INFINITI 534642TRecalls & reported problems
Cordis Corporation · CATH F6INF TL MP A2 100CM 2SH · Class 2
All models of INFINITI (22)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Cordis INFINITI diagnostic catheter is a thin tube used to look inside blood vessels. The reports mostly describe the catheter coming apart — parts detaching, and in several cases the shaft cracking so it split into separate pieces, sometimes discovered as soon as the package was opened before the procedure began. There are 71 reports, with no deaths and no injuries, and no recalls name this device. Reporting jumped sharply, from 2 in the earlier 12 months to 66 in the most recent 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 71
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 71 adverse event reports that list this exact device (DI 10705032014304), received between Jan 25, 2024 and Mar 25, 2026 — classified as 71 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (67 reports).
3 other recalls cover devices cleared under the same FDA submission (K970854); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component67
- Crack3
- Difficult to Open or Remove Packaging Material1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions71
Recent adverse event reports
The 25 most recent of 71 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 25, 2026Detachment of Device or Device Component
AS REPORTED, A FRACTURE WAS FOUND ON A 6F 100CM MULTIPURPOSE (MPA) 2 INFINITI DIAGNOSTIC CATHETER IMMEDIATELY UPON OPENING THE PACKAGE AFTER REMOVAL OF THE CATHETER FROM THE STERILE PACKAGING DURING PREPARATION. THE FRACTURE WAS LOCATED AT THE BODY NEAR THE TIP AND RESULTED IN COMPLETE SEPARATION OF THE DEVICE. ANOTHER 6F 100CM MPA 2 INFINITI DIAGNOSTIC CATHETER WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THE DEVICE WAS STORED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO DAMAGE TO THE PACKAGING PRIOR TO USE. THE CATHETER WAS REMOVED TOGETHER WITH THE CARDBOARD HOLDER. PRIOR TO OBSERVING THE FRACTURE, THE PREPARATION STEP PERFORMED WAS SHAPING THE TIP. THE DEVICE WAS NOT RETURNED AS EXPECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24690080 ↗MalfunctionMar 25, 2026Detachment of Device or Device Component
AS REPORTED, A FRACTURE WAS FOUND ON A 6F 100CM MULTIPURPOSE (MPA) 2 INFINITI DIAGNOSTIC CATHETER IMMEDIATELY UPON OPENING THE PACKAGE AFTER REMOVAL OF THE CATHETER FROM THE STERILE PACKAGING DURING PREPARATION. THE FRACTURE WAS LOCATED AT THE BODY NEAR THE TIP AND RESULTED IN COMPLETE SEPARATION OF THE DEVICE. ANOTHER 6F 100CM MPA 2 INFINITI DIAGNOSTIC CATHETER WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THE DEVICE WAS STORED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO DAMAGE TO THE PACKAGING PRIOR TO USE. THE CATHETER WAS REMOVED TOGETHER WITH THE CARDBOARD HOLDER. NO PREPARATION STEPS WERE PERFORMED PRIOR TO OBSERVING THE FRACTURE. THE DEVICE WAS NOT RETURNED AS EXPECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24689975 ↗MalfunctionMar 25, 2026Detachment of Device or Device Component
AS REPORTED, A FRACTURE WAS FOUND ON A 6F 100CM MULTIPURPOSE (MPA) 2 INFINITI DIAGNOSTIC CATHETER IMMEDIATELY UPON OPENING THE PACKAGE. THE DAMAGE WAS IDENTIFIED AFTER REMOVAL OF THE CATHETER FROM THE STERILE PACKAGING AND SHAPING THE TIP DURING PREPARATION. THE FRACTURE OCCURRED AT THE BODY NEAR THE TIP AND RESULTED IN A COMPLETE SEPARATION OF THE DEVICE. ANOTHER 6F 100CM MPA 2 INFINITI DIAGNOSTIC CATHETER WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THE DEVICE WAS NOT RETURNED AS EXPECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24689916 ↗MalfunctionMar 25, 2026Detachment of Device or Device Component
AS REPORTED, A FRACTURE WAS FOUND ON A 6F 100CM MULTIPURPOSE (MPA) 2 INFINITI DIAGNOSTIC CATHETER IMMEDIATELY UPON OPENING THE PACKAGE. THE DAMAGE WAS IDENTIFIED AFTER REMOVAL OF THE CATHETER FROM THE STERILE PACKAGING AND SHAPING THE TIP DURING PREPARATION. THE FRACTURE OCCURRED AT THE BODY NEAR THE TIP AND RESULTED IN A COMPLETE SEPARATION OF THE DEVICE. ANOTHER 6F 100CM MPA 2 INFINITI DIAGNOSTIC CATHETER WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THE DEVICE WAS NOT RETURNED AS EXPECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24689821 ↗MalfunctionMar 25, 2026Detachment of Device or Device Component
AS REPORTED, A FRACTURE WAS FOUND ON A 6F 100CM MULTIPURPOSE (MPA) 2 INFINITI DIAGNOSTIC CATHETER IMMEDIATELY UPON OPENING THE PACKAGE. THE DAMAGE WAS IDENTIFIED AFTER REMOVAL OF THE CATHETER FROM THE STERILE PACKAGING AND SHAPING THE TIP DURING PREPARATION. THE FRACTURE OCCURRED AT THE BODY NEAR THE TIP AND RESULTED IN A COMPLETE SEPARATION OF THE DEVICE. ANOTHER 6F 100CM MPA 2 INFINITI DIAGNOSTIC CATHETER WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24689742 ↗MalfunctionMar 25, 2026Detachment of Device or Device Component
AS REPORTED, A FRACTURE WAS FOUND ON A 6F 100CM MULTIPURPOSE (MPA) 2 INFINITI DIAGNOSTIC CATHETER IMMEDIATELY UPON OPENING THE PACKAGE. THE DAMAGE WAS IDENTIFIED AFTER REMOVAL OF THE CATHETER FROM THE STERILE PACKAGING AND SHAPING THE TIP DURING PREPARATION. THE FRACTURE OCCURRED AT THE BODY NEAR THE TIP AND RESULTED IN A COMPLETE SEPARATION OF THE DEVICE. ANOTHER 6F 100CM MPA 2 INFINITI DIAGNOSTIC CATHETER WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24689714 ↗MalfunctionMar 25, 2026Detachment of Device or Device Component
AS REPORTED, A FRACTURE WAS FOUND ON A 6F 100CM MULTIPURPOSE (MPA) 2 INFINITI DIAGNOSTIC CATHETER IMMEDIATELY UPON OPENING THE PACKAGE. THE DAMAGE WAS IDENTIFIED AFTER REMOVAL OF THE CATHETER FROM THE STERILE PACKAGING AND SHAPING THE TIP DURING PREPARATION. THE FRACTURE OCCURRED AT THE BODY NEAR THE TIP AND RESULTED IN A COMPLETE SEPARATION OF THE DEVICE. ANOTHER 6F 100CM MPA 2 INFINITI DIAGNOSTIC CATHETER WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24689671 ↗MalfunctionMar 25, 2026Detachment of Device or Device Component
AS REPORTED, A FRACTURE WAS FOUND ON A 6F 100CM MULTIPURPOSE (MPA) 2 INFINITI DIAGNOSTIC CATHETER IMMEDIATELY UPON OPENING THE PACKAGE. THE DAMAGE WAS IDENTIFIED AFTER REMOVAL OF THE CATHETER FROM THE STERILE PACKAGING AND SHAPING THE TIP DURING PREPARATION. THE FRACTURE OCCURRED AT THE BODY NEAR THE TIP AND RESULTED IN A COMPLETE SEPARATION OF THE DEVICE. ANOTHER 6F 100CM MPA 2 INFINITI DIAGNOSTIC CATHETER WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24689626 ↗MalfunctionMar 25, 2026Detachment of Device or Device Component
AS REPORTED, A FRACTURE WAS FOUND ON A 6F 100CM MULTIPURPOSE (MPA) 2 INFINITI DIAGNOSTIC CATHETER IMMEDIATELY UPON OPENING THE PACKAGE. THE DAMAGE WAS IDENTIFIED AFTER REMOVAL OF THE CATHETER FROM THE STERILE PACKAGING DURING PREPARATION. THE FRACTURE OCCURRED AT THE BODY NEAR THE TIP AND RESULTED IN A COMPLETE SEPARATION OF THE DEVICE. ANOTHER 6F 100CM MPA 2 INFINITI DIAGNOSTIC CATHETER WAS USED TO COMPLETE THE PROCEDURE. NO PREPARATION STEPS WERE PERFORMED PRIOR TO OBSERVING THE FRACTURE. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24689509 ↗MalfunctionMar 25, 2026Detachment of Device or Device Component
AS REPORTED, A FRACTURE WAS FOUND ON A 6F 100CM MULTIPURPOSE (MPA) 2 INFINITI DIAGNOSTIC CATHETER IMMEDIATELY UPON OPENING THE PACKAGE. THE DAMAGE WAS IDENTIFIED AFTER REMOVAL OF THE CATHETER FROM THE STERILE PACKAGING DURING PREPARATION. THE FRACTURE OCCURRED AT THE BODY NEAR THE TIP AND RESULTED IN A COMPLETE SEPARATION OF THE DEVICE. ANOTHER 6F 100CM MPA 2 INFINITI DIAGNOSTIC CATHETER WAS USED TO COMPLETE THE PROCEDURE. NO PREPARATION STEPS WERE PERFORMED PRIOR TO OBSERVING THE FRACTURE. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24689446 ↗MalfunctionMar 25, 2026Detachment of Device or Device Component
AS REPORTED, A FRACTURE WAS FOUND ON A 6F 100CM MULTIPURPOSE (MPA) 2 INFINITI DIAGNOSTIC CATHETER IMMEDIATELY UPON OPENING THE PACKAGE. THE DAMAGE WAS IDENTIFIED AFTER REMOVAL OF THE CATHETER FROM THE STERILE PACKAGING DURING PREPARATION. THE FRACTURE OCCURRED AT THE BODY NEAR THE TIP AND RESULTED IN A COMPLETE SEPARATION OF THE DEVICE. ANOTHER 6F 100CM MPA 2 INFINITI DIAGNOSTIC CATHETER WAS USED TO COMPLETE THE PROCEDURE. NO PREPARATION STEPS WERE PERFORMED PRIOR TO OBSERVING THE FRACTURE. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24689385 ↗MalfunctionMar 25, 2026Detachment of Device or Device Component
AS REPORTED, A FRACTURE WAS FOUND ON A 6F 100CM MULTIPURPOSE (MPA) 2 INFINITI DIAGNOSTIC CATHETER IMMEDIATELY UPON OPENING THE PACKAGE. THE DAMAGE WAS IDENTIFIED AFTER REMOVAL OF THE CATHETER FROM THE STERILE PACKAGING DURING PREPARATION. THE FRACTURE OCCURRED AT THE BODY NEAR THE TIP AND RESULTED IN A COMPLETE SEPARATION OF THE DEVICE. ANOTHER 6F 100CM MPA 2 INFINITI DIAGNOSTIC CATHETER WAS USED TO COMPLETE THE PROCEDURE. NO PREPARATION STEPS WERE PERFORMED PRIOR TO OBSERVING THE FRACTURE. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THE DEVICE WILL BE RETURNED FOR EVALUATION. THE DEVICE WAS STORED IN ACCORDANCE WITH THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO DAMAGE TO THE DEVICE PACKAGING PRIOR TO USE; HOWEVER, DAMAGE WAS IDENTIFIED AFTER REMOVAL OF THE CATHETER FROM THE STERILE PACKAGING DURING PREPARATION. THE CATHETER WAS REMOVED TOGETHER WITH THE CARDBOARD HOLDER. THE FRACTURE OCCURRED AT THE CATHETER BODY NEAR THE TIP AND RESULTED IN COMPLETE SEPARATION OF THE DEVICE. ANOTHER MPA (534642T) WAS USED TO COMPLETE THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24689316 ↗MalfunctionMar 23, 2026Detachment of Device or Device Component
AS REPORTED, THE 6F MP A2 INFINITI ANGIOGRAPHY CATHETER BROKE INTO THREE PIECES AS SOON AS IT WAS TAKEN OUT, SO WE REPLACED IT WITH ANOTHER CATHETER AND CONTINUED USING IT. THERE WAS NO REPORTED INJURY TO THE PATIENT. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24663688 ↗MalfunctionMar 12, 2026Detachment of Device or Device Component
AS REPORTED, TWO 6F MULTIPURPOSE (MP) A2 100CM INFINITI THRULUMEN DIAGNOSTIC CATHETER BROKE DURING SHAPING. THE STRAIN RELIEF PORTION OF THE DEVICE BROKE DURING SHAPING, RESULTING IN A FRACTURE AND SEPARATION INTO MULTIPLE PIECES. THE SURGEON IMMEDIATELY STOPPED USING THE CATHETERS AND REPLACED THEM WITH ANOTHER NON-CORDIS CATHETER TO CONTINUE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY. THE EVENT OCCURRED DURING A PATENT DUCTUS ARTERIOSUS (PDA) SURGICAL PROCEDURE AFTER AN UNKNOWN GUIDEWIRE HAD BEEN REMOVED. THE DEVICE WAS STORED AND PREPARED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). SHAPING WAS PERFORMED OUTSIDE OF THE PATIENT, AND THE BREAKAGE ALSO OCCURRED OUTSIDE OF THE PATIENT. THE DEVICES WILL BE RETURNED FOR EVALUATION. THE HOSPITAL REPORTED THIS CASE AS A SERIOUS INJURY ADVERSE EVENT TO THE CHINA NMPA.ADDENDUM: BOTH DEVICES WERE RETURNED WITH SEPARATIONS AT THE DISTAL TIP AND THE STRAIN RELIEFS FULLY INTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24583488 ↗MalfunctionMar 12, 2026Detachment of Device or Device Component
AS REPORTED, TWO 6F MULTIPURPOSE (MP) A2 100CM INFINITI THRULUMEN DIAGNOSTIC CATHETER BROKE DURING SHAPING. THE STRAIN RELIEF PORTION OF THE DEVICE BROKE DURING SHAPING, RESULTING IN A FRACTURE AND SEPARATION INTO MULTIPLE PIECES. THE SURGEON IMMEDIATELY STOPPED USING THE CATHETERS AND REPLACED THEM WITH ANOTHER NON-CORDIS CATHETER TO CONTINUE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY. THE EVENT OCCURRED DURING A PATENT DUCTUS ARTERIOSUS (PDA) SURGICAL PROCEDURE AFTER AN UNKNOWN GUIDEWIRE HAD BEEN REMOVED. THE DEVICE WAS STORED AND PREPARED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). SHAPING WAS PERFORMED OUTSIDE OF THE PATIENT, AND THE BREAKAGE ALSO OCCURRED OUTSIDE OF THE PATIENT. THE DEVICES WILL BE RETURNED FOR EVALUATION. THE HOSPITAL REPORTED THIS CASE AS A SERIOUS INJURY ADVERSE EVENT TO THE CHINA NMPA. ADDENDUM: BOTH DEVICES WERE RETURNED WITH SEPARATIONS AT THE DISTAL TIP AND THE STRAIN RELIEFS FULLY INTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24582551 ↗MalfunctionMar 12, 2026Detachment of Device or Device Component
AS REPORTED, A 6F MULTIPURPOSE (MP) A2 100CM INFINITI THRULUMEN DIAGNOSTIC CATHETER BROKE DURING SHAPING. THE CATHETER BODY FRACTURED AT THE JUNCTION BETWEEN THE DISTAL TIP AND THE CATHETER SHAFT DURING SHAPING, RESULTING IN A SEPARATION INTO MULTIPLE PIECES. THE SURGEON IMMEDIATELY STOPPED USING THE CATHETER AND REPLACED IT WITH ANOTHER NON-CORDIS CATHETER TO CONTINUE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY. THE EVENT OCCURRED DURING A PATENT DUCTUS ARTERIOSUS (PDA) SURGICAL PROCEDURE AFTER AN UNKNOWN GUIDEWIRE HAD BEEN REMOVED. THE DEVICE WAS STORED AND PREPARED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). SHAPING WAS PERFORMED OUTSIDE OF THE PATIENT, AND THE BREAKAGE ALSO OCCURRED OUTSIDE OF THE PATIENT. THE DEVICE WILL BE RETURNED FOR EVALUATION. THE HOSPITAL REPORTED THIS CASE AS A SERIOUS INJURY ADVERSE EVENT TO THE CHINA NMPA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24582418 ↗MalfunctionFeb 6, 2026Detachment of Device or Device Component
AS REPORTED, A 6F 100CM MULTIPURPOSE (MPA) 2 TWO SIDE-HOLE INFINITI DIAGNOSTIC CATHETERS FRACTURED AT THE STRAIN RELIEF, APPROXIMATELY 8.5 CM FROM THE DISTAL TIP DURING ROUTINE SHAPING AT THE TIP OF THE CATHETER, AND IT WAS FOUND THAT THERE WAS NO STEEL WIRE BRAID LAYER IN THE STRAIN RELIEF. THE DEVICE DID NOT ENTER THE PATIENT AND ANOTHER 6F 100CM MPA2 TWO SIDE-HOLE INFINITI DIAGNOSTIC CATHETER WAS USED AS A REPLACEMENT. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THE LUER HUB DID NOT SEPARATE FROM THE STRAIN RELIEF. THE DEVICE WAS STORED AND PREPARED PER THE INSTRUCTIONS FOR USE (IFU). THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24280048 ↗MalfunctionJan 14, 2026Detachment of Device or Device Component
AS REPORTED, A 6F .070-INCH MULTIPURPOSE (MP) A2 100CM INFINITI DIAGNOSTIC CATHETER WAS FOUND TO BE SEPARATED UPON OPENING, WHEN THE OPERATOR WAS PREPARING TO START THE PROCEDURE. UPON OPENING THE INNER PACKAGE, THE SEPARATION OF THE TUBE BODY WAS OBSERVED. THE REPORTED ISSUE DESCRIBED AS ¿BROKEN¿ REFERRED TO THE CATHETER HEAD END BEING SEPARATED FROM THE TUBE BODY WHEN THE PACKAGE WAS OPENED. ANOTHER INFINITI CATHETER WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY. THE INTENDED PROCEDURE WAS TRANSCATHETER AORTIC VALVE IMPLANTATION (TAVI). THE PRODUCT WAS STORED AND HANDLED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). UPON OPENING THE INNER PACKAGE, THE SEPARATION OF THE TUBE BODY WAS OBSERVED. THE INNER PACKAGE WAS TORN OPEN, AND THE CARDBOARD WAS ATTACHED TOGETHER WITH THE CATHETER. NO OTHER DAMAGE WAS NOTED. THERE WAS NO DIFFICULTY REMOVING THE DEVICE FROM THE STERILE PACKAGING. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24071542 ↗MalfunctionJan 14, 2026Detachment of Device or Device Component
AS REPORTED, A 6F .070-INCH MULTIPURPOSE (MP) A2 100CM INFINITI DIAGNOSTIC CATHETER WAS FOUND TO BE SEPARATED UPON OPENING, WHEN THE OPERATOR WAS PREPARING TO START THE PROCEDURE. UPON OPENING THE INNER PACKAGE, THE SEPARATION OF THE TUBE BODY WAS OBSERVED. THE REPORTED ISSUE DESCRIBED AS ¿BROKEN¿ REFERRED TO THE CATHETER HEAD END BEING SEPARATED FROM THE TUBE BODY WHEN THE PACKAGE WAS OPENED. ANOTHER INFINITI CATHETER WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY. THE INTENDED PROCEDURE WAS TRANSCATHETER AORTIC VALVE IMPLANTATION (TAVI). THE PRODUCT WAS STORED AND HANDLED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). UPON OPENING THE INNER PACKAGE, THE SEPARATION OF THE TUBE BODY WAS OBSERVED. THE INNER PACKAGE WAS TORN OPEN, AND THE CARDBOARD WAS ATTACHED TOGETHER WITH THE CATHETER. NO OTHER DAMAGE WAS NOTED. THERE WAS NO DIFFICULTY REMOVING THE DEVICE FROM THE STERILE PACKAGING. THE DEVICE WILL BE RETURNED FOR EVALUATION
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24067820 ↗MalfunctionJan 12, 2026Detachment of Device or Device Component
AS REPORTED, FORTY-FIVE 6F MP A2 INFINITI CATHETERS WERE FOUND TO BE BROKEN IMMEDIATELY UPON OPENING THE PACKAGE. WHEN UNPACKING, IT WAS FOUND THAT THE HEAD END AND THE TUBE BODY HAD BEEN SPITED. ALL 45 CATHETERS WERE CONFIRMED TO HAVE THE SAME MALFUNCTION. THERE WAS NO REPORTED INJURY TO THE PATIENT. THERE WAS NO DAMAGE TO THE PACKAGING. THE DEVICES WERE REMOVED BY TEARING THE INNER PACKAGE, AND REMOVING THE CATHETER AND CARDBOARD TOGETHER. THE DEVICES WILL NOT BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24050025 ↗MalfunctionJan 12, 2026Detachment of Device or Device Component
AS REPORTED, FORTY-FIVE 6F MP A2 INFINITI CATHETERS WERE FOUND TO BE BROKEN IMMEDIATELY UPON OPENING THE PACKAGE. WHEN UNPACKING, IT WAS FOUND THAT THE HEAD END AND THE TUBE BODY HAD BEEN SPITED. ALL 45 CATHETERS WERE CONFIRMED TO HAVE THE SAME MALFUNCTION. THERE WAS NO REPORTED INJURY TO THE PATIENT. THERE WAS NO DAMAGE TO THE PACKAGING. THE DEVICES WERE REMOVED BY TEARING THE INNER PACKAGE AND REMOVING THE CATHETER AND CARDBOARD TOGETHER. THE DEVICES WILL NOT BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24050015 ↗MalfunctionJan 12, 2026Detachment of Device or Device Component
AS REPORTED, FORTY-FIVE 6F MP A2 INFINITI CATHETERS WERE FOUND TO BE BROKEN IMMEDIATELY UPON OPENING THE PACKAGE. WHEN UNPACKING, IT WAS FOUND THAT THE HEAD END AND THE TUBE BODY HAD BEEN SPITED. ALL 45 CATHETERS WERE CONFIRMED TO HAVE THE SAME MALFUNCTION. THERE WAS NO REPORTED INJURY TO THE PATIENT. THERE WAS NO DAMAGE TO THE PACKAGING. THE DEVICES WERE REMOVED BY TEARING THE INNER PACKAGE, AND REMOVING THE CATHETER AND CARDBOARD TOGETHER. THE DEVICES WILL NOT BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24050004 ↗MalfunctionJan 12, 2026Detachment of Device or Device Component
AS REPORTED, FORTY-FIVE 6F MP A2 INFINITI CATHETERS WERE FOUND TO BE BROKEN IMMEDIATELY UPON OPENING THE PACKAGE. WHEN UNPACKING, IT WAS FOUND THAT THE HEAD END AND THE TUBE BODY HAD BEEN SPITED. ALL 45 CATHETERS WERE CONFIRMED TO HAVE THE SAME MALFUNCTION. THERE WAS NO REPORTED INJURY TO THE PATIENT. THERE WAS NO DAMAGE TO THE PACKAGING. THE DEVICES WERE REMOVED BY TEARING THE INNER PACKAGE AND REMOVING THE CATHETER AND CARDBOARD TOGETHER. THE DEVICES WILL NOT BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24050000 ↗MalfunctionJan 12, 2026Detachment of Device or Device Component
AS REPORTED, FORTY-FIVE 6F MP A2 INFINITI CATHETERS WERE FOUND TO BE BROKEN IMMEDIATELY UPON OPENING THE PACKAGE. WHEN UNPACKING, IT WAS FOUND THAT THE HEAD END AND THE TUBE BODY HAD BEEN SPITED. ALL 45 CATHETERS WERE CONFIRMED TO HAVE THE SAME MALFUNCTION. THERE WAS NO REPORTED INJURY TO THE PATIENT. THERE WAS NO DAMAGE TO THE PACKAGING. THE DEVICES WERE REMOVED BY TEARING THE INNER PACKAGE, AND REMOVING THE CATHETER AND CARDBOARD TOGETHER. THE DEVICES WILL NOT BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24049964 ↗MalfunctionJan 12, 2026Detachment of Device or Device Component
AS REPORTED, FORTY-FIVE 6F MP A2 INFINITI CATHETERS WERE FOUND TO BE BROKEN IMMEDIATELY UPON OPENING THE PACKAGE. WHEN UNPACKING, IT WAS FOUND THAT THE HEAD END AND THE TUBE BODY HAD BEEN SPITED. ALL 45 CATHETERS WERE CONFIRMED TO HAVE THE SAME MALFUNCTION. THERE WAS NO REPORTED INJURY TO THE PATIENT. THERE WAS NO DAMAGE TO THE PACKAGING. THE DEVICES WERE REMOVED BY TEARING THE INNER PACKAGE, AND REMOVING THE CATHETER AND CARDBOARD TOGETHER. THE DEVICES WILL NOT BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24049947 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K970854. One submission often covers many models, so check the product and lot details in each record.
- Z-0286-2024 · Sep 26, 2023 · Open, ClassifiedCordis ANGIOGRAPHIC CATHETERIZATION SETS, INFINITI THRULEMEN, 5F, REF CO0208, cardiac catheter
- Z-0720-2023 · Nov 7, 2022 · Open, ClassifiedCordis Angiographic Catheter Extensions. Used to transport fluid from the power injector to the catheter for injection into the patient.
- Z-1240-06 · Jun 26, 2006 · TerminatedCordis INFINITI 6 French Diagnostic Catheter. Catalog 534-6xxT. That is: 534614T, 534615T, 534617T, 534618T, 534619T, 534620T, 534621T, 534622T, 534623T, 534624T, 534625T, 534627T
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H74908526031Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page