Skip to content

INFINITI 534676TRecalls & reported problems

Cordis US Corp. · CATH F6INF TL 3DRC 100CM · Class 2

All models of INFINITI (20)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705032014489), received between May 4, 2018 and Oct 20, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Crack” (2 reports).

3 other recalls cover devices cleared under the same FDA submission (K970854); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Crack2

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 20, 2025Crack; Crack

    AS REPORTED, THE TIP OF A 6F 100 CM 3DRC INFINITI DIAGNOSTIC CATHETER FRACTURED WHILE A .035-INCH WIRE WAS BEING INTRODUCED THROUGH THE CATHETER. THE CATHETER DID NOT BECOME ENTRAPPED, AND THERE WAS NO DEVICE SEPARATION. A SECOND 3DRC INFINITI CATHETER WAS USED TO COMPLETE THE ANGIOGRAPHY PROCEDURE, AND A CONTRALATERAL APPROACH WAS NOT USED. THERE WAS NO REPORTED PATIENT INJURY. THERE WAS NO DAMAGE IN THE CATHETER PACKAGING. THE PRODUCT WAS STORED ON THE SHELF IN THE CATH. LAB AND OPENED IN STERILE FIELD. THE DEVICE WAS PREPPED PER THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO KINK IN THE DISTAL OF THE CATHETER WAS REMARKED. THE DEVICE DAMAGE WAS NOT NOTICED ONCE THE CATHETER WAS OPENED FROM THE PACKAGING. NO FORCE WAS APPLIED BECAUSE THE CATHETER DID NOT INTRODUCE THROUGH THE SHEATH. AT THE TARGET SITE, THE VESSEL WAS NON-CALCIFIED, MILDLY TORTUOUS, NON-BIFURCATING, AND THE STENOSIS PERCENTAGE AND PRESENCE OF AN ACUTE ANGLE WERE NOT AVAILABLE. AT THE ACCESS SITE, THE VESSEL WAS NEITHER CALCIFIED NOR TORTUOUS, AND DETAILS ON STENOSIS PERCENTAGE, ACUTE ANGLE, AND BIFURCATION WERE NOT AVAILABLE. THE DEVICE WAS NEITHER PULLED FROM THE PACKAGING BY THE HUB NOR TORQUED/STEERED BY THE HUB

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23335438 ↗
  • MalfunctionMay 4, 2018Crack; Crack

    BEFORE USE, THE TIP OF THE 6F INFINITI CATHETER WAS NOT ATTACHED AND WAS HANGING BY A PIECE OF PLASTIC. THERE WAS NO PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7487688 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K970854. One submission often covers many models, so check the product and lot details in each record.

  • Z-0286-2024 · Sep 26, 2023 · Open, Classified
    Cordis ANGIOGRAPHIC CATHETERIZATION SETS, INFINITI THRULEMEN, 5F, REF CO0208, cardiac catheter
  • Z-0720-2023 · Nov 7, 2022 · Open, Classified
    Cordis Angiographic Catheter Extensions. Used to transport fluid from the power injector to the catheter for injection into the patient.
  • Z-1240-06 · Jun 26, 2006 · Terminated
    Cordis INFINITI 6 French Diagnostic Catheter. Catalog 534-6xxT. That is: 534614T, 534615T, 534617T, 534618T, 534619T, 534620T, 534621T, 534622T, 534623T, 534624T, 534625T, 534627T

Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices

All CATHETER, INTRAVASCULAR, DIAGNOSTIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page