INFINITI 534622TRecalls & reported problems

Cordis US Corp. · CATH F6INF TL JL 5 100CM · Class 2

All models of INFINITI (20)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10705032014205), received between Aug 10, 2021 and Mar 31, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Crack” (1 report).

3 other recalls cover devices cleared under the same FDA submission (K970854); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Crack1
  • Degraded1
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 31, 2025Degraded; Degraded

    AS REPORTED, DURING INVENTORY REVIEW, A STAFF MEMBER NOTICED THAT THE UN-OPENED F6 INFINITI JL 5 DIAGNOSTIC CATHETER HAD A SIGNIFICANT AMOUNT OF BLUE COLORED PARTICLES SPRINKLED ALL THROUGHOUT THE INSIDE OF THE UNOPENED PACKAGE. THERE WAS NO REPORTED INJURY TO THE PATIENT. THE DEVICE HAS NOT BEEN REMOVED FROM ITS PACKAGE, SO NO OBSERVATIONS OF FLAKING COULD BE MADE. IT WAS STORED AND PREPPED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THE DEVICES ARE HUNG ON A HOOK IN THE CATH LAB AND RECEIVED NO DIRECT SUNLIGHT EXPOSURE, ONLY INDIRECT EXPOSURE TO NORMAL CATH LAB LIGHTING. THEY ARE NOT STORED IN THE WHITE OUTER BOX UNTIL USE BUT ARE KEPT IN CABINETS WITH TRANSPARENT OR DARK DOORS. THE SHIPMENT OF DEVICES WAS ALWAYS STORED IN A CLIMATE-CONTROLLED ENVIRONMENT BEFORE BEING PLACED IN THE LAB. THE DEVICE HAS NOT BEEN REPROCESSED OR RE-STERILIZED AND REMAINS IN ITS ORIGINAL CORDIS PACKAGING. IT HAS NOT BEEN EXPOSED TO UV LIGHTING OR ANY FORM OF STERILIZATION PROCESS AT ANY POINT. THE LAB DOES NOT USE ANY UV OR STERILIZATION PROCESSES IN GENERAL. OTHER DEVICES IN THE BOX WERE NOT EXAMINED, AS NO OTHER DEVICES WITH THIS LOT NUMBER REMAINED IN THE INVENTORY. THE DEVICE WILL BE RETURNED FOR

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21729871 ↗
  • MalfunctionMay 14, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    AS REPORTED, A 6F JUDKINS LEFT 5 INFINITI DIAGNOSTIC CATHETER BROKE OFF AT THE PROXIMAL BEND UPON INSERTING INTO A NON-CORDIS SHEATH. THERE WAS NO REPORTED PATIENT INJURY. THE DEVICE SEPARATED IN PATIENT. THE DEVICE WAS PULLED INTO THE SHEATH AND SEPARATED ONCE IN SHEATH. THE SHEATH WAS REMOVED WITH THE PIECE INTACT. THE PATIENT RECENTLY HAD AN ENDOVASCULAR ANEURYSM REPAIR (EVAR) AND THE OPERATOR OF THE DEVICE BELIEVES THE STENT-GRAFT MAY HAVE NOT BEEN FULLY ENDOTHELIALIZED, WHICH COULD HAVE CAUSED A CUT ON THE CATHETER THAT CAUSED IT TO BREAK. THE PRODUCT WAS STORED PROPERLY AND PREPPED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THERE WERE NO OTHER DAMAGES NOTED PRIOR TO INSERTING THE PRODUCT THE PRODUCT INTO THE PATIENT. A CONTRALATERAL APPROACH WAS NOT USED. THE DEVICE WAS PULLED FROM THE PACKAGING BY THE HUB AND SHAFT OF CATHETER. THE VESSEL WAS NOT ENGAGED DUE TO THE ISSUE EXPERIENCED. THE DEVICE WAS NOT USED FOR A CHRONIC TOTAL OCCLUSION (CTO). THE CATHETER DID NOT BECOME ENTRAPPED AT ANY POINT IN THE PROCEDURE. EXCESSIVE TORQUING WAS NOT REQUIRED. THE DEVICE DID NOT KINK IN THE AREA OF SEPARATION. THE PROCEDURE ENDED BECAUSE OF CATHETER SEPARATION. THERE WAS NO INJURY TO T

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16928420 ↗
  • MalfunctionAug 10, 2021Crack; Crack

    A 6F 100CM JL 5 INFINITI THRULMEN DIAGNOSTIC CATHETER HAD A CRACK IN THE DISTAL END AND WAS IT WAS NOTICED WHEN DEVICE WAS BEING PREPPED FOR USE. THE DEVICE NEVER ENTERED THE PATIENT, AND NO HARM WAS CAUSED. IT WAS INITIALLY REPORTED "WIRE LOOKS LIKE IT WAS DAMAGE DURING MANUFACTURING. THERE IS A SNIP OUT IF IT NEAR THE J END." THIS PRODUCT DOES NOT CONTAIN A WIRE. THE DEVICE WILL BE RETURNED FOR EVALUATION. 3 IMAGES WERE PROVIDED FOR REVIEW.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12301758 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K970854. One submission often covers many models, so check the product and lot details in each record.

  • Z-0286-2024 · Sep 26, 2023 · Open, Classified
    Cordis ANGIOGRAPHIC CATHETERIZATION SETS, INFINITI THRULEMEN, 5F, REF CO0208, cardiac catheter
  • Z-0720-2023 · Nov 7, 2022 · Open, Classified
    Cordis Angiographic Catheter Extensions. Used to transport fluid from the power injector to the catheter for injection into the patient.
  • Z-1240-06 · Jun 26, 2006 · Terminated
    Cordis INFINITI 6 French Diagnostic Catheter. Catalog 534-6xxT. That is: 534614T, 534615T, 534617T, 534618T, 534619T, 534620T, 534621T, 534622T, 534623T, 534624T, 534625T, 534627T

Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices

All CATHETER, INTRAVASCULAR, DIAGNOSTIC devices →