INFINITI 534560TRecalls & reported problems
Cordis Corporation · CATH F5 INF IM 100CM · Class 2
All models of INFINITI (22)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10705032013963), received between Jan 23, 2024 and Sep 17, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Degraded” (3 reports).
3 other recalls cover devices cleared under the same FDA submission (K970854); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Degraded3
- Crack1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 17, 2025Degraded; Degraded
AS REPORTED, BLUE PIECES AND CRACKING OF PLASTIC NEAR THE HUB WERE FOUND ON MULTIPLE DIAGNOSTIC CATHETERS, INCLUDING 5F JL3.5 SUPER TORQUE PLUS DIAGNOSTIC CATHETERS (CATALOG NUMBER 534-518T), 5F JL4 SUPER TORQUE PLUS DIAGNOSTIC CATHETERS (CATALOG NUMBER 534-520T), 5F JL4.5 SUPER TORQUE PLUS DIAGNOSTIC CATHETERS (CATALOG NUMBER 534-517T), 5F IM SUPER TORQUE PLUS DIAGNOSTIC CATHETERS (CATALOG NUMBER 534-560T), AND A 5.2F MP A2(I) 100CM SUPER TORQUE PLUS DIAGNOSTIC CATHETER (CATALOG NUMBER 533-556). THE AFFECTED PRODUCTS WERE IDENTIFIED ACROSS MULTIPLE PACKAGES DURING STOCK INSPECTION, AND THE PACKAGES HAD NOT BEEN OPENED. THERE WAS NO REPORTED PATIENT INJURY. THE DEVICES WILL BE RETURNED FOR EVALUATION. THE DEVICES WILL BE RETURNED FOR EVALUATION. FOUR IMAGES WERE REVIEWED AND THE SHOW CATHETERS INSIDE THEIR PACKAGING. POLYMER FLAKES CAN BE SEEN INSIDE THE PACKAGING, AND THEY APPEAR TO HAVE ORIGINATED FROM THE STRAIN RELIEF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23079212 ↗MalfunctionSep 17, 2025Degraded; Degraded
AS REPORTED, BLUE PIECES AND CRACKING OF PLASTIC NEAR THE HUB WERE FOUND ON MULTIPLE DIAGNOSTIC CATHETERS, INCLUDING 5F JL3.5 SUPER TORQUE PLUS DIAGNOSTIC CATHETERS (CATALOG NUMBER 534-518T), 5F JL4 SUPER TORQUE PLUS DIAGNOSTIC CATHETERS (CATALOG NUMBER 534-520T), 5F JL4.5 SUPER TORQUE PLUS DIAGNOSTIC CATHETERS (CATALOG NUMBER 534-517T), 5F IM SUPER TORQUE PLUS DIAGNOSTIC CATHETERS (CATALOG NUMBER 534-560T), AND A 5.2F MP A2(I) 100CM SUPER TORQUE PLUS DIAGNOSTIC CATHETER (CATALOG NUMBER 533-556). THE AFFECTED PRODUCTS WERE IDENTIFIED ACROSS MULTIPLE PACKAGES DURING STOCK INSPECTION, AND THE PACKAGES HAD NOT BEEN OPENED. THERE WAS NO REPORTED PATIENT INJURY. THE DEVICES WILL BE RETURNED FOR EVALUATION. THE DEVICES WILL BE RETURNED FOR EVALUATION. FOUR IMAGES WERE REVIEWED AND THE SHOW CATHETERS INSIDE THEIR PACKAGING. POLYMER FLAKES CAN BE SEEN INSIDE THE PACKAGING, AND THEY APPEAR TO HAVE ORIGINATED FROM THE STRAIN RELIEF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23077615 ↗MalfunctionSep 17, 2025Degraded; Degraded
AS REPORTED, BLUE PIECES AND CRACKING OF PLASTIC NEAR THE HUB WERE FOUND ON MULTIPLE DIAGNOSTIC CATHETERS, INCLUDING 5F JL3.5 SUPER TORQUE PLUS DIAGNOSTIC CATHETERS (CATALOG NUMBER 534-518T), 5F JL4 SUPER TORQUE PLUS DIAGNOSTIC CATHETERS (CATALOG NUMBER 534-520T), 5F JL4.5 SUPER TORQUE PLUS DIAGNOSTIC CATHETERS (CATALOG NUMBER 534-517T), 5F IM SUPER TORQUE PLUS DIAGNOSTIC CATHETERS (CATALOG NUMBER 534-560T), AND A 5.2F MP A2(I) 100CM SUPER TORQUE PLUS DIAGNOSTIC CATHETER (CATALOG NUMBER 533-556). THE AFFECTED PRODUCTS WERE IDENTIFIED ACROSS MULTIPLE PACKAGES DURING STOCK INSPECTION, AND THE PACKAGES HAD NOT BEEN OPENED. THERE WAS NO REPORTED PATIENT INJURY. THE DEVICES WILL BE RETURNED FOR EVALUATION. THE DEVICES WILL BE RETURNED FOR EVALUATION. FOUR IMAGES WERE REVIEWED AND THE SHOW CATHETERS INSIDE THEIR PACKAGING. POLYMER FLAKES CAN BE SEEN INSIDE THE PACKAGING, AND THEY APPEAR TO HAVE ORIGINATED FROM THE STRAIN RELIEF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23077471 ↗MalfunctionJan 23, 2024Crack; Crack
AS REPORTED, A 5F INFINITI INTERNAL MAMMARY 100CM CATHETER WAS FRACTURED. THERE WAS NO REPORTED PATIENT INJURY. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18563119 ↗
FDA clearance and classification
Cleared through 510(k) K970854 on Sep 30, 1997 (decision: Substantially Equivalent). Submitted by Cordis Corp. as “5 Fr And 6 Fr Infiniti Angiographic Catheter”. The FDA review took 204 days. FDA record ↗
The FDA regulates product code DQO (catheter, intravascular, diagnostic) under 21 CFR 870.1200 as a Class II device. 21 CFR 870.1200 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K970854. One submission often covers many models, so check the product and lot details in each record.
- Z-0286-2024 · Sep 26, 2023 · Open, ClassifiedCordis ANGIOGRAPHIC CATHETERIZATION SETS, INFINITI THRULEMEN, 5F, REF CO0208, cardiac catheter
- Z-0720-2023 · Nov 7, 2022 · Open, ClassifiedCordis Angiographic Catheter Extensions. Used to transport fluid from the power injector to the catheter for injection into the patient.
- Z-1240-06 · Jun 26, 2006 · TerminatedCordis INFINITI 6 French Diagnostic Catheter. Catalog 534-6xxT. That is: 534614T, 534615T, 534617T, 534618T, 534619T, 534620T, 534621T, 534622T, 534623T, 534624T, 534625T, 534627T
Common questions about the INFINITI 534560T
Has the INFINITI 534560T been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705032013963), based on the FDA recall file dated Nov 27, 2026. However, 3 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the INFINITI 534560T?
The FDA's MAUDE database lists 4 adverse event reports for this device, received between Jan 23, 2024 and Sep 17, 2025. In those reports, the most common specific device problems are “Degraded” (3) and “Crack” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the INFINITI 534560T?
None of the reports are classified as deaths, 0 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the INFINITI 534560T FDA approved?
It was cleared through 510(k) K970854 on Sep 30, 1997. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the INFINITI 534560T, and is it still sold?
The FDA's device database lists Cordis Corporation as the company responsible for it (brand INFINITI, model 534560T). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H74908526031Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page