Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10705032014397), received between Feb 23, 2021 and Aug 15, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Crack” (1 report).
3 other recalls cover devices cleared under the same FDA submission (K970854); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Crack1
- Degraded1
- Device Contamination with Chemical or Other Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 15, 2025Degraded; Degraded
AS REPORTED, THERE WAS CRACKING OF THE CASING NEAR THE HUB OF TWO 6F INFINITI PIGTAIL 110CM 6SH DIAGNOSTIC CATHETERS. THERE WAS NO REPORTED PATIENT INJURY. THE DEVICE WAS NEVER USED. THE CRACKING WAS NOTICED BY SCRUB TECH AFTER OPENING PACKAGE. THREE IMAGES WERE PROVIDED FOR REVIEW. ONE OF THEM SHOWS A CATHETER WITH A STRAIN RELIEF THAT SHOWS SIGNS OF DEGRADATION. THE DEVICE WAS NOT RETURNED AS EXPECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22822500 ↗MalfunctionJul 28, 2022Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
AS REPORTED, THERE WAS A HAIR PRESENT IN THE ¿INNER PRODUCT¿ OF A 6F INFINITI VENTRICULAR STRAIGHT PIGTAIL SIDEHOLES 110CM DIAGNOSTIC CATHETER. THE ISSUE WAS OBSERVED IN THE LSP CUSTOM BONDED WAREHOUSE DURING INSPECTION AND AT THE TIME OF PASTING OVER LABELLING (INNER AND OUTER) PROCEDURE. THE DEVICE WAS NOT USED IN THE PATIENT. THERE WAS NO REPORTED PATIENT INJURY. THE DEVICE WAS STORED PER INSTRUCTIONS FOR USE (IFU). IMAGE REVIEW WAS COMPLETED AND IMAGE ONE SHOWS THE .UTER BOX WITH LABEL STATING 6F 110CM INFINITI PIGTAIL CATHETER. THE SECOND IMAGE SHOWS THE TIP OF A PIGTAIL CATHETER INSIDE THE STERILE POUCH. A FOREIGN MATERIAL, WHICH LOOKS LIKE HAIR IS NOTED INSIDE THE POUCH. OTHER PROCEDURAL DETAILS WERE REQUESTED BUT ARE UNKNOWN, UNAVAILABLE, NOT ANSWERED, OR NOT APPLICABLE. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15121974 ↗MalfunctionFeb 23, 2021Crack; Crack
WHEN USING A 6F 110CM 6 SIDE HOLES (6SH) INFINITI PIGTAIL DIAGNOSTIC CATHETER, IT WAS FOUND THAT THERE WAS A FRACTURE IN THE MIDDLE SECTION. THEREFORE, THE DEVICE WAS REPLACED WITH ANOTHER UNKNOWN ANGIOGRAPHIC CATHETER TO COMPLETE THE OPERATION. THERE WAS NO REPORTED PATIENT INJURY. ANGIOGRAPHY FOUND THAT THE RIGHT CORONARY ARTERY (RCA) HAD MORE THAN 80% STENOSIS AND NEEDED A PERCUTANEOUS CORONARY INTERVENTION (PCI) TREATMENT. THE RCA HAD MODERATE CALCIFICATION, MODERATE TORTUOSITY, AND MODERATE ANGULATION. THE DEVICE WAS STORED AND PREPPED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THE USER WAS EXPERIENCED WITH THE DEVICE. THE DEVICE WAS OPENED IN THE STERILE FIELD. THE DEVICE WAS NOT RESTERILIZED. THE DEVICE WAS INSPECTED WHEN REMOVED FROM THE PACKAGE BEFORE IT WAS USED AND NO ANOMALIES WERE NOTED WHEN REMOVED FROM THE PACKAGE. THE DEVICE WAS NOT INSERTED THROUGH A STOPCOCK INSTEAD OF A HEMOSTATIC VALVE. THERE WAS NO SEPARATION THAT HAD TO BE SNARED OUT. THERE WAS NO RESISTANCE MET WHILE ADVANCING THE DEVICE OR WHILE ADVANCING THE DEVICE OVER THE GUIDEWIRE. THERE WAS NO EXCESSIVE TORQUEING REQUIRED. THE DEVICE DID NOT PREVIOUSLY KINK OR BEND IN THE AREA THAT CRACKED. THE DEVICE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11364155 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K970854. One submission often covers many models, so check the product and lot details in each record.
- Z-0286-2024 · Sep 26, 2023 · Open, ClassifiedCordis ANGIOGRAPHIC CATHETERIZATION SETS, INFINITI THRULEMEN, 5F, REF CO0208, cardiac catheter
- Z-0720-2023 · Nov 7, 2022 · Open, ClassifiedCordis Angiographic Catheter Extensions. Used to transport fluid from the power injector to the catheter for injection into the patient.
- Z-1240-06 · Jun 26, 2006 · TerminatedCordis INFINITI 6 French Diagnostic Catheter. Catalog 534-6xxT. That is: 534614T, 534615T, 534617T, 534618T, 534619T, 534620T, 534621T, 534622T, 534623T, 534624T, 534625T, 534627T
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H74908526031Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page