INFINITI 538476Recalls & reported problems
Cordis Corporation · CATH F4 INF 3DRC 100CM · Class 2
All models of INFINITI (22)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705032014922), received between Dec 9, 2019 and Mar 16, 2020 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Material Puncture/Hole1
Patient problems
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMar 16, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, THE DISTAL TIP OF 4F 100CM 3-DIMENSIONAL RIGHT CORONARY (DRC) INFINITI DIAGNOSTIC CATHETER WAS DETACHED FROM THE CATHETER BODY AFTER REMOVING THE DEVICE FROM A 4F NON-CORDIS SHEATH. THE SCRUB TECH INFORMED THE PHYSICIAN AND THE PHYSICIAN ASSUMED THAT THE "FOREIGN BODY" WAS STILL ON THE WIRE AND POSSIBLE INSIDE THE SHEATH. HOWEVER, AFTER DOING FLUOROSCOPY UP FROM THE FEMORAL ARTERY TO THE AORTA THEY WERE UNABLE TO FIND THE "FOREIGN BODY" IN BOTH FEMORAL AND IN THE AORTA. THEY CONTINUE DOING THE FLUOROSCOPY DOWN THE RIGHT ILIAC AND THEY "FOUND THE FOREIGN" BODY WAS SECURED IN THE PROFUNDA ARTERY. THEY THEN COMPLETED THE LEFT VENTRICULOGRAM (LVGRAM) AND REMOVED THE SHEATH, HOLDING A MANUAL COMPRESSION , PHYSICIAN WAS CALLED IN TO REMOVE THE CATHETER TIP. THE PHYSICIAN WENT IN THE LEFT COMMON FEMORAL ARTERY AND REMOVED THE FOREIGN BODY BY GETTING AN 0.014" WIRE THROUGH THE TIP. HE THEN RAN A SMALL BALLOON INTO THE TIP, INFLATED, AND SUCCESSFULLY REMOVED THE TIP. NO ADVERSE PATIENT EFFECTS, OUTSIDE OF THE SECONDARY SURGERY. THE PROCEDURE WAS LEFT HEART CATHETERIZATION AFTER A RIGHT CORONARY ANGIOGRAM. A 0.035" UNKNOWN GUIDEWIRE WAS INSERTED THROUGH THE INFINITI 3DRC CATHETE
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9834762 ↗MalfunctionDec 9, 2019Material Puncture/Hole; Material Puncture/Hole
AS REPORTED, A 4F 100CM RIGHT CORONARY 3-DIMENSIONAL (3DRC) INFINITI DIAGNOSTIC CATHETER WAS KINKED WITH A HOLE DURING PREP. IT WAS NOTICED THAT THE DEVICE WAS KINKED AFTER PRE-LOADING THE J-WIRE. A J-WIRE WHICH IS FRESH OUT OF PACKAGE WENT THROUGH HOLE. THEREFORE, THE PROCEDURE WAS COMPLETED USING ANOTHER UNKNOWN DIAGNOSTIC CATHETER. THERE WAS NO REPORTED PATIENT INJURY. THE DEVICE WAS STORED AND HANDLED AS PER THE INSTRUCTION FOR USE (IFU). THERE WAS NO DAMAGE NOTED TO THE PACKAGING OF THE DEVICE PRIOR TO USE. THE DEVICE WAS PREPPED ACCORDING TO THE IFU. THE DEVICE WAS OPENED IN THE STERILE FIELD. THE DEVICE WILL BE RETURNED FOR EVALUATION. ADDITIONAL PROCEDURAL DETAILS WERE REQUESTED BUT ARE UNKNOWN.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9438648 ↗
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H74908526031Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page