETERS (X2) FOR FLUSHING. THE INTENDED PROCEDURE WAS A RIGHT HEART CATHETERIZATION FOR BLOOD SAMPLING, TARGETING NORMAL PA¿S, WITH NO CALCIFICATION IN A PEDIATRIC PATIENT. THE VESSEL HAD NO TORTUOSITY. THE SYRINGE WAS NOT TORQUED AGAINST RESISTANCE WHEN ATTACHING TO THE CATHETER. AND THE SECOND DEVICE WILL NOT BE RETURNED FOR EVALUATION. THE FIRST DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21282352 ↗
Other Catheter, intravascular, diagnostic devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H74908526031Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page