INFINITI 534570TRecalls & reported problems
Cordis Corporation · CATH F5 INF RCB 100CM · Class 2
All models of INFINITI (22)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10705032013994), received between Jun 20, 2025 and Jun 23, 2025 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Degraded” (8 reports).
3 other recalls cover devices cleared under the same FDA submission (K970854); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Degraded8
Patient problems
- No Clinical Signs, Symptoms or Conditions8
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 23, 2025Degraded; Degraded
AS REPORTED, FOUR (4) 5F 100 CM INFINITI RIGHT CORONARY BYPASS (RCB) DIAGNOSTIC CATHETERS DETERIORATED NEAR THE HUB. IT IS POSSIBLE THAT UV LIGHT EXPOSURE OCCURRED THROUGH A CRACK IN THE CABINET. THE ONLY CATHETERS AFFECTED WERE THOSE HANGING ON THE MIDDLE RACK, WHICH WAS ALIGNED WITH THE CABINET OPENING. THERE WAS NO REPORTED PATIENT INJURY. THE CATHETERS WERE STORED IN A CABINET WITH ULTRAVIOLET (UV) PROTECTION AND ARE STILL IN THE STERILE PACKAGING. THERE WAS NO VISIBLE PACKAGING DAMAGE. NO PHOTOGRAPHS ARE AVAILABLE. THE DEVICES WERE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22306487 ↗MalfunctionJun 23, 2025Degraded; Degraded
AS REPORTED, FOUR (4) 5F 100 CM INFINITI RIGHT CORONARY BYPASS (RCB) DIAGNOSTIC CATHETERS DETERIORATED NEAR THE HUB. IT IS POSSIBLE THAT UV LIGHT EXPOSURE OCCURRED THROUGH A CRACK IN THE CABINET. THE ONLY CATHETERS AFFECTED WERE THOSE HANGING ON THE MIDDLE RACK, WHICH WAS ALIGNED WITH THE CABINET OPENING. THERE WAS NO REPORTED PATIENT INJURY. THE CATHETERS WERE STORED IN A CABINET WITH ULTRAVIOLET (UV) PROTECTION AND ARE STILL IN THE STERILE PACKAGING. THERE WAS NO VISIBLE PACKAGING DAMAGE. NO PHOTOGRAPHS ARE AVAILABLE. THE DEVICES WERE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22306431 ↗MalfunctionJun 23, 2025Degraded; Degraded
AS REPORTED, FOUR (4) 5F 100 CM INFINITI RIGHT CORONARY BYPASS (RCB) DIAGNOSTIC CATHETERS DETERIORATED NEAR THE HUB. IT IS POSSIBLE THAT UV LIGHT EXPOSURE OCCURRED THROUGH A CRACK IN THE CABINET. THE ONLY CATHETERS AFFECTED WERE THOSE HANGING ON THE MIDDLE RACK, WHICH WAS ALIGNED WITH THE CABINET OPENING. THERE WAS NO REPORTED PATIENT INJURY. THE CATHETERS WERE STORED IN A CABINET WITH ULTRAVIOLET (UV) PROTECTION AND ARE STILL IN THE STERILE PACKAGING. THERE WAS NO VISIBLE PACKAGING DAMAGE. NO PHOTOGRAPHS ARE AVAILABLE. THE DEVICES WERE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22306379 ↗MalfunctionJun 23, 2025Degraded; Degraded
AS REPORTED, FOUR (4) 5F 100 CM INFINITI RIGHT CORONARY BYPASS (RCB) DIAGNOSTIC CATHETERS DETERIORATED NEAR THE HUB. IT IS POSSIBLE THAT UV LIGHT EXPOSURE OCCURRED THROUGH A CRACK IN THE CABINET. THE ONLY CATHETERS AFFECTED WERE THOSE HANGING ON THE MIDDLE RACK, WHICH WAS ALIGNED WITH THE CABINET OPENING. THERE WAS NO REPORTED PATIENT INJURY. THE CATHETERS WERE STORED IN A CABINET WITH ULTRAVIOLET (UV) PROTECTION AND ARE STILL IN THE STERILE PACKAGING. THERE WAS NO VISIBLE PACKAGING DAMAGE. NO PHOTOGRAPHS ARE AVAILABLE. THE DEVICES ARE EXPECTED TO BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22306334 ↗MalfunctionJun 20, 2025Degraded; Degraded
AS REPORTED, FOUR (4) 5F 100 CM INFINITI RIGHT CORONARY BYPASS (RCB) DIAGNOSTIC CATHETERS DETERIORATED NEAR THE HUB WHILE STILL IN THE PACKAGING. IT IS POSSIBLE THAT UV LIGHT EXPOSURE OCCURRED THROUGH THE CRACK IN THE CABINET. THE ONLY CATHETERS AFFECTED WERE THOSE HANGING ON THE MIDDLE RACK, WHICH WAS ALIGNED WITH THE CABINET OPENING. THERE WAS NO REPORTED PATIENT INJURY. THE CATHETERS WERE STORED IN A CABINET WITH ULTRAVIOLET (UV) PROTECTION. THERE WAS NO VISIBLE PACKAGING DAMAGE. NO PHOTOGRAPHS ARE AVAILABLE. THE DEVICES WERE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22301741 ↗MalfunctionJun 20, 2025Degraded; Degraded
AS REPORTED, FOUR (4) 5F 100 CM INFINITI RIGHT CORONARY BYPASS (RCB) DIAGNOSTIC CATHETERS DETERIORATED NEAR THE HUB WHILE STILL IN THE PACKAGING. IT IS POSSIBLE THAT UV LIGHT EXPOSURE OCCURRED THROUGH THE CRACK IN THE CABINET. THE ONLY CATHETERS AFFECTED WERE THOSE HANGING ON THE MIDDLE RACK, WHICH WAS ALIGNED WITH THE CABINET OPENING. THERE WAS NO REPORTED PATIENT INJURY. THE CATHETERS WERE STORED IN A CABINET WITH ULTRAVIOLET (UV) PROTECTION. THERE WAS NO VISIBLE PACKAGING DAMAGE. NO PHOTOGRAPHS ARE AVAILABLE. THE DEVICES WERE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22301647 ↗MalfunctionJun 20, 2025Degraded; Degraded
AS REPORTED, FOUR (4) 5F 100 CM INFINITI RIGHT CORONARY BYPASS (RCB) DIAGNOSTIC CATHETERS DETERIORATED NEAR THE HUB WHILE STILL IN THE PACKAGING. IT IS POSSIBLE THAT UV LIGHT EXPOSURE OCCURRED THROUGH THE CRACK IN THE CABINET. THE ONLY CATHETERS AFFECTED WERE THOSE HANGING ON THE MIDDLE RACK, WHICH WAS ALIGNED WITH THE CABINET OPENING. THERE WAS NO REPORTED PATIENT INJURY. THE CATHETERS WERE STORED IN A CABINET WITH ULTRAVIOLET (UV) PROTECTION. THERE WAS NO VISIBLE PACKAGING DAMAGE. NO PHOTOGRAPHS ARE AVAILABLE. THE DEVICES WERE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22301598 ↗MalfunctionJun 20, 2025Degraded; Degraded
AS REPORTED, FOUR (4) 5F 100 CM INFINITI RIGHT CORONARY BYPASS (RCB) DIAGNOSTIC CATHETERS DETERIORATED NEAR THE HUB WHILE STILL IN THE PACKAGING. THERE WAS NO REPORTED PATIENT INJURY. THE CATHETERS WERE STORED IN A CABINET WITH ULTRAVIOLET (UV) PROTECTION. THE DEVICES ARE EXPECTED TO BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22301463 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K970854. One submission often covers many models, so check the product and lot details in each record.
- Z-0286-2024 · Sep 26, 2023 · Open, ClassifiedCordis ANGIOGRAPHIC CATHETERIZATION SETS, INFINITI THRULEMEN, 5F, REF CO0208, cardiac catheter
- Z-0720-2023 · Nov 7, 2022 · Open, ClassifiedCordis Angiographic Catheter Extensions. Used to transport fluid from the power injector to the catheter for injection into the patient.
- Z-1240-06 · Jun 26, 2006 · TerminatedCordis INFINITI 6 French Diagnostic Catheter. Catalog 534-6xxT. That is: 534614T, 534615T, 534617T, 534618T, 534619T, 534620T, 534621T, 534622T, 534623T, 534624T, 534625T, 534627T
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H749085263001Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page